Checkline CP-425-KIT Manual
Text of the manual
TABLE OF CONTENTS
1.0
General ……………………………………………..…………………
2
2.0
Test procedure ……………………………………………………..…
2
3.0
Possible reasons for error ………………………………………..
4
4.0
Storage …………………�
The pins of the CP-425-KIT have to be free from dust, dirt, oil and dampness.
The ideal test temperature was defined at 23°C. However, a test temperature
between 20°C and 26°C is sufficient. Air humidity has to be between 30% r.h. and
80% r.h.
The device as well as the CP-425-KIT has to be stored in the same surroundin
Now your device should show the value 22,0% (+/-1,0%), as shown in the picture
below.
If the measuring value is blinking in grey, the value is beyond the stated
specifications (please read „possible reasons for errors). If the measuring value is
shown in black, the value is o.k.
Put the black stick (CP-425-2) into the electrode hol
POSSIBLE REASONS FOR ERRORS
Discrepancy in temperature between TMT-425(USB) and the CP-425-KIT
Please ensure that the device and the test block are being stored at
nearly the same temperature before measuring. A high temperature difference has a
negative effect on the stability of the measurement results.
Contact pressure is too