Pintek DP-35 使 用 說 明 書
HIGH VOLTAGE DIFFERENTIAL PROBE
差 動 測 試 棒
DP-35
Pintek DP-35 TABLE OF CONTENTS
FEATURES--------------------------------------------------------------- 5
SPECIFICATIONS-------------------------------------------------------- 5
PANEL DESCRIPTION------------------------------------------------------6
OPERATING ENVIRONMENTAL CONDITIONS----------------7
OPERATING PROCEDURE------------------------------------------7
EXT. POWER SOURMAINTENANCE---------------------------------
Read the instructions before using the instrument:
1. Must acquire a differential voltage probe & get the best service from
instrument.
2. Read carefully the USER MANUAL.
3. Respect the safety precautions.
SAFETY PRECAUTIONS
Warning, Risk of Electric Shock,
Respect the max input voltages.
1. Max differential voltage: 1600 V (DC + AC peak) or 560V RMS
2. Max voltage between each input terminal and ground: 600 V RM
Do no
The DP-35 differential voltage probe provides a safety means of
measuring floating potentials for all models of oscilloscopes incomplete
safety.
It converts the high differential voltage (≦ 1600 peak) into a low
voltage (≦ 8.0 V) with reference to the earth for display on the
oscilloscopes.
The BNC output is designed to operate on an input with an impedance
of 1 MΩ. It is 2 times of the 50 Ω.
Recommend to use PINTEK PL-1
(8) Output Impedance: 50 Ω
(9) Rise Time: 10 ns
(10) Rejection Rate on Common Mode:
60 Hz: >80 dB ; 100 Hz: >60 dB ; 1 MHz: >50 dB
(11) Power Supply:
External 5V DC power supply.
(12) Consumption: 300 mA about [ 5 V / 2 A DC(100~240V : ASW-01) ]