biodex ATOMLAB 500 Quick Reference Procedures
Text of the manual
FN: 13-305 11/13
AtomlAb 500
Quick RefeRence PRoceduRes
1.
Procedure to add Ra-223 to the isotope library and create a dial value setting
in the Atomlab 500
2.
Procedure to set a dial value for Ra-223 for the Atomlab 500
3.
Procedure to adjust an isotope dial setting in the Atomlab 500 Dose Calibrator
4.
Procedure to change one of the button isotopes to be
1. Procedure to add Ra-223 to the isotope library and create a dial value
setting in the Atomlab 500
Ra-223 is not listed in the current isotope library. The user will have to add it to the isotope list.
1.
Press the HOME key
2.
Press SELECT UTILITIES
3.
Select DOSE CALIBRATOR CONFIGURATION
4.
Select ISOTOPE LIST – you must now enter the password
5.
The I
3. Procedure to adjust an isotope dial setting in the
Atomlab 500 dose calibrator
1.
Press the HOME key
2.
Press SELECT UTILITIES
3.
Select DOSE CALIBRATOR CONFIGURATION
4.
Select ISOTOPE LIST – you must now enter the password
5.
The ISOTOPE LIST screen is now displayed
6.
Select CHANGE DIAL VALUES; the Change Dial Values Isotope List is now displayed
7.
U