The DT-47FT is Digatron’s fuel indicator that performs two electrical tests for preliminary screening of

fuels. This is done by comparing the dielectric constant and the direct current conductivity of a competitor’s fuel to a known base. Fuel with added oxidizers will cause the instrument to register more positive than the base. This indicates a possible additive to the competitor’s fuel. This instrument is an excellent first screenin

A. Test Mode

When your instrument is first turned on, it is in Test mode and ready to perform fuel test #1 (FT1). This is the mode where all the fuel tests are performed. The displayed outputs of both fuel test 1 and 2 are non-linear scales which give very sensitive readings near zero and then expand to lower resolution as the number gets larger. This allows a large range to be covered while maintaining good sensitivity in the main area of interest. The

Fuel Test 2: FT2 measures the direct current (DC) conductivity of the fuel. Keeping the sensor in the fuel,

press and hold the DC button to perform FT2. The results of the test will be displayed while the button is held down. A 2 will also appear on the left of the display to signify test 2. FT2 should read 0 to +1. A reading greater than +1 indicates possible additives in the fuel. The more positive the number, the greater the possibility that the fuel

When you are finished calibrating FT1, press the Next button to calibrate FT2 or press the Exit button to

return to Test mode. Manual Calibration of FT2 (optional): In Calibration mode, press the Next button to calibrate FT2. Be sure the sensor is covered by the Cyclohexane and let the sensor stabilize (the numbers on the display of your instrument will stop changing). Use the arrow buttons to set the base number to the desired number, we recommend 0. To

Digatron DT-47FT The DT-47FT Uses One AAA Battery

The DT-47FT is powered by one AAA battery, which allows the instrument to run for about 500 hours. LOB will be displayed when the batteries are within a few hours of going dead. Replacing the battery is easy. Remove the battery door on the back of the instrument. When replacing the battery, be sure to observe battery polarity. Button Functions Defined DC/← Calibration: Decreases the value of the calibration number being