Flomag 3000 Installation And Operation Manual

Text of the manual

CZ-602 00 Brno

Czech Republic tel: +420 541212539 fax: +420 549240356 e-mail: [email protected] www.flomag.com Magnetic Flowmeter FLOMAG®3000 Installation and Operation

FLOMAG 3000 - Installation and Operation Manual

3 An magnetic flowmeter is used for volume flow measurement of electrically conductive liquids. Measurement principle is based o n F a r a d a y l a w o n electromagnetic induction. A sensor consists of a non- magnetic tube with non- conductive lining, measuring electrodes and two coils generating electromagnet

FLOMAG 3000 - Installation and Operation Manual

4 min 5 x DN min 3 x DN The flowmeter will give the best results when flow of liquid is steady. Therefore a few basic recommendations should be observed for its locating in a pipeline. There should be no transitions between the sensor and the adjacent pipeline that could be a source of turbulence. Correct axial

FLOMAG 3000 - Installation and Operation Manual

5 Correct function of the magnetic flowmeter requires perfect electrical connection between the sensor and the adjacent pipeline, grounding potential and the power supply protective wire. For the flanged sensor with the adjacent conducting pipeline, flanges should be electrically connected and the pipeline grou

FLOMAG 3000 - Installation and Operation Manual

6 Converter is capable to detect flow rates as low as 0.1 m/s. Upper limit is determined by capability of liquid to maintain continuous flow at higher velocities. This is usually true for flow rates up to 12 m/s. Measurement error rapidly increases for too low flow rates, as can be seen in the diagram. It shows

FLOMAG 3000 - Installation and Operation Manual

7 Volba výstelky snímače Block diagram of the flowmeter Main advantage of the m a g n e t i c f l o w m e t e r FLOMAG3000 is its significant variability. Flowmeter converter in basic version consists only of power supply, microcomputer and sensor input module (module 1). Display, outputs and other optional

FLOMAG 3000 - Installation and Operation Manual

8 For the remote sensor, there is a terminal box in its tapping (see figure 21). The sensor should be connected to the converter using a double shielded cable. You can use our special sensor cable PAAR- LiYCY-CY[1X(2X0,25 LiYCY) +1X(2X0,75 LiYCY)+1X0,75] CY (length up to 200 m) or Maximum length of the cable be