Vapotherm 2000i Operating Instructions Manual

Text of the manual

Section 1 Indications and Contraindications

Primary Indications: Used to warm and humidify breathing gases, generally prescribed during oxygen therapy where concentrations of oxygen greater than ambient air are utilized to treat symptoms and manifestations of hypoxia including: • Documented hypoxemia: decreased Pa02 in blood below normal range • Acute ca

Section 2 Warnings and Cautions

General Warnings (cont.) • The Vapotherm® 2000i must be disinfected between each patient use and after 30 days of use on the same patient. • Gas flow delivered by this device is limited to 40 liters per minute. Maximum operational flow rate should not be exceeded. • Prior to use machine should be positioned and secured

Operating Instruction Manual

Page 7 Mounting pole Flowmeter Air line Oxygen line Vapotherm 2000i Delivery tube Sterile Water Bag Simple, easy to use touch control panel and information screen. Vapor transfer cartridge Note: System specification and front panel layout may vary. VSS-1 Spike set Air/Oxygen Blender Section 4 Set-up 4.1 The 2000i Unit 1. The ba

Operating Instruction Manual

Page 9 Fig. 4 Fig. 5 Fig. 6 Section 4 Set-up 4.5 Vapotherm Spike Set (VSS-1): for connecting Sterile Water Bag 1. Hang a sterile water bag from IV Pole. 2. Connect VSS-1 to the water inlet port on back of unit and make sure it locks into place (Fig. 5). 3. Ensure the VSS-1 is clamped then remove spike cap. Wipe spike with disinf

Section 5 Operation of The Vapotherm™ 2000i

5.5 Setting the Temperature and Warm-Up 1. The Vapotherm 2000i displays the actual temperature of the circulating water. Press and release the up or down arrow on the front of the unit to display temperature setting for 3 seconds. 2. To adjust the temperature setting of the Vapotherm 2000i, press and hold the u

Operating Instruction Manual

Page 13 Alarm Cause Action indication NOTE: To restart after a system failure, the unit must be reset by a momentary pressure on the Power button. Do not hold the Power button. The alarm will shut off after a delay of about a second, and the unit can then be restarted normally. S S Water is not filling system properly Low Water P

Operating Instruction Manual

Page 15 Fig. 1 Fig. 3 Section 6 Alarms, Trouble Shooting and Component Change-Outs 6.3 Component Change Outs WARNINGS: The vapor transfer cartridge, patient delivery tube, and VSS-1 spike set are all single use only and should be discarded after removal from the Vapotherm 2000i. The Patient Delivery Tube, VSS-1 and Vapor Transfe

Operating Instruction Manual

Page 17 Fig. 1 Section 8 Routine Disinfecting Protocol 8.1 Disinfection Supplies 1. DK-301 (a-e included in kit) a. Disinfection Bag A b. Disinfection Bag B c. Disinfection Tube d. Cartridge Bypass Tubes e. Y-Spike Assembly 2. Gloves 3. Safety Glasses 4. Disinfectant wipes, 70-90% isopropyl alcohol 5. Approved Disinfectant 6. 10

Section 8 Routine Disinfecting Protocol

8.4 Set-up (cont.) 6. Hang Bag A with 200 ml of approved disinfecting solution from IV pole hook. 7. Wipe blue end of disinfection tube with an approved disinfectant wipe. Insert into bottom port of Vapotherm™ 2000i system. Press firmly into place, rotate 1/4 turn clockwise, and pull down slightly to lock in positio