Apera Instruments EC20 Instruction Manual
Text of Apera Instruments EC20 Instruction Manual
EC20 Pocket Conductivity Tester
Instruction Manual
APERA INSTRUMENTS (Europe) GmbH
www.aperainst.de
v2.2
EC20 Pocket Conductivity Tester Instruction Manual
1. Battery Installation
Please install batteries according to the following steps. *Please note the correct
direction of battery installation:
The Positive Side (“+”) OF EVERY SINGLE Battery MUST FACE UP.
(WRONG INSTALLATION OF BATTERIES WILL CAUSE DAMAGE TO THE
TESTER AND POTENTIAL HAZARDS!)
2. Keypad Functions
■Short press------- < 2 seconds
■Long press-------- > 2 seconds
1. Short press to turn
3. Complete Kit
(batteries are pre-installed)
4. Calibration
4.1 Rinse the probe in distilled water and dry it with a clean
tissue. Short press to turn on the tester.
4.2 Long press to enter calibration mode; Short press
to exit.
4.3 Dip the probe into 1413 µS/cm calibration solution. Stir gently, leave it to
stand, LCD displays (Diagram 3), short press to complete 1-point
calibration. The tester returns to measuri
5. Conductivity Measurement
5.1 Short press to turn on tester. Rinse probe in distilled water and dry it
with a clean tissue to remove excess water.
5.2 Stir the probe in the sample solution gently, leave it to stand. Get readings
after the smile icon comes up and stays.
5.3 Rinse off the tester thoroughly in distilled water after each test.
6. Notes
6.1 If it’s first time use or the tester hasn’t been used for a long time, we
recommend soaki
6.4 The tester has already been calibrated after manufacture. Usually users can
use the tester right away, or test it in the calibration solutions to test its
accuracy. When error is large, calibrate it before using.
7. Parameter Setting
7.1 Parameter setting reference chart:
Symbol
Parameter Setting content
Code
Factory Default
P1
Select conductivity ranges
Aut-µS-mS
Aut
P2
Select Temperature Unit
°C - °F
°C
P3
Restore to factory default
No – Yes
No
8. Technical Specifications
Conductivity
Range
0 – 200.0 µS/cm, 0 – 2000 µS/cm, 0 –
20.00 mS/cm
Resolution
0.1/1 µS/cm, 0.01 mS/cm
Accuracy
±1% F.S±1 digit
Calibration points
1-2 points auto calibration
Automatic Temp.
Compensation
0-50˚C
Temp.
Range
0-50˚C
Resolution
0.1˚C
Accuracy
±0.5˚C
9. Other Functions & Parameters
Indication of
calibration points
Auto Power-
off
Power-off in 8
minutes if no
operation
Indication of stable
measurements