Thank you for selecting the ECscan series pocket conductivity tester. This manual provides a step-by-step guide to help you operate the instrument,

please carefully read the following instructions before use. Installing the Batteries 1. Remove the battery compartment lid. 2. Insert the three LR44 batteries into the battery compartment, note

The ECscan10 conductivity tester allows 1 point calibration. For better accuracy, we recommend to calibrate the tester regularly. The following

table shows acceptable conductivity range of standard solution for each tester. Model Calibration Solution Range Default ECscan10 L 100 to 180 µS/cm 146.5 µS/cm ECscan10 M 1000 to

Press the oC key, the tester shows the temperature readings.

• Press the Meas key to return to the conductivity measurement. Hold Function This function lets you lock the conductivity reading for viewing and recording. • Press the Hold key, the tester locks the measurement, the icon appears on the display. • Press

1.0 to 199.9 μS/cm

10 to 1999 μS/cm 0.1 to 19.99 mS/cm Resolution 0.1 μS/cm 1 μS/cm 0.1 mS/cm Accuracy ±1% F.S. Calibration Point 1 point Calibration Solution 146.5 μS/cm 1413 μS/cm 12.88 mS/cm Temperature Range 0 to 50°C Resolution 1°C Accuracy ±1°C Other Specification