1.0 INTRODUCTION
The Checkline model TI-25LTX Ultrasonic Thickness Gauge accurately measures the
thickness of steel. This gauge uses the “pulse-echo” principle of ultrasonic testing where
a short ultrasonic signal is transmitted from the probe. The signal travels through the
measurement sample until it is reflected back towards the probe from
– 3 –
2.0 KEYPAD, MENU, DISPLAY & CONNECTORS
The Keypad
2.1 ON/OFF/ENTER Key
The ON/OFF/ENTER key powers the unit ON or OFF. Since the same key is
also used as an ENTER key, the gauge is powered off by pressing and holding
down the key until the unit powers off.
Once the gauge is initially powered on, this key will function as the ENTER key, similar
t
– 4 –
2.3 LIGHT Key
The LIGHT key accesses the backlight setting of the LCD display. The backlight
has three setting options; ON, OFF, AUTO. The auto option will only illuminate
the display when the TI-25LTX is measuring, or receiving an echo. If either ON or
AUTO are selected, there are three brightness options, LO, MED, HI, to select a
preferred ove
– 5 –
D. Stability/Repeatability Indicator: This is used in conjunction with the thickness
measurement as a reference for the validity of the measurement. The TI-25LTX
takes multiple measurements per second, and when all the vertical bars are
illuminated, it’s a reference that the same thickness value is reliably being measured
multiple times per seco
– 6 –
The transducer must be used correctly in order for the TI-25LTX to produce accurate,
reliable measurements. Below is a short description of the transducer, followed by
instructions for its use.
This is a bottom view of a typical transducer. The two semicircles of the wear face
are visible, as is the barrier separating them. One of the semicircles i
– 7 –
beneath the transducer, and that the transducer is seated flat against the material. If the
condition persists, it may be necessary to select a different transducer (size or frequency)
for the material being measured. See section 4.1 for information on transducer selection.
While the transducer is in contact with the material that is being measured