MRC DW-6163 Operation Manual

Text of the manual

6000 W x 1 W, PF, ACV, ACA

WATT METER Model : DW-6163 Your purchase of this WATT METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and delicate instrument, its durable structure will allow many years of use if p r o p e r o p e r a t i n g t e c h n i q u e s a r e developed. Please read t h e f o l

* Risk of electric shock !

Caution : * Do not apply the overload voltage, current to the input terminal ! * Remove test leads before open the battery cover ! * Cleaning - Only use the dry cloth to clean the plastic case ! Environment Conditions * Installation Categories III-600V. * Pollution Degree 2. * Altitude up to 2000 meters. *

TABLE OF CONTENTS

1. FEATURES....................................................................... 1 2. SPECIFICATIONS........................................................... 1 2-1 General specifications.............................................. 1 2-2 Electrical specifications................................................ 2 3. FRONT PANEL DESCRIPT

* Professional precision WATT meter with 1 W resolution

digital display , battery operated. * LSI - circuit provides high reliability and durability. * Directly operation. * Measurement : WATT ( AC ) : 6000W x 1W. ACV : 600V x 0.1 V . ACA : 10 A x 10 m A . PF : 1.00 x 0.01 * True Power and Wide range 0 W to 6000 Watt. * True RMS measurement of A

Freeze the display reading .

Memory Recall Maximum & Minimum value . Data output RS232 / USB PC Computer interface. * Connect the optional RS232 cable UPCB - 02 will get the RS232 plug. * Connect the optional USB cable USB - 01 will get the USB plug. Operation 0 to 50 ( 32 to 122 ). ℃ ℃ ℉ ℉ Temperature Operation Less than 80% RH Humidity

V/A ( TRUE RMS )

Range Resolution Accuracy ACV 600 V 0.1 V ± ( 0.3 % + 0.3 V ) ACA 10 A 0.01 A ± ( 0.3 % + 0.03 A ) * Auto range. * ACV, ACA accuracy is test under input signal is sine wave, 50/ 60 Hz. * ACV ACA frequency response is from 40 to 400 Hz. * Max. input value, AC voltage : 600V, AC current : 10 A. * Accuracy @ 23 ± 5 ℃ ℃ PF ( Power fact

3. FRONT PANEL DESCRIPTION

Fig. 1 3-1 Display 3-2 POWER / Backlight button 3-3 FUNCTION button 3-4 HOLD button 3-5 RECORD button ( MAX. / MIN. ) 3-6 LOAD terminal 3-7 POWER SOURCE terminal 3-8 RS-232 output terminal 3-9 DC 9V Power Adapter input socket 3-10 Battery Cover / Compartment 3-11 STAND 4

4. MEASURING PROCEDURE

4-1 Power ON/OFF 1)Press the POWER button ( 3-2, Fig. 1 ) more than 1.5 sec to Turn ON the meter. 2)Press the POWER button ( 3-2, Fig. 1 ) more than 1.5 sec to Turn OFF themeter. 4-2 ACV measurement procedure 1)Press the POWER button ( 3-2, Fig.1 ) more than 1.5 sec to Turn ON the meter. 2)Push the FUNCTION button ( 3-3, Fig. 1 ) once

4-4 Watt measurement procedure

1)Press the POWER button ( 3-2, Fig. 1 ) more than 1.5 sec to Turn ON the meter. 2)Make sure your equipment work voltage, such as 110 V or 220 V. 3)Turn off the POWER SOURCE. 4)Plug in the LOAD ( equipment ) to V terminal and COM terminal. 5)Plug in the power source to WATT terminal and 10A terminal. 6)Push the FUNCTION ( 3-3,