ImpediMed SFB7 Use And Cleaning

File Specifications

Brand: ImpediMed
Category: Medical Equipment
Model: SFB7 (Use And Cleaning), prQGtjIDeEiPQ
File Type: PDF
File Size: —
Pages: 10 pages

Download PDF

Download ImpediMed SFB7 Manual (10 pages)

Read Online:

Read the ImpediMed SFB7 Use And Cleaning online using the built-in PDF viewer.

Key Pages of Document

Printed copies are not controlled.

Page 1 of 10 PC-POD-CP-010-v02 PERFORM Operating Document Use and Cleaning of ImpediMed SFB7 PC-POD-CP-010-v02 Revision History Version Reason for Revision Date 02 Created a more consolidated version May 7, 2020 1. Introduction The content of this PERFORM Operating Document (POD) provides guidelines for us ...

Printed copies are not controlled.

Page 2 of 10 PC-POD-CP-010-v02 2. Unit Components 3. Unit Calibration Please see user manual for calibration instructions. 4. Modes & Predetermined Measurement Settings This unit offers two modes of functions: bioimpedance spectroscopy (BIS), and selected frequencies (SFBI). a. BIS Mode Reactance vs. Resis ...

Printed copies are not controlled.

Page 3 of 10 PC-POD-CP-010-v02 This mode measures bioimpedance parameters over a frequency range of 4-1000 kHz with 256 data points defining the BIS mode as a true BIS. On-screen graphs display the measured data in the form a cole-cole plot. In addition, the characteristic frequency for the subject is det ...

Printed copies are not controlled.

Page 4 of 10 PC-POD-CP-010-v02 6. Preliminary Instructions Participant should respect the following instructions in order to minimize changes in the body’s hydration status (as this can over or underestimate values): a. Excessive exercise should be avoided 2 hours prior to measurement to avoid changes in ...

Printed copies are not controlled.

Page 5 of 10 PC-POD-CP-010-v02 NOTE: Measurements can be done on either the right or left side, but do not mix sides. Consistent measurement practices and electrode placement are important to obtain accurate results. h. Clean the sites with an alcohol swab. i. Allow sites to dry for 30 seconds before placi ...

Printed copies are not controlled.

Page 6 of 10 PC-POD-CP-010-v02 p. Predetermined measurement settings (e.g. number of measures to be taken and at what interval) may be set to the following, but can be modified at this stage: Measurement: interval Interval(s): 10 Number: 3 q. Click “Measure”. r. Click “Start”. Once measurements are ...

Printed copies are not controlled.

Page 7 of 10 PC-POD-CP-010-v02 5. Select your language. 6. Click the Set up the USB Ethernet adapter button. 7. This window will display all USB-Ethernet adapters that have been installed on your computer. Click the USB to fast Ethernet converter inside the network adapters found window and click the Next ...

Printed copies are not controlled.

Page 8 of 10 PC-POD-CP-010-v02 1. Patients with active implanted medical devices, e.g cardiac pacemakers, defibrillators or patients connected to electronic life support devices. 2. Patients undergoing external defibrillation. iii. Do not plug the unit leads to any mains power outlet/point. iv. Do not use ...

Printed copies are not controlled.

Page 9 of 10 PC-POD-CP-010-v02 3. Unnecessary removal of the leads from the unit may reduce lead life. 4. Ensure that the lid of the case does not close on the leads. 5. Do not wind the leads tightly, crinkle, or twist the leads, as this may cause the fine wires inside to break. 6. If the leads appear dama ...

Printed copies are not controlled.

Page 10 of 10 PC-POD-CP-010-v02 Internal software fault. The device stops responding to touch screen commands. Power light is blinking. Or Power light is red. Or Power light is orange. Normal behaviour. Refer to section "Battery Status Indicator." 1. Power the device off. Power the device on. 2. Report the ...

⚠️ Download Safety

The Medical Equipment Use and cleaning (10 pages) is completely safe to download (last scan date: 06/08/2026).

Related Devices for ImpediMed SFB7: