biodex ATOMLAB 500 Quick Reference Procedures

Text of the manual

FN: 13-305 11/13

AtomlAb 500 Quick RefeRence PRoceduRes 1. Procedure to add Ra-223 to the isotope library and create a dial value setting in the Atomlab 500 2. Procedure to set a dial value for Ra-223 for the Atomlab 500 3. Procedure to adjust an isotope dial setting in the Atomlab 500 Dose Calibrator 4. Procedure to change one of the button isotopes to be

1. Procedure to add Ra-223 to the isotope library and create a dial value

setting in the Atomlab 500 Ra-223 is not listed in the current isotope library. The user will have to add it to the isotope list. 1. Press the HOME key 2. Press SELECT UTILITIES 3. Select DOSE CALIBRATOR CONFIGURATION 4. Select ISOTOPE LIST – you must now enter the password 5. The I

3. Procedure to adjust an isotope dial setting in the

Atomlab 500 dose calibrator 1. Press the HOME key 2. Press SELECT UTILITIES 3. Select DOSE CALIBRATOR CONFIGURATION 4. Select ISOTOPE LIST – you must now enter the password 5. The ISOTOPE LIST screen is now displayed 6. Select CHANGE DIAL VALUES; the Change Dial Values Isotope List is now displayed 7. U