Dynamic DX-SLM Installation Manual
Text of Dynamic DX-SLM Installation Manual
Servo Steering and Lighting Module
DX-SLM INSTALLATION MANUAL
GBK60027: Issue 5 - March 2014
DX-SLM
Servo Steering and Lighting
Module
THE
ULTIMATE
POWERCHAIR
CONTROL
SOLUTION
M
T
M
T
GBK60027 DX-SLM Installation Manual Issue 5
Contents
1
Introduction ............................................................. 3
2
Related Documentation ......................................... 4
3
General Description ............................................... 5
3.1
General Features ........................................................................ 6
3.2
Safety and Protection Features ............................................... 6
4
Speci
GBK60027 DX-SLM Installation Manual Issue 5 7.4
Calibrating the SLM using the HHP ........................................ 40
7.4.1 Initial Operation ............................................................. 40
7.4.2 Calibrating Parameters ................................................ 41
8
Diagnostics ............................................................ 43
8.1
Troubleshooting ..................................................................
GBK60027 DX-SLM Installation Manual Issue 5 1 Introduction
The Servo Steering and Lighting Module (SLM) controls the steering and lighting for
servo steered power wheelchairs. It is part of a DX System also comprising a DX Power
Module and a DX Remote as a minimum system.
Example of Minimum DX System with the SLM
The servo wheelchair has one motor for steering, and either one or two motors to
provide the driving power of the wheelchair. The DX Remote contr
GBK60027 DX-SLM Installation Manual Issue 5 2 Related Documentation
A DX based wheelchair control system may comprise between two and sixteen DX
compatible modules depending on the application. Each DX compatible module
has its own Installation Manual, which describes the installation requirements of that
particular module.
This manual describes the installation of the SLM only and must be read in conjunction
with the:
• DX System Manual
• Relevant DX
GBK60027 DX-SLM Installation Manual Issue 5 3 General Description
The SLM is designed to enable the DX System to run on a servo steered power
wheelchair, with lights. The wheelchair will have: a servo motor, a position sense
potentiometer, a steering release microswitch, one or two drive motors with integral
park brakes, a 24 V battery supply, and optional 24 V lights.
The DX system will comprise of a minimum of a DX Remote, a DX Power Module, a
SLM and ma
GBK60027 DX-SLM Installation Manual Issue 5 3.1 General Features
The SLM has the following general features:
• Servo output capable of supplying 5 A continuous and 30 A peak to a 24 V
servo actuator.
• Fully DX BUS compatible.
• Two identical DX BUS sockets for Daisy Chain connection to the DX System.
• Powered by a 24 V wheelchair battery via the DX BUS connection, or directly
from the battery for high current applications.
• Fully programmable
GBK60027 DX-SLM Installation Manual Issue 5 4 Specifications
4.1 Electrical Specifications
Power Supply Input
Symbol
Parameter
Min
Nom
Max
Units
VBAT
Battery voltage
18.0
24.0
32.0
V
IQ
Quiescent Current (No servo load)
150
mA
IO
Operating Current (+ servo load)
250
mA
IDXBUS
DX BUS Current (RMS Continuous to entire
system)
12
A
Servo Motor + / - Output
Symbol
Parameter
Min
Nom
Max
Units
SM+
Output Voltage (when not in current limit)
VBAT-1.2V
VBAT
V
IAV
C
GBK60027 DX-SLM Installation Manual Issue 5 Lighting Specification
Symbol
Parameter
Min
Nom
Max
Units
Turn Indicator Outputs (each output)
VTIO
Output Voltage
VBAT-1 VBAT-0.1 VBAT
V
ITIO
Continuous Output Current
2.3
2.5
A
Sidelight (Head / Tail Light) Output
VSLO
Output Voltage
VBAT-1 VBAT-0.1 VBAT
V
ISLO
Continuous Output Current
2.3
2.5
A
Page 8
M
T
GBK60027 DX-SLM Installation Manual Issue 5 4.2 Mechanical Specifications
Size:
210 * 123 * 38 mm
Weight:
0.530 Kg
Mounting:
M5 screws four places, or placed in tray
Case material:
Aluminium sheet, powder coat finish
Case sealing:
Tamper proof, IP54 if mounted as per mounting instructions
SLM Configuration
4.3 Environmental Specifications
Symbol Parameter
Min
Max
Units
Operating ambient temperature range
-25
50
°C
Storage temperature range
-25
70
°C
Oper
GBK60027 DX-SLM Installation Manual Issue 5 5 Installation
5.1 Introduction
Installing a SLM requires the following steps:
1.
Mounting the SLM
Refer Section 5.2
2.
SLM Connection with the DX system
Refer Section 5.3
3.
SLM 21 Way Connector
Refer Section 5.4
4.
SLM Servo Connector
Refer Section 5.5
5.
Programming the SLM related parameters
Refer Section 7.0
5.2 Mounting
Optimum mounting orientation
Fit the SLM with the top label facing up.
Unacceptable Mo
GBK60027 DX-SLM Installation Manual Issue 5 Securing the SLM
The SLM can be mounted using the four mounting holes and mounting screws
provided, or placed in a suitable tray. Do not use screws that protrude into the SLM
case by more than 12 mm.
The SLM must be mounted in a position that offers the maximum protection from
water and mechanical abuse. Since there are no user accessible controls on the SLM,
it can be mounted in a position which is inaccessible
GBK60027 DX-SLM Installation Manual Issue 5 5.3 SLM Connection with the DX System
5.3.1 DX BUS Connections
Note:
If only one DX BUS connector is used on the SLM and the remaining
connector is accessible to the wheelchair user, a GPL65009 DX BUS
Connector Cover should be fitted to the unused connector. This complies
with the ISO 7176-14 standard.
Like all other DX Modules, the SLM is connected to the DX system using the DX BUS.
Refer to the DX System Manual
GBK60027 DX-SLM Installation Manual Issue 5 5.3.2 DX Module Interconnection Topology Options
The battery and DX Power Module combination are always considered the heart of a
DX system. Other DX Modules can be arranged in several ways:
Star DX BUS Topology
In-line DX BUS Topology
Mixed DX BUS Topology
DX
Power
Module
DX
Module
DX
Module
DX
Module
DX
Module
DX
Module
DX
Module
DX
Module
DX
Module
DX
Module
DX
Module
DX
Remote
Battery
24V
D X
M o d u l e
D X
GBK60027 DX-SLM Installation Manual Issue 5 DX modules normally have one or two DX BUS sockets for system interconnections.
Smaller DX modules may have a permanently mounted cable terminated in a DX BUS
plug, rather than DX sockets.
The star and mixed topologies both require the use of one or more DX BUS Expanders
(DX-SKT-X3). The DX-SKT-X3 may be purchased from Dynamic Controls or a Dealer.
For lowest cost and simplicity the in-line topology is generally
GBK60027 DX-SLM Installation Manual Issue 5 Example:
Consider a Power Module connected to an SLM via five other DX Modules using 1
metre cables.
Example of DX Module interconnection
The total resistance of the 0 V return path, between the Power Module and SLM is:
6 X (2 * Rct + Rca) = 111 mOhms
This means that the maximum load that the SLM can drive and not exceed the1.0 V
drop requirement is 1 / 0.111 = 9 A.
If, for example, the servo motor and lighting
GBK60027 DX-SLM Installation Manual Issue 5 This favours topologies such as:
Rationalised In-line Topology
Two-way star Topology
Multi-star Topology
D X
P o w e r
M o d u l e
B a t t e r y
2 4 V
H i g h C u r r e n t M o d u l e s
N e a r e s t P o w e r M o d u l e
L o w C u r r e n t M o d u l e s
F u r t h e s t a w a y f r o m
Power Module
H
H
H
L
L
D X
P o w e r
M o d u l e
B a t t e r y
2 4 V
H
L
L
H
L
L
D X
P o w e r
M o d u l e
B a