Section 1 Indications and Contraindications
Primary Indications:
Used to warm and humidify breathing gases, generally prescribed during oxygen therapy where concentrations
of oxygen greater than ambient air are utilized to treat symptoms and manifestations of hypoxia including:
• Documented hypoxemia: decreased Pa02 in blood below normal range
• Acute ca
Section 2 Warnings and Cautions
General Warnings (cont.)
• The Vapotherm® 2000i must be disinfected between each patient use and after 30 days of use on the
same patient.
• Gas flow delivered by this device is limited to 40 liters per minute. Maximum operational flow rate
should not be exceeded.
• Prior to use machine should be positioned and secured
Operating Instruction Manual
Page 7
Mounting pole
Flowmeter
Air line
Oxygen line
Vapotherm
2000i
Delivery tube
Sterile
Water
Bag
Simple, easy
to use touch
control panel
and information
screen.
Vapor transfer
cartridge
Note: System specification
and front panel layout
may vary.
VSS-1
Spike set
Air/Oxygen
Blender
Section 4 Set-up
4.1 The 2000i Unit
1. The ba
Operating Instruction Manual
Page 9
Fig. 4
Fig. 5
Fig. 6
Section 4 Set-up
4.5 Vapotherm Spike Set (VSS-1): for
connecting Sterile Water Bag
1. Hang a sterile water bag from IV Pole.
2. Connect VSS-1 to the water inlet port on back of unit and make
sure it locks into place (Fig. 5).
3. Ensure the VSS-1 is clamped then remove spike cap. Wipe spike
with disinf
Section 5 Operation of The Vapotherm™ 2000i
5.5 Setting the Temperature and Warm-Up
1. The Vapotherm 2000i displays the actual temperature of the circulating water. Press and release the
up or down arrow on the front of the unit to display temperature setting for 3 seconds.
2. To adjust the temperature setting of the Vapotherm 2000i, press and hold the u
Operating Instruction Manual
Page 13
Alarm
Cause
Action
indication
NOTE: To restart after a system failure, the unit must be reset by a momentary pressure on the Power
button. Do not hold the Power button. The alarm will shut off after a delay of about a second, and
the unit can then be restarted normally.
S
S
Water is not filling system properly
Low Water P
Operating Instruction Manual
Page 15
Fig. 1
Fig. 3
Section 6 Alarms, Trouble Shooting and Component Change-Outs
6.3 Component Change Outs
WARNINGS:
The vapor transfer cartridge, patient delivery tube, and VSS-1
spike set are all single use only and should be discarded after
removal from the Vapotherm 2000i.
The Patient Delivery Tube, VSS-1 and Vapor Transfe
Operating Instruction Manual
Page 17
Fig. 1
Section 8 Routine Disinfecting Protocol
8.1 Disinfection Supplies
1. DK-301 (a-e included in kit)
a.
Disinfection Bag A
b. Disinfection Bag B
c.
Disinfection Tube
d. Cartridge Bypass Tubes
e.
Y-Spike Assembly
2. Gloves
3. Safety Glasses
4. Disinfectant wipes, 70-90% isopropyl alcohol
5. Approved Disinfectant
6. 10
Section 8 Routine Disinfecting Protocol
8.4 Set-up (cont.)
6. Hang Bag A with 200 ml of approved disinfecting solution from IV
pole hook.
7. Wipe blue end of disinfection tube with an approved disinfectant
wipe. Insert into bottom port of Vapotherm™ 2000i system. Press
firmly into place, rotate 1/4 turn clockwise, and pull down slightly
to lock in positio