GDS XDI-XDIwin/30J Technical Sheet
Read Online
Read the GDS XDI-XDIwin/30J Technical Sheet online using the built-in PDF viewer.
Key Pages of Document
E: [email protected] T: 0113 286 0166 F: 0113 287 8178 GDS TECHNOLOGIES LTD | FUSION POINT | ASH LANE | GARFORTH | LEEDS | LS25 2GA | UK www.gds-technologies.co.uk Technical Sheet ref C1518B XDI-XDIwin / 30J PID SENSOR Set-up procedure: 292 D6C Issue Bv5 • NEW SENSORS ARE SUPPLIED READY TO CONNECT AND GO BUT NEED 15 MINUTES POWER STABILISATION. The following procedure is for full set up using a PC but onsite calibr ...
SPACE to set and exit. 12. Before using span gas, ensure that an autozero is not about to take place. This is indicated by the MPU LED flashing at 1 Hz. The MPU LED on the PCB will flash normally at 1 Hz but when there is less than 30 seconds to an autozero flash rate will increase to 2Hz i.e faster flash. 13. Apply a test gas to the senso ...
PCB UP and DOWN push buttons. LOW ALARM is the next menu and left and right magnets increase and decrease this value. HIGH ALARM is next followed by OVER RANGE alarm. The direction of the alarms is displayed as ^ for rising and v for falling but these can be changed using left and right magnets together. Appendix 1 Hyperterminal Display ...
E: [email protected] T: 0113 286 0166 F: 0113 287 8178 GDS TECHNOLOGIES LTD | FUSION POINT | ASH LANE | GARFORTH | LEEDS | LS25 2GA | UK www.gds-technologies.co.uk This document is not contractual and the equipment specification may be modified at any time without prior notice. Technical Sheet ref C1518B XDI-XDIwin / 30J PID SENSOR Set-up procedure: 292 D6C Issue Bv5 Figure 1 Figure 2 Figure 3 Appendix 2 The PID sens ...