INSTRUCTION MANUAL
THERMOCOUPLE TRANSMITTER
(field- and PC-configurable)
MODEL
M3LT
BEFORE USE ....
Thank you for choosing M-System. Before use, please check
contents of the package you receiv
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: [email protected]
EM-2661-A Rev.2 P. 2 / 7
COMPONENT IDENTIFICATION
Specifications
(side)
Body
SW Setting
■
■HOW TO OPEN THE COVER WHEN SETTING DIP SW
Hold at the top and bottom of the unit as shown below and
slide the housing
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: [email protected]
EM-2661-A Rev.2 P. 3 / 7
TERMINAL CONNECTIONS
Connect the unit as in the diagram below or refer to the connection diagram on the side of the unit.
■
■EXTERNAL DIMENSIONS unit: mm (inch)
10
11
12
7
8
9
4
5
6
1
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: [email protected]
EM-2661-A Rev.2 P. 4 / 7
EXTERNAL & INTERNAL VIEWS
Configuration DIP SW
4
5
6
■ FRONT VIEW
■ SIDE VIEW
8
SW2
SW1
SW3
7 6 5 4 3 2 1
ON
OFF
ON
OFF
4 3 2 1
ON
OFF
4 3 2 1
10
11
12
7
8
9
Configurator Jack
DOWN But
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: [email protected]
EM-2661-A Rev.2 P. 5 / 7
■
■OUTPUT TYPE (SW2 & 1)
Table 6
OUTPUT
SW2-3
SW2-2
SW1-4
SW1-3
SW1-2
SW1-1
0 – 20 mA
OFF
OFF
OFF
ON
OFF
OFF
-2.5 – +2.5 V
OFF
ON
ON
OFF
OFF
ON
-10 – +10 V
ON
OFF
ON
OFF
ON
OFF
�
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: [email protected]
EM-2661-A Rev.2 P. 6 / 7
■
■PREPARATION (e.g. M3LT-R4/A)
1) Mount the DIP-SW-configured M3LT on to a DIN rail.
2) Connect the M3LT to a simulator and a multimeter and
to a DC power source as shown below.
3) Tu