Analog Tension Measuring Amplifier Version 2.03 04/08 st This operation manual is also available in German. Please contact your local representative. Diese Bedienungsanleitung ist auch in Deutsch erhältlich. Bitte kontaktieren Sie die Vertretung im zuständigen Land. © by FMS Force Measuring Systems AG, CH-8154 Oberglatt – All rights reserved.

Operating manual EMGZ308

2 1 Safety Instructions 1.1 Description conditions a) High danger of health injury or loss of life Danger This symbol refers to high risk for persons to get health injury or loss life. It has to be followed strictly. b) Risk of damage of machines Caution This symbol refers to informations, that, if ignored, could cause heavy mecanical damage. This warning has to be followed absolutely. c) Notice for proper function Notice This symb
Operating manual EMGZ308 3 Table of contents 1 Safety Instructions .................................................................................... 2 1.1 Description conditions 2 1.2 List of safety instructions 2 2 Definitions .................................................................................................. 4 3 System components................................................................................... 4 4 System description ..
Operating manual EMGZ308 4 2 Definitions Offset: Correction value for compensation of the zero point difference. Thanks to the offset, it is ensured that a force of 0N will generate a signal of 0V exactly. Gain: Amplification factor for the measuring signal. Use of proper value will set the measuring range of the sensor exactly corresponding to the signal output range (0...10V). Strain gauge: Electronic component that will change its resistance while its l
Operating manual EMGZ308 5 4 System description fig. 1: Basic structure of the EMGZ308 Tension Measuring Amplifier (picture shows single channel version) E308001e 4.1 Functional description The force sensors measure the tension force in the processed material and transmit the measuring value as a mV signal to the hybrid module in the measuring amplifier EMGZ308. The measuring amplifier amplifies the mV signal depending on configuration. The resulting feedb
Operating manual EMGZ308 6 Fixed setting of gain and offset Fixed setting is used when maximum thermal stability and vibration-proof conditions are needed. The signal is pre-amplified with a fixed value and has to be evaluated digitally (for ex. by a PLC), so that offset and gain can be calculated. 4.4 Block diagram fig. 2: The block diagram shows the single channel version. E308002e
Operating manual EMGZ308 7 5 Dimensions 8 x M16 x 1.5 EMGZ308 SERIES CHANNEL 1 CHANNEL 2 fig. 3 E308013e
Operating manual EMGZ308 8 6 Installation and wiring Caution Proper function of the Tension Measuring Amplifier is only guaranteed with the recommended application of the components. In case of other arrangement, heavy malfunction can be the result. Therefore, the installation instructions on the following pages must be followed strictly. Caution Local installation regulations are to preserve safety of electric equipment. They are not taken into considerat
Operating manual EMGZ308 9 6.2 Mounting the force sensors Mounting of the force sensors is done referring to the FMS Installation manual which is delivered together with the force sensors. Wiring to the terminals of the electronic unit is done according to wiring diagram (fig. 5). Notice Connecting the shield of the signal cable to the electronic unit and to the force sensor may cause ground circuits which may interfere the measuring signal massively. Malf
Operating manual EMGZ308 10 7 Setting into operation 7.1 View of the setting elements fig. 6: Setting elements E308008e 7.2 Configuring gain and offset Gain and offset are configured by jumpers. The settings can be made individually for each channel. Gain and offset with infinite setting If gain and offset are adjusted with the trimmers, the jumpers have to be set according to fig. 7. Gain and offset then may be adjusted using the trimmers (refer to „7.5
Operating manual EMGZ308 11 Gain and offset with fixed setting If the signal is processed by a control unit (i.e. a PLC), configuration with fixed gain and offset is recommended (offset = 0; gain = 1111). Therefore, the jumpers have to be set according to fig. 8. With this configuration, the good thermal characteristics of the hybrid module are best utilized. The output signal will be 20mA when the nominal force of the sensor is applied. 7.3 Configuring th
Operating manual EMGZ308 12 7.4 Configuring the lowpass filter The measuring amplifier provides 2 lowpass filters per channel. They are used to eliminate faulty signal variations which may be caused by unbalanced rollers, vibrations of the machine, or equivalent. Lowpass U acts only to the tension output Lowpass I acts only to the current output The lower the cut-off frequency, the more sluggish the signal will be. The configuration is made by soldering a
Operating manual EMGZ308 13 7.5 Calibrating the measuring amplifier Proceed the calibration for each channel as follows: • Connect gauge to the tension or current output. • Connect the first force sensor. • Check if a positive output signal is appearing when loading the sensor in measuring direction. If not, exchange the two signal wires of the referring force sensor in the terminal block (+Signal and –Signal). • If used, connect the second force
Operating manual EMGZ308 14 8 Trouble shooting Error Cause Corrective action Output shows signal > 0 although material has no tension Offset setting is bad Re-adjust offset (refer to „7.5 Calibrating the measuring amplifier“) Current output is configured for 4...20mA Change jumper setting if required Output signal is not stable although material tension doesn’t change Cut-off frequency too high Adjust cut-off frequency (refer to „7.4 Configuring th
Operating manual EMGZ308 15 9 Technical Data Sensor excitation 5VDC ±20ppm/K max. 30mA [±11ppm/°F] Gain factor range fixed: 1111; variable: 500...5000 Offset range fixed: 0mV; variable: ±9mV Temperature drift offset fixed: < 20ppm/K [11ppm/°F]; variable: ca. 50ppm/K [28ppm/°F] Linearity error < 0.1% Tension output ±10V min. 1kΩ Current output 0/4...20mA max. 500Ω Lowpass cut-off frequency adjustable, ca. 1...1000Hz Power supply 24VDC or 100...24
Operating manual EMGZ308 16 FMS Force Measuring Systems AG Aspstrasse 6 8154 Oberglatt (Switzerland) Tel. 0041 44 852 80 80 Fax 0041 44 850 60 06 [email protected] www.fms-technology.com FMS USA, Inc. 925 East Rand Road Suite 207 Arlington Heights, IL 60004 (USA) Tel. 847 392 7872 Fax 847 392 7873 [email protected] FMS (UK) Highfield, Atch Lench Road Church Lench Evesham WR11 4UG (Great Britain) Tel. 01386 871023 Fax 01386 871021 fmsuk@fms-te