REA VeriCube Operating Manual
Text of the manual
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 1 of 119
REA VeriCube
Verifier for Matrix Codes and
Barcodes
Operating Manual
Document Version 2.16
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 2 of 119
Credits
Version:
2.1.6 English
© 2020 REA Elektronik GmbH, all rights reserved
Any reproduction of these Operation Instructions either in part or as a whole, including
electronic processing, reproduction, or dissemination, without written permission from REA
Elektronik GmbH is prohibited.
Although this document has been prepared with utmost care and caution, there is always a
possibility that it may nonetheless contain erroneous figures and inaccuracies. REA Elektronik
is not liable for any damages which may arise from inaccuracies in the document.
REA Elektronik constantly strives to improve its products and therefore reserves the right to
make changes to the figures and data contained in this document on an ongoing basis
without prior notice.
REA Elektronik GmbH
Teichwiesenstraße 1
D-64367 Mühltal
Tel.: +49 (0)6154-638-0
Fax.: +49 (0)6154-638-195
E-Mail: [email protected]
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 3 of 119
Document history
Version
date
description
author
2.01
2018-02-09 Table with tolerances for calibration
S.Liebler
2.10
2018-02-16 Safety instructions, license conditions
S.Liebler
2.11
2018-05-11 Terms of guarantee, datasheet, accessories,
temporary IP address
S.Liebler
2.12
2019-05-13 Revision adjustment and calibration,
adjustment with Dnom=0.75, device
maintenance/cleaning, report radiation
emission
S.Liebler
2.13
2019-08-20 EU declaration of conformity, 11.10 View of
the reports, 12.1 Evaluation, Techn.data
camera modules
S.Liebler
2.14
2019-11-26 Screenshots for adjustment changed 9.1, new
icons for mirror live image and manual
brightness of live image (CD9.09)
S.Liebler
2.15
2020-08-16 Contact in 13, Technical data, system
requirements, part-no. calibration card, new
functions automatic measurement +
behaviour with screensaver from CD9.15
S.Liebler
2.16
2020-11-03 New calibration function from CD9.17
VeriCube-UV, IR added
S.Liebler
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 4 of 119
Table of contents
1
Introduction ............................................................................................... 7
2
Delivery scope ............................................................................................ 7
2.1
Basic configuration ................................................................................ 7
2.2
VeriCube variations ............................................................................... 8
2.3
Optional accessories .............................................................................. 8
2.4
Software options ................................................................................. 11
3
Transport, storage and operating conditions ................................................. 12
4
Safety Instructions .................................................................................... 13
4.1
General safety .................................................................................... 13
4.2
Additional safety instructions for REA VeriCube UV .................................. 15
4.3
Electrical hazards ................................................................................ 16
5
REA VeriCube Version ................................................................................ 17
5.1
REA VeriCube Standard ........................................................................ 17
5.2
REA VeriCube Pharma .......................................................................... 17
5.3
REA VeriCube UV ................................................................................ 18
5.4
REA VeriCube IR ................................................................................. 20
6
Commissioning ......................................................................................... 20
6.1
Power supply ...................................................................................... 20
6.2
System requirements .......................................................................... 21
6.3
Temperature change ........................................................................... 21
6.4
Connecting ......................................................................................... 22
6.4.1
Own Ethernet port ......................................................................... 23
6.4.2
Network operation ......................................................................... 24
6.4.3
PoE in switch ................................................................................. 24
6.4.4
MAC addresses .............................................................................. 24
6.4.5
Firewall and network type ............................................................... 24
6.4.6
Firewall and Network ports ............................................................. 25
6.4.7
Using a gateway ............................................................................ 25
6.4.8
Fixed IP Addresses ......................................................................... 25
6.5
Software ............................................................................................ 26
6.5.1
Software installation ...................................................................... 26
6.5.2
Software and device options ........................................................... 28
6.6
Mounting and exchanging the camera lens module .................................. 34
7
Correct handling ....................................................................................... 35
7.1
General considerations ......................................................................... 35
7.2
Checklist for measuring........................................................................ 36
8
Operating the device ................................................................................. 38
9
Adjustment and Calibration of the REA VeriCube ........................................... 40
9.1
Adjustment of REA VeriCube ................................................................. 42
9.1.1
Level correction ............................................................................. 47
9.1.2
Reflectance adjustment .................................................................. 50
9.1.3
Scale adjustment ........................................................................... 51
9.2
REA VeriCube calibration ...................................................................... 52
9.2.1
Admissible device tolerances for calibration with red light ................... 52
9.2.2
Calibration for white light ............................................................... 54
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 5 of 119
9.2.3
Calibration with REA calibration card ................................................ 55
9.2.4
Calibration with extended calibration cards ....................................... 60
10
Synthetic aperture / measuring opening ................................................... 62
11
Execution of the measurements ............................................................... 63
11.1
Positioning of the device .................................................................... 63
11.1.1
Camera downwards ..................................................................... 63
11.1.2
Camera upwards ......................................................................... 64
11.1.3
Camera sideways ........................................................................ 65
11.1.4
Turning the keyboard .................................................................. 66
11.1.5
Testing 3D parts ......................................................................... 66
11.2
Measuring process using the Windows software. ................................... 67
11.3
Automatic measurement mode ........................................................... 69
11.4
Automatic ROI ................................................................................. 70
11.5
Mirror live image .............................................................................. 70
11.6
Setting the brightness of the live image .............................................. 71
11.7
Recording the image ......................................................................... 71
11.8
Opening the image file ...................................................................... 71
11.9
Saving the image file ........................................................................ 72
11.10
Static reflection measurement ............................................................ 72
11.11
Preset the measurement image size ................................................... 73
11.12
Rotating the image ........................................................................... 73
11.13
Behaviour of the software TransWin32 when locking the PC ................... 74
11.14
View of the Reports .......................................................................... 74
11.15
Details in the test reports .................................................................. 78
12
Configuring the evaluation and symbologies .............................................. 80
12.1
Evaluation ....................................................................................... 81
12.2
Symbology selection ......................................................................... 82
12.3
Lighting and aperture ........................................................................ 84
12.4
Profile management .......................................................................... 86
12.5
File Saving Options ........................................................................... 87
12.6
Device data ..................................................................................... 89
12.7
User management ............................................................................ 90
12.7.1
Simple user name ....................................................................... 90
12.7.2
Simple password ........................................................................ 90
12.7.3
Setting up the user management .................................................. 91
12.7.4
User management during operation .............................................. 94
12.7.5
Transwin32 configuration protection .............................................. 96
13
Trouble shooting .................................................................................... 97
14
Maintenance and cleaning ..................................................................... 102
15
Intellectual property reserve ................................................................. 102
16
Software update .................................................................................. 103
17
Protected trademarks ........................................................................... 104
18
License conditions ................................................................................ 105
18.1
License Agreement ......................................................................... 106
18.2
Limited support and fault / error rectifications .................................... 106
18.3
Export limitations ........................................................................... 107
18.4
Maintenance .................................................................................. 107
18.5
Liability ......................................................................................... 107
19
Certificate of conformity for REA VeriCube............................................... 108
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 6 of 119
20
Certificate of conformity for the REA calibration cards ............................... 109
21
EU declaration of conformity for the REA VeriCube ................................... 110
22
Technical data ..................................................................................... 111
23
Measurement Report radiation emission .................................................. 114
24
Terms of guarantee .............................................................................. 114
25
Attachment ......................................................................................... 116
25.1
General symbols used in this manual ................................................ 116
25.2
Abbreviations used ......................................................................... 117
25.3
Handling ....................................................................................... 117
25.4
Manager’s Obligations ..................................................................... 118
25.5
Obligation of the user ..................................................................... 118
25.6
Intended use ................................................................................. 118
25.7
Recycling ....................................................................................... 119
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 7 of 119
1
Introduction
The REA VeriCube is a measurement device for the quality control of printed 2D
matrix codes and 1D barcodes. This is a stationary device, which has to be connected
to a PC via an Ethernet network. The application software along with the device
driver and the display of the test reports run on a computer running Microsoft
Windows.
In contrast to a scanner for barcodes and matrix codes, REA VeriCube is an optically
operating measurement device. Measurement devices are subject to a regular
adjustment and calibration (monitoring of testing equipment). Adjustment and
calibration (monitoring of testing equipment) can be performed by the operator and
as long as calibration is successful, there is no need to return the unit to the
manufacturer.
An overview of all Verifier variants and brochures can be found on our website:
https://www.rea-verifier.com
2
Delivery scope
The REA VeriCube in its basic configuration is delivered in cardboard packaging and
fitting foam padding with the following parts:
2.1
Basic configuration
1 REA VeriCube, test unit with operating unit as per delivery note (see following
chapter for available types)
1 optic module, exchangeable, according to the delivery note (built into the unit)
1 Plexiglas dust-protection cover – built-in in transport (must be replaced with a
glass plate during commissioning)
1 anti-reflective glass plate (08316280) (separately packaged)
1 GigaBit PoE power supply unit (100–240V, 30W) (06113003)
1 power cable for the ordered plug standard (Euro, UK, US)
2 network cables, highly flexible, RJ45 CAT7 (08316608)
1 printed operating manual (this document)
1 USB stick with software and documentation (08316300)
1 calibration card, serialized, in a protective envelope (90031702)
1 test report of the reference code of the calibration card, created with the delivered
VeriCube for documenting the delivery status
1 test report of the reference code (req. for test equipment monitoring)
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 8 of 119
1 white level correction calibration card (protected in new condition with blue
protection foil) in a protective envelope (90031703) (can be cleaned with 3M Novec
71IPA)
1 field of view cover as a black PVC plate (07601403)
1 field of view cover as a black cloth (07799084)
2.2
VeriCube variations
VeriCube
Type
description
Artikel-Nr.
Standard
with threaded mounting holes for mounting customer-
specific devices or stops
030.035.000
Pharma
as standard, but with closed plate without holes
030.035.001
Inline
as standard, but with PLC interface with 24V digital I/O 030.035.004
Diffuse
Standard
as standard, but with diffuse lighting
030.035.050
Diffuse
Pharma
as pharma, but with diffuse lighting
030.035.051
DPM
as pharma, but with dome lighting
030.035.061
UV
as pharma, but with lighting 660/365nm
030.035.011
IR 945 nm
as pharma, but with lighting 660/945nm
030.035.141
IR 855 nm
as pharma, but with lighting 660/855nm
030.035.131
2.3
Optional accessories
Optional parts are only included if they have been ordered and are listed separately
on the delivery note. License keys are not necessary. In devices that are already in
operation, optional functions can be activated later by means of license keys.
Camera modules
There is a variety of different camera modules available. Depending on the range of
codesizes and module sizes that have to be measured, the appropriate camera
module must be selected. For UV and IR, a camera module set up for this purpose
must be used, because with this lighting option the focus position shifts relative to
that for red and white light.
At least one camera module must be specified when ordering.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 9 of 119
Camera modules for red and white lighting:
Measuring plane 0mm (on the glass pane):
030.037.100 Camera module for matrix codes typical>= 0.44 mm
(focal length 8 mm, field of view 114 x 71 mm)
030.037.101 Camera module for matrix codes typical >= 0.31 mm
(focal length 12 mm, field of view 80 x 60 mm)
030.037.102 Camera module for matrix codes typical >= 0.25 mm
(focal length 16 mm, field of view 64 x 47 mm)
030.037.103 Camera module for matrix codes typical >= 0.14 mm
(focal length 25 mm, field of view 37 x 27 mm)
030.037.104 Camera module for matrix codes typical >= 0.042 mm
(focal length 50 mm, field of view 9 x 6 mm)
Measuring plane +15mm (Inline application or with stand):
030.037.112 Camera module for matrix codes typical >= 0,27 mm
(focal length 16 mm, field of view 68 x 51 mm)
030.037.113 Camera module for matrix codes typical >= 0,16 mm
(focal length 25 mm, field of view 40 x 30 mm)
030.037.114 Camera module for matrix codes typical >= 0,05 mm
(focal length 50 mm, field of view 10 x 7 mm)
Measuring plane +45mm (Inline application or with stand):
030.037.123 Camera module for matrix codes typical >= 0,2 mm
(focal length 25 mm, field of view 47 x 25 mm)
Camera module for UV lighting:
Measuring plane 0mm (on the glass pane):
030.037.202 Camera module for UV - matrix codes typical >= 0,27 mm
(focal length 16 mm, field of view 68 x 51 mm)
Camera module for IR lighting:
Measuring plane 0mm (on the glass pane):
030.037.301 Camera module for IR 855nm - matrix codes typical >= 0.31 mm
(focal length 12 mm, field of view 80 x 60 mm)
030.037.302 Camera module for IR 855nm - matrix codes typical >= 0.25 mm
(focal length 16 mm, field of view 64 x 47 mm)
030.037.401 Camera module for IR 945nm - matrix codes typical >= 0.31 mm
(focal length 12 mm, field of view 80 x 60 mm)
030.037.402 Camera module for IR 945nm - matrix codes typical >= 0.25 mm
(focal length 16 mm, field of view 64 x 47 mm)
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 10 of 119
98031601 REA VeriCube compact case
09930162 REA VeriCube High quality aluminium case incl. packaging
Optional test cards for extended monitoring of the testing equipment
90031705 (replaces 98320009) Calibrated test card with EAN/UPC codes (AI)
90031704 (replaces 98320008) Calibrated test card Data Matrix Codes (AI)
90031706 Calibration card GS1-128 (AI)
90031707 Kalibrierkarte GS1-Databar EAN/UPC (AI)
08316223 Focal distance tool 15mm & 45mm
99031621 Spacer 15mm (only required for 15mm distance between optics, please
order separately)
98031600 30+45mm spacer set Accessories for spacer 99031621
99031640 Adjustable positioning stop VeriCube
08316222 Protective Cover VeriCube
030.037.200 Measuring table – one measuring table as a table for parts / cans which
can neither be left on the REA VeriCube for measuring nor placed on the REA
VeriCube.
99031620 Stand – REA VeriCube can be mounted into our height-adjustable stand
and used with the camera modules with focus position +15mm. This combination is
mainly used for checking codes on components/products (circuit boards, brake discs,
hoses, surgical instruments, etc.
08250330 USB Gigabit Ethernet Adapter
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 11 of 119
2.4
Software options
030.207.101 Optional symbologies (internal software option)
08300920
ScanLink – testing the structured code content (additional installation)
030.207.160 Comparator function (internal software option)
030.207.170 Article database32 (additional installation)
030.035.180 Code analysis (internal software option) Software extension for client-
specific analysis of data structures using a definition table
030.035.181 DPM -ISO/IEC TR 29158 (AIM DPM Guideline 2006) (internal software
option) Measurement evaluation for direct part marking, limited to
45° illumination
030.035.183 Audit Trail software option (new installation necessary)
When ordering internal software options, these are activated by entering an
activation code. If a software option is purchased subsequently, it is supplied with an
activation key only, which must be entered in TranWin32.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 12 of 119
3
Transport, storage and operating conditions
The REA VeriCube can only be transported in the supplied
case or transport packaging.
Transport: The REA VeriCube is delivered in a case or in a cardboard
box with adequate foam padding. This package should be preserved
for possible transportation. The REA VeriCube should only be
transported in its original packaging, especially with parcel services
and similar means. In case, it is shipped without the original
packaging though, it should be packaged with good padding, so that
the device cannot move in the cardboard shipment box.
The glass pane must always be removed for transport and the plastic
pane must then be used as a protection. The glass cover must always be removed
prior to the transportation and then the plastic cover must be used as protection.
After transportation, the device must be warmed up to room temperature because
the optical parts can be fogged by the temperature change.
Storage: The device should always be stored in dry, dust-free rooms and at room
temperature. When not in use, the device should be covered or stored in the
transport packaging to protect it from dust deposits. For storage, remove the glass
pane and insert the plastic pane.
Dust on the optically effective surfaces leads to misalignment of the optical
measuring paths. This leads to incorrect measurement results.
Operation: Humidity, cold, heat and mechanical impacts incurred when the
equipment is dropped, can lead to a situation where the measuring chamber cover as
well as the mechanical and electronic parts can be damaged or destroyed.
If the device still provides measurement results with these damages, there is a high
risk that these results are incorrect or deviate too far from the correct values.
The device is delivered with a plastic cover and a top plate, which protect the
measuring chamber. The plastic cover serves as dust protection during the storage
and transportation of the REA VeriCube. The glass plate serves as a top plate in the
event that the codes to be measured need to be placed on the surface. The glass top
plate must always be removed prior to the transportation and the plastic cover used
as protection.
ATTENTION
N
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 13 of 119
4
Safety Instructions
4.1
General safety
Before setting up the REA product and using it, please read and strictly observe
these operating instructions very carefully, especially the safety instructions.
Supplementary notes can be found in Annex 24. Symbols are described in 24.1.
The REA Verifier tester REA VeriCube is built according to state of the art and
recognized safety regulations. Nevertheless, its use risks to life and limb of the user
or others, or damage to the REA VeriCube or other property may arise.
The REA VeriCube is
to use in accordance with the intended use.
to put into operation in absolutely perfect condition.
Disorders that may affect safety must be rectified immediately.
(See chapter "Intended Use, Chapter 24.6").
In case of breakdown, the REA VeriCube must shut down immediately and must only
be taken back into operation after proper repair!
Warning:
Commissioning, operation and any subsequent applicable
service may only be performed by trained personnel.
Warning:
Before maintenance work or other interventions, it is
obligatory to turn off the verifier and unplug the power
cord. Otherwise there is a risk of damage on the device or
even personal injury because of electrical voltage.
Warning:
Don’t cover the power supply (PoE injector). Risk of
overheating and danger of fire.
Warning LED lighting:
In the REA VeriCube, LEDs with red light 660nm and white
light 4000K are used. It is essential to avoid looking directly
into the LEDs, otherwise the eyes might be damaged.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 14 of 119
Warning:
Be careful with the glass cover. When the VeriCube is tilted,
it could slip out and fall down. Caution, danger of injury.
Warning:
Be careful with the glass cover. After a glass breakage, be
careful with possible glass splinters, danger of cutting.
Warning:
Caution when handling the VeriCube: Cutting danger of injury at
the corners and edges of the device.
Cleaning
If necessary, clean the housing with a damp (not wet) cloth.
Clean glass cover with a damp cloth (e.g. microfiber as attached with
shipment).
If necessary, clean the lense of the camera module and the mirror only
with special auxiliary means for optical lenses.
Attention: cleaning with solvents will destroy the device
Dust may only be removed with compressed air, if it is completely free of
oils and other contaminants. In dusty environments, it is recommended
to protect the device and cover it if possible.
Electrostatic sensitive parts
Electrostatic discharges should be avoided as they can destroy the device.
Functional test
It is recommended to do a monthly review (calibration) of the accuracy of
the verifier by measuring the supplied test code.(See Section 11 Calibration)
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 15 of 119
4.2
Additional safety instructions for REA VeriCube UV
In addition to the safety instructions in chapter 4.1General safety the following must
be observed when operating the REA VeriCube UV:
Attention! The device works with UV radiation. It must only be used with
appropriate protective measures in accordance with EU Directive 2006/25/EC
In countries outside the EU, it is obligatory to comply with the occupational
safety regulations that are in force in that country.
The warning sign "UV radiation”
must not be removed from the device or concealed or made
unrecognisable in any other way.
UV protective glasses are supplied with the unit. It is recommended that these
glasses be put on before the installation or commissioning process and before each
power-up.
The unit switches on the UV-LED's during continuous operation. The unit should
therefore only be installed in a room where it is not possible to enter without taking
protective measures (wearing UV glasses).
With REA VeriCube UV the irradiation values according to EN 62471:2008 for the skin
directly on the device are exceeded. For longer-term work at the radiation source it is
therefore recommended to wear opaque clothing and gloves. During operation,
please always cover the field of view with the black plate or black cover cloth after
the object has been positioned.
The use of the device is the sole responsibility of the user. The
manufacturer REA Elektronik GmbH is not liable for any damages
caused by the use of the device. This exclusion of liability is only limited
by legal regulations that apply to contracts or transactions between
companies (BtB).
The manufacturer REA Elektronik GmbH has designed this product, REA
VeriCube with UV LED's 365nm, exclusively for commercial use with
appropriate protective devices. Use in private areas and / or without
protective equipment is NOT intended use. The liability for damage
resulting from such improper use, lies exclusively with the user.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 16 of 119
4.3
Electrical hazards
Warning:
When mains plug is unplugged or a main switch is switched off
during cleaning, maintenance or repair work, appropriate measures
must be taken to prevent unauthorized re-establishment of the mains
connection. This applies to the power supply. The verifier itself falls
under the category of Low Voltage.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 17 of 119
5
REA VeriCube Version
5.1
REA VeriCube Standard
The REA VeriCube, in the standard design, has 13 screws on the upper side. With the
exception of any two screws, all the others can be removed.
The plate is either in blue or silver in color. The screws have Torx T10 heads.
The screws are there to close the threaded holes. The threaded holes are provided
for fastening the used fixtures when, for example, auxiliaries and other measurement
aids must be affixed to the device. The advantage of this is that frequently made
measurements of identical products can be performed by the device without the
positioning aided by the live image function.
5.2
REA VeriCube Pharma
The REA VeriCube in the Pharma design has a closed plate at the top without screw
holes.
This design is intended for use in areas in which it must be possible to clean the
surface easily and quickly.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 18 of 119
5.3
REA VeriCube UV
The REA VeriCube UV has an ultraviolet illumination as an alternative to the red light
illumination with 660 nm. For this alternative illumination there are arranged 16 UV-
LEDs with a wavelength of 365 nm in the unit.
This applies to VeriCube devices with UV illumination
Devices can be identified by a warning sign:
Please read this manual completely before you start using the device!
Please pay particular attention to the safety instructions for the REA
VeriCube UV in chapter 4.1General safety + 4.2Additional safety
instructions for REA VeriCube UV
With REA VeriCube UV the irradiation values according to EN 62471:2008 for the skin
directly on the device are exceeded. For longer-term work at the radiation source it is
therefore recommended to wear opaque clothing and gloves. During operation,
please always cover the field of view with the black plate or black cover cloth after
the object has been positioned.
The Level correction when adjusting this type of test instrument differs from the
standard procedure. The greyscale card must be used here instead of the luminance
card. Please refer to the note in chapter Level adjustment procedure for this type of
verifier.
Attention should be paid to the fact, that upon UV illumination, dirt will get visible,
that was invisible with natural light before, like for example thin layers of fat,
fingerprints etc. Extremely precise attention to cleanliness is a prerequisite
for reproducible measurements.
It is recommended to always wear gloves (e.g. Monochrom article no.: 5241L) when
handling the device in order to avoid contamination of the verifier, the calibration /
adjustment cards and the measuring objects.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 19 of 119
For the reflectance adjustment the blue component of the UV-LED light is used,
which is within the sensitivity range of the camera. An absolute reference to national
standards is not established in this case, but only a reference to the reflectance
values measured with red light as a reference. This adjustment is used to obtain
repeatable, stable measurement results that refer to a defined standard.
The typical measurement objects are markings with dyes / inks / printing colors that
are invisible or completely transparent in visible light. In UV light, these markings
generate fluorescence in the visible range of the light spectrum. This light, which is
generated by fluorescence is evaluated as if it were a reflected light from the surface.
Due to the adjustment, described above, the UV light source has an always identical
intensity. Thus, the fluorescence is evaluated with stable lighting values. The
evaluation must be in accordance with ISO/IEC TR 29158 (DPM).
Attention:
The focusing plane of the camera module in UV is shifted in relation to
that in red light. If the REA VeriCube UV is to be used for measurements
with red light, it is essential that a suitable camera module is then used
for red light.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 20 of 119
5.4
REA VeriCube IR
The REA VeriCube IR has an infrared illumination as an alternative to the red light
illumination with 660 nm.
REA has two different versions for infrared lighting in its portfolio.
The REA VeriCube IR can be purchased as a variant with 855 nm or with 945 nm.
With these types of device, IR is always the alternative lighting to red light. For this
alternative lighting, the device contains 16 IR LEDs with a wavelength of 855 nm or
945 nm.
Handling, adjustment and calibration are the same with the REA VeriCube IR as with
the standard device. However, it should be noted that due to unavoidable reflections
through the glass pane, a warning appears during Level correction: "Level correction
finished with a warning. Warning: the calibration card seems dirty. Please clean it.
However, on this occasion the adjustment result has been accepted. [0x20081285].“
This warning can be noted in this case, but work can still continue normally.
The reflections can only be prevented if both the adjustment and the measurements
afterwards are performed without the glass pane. This is also permitted. However, it
must be taken into account that the mirror in the device and the camera lens will
then become contaminated by dust more quickly and therefore have to be cleaned
regularly. As long as the device is not in operation it should be protected against
dust.
Attention:
The focusing plane of the camera module with IR lighting is shifted in
relation to that in red light. If the REA VeriCube IR is to be used for
measurements with red light, it is essential that a suitable camera module
is then used for red light.
6
Commissioning
6.1
Power supply
The power for the VeriCube is supplied via a network cable connected to the device.
This type of power supply is designated as Power over Ethernet (PoE) (power supply
over a network cable). The delivered power supply unit provides power to a
connected device only when the device is designed for a power over Ethernet supply.
It is not recommended to connect a non PoE-enabled power supply device to the OUT
connection of the power supply unit. A second cable or additional separate power
supply unit is not necessary. The delivered network cables are, apart from the
increased flexibility, standard off-the-shelf network cables.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 21 of 119
6.2
System requirements
For the device power supply a mains connection in the range of 100 - 240 V AC,
between 50 and 60 Hertz is necessary.
The Windows software TransWin32 needs a standard off-the-shelf computer with a
Microsoft Windows operating system,
-
Windows 10 Pro/Home, 64-bit.
-
Microsoft .net framework from version 4.5 or newer
-
One or optionally two Ethernet network interfaces
(2 Ethernet network interfaces are recommended
(at least 1Gbit/s, Auto detect between normal and Crossover)
-
Main memory: The minimum requirements of the used operating system must
be respected. It is usually 4GB or more
-
Hard disk: The minimum requirements of the operating system must be
respected
-
CPU: The minimum requirements of the operating system must be respected
(a slow CPU increases the cycle time).
-
Mouse (with mouse wheel) / keyboard (supplementary touchscreen, if
necessary)
-
Screen / graphics: Non-glare, minimum 48cm (19“),
at least 1280 x 1024 pixels in resolution, color
-
Optionally: PDF program for the creation of test reports as PDF files (as printer
simulation)
-
PDF display program
-
Firewall. The firewall must permit the network operation of the REA VeriCube
The REA software TransWin32 is a single user software. The license is connected to
the REA Verifier and not to the PC installation. The software does not need a dongle
and can, if needed, be installed multiple times.
Operation on a server is not a tested and verified installation by the manufacturer. As
long as the server offers the REA program TransWin32 an environment, which reacts
as a single workplace for the program, the server installation should work properly
(virtual environment, sandbox, etc.).
An installation under Windows XP does not function and is not supported.
The minimum computer requirements are sufficient to run the system,
however higher specifictions like the recommended configuration of the
appropriate operating system in use, should be preferred.
6.3
Temperature change
When the REA VeriCube is subjected to temperature changes (e.g. when it is utilized
in heated rooms after transport in winter), condensation can occur on the optical
parts. In such an instant, wait for about 15 min before the REA VeriCube can be
used.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 22 of 119
6.4
Connecting
The REA VeriCube is delivered with two identical network cables.
Picture of PoE power supply and network cables
The REA VeriCube is connected with one of the two Ethernet cables to the PoE power
supply unit. The socket, which is marked on the power supply unit with "Data&Power
Out”, is intended for the connection to the device. When the other "Data In" socket is
used, the device does not receive power and will not power up.
The "Data/In" socket on the PoE power supply unit is connected with the second
network cable and with the computer (directly or indirectly).
The picture shows an example of the connection of the Data-In socket to the network
installation (here shown by the Ethernet switch).
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 23 of 119
The power supply unit has an IEC socket for connecting to the electricity supply. The
mains cable is delivered adapted to the specified country standard according to the
part number supplied.
The socket type is C14 (IEC-60320). The plug type is C13.
Picture of a socket, low-power mains plug type
The power supply through the PoE is characterized by the fact that the power supply
unit first checks whether a connected device is designed for a PoE power supply. In
this test period, the device cannot yet be turned on.
As long as the testing takes place, no LED lights up on the Ethernet socket of the
REA VeriCube. As soon as the device receives power, the left LED lights up (light-
orange)
Picture of LED OFF
Picture of LED ON
ATTENTION: REA has a complete range of code verifiers in its product line.
In all products, the power supply utilises PoE power. The REA VeriCube
requires 30W type POI-3000 (or POI-3002 or POI-3004). The other devices
use the 15W type POI-2002. When these power supply units are
interchanged, the REA VeriCube gives the appearance of functioning.
However, the measurement and the lighting do not function correctly.
Therefore, the power supply units should not be interchanged.
6.4.1
Own Ethernet port
The PC has one or two network connections, one of which is used exclusively for the
operation of the VeriCube. The "Data/IN" socket of the power supply unit is
connected directly with a cable to this socket.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 24 of 119
6.4.2
Network operation
The PC is connected to a company network LAN. The REA VeriCube is connected from
the "Data/IN" connection with an available and connected (patched) network socket.
The server (DHCP) in the network automatically assigns the network addresses.
When the company network is subjected to heavy network traffic, disruptions (the
device shows network errors) may occur during operation. When these disruptions
become too frequent, the network must be isolated by the network administrator or
alternatively the REA VeriCube needs to be connected to the computer with a
separate network connection (see the previous section).
6.4.3
PoE in switch
Operation using an Ethernet switch, which supplies PoE power (instead of using the
supplied PoE power supply) is possible. A connection, which can provide 30W power,
is required. This mode of operation is the responsibility of the User and cannot be
certified by the manufacturer. The PoE power supply is an industry standard and is,
therefore, usually non-critical as long as a PoE switch is correctly specified and is not
overloaded by too many devices.
6.4.4
MAC addresses
Every network-enabled device receives a MAC address from the manufacturer. The
MAC address is provided on the type label. Some network environments are
configured in such a way that only known devices are allowable. In this case, please
inform your IT department of the unit’s MAC address.
6.4.5
Firewall and network type
When a firewall is active, it must be configured in such a way that the network
operation between the PC and the REA VeriCube can pass safely through the firewall.
During initial installation, the security rules are usually required.
When the operating system recognizes a new network, it is asked whether this is a
public network. This must be answered “NO” because Public networks are WLAN (at a
railway station, airport, etc.) and not applicable for this device.
When there are particularly strict security rules in a company or if the network is
heavily loaded with data traffic, REA recommends using a second Ethernet port on
the PC which is dedicated to the REA VeriCube.
Supplied with the delivered unit, there is a document which describes the network
integration in detail and also designates port details.
In difficult installations, please refer to this document.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 25 of 119
6.4.6
Firewall and Network ports
The installation contains several modules (see chapter 6.5). Communication between
modules is performed via network components (Sockets, Ports). Ports that are
utilized are listed in the following table. The Firewall must open the ports and let the
program module communication pass through the Firewall.
Transwin32.exe
Port 127.0.0.1:23945
ServiceTool.exe
Port 127.0.0.1:21947
MorpheusHost.exe
Port 127.0.0.1:21946
MLV_Control_PC.exe
Port 127.0.0.1:10061
VeriCube_Control_PC.exe
Port 127.0.0.1:10051
Morpheus Broadcast
PC-IP/device-IP: 10049
readb.exe (Article database) n/a
If any of the program modules or ports are blocked then the application will not work
or show odd behavior.
6.4.7
Using a gateway
The REA VeriCube is using the camera protocol GigE Vision. This is primarily specified
for a direct communication between two devices (PC and Camera). Operation over a
Gateway is not recognized and does not work.
6.4.8
Fixed IP Addresses
In some network environments fixed IP Addresses are demanded or required. This is
usually solved by a manual configuration of the network settings of the PC and the
VeriCube. A manual network configuration restricts the flexibility of use of the
verifier. Therefore the use of a DHCP server is always recommended. The DHCP
server manages devices which are identified by their MAC addreses. For each device
requiring a fixed IP address the server is configured to provide such devices always
and automatically the same (fixed) IP address.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 26 of 119
6.5
Software
The REA TranWin32 software is necessary for the operation of the VeriCube. Only
one REA VeriCube can be used by a PC at any one time. Other REA Verifier products
can be used on this PC at the same time but not a second REA VeriCube.
6.5.1
Software installation
When the system prerequisites are met (see chapter 6.2
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 27 of 119
System requirements), the TransWin32 program can be installed. For this purpose,
the "Setup.exe" is launched. Depending on the version, the TransWin32 installation is
offered directly or the TransWin32 button comes first and then "Install TransWin32".
Administrator rights are required for the installation to be completed successfully. For
normal operation, the normal user rights are sufficient. When the software for all
users is installed, the "All users" documents folder is used. This general application
folder must be available to all users without administrator rights.
It is strongly recommended that all users have full access to the set directory for
document storage. Restriction to read and write access may result in restrictions to
some functions.
The most secure and trouble free installation allows the user account for the
installation to be an administrator account. In this instance the installation is
performed only for this user and therefore only the Documents folder of the logged-
in user is used. When the installation is carried out with "Execute as administrator",
the installation is performed with the administrator account and not with the user
account. In this case, it should be installed for all users and it should be ensured that
all users have read and write permission to the "All users" document folder.
In case of communication trouble (disconnects, error messages like Sensor errors)
two basic solutions are available:
1. Within the TransWin32 menu “Tool” the installation of the Vericube Network
driver is performed
2. A second Ethernet port is added to the PC and the REA VeriCube is exclusively
connected to this port.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 28 of 119
6.5.2
Software and device options
Select the desired language for the installation process
The next step is to agree to the license terms.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 29 of 119
The following dialog offers a selection of the components to be installed
TransWin32
The main program for displaying and managing the test reports
Check ER
The REA Verifier model REA Check ER is used
PC Scan LD3
The REA Verifier model REA PC-Scan/LD3 is used
ScanCheck 3n The REA Verifier model REA ScanCheck 3n is used
MLV-2D
The REA Verifier model REA MLV-2D is used
VeriCube
The REA Verifier model REA VeriCube is used
The installation option ADB Client installs the REA article database software. The
option ScanLink Assistant is an option for installing and using the ScanLink software
for extended analysis and evaluation of ISO standardized data structures.
These two software options should be installed only when these software options are
purchased with the REA Verifier.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 30 of 119
Selecting the installation type gives you several installation options, depending on
the desired type of installation.
a. "install the programs for all users", all users of the computer can use the
software. The configuration and adjustment data are stored by default in the
directory C:\ProgramData. The users need read and write access to this
directory.
i.
With activation of “Restore the backup of old configurations”, the
following data will be restored for REA VeriCube:
1) TW32 settings (all items in menu >configure >options
2) Printer settings (the printer that was used, settings of the printer)
3) The verifiers that were connected in last session.
4) Serial number of measurements on one day
5) All settings of VeriCube, profiles, adjustment data
b. “install the programs individually for each user”, only the respective user can
use the software. The configuration and adjustment data are stored in the
directory of the respective user under %AppData%. This means that each user
then has his or her individual adjustment and configuration data.
c. “Install the programs as Audit Trail System” (this is a chargeable option)":
with this option all relevant actions are recorded and the data is stored in a
secure area, the DataServer. This installation requires that the license is
activated in VeriCube or that a license key is ordered from REA. Please also
refer to the separate Audit Trail manual.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 31 of 119
Next, installation paths are displayed and can be modified, if desired.
The programs are installed in the Programs (or Program Files) folder.
The path C:\ProgramData is the file system for installation and programrelevant
data, which normally remain invisible for the user. This path should be changed only
when there are specific administrative reasons for that. The actual user data folder is
always in the Documents section of the user or in the one for all users. The folder for
saving the reports can be changed in the saving settings in TransWin32.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 32 of 119
At the end, the installation can be tested again.
If at this point the installation is cancelled, this means that the old version has
already been removed and the new one has not yet been installed.
After clicking Install, the software is installed and the process is displayed in the
window with a progress bar.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 33 of 119
When the installation was successful, this is shown as follows.
Depending on the selected installation, the following icons can then be found on the
desktop. In the actual release only the TransWin32 and maybe the REA ADB icon is
there. The other two icons are not required because these programs are started and
stopped automatically together with the TransWin32 program.
With the icon TransWin32, the REA Verifier software is launched. "VeriCube Barcode
Verifier" and "MLV-2D Barcode Verifier" are programs which are launched as device
service (no window, only an icon in the task bar). Both are launched automatically at
the start of TransWin32 and are also closed when TransWin32 is closed.
REA ADB is the article database; it is used for management, recording and data
import, which can be used with the REA Verifiers. The program is launched by the
operator when needed. The article database (REA ADB icon) is optional and can only
work with the REA Verifier when the program license is bought and also activated
with the device.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 34 of 119
6.6
Mounting and exchanging the camera lens module
Four different camera lens optic modules can be obtained for the testing of very
small or very large codes. The optic modules are interchangeable and can be
exchanged by the operator without the need for any tools. The different variants with
focal length, minimum code size and field of view size are described in section 2.2.
When changing the camera lens module, the device must be turned off. The knurled-
head screw is unscrewed and the module can be removed.
Picture of an camera lens module
The black dust-protection cap on the lens must be removed prior to mounting. For
each optic module type, one calibration dataset is saved. When a given camera lens
module is used for the first time, the calibration must be performed prior to any
measurement.
ATTENTION: The lens of the optic module is optically measured and is
preset. The alignment shall not be changed and the lens shall not be
mounted on other electronics.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 35 of 119
7
Correct handling
7.1
General considerations
The REA VeriCube is an optical measurement device with a camera and built-in
lighting. The device should be lifted only by the recessed grips provided for this
purpose. The camera module should never be used as a handle.
Falls and impacts will damage the REA VeriCube
Picture Recessed grip
Picture of a built-in camera module
The network cable is locked in the socket. The lock is loosened with a small plastic
lever on the plug for removal. This socket/plug combination is relatively sensitive
mechanically (robust industrial plugs were deliberately not used because they are
very large and require expensive and special cables, which are hard to source). The
cable should not be pulled without releasing the lock. The plug should not be
subjected to any pulling, pressure or other impact. Any lateral strains lead to high
risk of damage. The pulling strains can lead to damage or destruction of the cable
and, additionally, of the socket installed in the device.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 36 of 119
7.2
Checklist for measuring
The following points must always be observed for performing correct measurements:
1. The distance between the code and the device should vary only slightly. The
depth of field is contingent on the small distance between the lens and the
inspected area, limited to +/- 1 mm or 39,37 mils..
2. The code should always be positioned centrally in the middle of the field of
view.
3. The code should be aligned, as much as possible, in parallel or at an angle of
90° to the field of view in order to minimize any imprecisions due to staircase
effects at the code edges.
4. The automatic determination of the position defines the measuring area for 2D
codes and barcodes. The measuring area in the case of 2D codes is shown by
a superimposed green square and in the case of barcodes, by red
measurement lines. The square must surround the code in parallel to the code
sides.
The measurement lines must cross the bars of the code at an angle of 90°.
If this is not the case, the measuring area must be determined manually. (See
section 11.2)
Picture: 2D code with manual ROI (region of interest = determining the
measuring range)
Picture: Barcode with manual determination of the ROI
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 37 of 119
5. When determining the measuring area manually, the parallel alignment or the
90° angle must be observed. When the barcodes are positioned diagonally in
the field of view of the camera (when they are otherwise too wide), the
measuring area must be determined at an angle of 90° with respect to the
bars and not in parallel to the edges of the field of view.
Picture: Manual ROI at a more slanted position of the code
6. It should not be measured in direct sunlight, under direct (spot) light rays or
close to infrared sources (heat emitters). Ambient light will lead to changes in
the measured results and the sunlight leads to unusable results
7. The test sample must be positioned flat on the measurement surface. If this is
not observed, there are distortions in the measured picture level, which lead to
incorrect measurement results. The risk is particularly high in foils and labels,
which are processed as rolls. From the device position, the camera must
always record the code in perpendicular and the lighting must always be at an
angle of 45° (only the DPM measurements have particularities)
8. The transparent materials must always be measured on a flat, unchanging and
impermeable black base (usually with a reflection of maximum 5%). When
different measurement bases are used, the measurement results are also not
comparable
9. The synthetic aperture should always be in the range of about 80% of the
module width in order to obtain good decoded results. In case of 2D code
measurements according to ISO/IEC 15415 or DPM (ISO/IEC TR 29158), the
synthetic aperture is automatically determined
10.When there are several codes in the field of view and a code, which is not the
one centered in the middle, is recorded, the codes that are not required must
be covered.
11.The REA VeriCube must be regularly adjusted and has to be controlled with the
help of test equipment monitoring for measurement accuracy (calibrated)(see
sections 9 and 10)
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 38 of 119
12. Automatic recognition and decoding may recognize the codes incorrectly or
not at all. In such a case, the operator must evaluate whether the result is
valid or not and, if necessary, he should repeat the measurements with
corrected settings or perform them several times.
8
Operating the device
In some basic functions, the REA VeriCube can be operated with the device buttons
The on/off button
With this button, the device is turned on and off. The PoE power supply
needs to communicate with the connected device, this results in some
turn-on delay. It ends as soon as the left orange LED on the Ethernet
socket of the device lights up. Prior to that, the device cannot be turned
on.
The Scan button
A measurement is performed by observing the ROI, which is determined
(manually or automatically) on the PC
The F1 button
The F1 button turns on the live picture display so that the code can be
positioned. The live picture has a lower resolution than the
measurement picture, and the brightness is adjusted automatically so
that something becomes visible in the picture. This adjustment of the
brightness can be completely different than the adjustment for the
measurement with calibrated exposure settings.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 39 of 119
The F2 button
The F2 button toggles between red and white light illumination.
The F3 button
The F3 button is provided for subsequent extension, e.g. a function, which can be
assigned to the button by the operator through the TransWin32 software.
The LEDs above the buttons have the following meaning:
ON button LED
The LED lights up permanently as soon as the device is ready for operation.
SCAN button LED
The LED shows the testing result. Red means "failed" and green means OK/ passed.
F2 button LED
The LED shows the type of light used for measurement. It is either white or red.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 40 of 119
9
Adjustment and Calibration of the REA VeriCube
The terms “adjustment” and calibration are used in the sense of DIN 1319-1 and
JCGM 200:2012.
Calibration: The device measuring accuracy is checked against a known reference.
For this process the device adjustment shall not be changed.
Adjustment: The device settings (Exposure, white level, compensation of lens
distortions) are changed until a specific result is achieved. The following calibration
checks if the required measuring accuracy is achieved.
The REA VeriCube must be calibrated in regular intervals of between 4 and 12 weeks.
In the setting >configure >configure calibration and adjustment
the validity period of the calibration can be entered from 1 to 90 days.
(Up to and including TransWin32 version of CD9.16a (16607r) the validity of the
adjustment was entered here. It had to be adjusted via the software within this
entered period. Calibration with the REA calibration card was carried out in a further
manual step, just like the calibration of additional test cards.)
The REA VeriCube must necessarily be adjusted after installation of TransWin32 and
first connection to the device. Afterwards it must only be adjusted if the calibration
cannot be completed successfully, if a software update has been made or if you are
working with another computer.
In this case the menu item
>Configure
>Adjustment and calibration
must be selected and the complete adjustment procedure as described in the next
chapter must be performed. At the end, a calibration report is generated, which can
also be saved as a pdf, printed and signed.
If the calibration still fails after this, REA has to be contacted under
email [email protected] , possibly there is a hardware error which has to be
repaired.
A new calibration is always necessary if the ambient and lighting conditions change
significantly in the meantime or if the last calibration has expired.
It can be activated, that the operator won’t be able to go on with evaluation if the
calibration is not valid anymore.
This setting can be found in
>configure
>evaluation:
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 41 of 119
If the function “Only evaluate images with valid adjustment and calibration” is not
activated, the user only receives a warning when the calibration has expired. This
information is then also printed on the test report.
Note: this setting is saved individually in each profile
The instructions for performing a correct measurement must be observed particularly
closely during calibration (see section 7.2).
The adjustment and calibration data of the REA VeriCube are, in contrast to
the other REA Verifiers, saved in the PC – not on the device.
Therefore, software installations on other PCs must always be adjusted
again or the adjustment data must be imported from the first PC (see section
12.6) if the device is swapped to another PC.
The REA calibration cards are used for adjustment and calibration.
The following instructions for handling the calibration cards must be observed.
The calibration cards must not be dirty or scratched and their surface should
not be touched.
The card for adjustment of the white level correction is protected in the
delivered condition with a blue plastic foil. It must be removed prior to
adjustment.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 42 of 119
If the card is damaged or contaminated, it must be replaced.
The date of the first use must be noted. This date determines the start of the
lifespan of 2 years.
Before, the cards should not be stored unused for more than 2 years.
No calibration and adjustment can be performed with a photocopy of the
original calibration card because as a result of the copying, the metrical and
reflective properties change significantly. For the same reason, the calibration
card shall not be laminated.
The calibration cards may be cleaned.
The recommended cleaning agent is 3M Novec 71IPA
Adjustment and calibration must be performed in the same manner employed
for measurement. If a mounted glass top plate is used for measurement,
calibration and adjustment must also be performed with a mounted glass
cover. When work is conducted without a glass top plate, calibration and
adjustment shall also be executed without a glass top plate.
The adjustment and calibration data are stored for each camera module depending
on the respective s/n. For each camera module, adjustment and calibration must
therefore be performed first. In the following they must be calibrated regularly. After
changing to another optics module or after the respective calibration interval has
expired, a new calibration is required.
The calibration cards show a black reverse side. This is necessary in order to be able
to perform calibration and adjustment without influence from the reflectance on the
work table (the normal materials let the light penetrate). During the operation of an
adjustment step, the device should not be moved.
9.1
Adjustment of REA VeriCube
The device adjustment is initiated through the REA TransWin32 software.
For this purpose, the menu item >configure, >Adjustment and calibration
must be selected in TransWin32.
Make sure that the adjustment does not take place directly after switching on the
LEDs. You should start TransWin32, connect the device, make sure that the LEDs are
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 43 of 119
turned on and leaved them turned on for about 5 minutes, before starting the
adjustment procedure.
The REA VeriCube is delivered with a transparent plastic top plate, which covers the
inside space of the camera, as well as with a glass top plate.
The plastic top plate simply serves as dust protection during storage and
transportation. The glass top plate is meant for measurements.
Adjustments and measurements can be done with or without a top plate.
Be aware that the measurement is only valid, if you measure in the same
combination of verifier device and top plate, as you have adjusted the device.
Adjustment is performed in three consecutive steps:
1. Level correction (level correction calibration card Art. No. 90031703)
2. Reflectance adjustment (calibration card Art. No. 90031702)
3. Scale adjustment (calibration card Art. No. 90031702)
The adjustment status is shown after selecting the meue
>configure
>adjustment and calibration
If an individual step is not executed correctly, there is a possibility to repeat this
individual step separately. If an adjustment step was not successful, its old data are
retained for this step. In this way, the device always remains functional.
With the icon “Next”, it will be checked if there are valid target values for calibration
available.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 44 of 119
If the target values are missing, the error message above will pop up and with the
icon “OK”, you get into the input window.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 45 of 119
Explanation:
Point 1 and point 4 must be taken from the calibration card:
Point 2: here you enter the day when the card was unpacked and used for the first
time.
Point 3: the best before date is automatically calculated by the system according to
the production date and the day of first use.
Point 5: the 7 parameter values used for calibration must be transferred from the
reference test report. Please refer to chapter 9.2.2 9.2.2Calibration for white light
Attention: It is also possible to enter the parameters for several reference symbol
sizes if a reference test report is available for them. For each reference symbol,
however, all entries must be complete so that no error message will come up. The
calibration of several symbols may be necessary if additionally e.g. an 8mm camera
module or 50mm camera module is to be used and calibrated.
The reference values entered for adjustment and calibration remain saved
afterwards. However, it is essential to ensure that when using another calibration
card (e.g. after a card has expired) the new values of the new calibration card are
entered and afterwards an adjustment and calibration must be performed.
After the input of the reference values is completed, an overview of the entered data
appears after pressing the OK button.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 46 of 119
It is recommended to always make the adjustment for both types of illumination.
This means that both red and white light should be checked. It is possible to do
adjustment and calibration only for one type of illumination though.
By selecting the icon
the input window for the reference values appears again. This becomes necessary
when a new calibration card is inserted.
The "Next" button takes you to the menu for the adjustment:
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 47 of 119
9.1.1
Level correction
For the level correction, the white card with article number 90031703 is used.
Exception: The level adjustment with REA VeriCube UV
The white card (90031703) is not suitable for UV light. Instead, a white area of the
calibration card is used (90031702).
The card is positioned so that a completely unprinted area of the map is in the field
of view of REA VeriCube. The blue square in the image below shows the position that
should be shown in the field of view.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 48 of 119
After the white card is placed on the REA VeriCube, the level correction is started by
selecting the OK icon.
Picture of a level correction card (original size 22 x 25 cm)
Picture: The level correction card on the device
The level correction corrects the uneven lighting of the measurement field
(tolerances of the LEDs, construction, optics, etc.).
Level correction adjustment can fail, can give a warning or is successful.
If the highlight correction fails due to the fact that the adjustment data cannot be
saved, it must be checked whether the Windows user also has access rights to the
folder in which the adjustment data is saved.
By default, these data are stored under
a. Installation of the software for all users:
C:\Programdata\REA Elektronik GmbH\Verifier\ MorpheusVeriCube
b. Installation of the software for individual user:
%APPDATA%\REA Elektronik GmbH\Verifier\MorpheusVeriCube
In this folder, including subfolders, all users who perform an adjustment must have
full access.
The adjustment data are saved by:
C:\Program Files (x86)\REA Elektronik GmbH\REA
Verifier\MorpheusVeriCube\VeriCube_Control_PC.exe
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 49 of 119
If level correction fails or if there is a warning, the glass pane, the mirrow and the
lens of the objective in the optic module must be cleaned (make sure you use optics
cleaning cloth for glasses and lenses only, please see chapter 14). It may also be
necessary to clean the level correction card with a slightly damp cloth and then to
repeat level correction. If level correction does not function and a defect in the device
can be excluded, a new level correction card must be provided. This can be checked
with a simulated level correction using a new high-quality photographic paper. If this
adjustment is accepted, the card must be replaced. If not, there is possibly a defect
in the device and the device must be checked at the manufacturer or by an
authorized service center and repaired if necessary.
This does not apply to REA VeriCube IR. With this device the warning can only be
avoided by making adjustments and subsequent measurements without a disc.
However, it is also possible to continue working with this device despite the warning.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 50 of 119
9.1.2
Reflectance adjustment
Please use calibration card (90031702) for this step.
After level correction is finished, the menu item “reflectance adjustment” is
automatically selected. However, it can be switched in between the different types of
adjustment by selecting the appropriate icon.
The device must be positioned with the help of the live video representation in such a
way that the red cross hair is in the middle of the picture of the four reflectance
fields, which are in the middle of the "Reflectance calibration field" field. The software
automatically detects in which quadrant the bright field lies so that the direction is
not relevant.
Picture: Cross hair in the gray value field
Then the Reflectance adjustment is started with the button.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 51 of 119
During the reflectance adjustment, a progress window appears. The adjustments for
the red and white lighting are always performed one after the other and last for
about 2-3 minutes.
After finishing this process, the scale adjustment will automatically be selected:
9.1.3
Scale adjustment
As a third step, the scale adjustment is performed.
Picture: position of calibration card for scale adjustment
With the help of the live video representation, the Data Matrix code, which is
designated with Dnom = 0.5, is now positioned as precisely as possible in the middle
of the cross hair and perpendicular to the image section. With OK button the scale
adjustment can be started.
Exception:
Camera modules with 8mm and with 50mm focal length have to be adjusted
and calibrated with the other symbol sizes:
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 52 of 119
Camera modules with 8mm focal length:
Latest calibration cards contain a REA reference test report for
the symbol
Dnom = 0.75 and for Dnom = 0.5. For adjustment and
calibration of the 8mm lens the symbol with Dnom = 0.75
should be used and thus the reference values from the test
report of the 0.75 symbol should be transferred to the
software.
Nevertheless, it is also allowed to use Dnom = 0,5 for the 8mm camera module.
Camera modules with 50mm focal length:
For adjustment and calibration of the 50mm lens, the symbol with
Dnom = 0.25 must be used and thus the reference values from the
test report of the 0.25 symbol must be transferred to the software.
When ordering a new calibration card, it is absolutely necessary to
specify if a reference test report is required for the symbol
Dnom = 0.25, because this is not part of the standard scope of
delivery.
The scale adjustment sets the pixel size automatically in such a way that exactly the
known correct size of the calibration symbol appears in the measurement result.
9.2
REA VeriCube calibration
Calibration (Test equipment monitoring) is a process with which the values measured
by the device are compared with known reference results of a certified test code. The
values measured by the device shall deviate from the reference values only within
the specified tolerances.
Calibration should be repeated in regular intervals (e.g. monthly). Calibration shows
if the device is misaligned since the last calibration. If results show values out of
tolerance all measurements since the last calibration needs to be checked and re-
evaluated. The amount of deviation is taken into account.
For extended calibration, the calibrated GS1-DataMatrix test card (Art. No.
90031704) and / or the calibrated test card for the EAN/UPC Code Verifier (Art. No.
90031705), which can be obtained as optional accessories, can be used in addition.
The interval for calibration with these extended cards must be specified by the user.
A detailed description can be found in chapter 9.2.4
9.2.1
Admissible device tolerances for calibration with red light
The individual measurement values of the control measurements must be compared
with the corresponding reference values. The requirements for the precision of the
measurements are specified in the standards ISO/IEC 15426-2 for matrix code
verifiers and ISO/IEC 15426-1 for barcode verifiers.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 53 of 119
The following data must be examined in order to make sure that the correct symbol
with the settings for the test equipment monitoring has been used. Further
measurement data and threshold values, which are shown, in addition, in
the test report in the context, shall be ignored for this testing.
Aperture1
16 mil, 0.406 mm
Light type
660 nm - 45° from all four sides
Symbol
Passed 4.0
Symbology
Data Matrix ECC200
Code content
30Q324343430794
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 54 of 119
Further testing criteria
Measurement
parameter
Unit
Tolerance
Reference
Symbol contrast
%
+/- 8
(calculated from
Rmax and
Rmin)
Measurement
values on the
stamped
reference test
report
Modulation
Grade
+/- 0
Module size
µm
+/- 5
Print growth
µm
+/- 10
9.2.2
Calibration for white light
Together with the calibration card only one reference report created by using red
light is supplied. This report is created at the end of the manufacturing process.
In case white light is used and white light shall also be calibrated, then the process
described in chapter 9.2.2 is used too. The reference is also in this case the report
created by using red light. This simplification is possible because black and white is
supposed to provide the same reflectance values under red and white light. To prove
that the method is working the performance has been checked by using calibrated
reflectance standards (barium sulfate) which provide reflectance value for the visible
light range from 400 to 700nm wavelength.
The following data must be examined in order to make sure that the correct symbol
with the settings for the test equipment monitoring has been used. Further
measurement data and threshold values, which are shown, in addition, in
the test report in the context, shall be ignored for this testing.
Aperture5
16 mil, 0.406 mm
Light type
4000°K - 45° from all four sides
Symbol
Passed 4.0
Symbology
Data Matrix ECC200
Code content
30Q324343430794
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 55 of 119
Grid non uniformity (GN)
%*
+/- 6
reference test
report
Axial non uniformity (AN)
%*
+/- 2
Contrast uniformity
(see Modulation, text below)
%
+/- 10
Fixed Pattern Damage (FPD) Grade
+/- 0
*In the standard ISO/IEC 15426-2, GN and AN are described as absolute values with
tolerances +/-0,06 and +/-0,02. For better understanding, we report these
parameters in %. So we do have a tolerance of +/-6 and +/-2.
In the standard ISO/IEC 15426-2, Modulation is specified instead of contrast
uniformity. This is also correct in the sense of the control, but difficult to follow. A
footnote in ISO/IEC 15426-2 states that it is a per cent value. This per cent value
refers to the Contrast Uniformity value specified for the test symbol 5 (see the card
with the REA article number 90031704). Contrast uniformity is one detail value of
many used to determine Modulation for the entire symbol.
Further testing criteria
Measurement
parameter
Unit
Tolerance
Reference
Symbol contrast
%
+/- 8
(calculated from
Rmax and
Rmin)
Measurement
values on the
stamped
reference test
report
Modulation
Grade
+/- 0
Module size
µm
+/- 5
Print growth
µm
+/- 10
9.2.3
Calibration with REA calibration card
Calibration is performed by selecting the menu >configure, >Calibrate device..:
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 56 of 119
If the calibration fails, i.e. one or more parameters are outside the permissible
tolerances, the "Adjustment and Calibration..." function must be performed. If the
calibration still fails, the glass pane, mirror and objective lens must be cleaned and
then "Adjustment and Calibration..." must be selected again. If the calibration still
fails, please contact REA Support.
By selecting the icon
the last calibration report can be viewed again and if necessary a pdf can be created.
Via the button "Next" the overview with reference values will pop up:
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 57 of 119
The reference values can be edited and modified by
“Edit reference values”
A test report with the reference results of REA is supplied with all calibration cards.
The test reports with the reference results carries the following stamp:
Picture: Stamp on the reference test report
To avoid confusions, the certified calibration card has a serial number which is
mentioned in the reference test report.
Picture: User comment with the serial number of the calibration card
The serial number and maybe also the verifier type indicated on the reference test
report are different from those of the customer instrument. This is because the
reference test reports are generated with a REA reference instrument.
For calibration, the parameters listed in chapter 9.2.1 have to be checked.
For 8mm camera modules, the code Dnom=0.75 should be used. (Dnom=0.5 is also
possible though)
For 12mm, 16mm, 25mm, 35mm camera modules, the code Dnom=0.50 must be
used.
For 50mm camera modules, the code Dnom=0.25 must be used.
The values from the actual measurement report are compared with the target values
from the reference test report. The deviations of the current measured values from
the reference values must not exceed the maximum permissible instrument
tolerances. This comparison is done automatically by calibration function.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 58 of 119
The parameters to be compared, which must also be entered as reference values in
the calibration menu, are marked in the example report below:
With the icon “Next”, the calibration routine is edited:
The appropriate reference symbol (depending on the camera module that is to be
calibrated) is positioned in the crosshairs at 90° and calibration is started with "OK":
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 59 of 119
A calibration protocol is then automatically created and opened:
If both types of illumination have been adjusted and calibrated, an overview sheet is
obtained and subsequently the results of the calibration for both types of
illumination.
If only one type of illumination is adjusted and calibrated, the overview sheet is not
created as not necessary.
The calibration report differs from the measurement report for quality measurement,
as the tolerances from ISO/IEC 15426-2 are used as a basis here.
The calibration report can now be exported to pdf, then printed, signed and filed via
Windows if required.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 60 of 119
Calibration card 90031704
Calibration card 90031705
9.2.4
Calibration with extended calibration cards
For this purpose, the extended calibration uses test cards, which are sold by REA and
its authorized resellers, but are otherwise manufactured fully independently of REA
and are measured with other reference devices (traceable back to national
standards).
Optional accessories for extended calibration are the calibrated GS1-DataMatrix test
card (Art. No. 90031704) and / or the calibrated test card for the EAN/UPC Code
Verifier (Art. No. 90031705.
The interval for calibration with these extended cards must be specified by the user.
Each symbol on the extended calibration card is optimized in print for the parameters
listed on the front panel.
For the extended calibration, only those parameters are compared that are specified
directly on the calibration card.
This means that the reference test reports supplied with the cards, which were
measured with a high-resolution camera system, are not required for calibration.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 61 of 119
As an example, the contrast uniformity according to ISO/IEC 15426-2 is calibrated
with a certain symbol with a dark module, completely surrounded by light modules
("dark widow"). This is found in symbols #6 and #9 and thus the contrast uniformity
is only calibrated for #6 and #9. The Symbol Grade is not calibrated and may differ,
because some symbol parameters are close to the threshold between two gradings.
Permissible device tolerances from chapter 9.2.1 also apply with these cards.
For most users the calibration with our calibration card is completely sufficient.
The more reference codes are used for calibration, the safer it can be said that the
measuring instrument actually measures accurately. However, it should also be
considered that this requires more time for calibration.
For this reason, we have listed various "safety levels" in the following as a guide to
make it easier for the user to decide for himself how to calibrate.
Additionally, it must be considered whether the built-in camera module is even
suitable for measuring a code as small as the symbols #5, #10, #11 and #12 with
X=0,200. A resolution of approx. 10 pixels per module is recommended.
Level 1:
Calibration with REA calibration card with DataMatrix reference test
code: part-no. 90031702
Level 2:
Calibration with REA calibration card with DataMatrix reference test
code: part-no. 90031702
+ calibrated conformance test card for Data Matrix and GS1-DataMatrix
part-no. 90031704, Code #1, #4 und #6 only.
Level 3:
Calibration with REA calibration card with DataMatrix reference test
code: part-no. 90031702
+ calibrated conformance test card for Data Matrix and GS1-DataMatrix
part-no. 90031704, Code #1 to #4 and #6 to #8.
Level 4:
Calibration with REA calibration card with DataMatrix reference test
code: part-no. 90031702
+ calibrated conformance test card for Data Matrix and GS1-DataMatrix
part-no. 90031704
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 62 of 119
10 Synthetic aperture / measuring opening
Under aperture or opening we normally understand the diaphragm opening of a
camera lens. In this context, this is not the case. The diaphragm opening of the lens
of the REA VeriCube has a fixed presetting from the plant, which achieves the best
compromise between the maximum possible depth of field and the minimum possible
diffraction effects.
The synthetic aperture simulates the spot of light of a laser scanner. This takes place
through mathematical filtering of the recorded measurement image. Depending on
the aperture used in the code testing, the measurement results can vary. The
synthetic aperture is always automatically set to 80% of the module width of the
tested 2D matrix codes.
Alternatively, the synthetic aperture can also be set manually
The synthetic aperture forms an optical filter from the picture processing view, which
causes certain blurring. This makes sense for suppressing the smaller and up to the
smallest distortions in the picture, which are not relevant for verification.
The symbols, which show many distortions, can disrupt the automatic determination
of the synthetic aperture. When this is the case, the measurement results oscillate or
there are invalid measurements. In this case, a manual determination of the
synthetic aperture is recommended (under the condition that the symbols always
have the same module size).
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 63 of 119
11 Execution of the measurements
11.1
Positioning of the device
The REA VeriCube can be operated in different measuring positions.
11.1.1 Camera downwards
In this mode, the device is positioned, for example, on a printed sheet or on another
flat product with the camera pointing downwards. On a printed sheet, the device can
be moved from code to code and the measurements are made in this position. In this
case, the glass top plate must be removed, so that the cover is not contaminated or
scratched by moving on the product or the printed sheet. As in this case the device
largely covers the measurement area, the effect of the ambient lighting is negligible.
Nevertheless, intensive light sources such as direct sunlight must be prevented.
Picture: Camera downwards on the printed sheet
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 64 of 119
11.1.2 Camera upwards
In this mode, the product is positioned on top of the measurement window of the
device and the product is moved into the measurement field. For this type of
measurement, the glass top plate should be used, so that the inside space of the
camera (mirror, lens) is not contaminated. The ambient light influences the
measurement and, therefore, the measurement area must be covered with the
delivered black cover plate or with the black cover cloth in order to protect it from
the ambient light.
Picture: Camera upwards with the product on the glass top plate
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 65 of 119
11.1.3 Camera sideways
The device stands either on its long side or on its short side. Plastic strips are
inserted in the side walls, so that they are not scratched when the device is moved
on the table. The codes to be measured are, for example, on cans, bottles or parts,
which can be positioned in front of the device, so that the measurement conditions
are maintained correctly. Since the long sides and the short side change the distance
from the floor space, the respectively optimal position can be selected in this way. If
this is not sufficient, the product or the REA VeriCube and the product can
additionally be placed on a lifting table and thus the optimal measurement position
can be set.
Picture: Device positioned sideways with the product in front of the device
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 66 of 119
11.1.4 Turning the keyboard
The keyboard of the device can be taken out and then replaced in an appropriate
position with respect to the measurement position or the position of the device.
Attention: The keyboard is connected to the main electronic using zero force
connectors and a flat cable. Be careful when you pull out the keyboard.
It is not allowed to pull at the cable! Loss of warranty, if you disconnect the flat cable
from the connector!
11.1.5 Testing 3D parts
It is possible to measure 3D parts (eg. automotive, electronics, machine building,
and aerospace industry, etc.) if they shaped in irregular geometries. In some cases,
these parts can be measured with one of the three or four standard positions. In
other cases, this cannot be done.
For such cases, a measurement fixture must be manufactured, which ensures that
the code to be measured can be positioned correctly according to the handling
instructions (angle, distance, shadows from components). Those parts are usually
direct marked parts, marked by laser or other methods, for such DPM parts, the
additional software license for DPM verification is needed (please see chapter 2.3).
In simple cases the fitting of the glass top plate is used for a fixture. A part fitting
instead of the glass top plate is manufactured and this holds a part exactly in the
correct position. The advantage is the ease with which you can replace and modify
such a fixture.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 67 of 119
11.2
Measuring process using the Windows software.
A prerequisite for performing the measurement is the observation of the handling
instructions (see section 7) as well as the selection of the desired position of the
device (see section 11.1)
The live video is turned on for the positioning. For this purpose, the F1 button on the
device can be used or the symbol button for the live video on the PC screen.
Start the live video
Stop the live video
The measurement area can be set in various ways. Most simply, only the red cross
hair is displayed. The code is moved in the middle of the cross hair and then the
measurement is started. To start the measurement, either the "Scan" button on the
device or the "Evaluation" button in the Windows software can be used.
Cross hair in the barcode
Cross hair in the 2D code
The evaluation is started with the "VERIFY" button.
Pushing the evaluation button stops the live video and starts taking the measuring
image in the full camera resolution. Afterwards, the evaluation is initiated. In a
normal case, the symbology and the code type (1D or 2D) are recognized
automatically.
If the automatic recognition fails, the measurement area must be set manually.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 68 of 119
The icon with the blue square is the default for the measurement of a 2D code
If the function is active, the three corners of the matrix code in the measuring image
are clicked on by the mouse one after the other. This causes a blue square to
appear, with which the measurement area (Region of Interest – ROI) for the code is
determined. The thickness of the blue frame is not relevant for the measurement; it
can be changed by turning the PC mouse wheel. Only the outside contour is taken
into account. By changing the blue frame (the size, the precision with which the code
is targeted), the stability of the evaluation can be improved in difficult measurement
cases.
If a barcode is being measured, then a red measurement area is positioned
perpendicular to the barcode lines. The height of the bar determines the
measurement area through the height of the barcode.
The measuring width is determined automatically (up to Version 1.0.20.1) Version
1.0.20.1/13914r and later limmits the measuring width to the left and right position
points.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 69 of 119
11.3
Automatic measurement mode
If the automatic mode is activated, the measurement does not have to be started
manually via the "Verify" button in TransWin32 or via the "Scan" button on the
verifier. In this mode the user positions the code so that it is in the crosshairs and
the software recognizes it. As soon as the software recognizes the code, this is
represented by a green square or red bars:
After a delay time, which is specified in the settings, the measurement is then
automatically triggered. The result is then displayed and after a further specified
waiting time, the live image is shown again so that the next sample can be
positioned.
In automatic mode, the test reports are always saved automatically in the file path
set in menu >Configure >Options >File Saving "Default data path" and in
TransWin32 Explorer only the last measurement is displayed ("single mode active" in
the display options). In order not to disturb the automatic process, requests for job
number, printing or saving are not displayed, even if this is selected in the settings.
These settings with prompts apply only to manual operation.
In the menu
>Configure, >Others, >Automatic mode, the mode is activated and the time settings
are made:
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 70 of 119
With the buttons you can pause and restart the automatic mode.
pause:
restart:
11.4
Automatic ROI
The setting "Automatic recognition of the measurement area" is made with the icon:
The 2D codes are framed automatically. In the case of barcodes, the measurement
lines are determined automatically. In contrast to the mode with the cross hair, the
measurement area is limited to the code size. This reduces the amount of data to be
processed and increases working speed.
11.5
Mirror live image
With this function the live image can be mirrored. This function can facilitate the
positioning of the code in the crosshairs.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 71 of 119
11.6
Setting the brightness of the live image
With some samples, the live image is very dark. To make it easier to position the
code into the crosshair in these cases, you can manually adjust the brightness for the
live image. This setting does not influence the measurement.
11.7
Recording the image
The REA VeriCube can take an image without making an evaluation.
Icon.
Old icon
New icon
11.8
Opening the image file
With the icon
Old icon
New icon
BMP, PNG and JPG image and several other formats can be processed. The loaded
picture can be evaluated and thus a rough assessment of the code in the image can
be made. The image data read in this way is not calibrated and no calibration data is
included in such images.
Similary, the test reports in TWS format or the images in XML format can be opened.
The images can be saved again for documentation purposes, for example, as BMP or
JPG files but because the previously set ROI as well as the calibration data are
contained within the image, a full-value evaluation is possible afterwards.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 72 of 119
11.9
Saving the image file
The images, which have been loaded and which have been obtained by measuring or
by recording, can be saved (for analytical or documentation purposes)
This takes place with this ICON
Old icon
New icon
The images in PNG, BMP and JPG format can be used for documentation purposes
only whilst the images which are saved in XML format contain the calibration data
and can therefore, when opened again, provide a full-value testing evaluation. This is
useful for comparing different software releases or for support purposes.
11.10 Static reflection measurement
In a given measurement image (without evaluation, see section 11.7), a static
reflection measurement is performed
A red measurement point appears in the image
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 73 of 119
If the mouse pointer points to the measurement point, the point can be changed to
the desired size with the help of the mouse wheel in order to obtain an adequate
measurement section. With the help of the function, the reflection values can be
determined in order to see in advance whether the contrast between the bright and
dark areas (or the color selection between bright or dark) is sufficient. Depending on
the type of the light (white or red), the measurement values can change.
The displayed value of 26.3 is the reflection value that corresponds to the Rmin or
Rmax reflection values. The X and Y data are the pixel positions in the image. D is
the diameter of the simulated light point adjusted by the user. The usual synthetic
apertures (the simulated light point diameter) are between 100µm (4mil) up to
500µm (20mil). This area is exceeded above or below in exceptional cases only.
11.11 Preset the measurement image size
The measurement image can be presented considerably magnified or reduced in size.
With the help of the buttons
the display is changed. The piece of data in percentage shows the magnification. The
value 45.8µm/pixel depends on the focal length of the lens used and is set by the
adjustment.
If the mouse pointer is positioned in the image, the image can be displayed smaller
or larger by turning the mouse wheel. It can be magnified to such an extent that the
individual pixels of the camera chip are visible and distinguishable.
11.12 Rotating the image
Depending on the orientation of the REA VeriCube, the movement of the image does
not follow what is maybe intuitively expected. For this reason, the image can be
rotated with the help of the green arrow symbols
until the movement of the image moves as expected.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 74 of 119
11.13 Behaviour of the software TransWin32 when locking the PC
In the settings of the software you can specify how TransWin32 should behave, if
during a session in TransWin32 the PC is locked or the screen saver is activated.
In the menu >Configure >Options >General Settings it can now be selected, if the
TransWin32 session should be closed automatically when the PC is locked or if the
event should be ignored and thus the TransWin32 session remains:
When a Windows user logs off, the TransWin32 software is always closed.
If unsaved measurements were previously open, they are automatically saved in the
default path from “file saving”.
11.14 View of the Reports
Depending on the preset values in the options, the level of detail of the individual
test reports displayed can be adjusted and changed based on the individual
requirements.
After a new installation of the software, the simple report will be displayed by default
with every measurement, like shown in the following:
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 75 of 119
Screenshot1: simple report
For the user who only performs measurements, it is easy to see whether the test
result is good or bad without being loaded with too much information on the screen.
There are two ways to change this initial view:
1. subsequent modification of the view to see more information about the
measurement parameters
2. change the default setting, which view is always displayed after a
measurement
ref. 1: the selection of the appropriate view is done with these icons:
The icon on the left shows the detailed graphical test report on the PC monitor
(graphic view). This view is designed for qualified personnel with appropriate
background knowledge and shows all test parameters in detail, partly in editable
sub-windows.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 76 of 119
Screenshot2: Graphic view
The second icon shows the evaluation comparable with a printout of the report (Text
Report, long printout). Here the details are listed one after the other.
Screenshot3: Text report
The third icon shows a short view. This view provides all testing results in a
compressed form, which can be used by the trained routine personnel.
(Screenshot1)
The fourth icon shows a report with GS1 format. This view is available for GS1
barcode symbologies only, not for 2D codes.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 77 of 119
Screenshot4: GS1-Report (only selectable subsequently)
ref.2:
In the settings (>Configure >Options >View), the default view can be set for single
measurements and for multiple measurement. This view will be shown after each
verification. The icons described before change the view on demand without changing
the default setup.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 78 of 119
11.15 Details in the test reports
In the graphical view of a given test report, the following icons are available:
Binarization
With this icon, everything in the image which is evaluated as dark is highlighted in
yellow. This makes it possible to recognize the places where shadings or too bold /
thin prints shift the thresholds to such an extent that the individual code elements
become invisible
Edge grid
With the help of the edge grid, the reconstructed grid matrix is displayed. In this
way, it can be recognized where the matrix cells lie and whether the symbol has
been recognized correctly. The symbols with irregular brightness or with distortions
(for example, through rounding) are subject to the risk that the matrix is
inadequately recognized. With the help of this highlighting, it can be checked very
quickly as to whether the recognized matrix corresponds to the actual code picture.
If there are large deviations, this is an indication of invalid erroneous measurement,
which must be repeated.
Center grid
In this grid representation, the lines cross in the middle of the matrix cells. Apart
from that, the function and the meaning are the same as in the edge grid.
The light margin (quiet zone)
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 79 of 119
The light margin (also called a white field, bright field or quiet zone) is indicated by
the aid of this icon. This is in order to be able to easily recognize whether this area
contains spots or other disorders which should not be there (homogeneous white in
positive symbols or homogeneous black in negative symbols is necessary)
Details measurement
With this icon, the details measurement is turned on. A red point appears in the
measured image, whose diameter shows the size of the synthetic aperture. The point
can be moved with the mouse to different places of the code. The display then
provides information about the code word number, bit number, reflection values and
modulation values at the selected location.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 80 of 119
Matrix middle points
The middle points of the matrix cells are highlighted with green dots
Warnings
With this button, all matrix cells (when active) are marked yellow, which, in relation
to the modulation and the contrast measurement values, are close to the limit
between two grades. The more such modules appear, the higher is the risk for a
given measurement result to oscillate between two grades.
Error indication
All matrix cells which can be handled by the error correction are highlighted with red
dots. This always occurs when a given matrix cell is recognized as dark instead of
bright or vice versa. Depending on the symbology and code size, the error correction
can recognize these errors and facilitate the correct decoding to a certain limit. When
the capacity is exhausted, it is attempted to obtain a result with a non-standard
decoding (error: Reference decode failed). If this does not aid the decoding, then no
further decoding is possible.
12 Configuring the evaluation and symbologies
In this section, the essential REA VeriCube configurations are described. This
configuration is performed with the help of the Windows software TransWin32. The
detailed description of TransWin32 is in a separate operating manual.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 81 of 119
12.1
Evaluation
The presetting for the evaluation is done in the menu Configure / Evaluation.
The important setting is "selected grade". For this purpose, the desired minimum
quality in the area of the grade Sufficient (Grade 1), Satisfactory (Grade 2), Good
(Grade 3) or Very good (Grade 4) is determined according to the respective test
standard. A measurement result which has achieved, for example, Grade 2 – the
grade Satisfactory, is thus evaluated in the previously selected Grade 1 as "Passed"
and in the selected Grade 3 as "Failed".
The barcode evaluation (1D) permits the choice between the ISO/IEC 15416 and
ANSI X3.182 evaluation (this ANSI standard has subsequently been withdrawn.
The evaluation and the measurement result do not change depending on the setting
of ISO or ANSI. It is just the presentation of the measurement result that changes
from numerical data to letters for the class (A instead of 4, B instead of 3, C instead
of 2, D instead of 1 and F instead of 0). ISO/IEC 15416 is available in the version
from year 2000 (integer grading) and alternatively in the version from 2016
(continuous grading with one decimal).
The matrix code evaluation (2D) permits the choice between the method according
to ISO/IEC 15415 for printed codes, for example on labels, or the DPM method
according to ISO/IEC TR 29158, which is used, for example, for etched codes, dot
peen marking or laser marked on metal (remember DPM software and/or model is
optional).
Note: The standard REA VeriCube has 45 °lighting from all four sides. This
illumination type shall only be modified for DPM codes.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 82 of 119
Also the two transformation options shall be enabled for DPM codes only. This setting
is not according to ISO/IEC 15415.
Coordinate Transformation:
Activating this function is useful for 2D codes applied to cylindrical parts. A distortion
of the codes caused by the radius of the part is compensated by coordinate
transformation. It is planned to use this function for checking in production, as
inaccuracies in the alignment of the code may occur. The approval of printing
layout/laser marking should be based on a measurement without activating the
coordinate transformation.
Reflectance Transformation:
If enabled, bright spots on very glossy metal material will be ignored for verification.
12.2
Symbology selection
The symbology selection determines which of the symbologies and with which
respective properties they are verified. The selection and the properties determine
the respective use.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 83 of 119
The left column shows the symbologies available in the software version. The right
column shows the active symbology selection. To the right of this there are the
properties of the respectively selected code. In this case, it is "GS1 Data Matrix ECC
200". This designation of the symbology is a short form for practical use, which,
correctly formulated, would be "Data Matrix Code according to ISO/IEC 16022 with
the error correction method ECC200 for the applications which are specified by the
organization GS1-International". However, this is much too complicated.
Only symbologies which have also been activated in the pre-settings can be
tested and evaluated. All other non-activated symbologies are evaluated
with a "Not decoded" error message and cannot be recognized and
evaluated!
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 84 of 119
12.3
Lighting and aperture
The lighting setting is in the menu Configure / Device.
As light type (color of the lighting), red light with 660nm wavelength and white light
with 4000°K can be selected. The normal setting for the direction of the lighting is
the 45° quadripartite group. In certain cases, a single-sided lighting can be used.
However, in the ISO/IEC 15415 mode, these cases are not calibrated and thus they
can only be reasonably used in DPM mode.
For testing purposes or for in-depth analyses, the adjustment can be turned off. As
soon as the check mark next to "Use adjustment" is removed, the manual exposure
setting is active.
ATTENTION: Changing the manual exposure setting leads to changed and
non-calibrated measurement results
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 85 of 119
The DPM lighting is performed according to a preset algorithm of ISO/IEC TR 29158
when "DPM Automatic Exposure" is turned on. When "DPM Automatic Exposure" is
turned off, a manual exposure is active, which can also lead to measurement results
which are different to the ones with the automatic exposure according to the ISO/IEC
TR 29158 specification. The manual setting can be useful if the automatic is not set
up as desired or leads to no decoding. The "Duration" entry field is the manual
exposure setting.
In addition, the aperture can also be changed here. In a normal case, the aperture
will stay at the automatic setting. In this case, the aperture is set at 80% of the
module width. For all application specifications the aperture is used according to the
appropriate specification (e.g. GS1 specifications or PPN specifications for barcodes).
The aperture entry takes place in the "mil" measurement unit. Mil means milli-
inches. One inch is equal to 25.4 mm. Milli means 1/1000. Therefore, 1 mil is 25.4
µm (corresponds to 0.0254mm). The typical apertures, such as 6 mil or 10 mil,
correspond to 0.15mm or 0.25 mm.
The aperture is a subsequent formation of a light spot created by a laser. Since it is a
subsequent formation, the aperture is called "synthetic aperture".
The synthetic aperture is an image processing filtering whose purpose is to suppress
the small, irrelevant disruptions in the code. Further details about the opening can be
found in the REA training manual.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 86 of 119
12.4
Profile management
The settings for evaluation, symbology selection, minimum quality and aperture
setting are extensive and too complicated for most standard users. For this reason,
settings can be stored in profiles. (Configure / Others / Manage profile)
Up to 80 profiles can be defined. In practical use (e.g. production), a user selects the
matching profile for a production job. Selection and activation of a profile sets a
complete device configuration within one step.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 87 of 119
Profile selection is carried out using the icon in the icon bar.
Along with the profile selection, password protection
can also be used. With password protection, no settings can
be
changed. The task-related selection of the settings profile and
the
entry of the order number are excluded.
12.5
File Saving Options
The test reports can be easily identified with the help of the entry of the order
number. The order number is entered under Configure / Order number or by using
the appropriate icon in the icon bar.
This icon opens the window for inquiring the order number to appear automatically
before or after any measurement, allowing the user to confirm or change the order
number. The intention is to prevent users to forget the entry of new order numbers.
The last order number that was in use will be shown in the text. This allows the user
just to confirm with the OK button the last order number.
Under Configure / Options / File Saving, it can be specified how the filename of the
reports is to be given.
For example, an order number or user name can be automatically added as a prefix
in the file name of the test report. This function is valid for both manual and
automatic saving.
Additionally, a part of the decoded code content can be included into the report
filename. For this mode the configuration “Configure / Options / Saving” needs to
enable this function and to specify the part of the code content. The part of the code
content defines the first digit and the data length. Counting of the first digit can be
configured from the end or the beginning of the code (left to right or opposite).
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 88 of 119
In addition the configuration has a setting which ensures that each order number
creates a subfolder in which reports are stored into.
If a test report is converted to a PDF document through a PDF Writer, the generated
file name (including the automatic numbering) is transferred to the PDF program.
Thus, if the order number is used consistently, manual entry of file names is not
necessary for saving and creating PDF documents.
Saving measuring reports as PDF files can be configured to work in a background
mode. This means that no user dialog is required to create the PDF files. All that is
required is a third party PDF writer program. Measuring reports are in this case
always available as PDF files without any action required to be done by the routine
user.
Attention: It is strongly recommended that all users have full access to the set
directory for document storage. Restriction to read and write access may result in
restrictions to some functions.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 89 of 119
12.6
Device data
The calibration and service data can be saved by using Configure / Others / Collect
device data. The calibration data can be re-imported on another PC with the
TransWin32 installation for a REA VeriCube device. In this way, a new calibration can
be avoided if one device is used alternately at different PCs.
The device data also contains the settings
The service data is required for support, bug fixing and device failure analysis.
The import of the data can be executed with the "Configure / Others/ Import
Vericube device data" menu item.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 90 of 119
12.7
User management
12.7.1 Simple user name
The setting "TransWin32 User …" is a pure text replacement for the names of the
respectively logged-in Windows users. If the user is not entered in the TransWin32
setting, the user name of the Windows user account appears in the printed test
report. If there is an entry, this setting is always included as user in the test reports.
In a next software version, this menu item will be moved into the "User and
password" dialog box.
12.7.2 Simple password
Independently of the simple user names, there is a simple password protection.
The setting can be accessed under "Configure / Others / User and Password Setting".
In the displayed setting, the simple password protection is active. In this mode, all
settings, except for the profile selection, are inaccessible for the user without a
password.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 91 of 119
12.7.3 Setting up the user management
After an update of a device without user management or after the first installation,
neither the user management nor the password protection is active.
The user management must first be activated before becoming effective. It can be
temporarily deactivated at any time or completely turned off by deleting all users
(see below).
The deactivation of the user management changes the following:
A list of users with their passwords and their rights status is managed. The
management of the list of users, the logging in and logging out of the users takes
place centrally on the device. The respective user must authenticate himself with a
password prior to using the device.
There are two user types: normal users and administrators. The administrator has
access and permission to do everything; the normal user can only do tasks which are
possible without entering a password when the password protection is active (e.g.
performing a measurement, selecting a profile, ...).
The name of the logged-in user is stored in a log file. The following actions are stored
in the log file:
Calibration
Creating a new user, deleting a user
The user management can be applied through two alternative approaches:
User management through TransWin32 (only if the SW option is available)
-
User management through the service program (relevant for users of REA
Check ER and REA ScanCheck 3 without a TransWin32 license)
The actual user management dialog and also the subsequent work procedure are the
same in both cases. The ways to access the user management dialog are different. In
the TransWin32 software, the user management is accessed through the "Configure -
> Others -> User and password setting ..." menu item. The service program is
started either as a separate program or from TranWin32 under the "Tools" menu. In
the service program, the "tab" "Profile and User Management" is selected. The
desired device is selected with the search function of the device. With the "User
Management..." button, the user dialog is accessed.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 92 of 119
At the first, the user dialog looks as follows:
Picture: Selection of the setting "Protection of the device configuration with User ID
and Password"
When neither user nor password has been assigned, the dialog is confirmed with OK.
A dialog appears with the password prompt of the preset administrator (Admin
default). If a user has already been created, the dialog for user management can be
started directly with the "User Management" button.
The preset password is 739485.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 93 of 119
After logging in, the following dialog appears:
If "Add user" is selected, the next dialog box pops up:
A unique name is assigned as user name. The password is entered below. The user
type selection list determines whether the new user is an "Administrator" or a
"Normal User". The Administrator has all rights and permissions. The "Normal User"
can essentially perform measurements and select predefined profiles.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 94 of 119
Attention: The first user that is entered must have administrator rights. As a
result, the Admin default becomes invisible, i.e. it does not appear anymore
in the list of the users.
Afterwards, the user accounts for the operators can be created.
The Admin default user disappears after the creation of further user accounts.
Using "Delete Users", a users can be removed with the exception of the logged-in
user.
After the end of the creation of users, the administrator must log out. This is done in
the main display with the Log Out button.
If the password is no longer accessible (forgotten, misplaced), a three-time login
with an incorrect password takes place. After this, the login can be attempted once
more with the Admin default user.
12.7.4 User management during operation
If the device is started with active user management, the user ID is always
requested first (selection from the list of user names), together with the password
belonging to it. Depending on the selection, the use as administrator or as normal
user is possible.
After logging in, the REA device is ready for measurement, or if an administrator has
logged in, it is ready for measurement and configuration.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 95 of 119
Every user must log out after completing their tasks as automatic logout is not
provided.
A logged-in normal user can only change his own password in the user management.
The dialog is accessed through the
User management icon or through the menu Configure / Others / User and password
setting
The following input form appears.
The data in parenthesis (here (ww) ) shows that the logged-in user has the name
"ww".
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 96 of 119
When an administrator calls up the user management, the full dialog appears:
12.7.5 Transwin32 configuration protection
The selection "Transwin32 configuration protection" just locks down settings of the
Transwin32 software. Eg. print layout, data saving path and data export settings.
Settings related to evaluation, measurement, reporting and calibration are not
locked.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 97 of 119
13 Trouble shooting
In this manual, only the device operation and the configuration are described. The
testing results, their meaning and the search for errors in the case of poor results are
not contained in this manual. The measurement results and their interpretation are
introduced in the REA training document.
Network error
Prior to a further search for errors, the cables must be checked as well as the
network sockets on the REA device. If the locking mechanism on the plug is broken,
the contact is poor. If the socket on the device is defective, the contact may be poor.
In this cases, a corresponding repair or replacement of the cable is necessary.
Recommendation: Only network hardware supporting 1Gbit/s should be used.
Unshielded (UTP) network cable types should not be used.
Same addresses
The PC and the REA Verifier device have identical network addresses (IP addresses).
They must be different or there will be an IP address conflict.
Several network cards
If a given PC has several network cards, they must be separated with the help of a
logical setting. This is done with the help of an IP mask and basic selection
Example of a bad configuration
Adapter 1
Address:
192.168.111.11
Mask: 255.255.255.0
Adapter 2
Address:
192.168.111.12
Mask: 255.255.255.0
Both network adapters (network cards, including WLAN) work in the same network.
The REA Verifier on adapter 1 sends the data to the PC and the PC can send back the
data through either adapter 1 or 2. If adapter 2 was chosen, the data no longer
comes to the REA Verifier as it is connected to adapter 1. In this case, as an
example, adapter 1 should be set to 192.168.113.11 and the REA device to
192.168.113.12. The Adaptor 2 setting remains unchanged. With this setting,
adapters 1 and 2 are separated.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 98 of 119
Different logical networks
If the REA testing device and the PC with the TransWin32 are in different networks,
the connection does not function anymore. The device search is successful and then
an error message follows. In this case, please follow the instructions in “temporary
change of the IP address”.
Example
Device:
Address:
192.168.1.12
Mask: 255.255.255.0
PC
Address:
192.168.2.13
Mask: 255.255.255.0
Sometimes the communication is realized by using a network gateway in such cases.
However this does not work with the VeriCube because the GigE Vision network
protocol doesn’t support communications via gateways.
Speed
The REA VeriCube has a network connection with max. speed of 1 Gbit/s. If the REA
device is connected directly to a PC with a 100Mbit/s network interface, the speed is
automatically restricted to 100Mbit/s. If the REA device and the PC are connected
through an Ethernet switch with a speed of 1Gbit/s and the PC can, as before,
process max. 100Mbit/s., no automatic speed adaptation takes place because the
Ethernet switch connects to the PC with 100Mbit/s and the REA VeriCube with 1Gb/s.
In this case, the REA VeriCube sends the data 10 times faster than the PC network
interface can process.
Connection interruptions
Connection interruptions can happen when the REA Verifier is integrated together
with the PC in a company network. A high network load can lead to communication
errors which are displayed as network errors or sensor errors. In this case, a second
network card, which is used for the REA VeriCube only, should be installed in the PC.
In some cases, problems might arise if the network adaptor is built with components
of the manufacturer Realtek. In this case, a second network card, which works with
the chip set of another manufacturer, can help.
Version 1.0.20.1/13915r and later offers under Tools / Install or Uninstall VeriCube
Network driver an installation of a network filter driver. This driver improves
communication reliability.
If both, the computer and the REA VeriCube are connected to an Ethernet switch
with 100Mbit/s as highest speed communication errors can happen. In such an
installation a 1Gbit/s Ethernet switch should be used.
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 99 of 119
Sensor error
The correct positioning of the camera modules must be checked. Sometimes the
sensor error is also reported if the network communication is not stable (see
connection interruptions)
Firewall
The majority of computers use firewall software which, during day-to-day operation,
protects the data from theft by outside parties or from the installation of undesired
software from outside. This function is mandatory for all computers which are
connected to the Internet.
The REA VeriCube works through the network. The firewall must permit the data
packages of the device to pass through. In particular, in very strictly administrated
networks, the firewall is often set up so restrictively so that no communication is
possible. The firewall must allow the GigE Vision network traffic based on UDP to
pass.
Temporary change of the IP address
If a fixed IP has been assigned to REA VeriCube, but this is no longer permitted in
the network, then the device can no longer be connected, but is still found. This will
also happen, if the REA VeriCube and PC are in different logical networks.
The current IP address of the REA VeriCube can be read after selecting “browse
device”
If the REA VeriCube network settings are set to DHCP ON, the address should be in
the following range 169.254.xxx.xxx
After selecting the device in the browse window above, the following message will
come up
To change the IP address temporarily, select “Ja”
In the background, usually behind the TransWin window, a new window will start up:
© REA Elektronik GmbH, Teichwiesenstraße 1 D-64367 Muehltal
REA VeriCube Operating Manual Version 2.16 – 11/2020
Page 100 of 119
tick on "show unreachable network devices" , select the appropriate unit and click on
the icon on bottom left side "Set IP address"
Now, an IP address can be set, that is reachable by your PC. But please be aware,
that this IP will only be active in your VeriCube until next restart of the device!!!
This means, don't turn off the Vericube, but connect it by TransWin32 and change
the network settings of the VeriCube in the menue
Configure
Others
Device system