UNI-T UT30B Operating Manual

Brand: UNI-T
Category: Multimeter
Model: UT30B (Operating Manual)
File Type: PDF
File Size: 0.8 Mb
Pages: 32
Text of UNI-T UT30B Operating Manual

UNI-T UT30B B. Safety Information

C. Rules for safe operation D. International Electrical Symbols E. General Specifications F. Specification G. Operation plate H. Make Measurements I. Fuse and battery replacement J. Accessories Remark: The model of UT30B,UT30C,UT30D will be marked as UT30BL,UT30CL UT30DL,if they have backlight function. 1 Model UT30B/C/D/F: OPERATING MANUAL

UT30 series Multimeter is 3 1/2 digits with steady operations,fashionable

structure and highly reliable hand-held measuring instrument. The meter can measure DC/AC voltage, DC/AC current, Resistance, Frequency, Temperature, Diode, Transistor hFE, and Continuity. It is an ideal tool for maintenance. A. Introduction 2 Model UT30B/C/D/F: OPERATING MANUAL

This Meter complies with the standards IEC61010-1: in pollution degree 2, overvoltage

category (CAT I 600V, CAT II 300V) and double insulation. CAT. I: Signal level, special equipment or parts of equipment, telecommunication, electronic, etc., with smaller transient overvoltages than overvoltages CAT. II. CAT. II: Local level, appliance, PORTABLE EQUIPMENT etc., with smaller transient overvoltages than CAT. III Use the Meter only as specified in this opera

C.Rules for safe operation(1)

l Use the meter only as the rules specified in this manual, otherwise, the protection provided by the meter may be impaired. l Do not operate the meter unless the bottom case has been closed as terminal can carry voltage. l Inspect the insulation of the test leads and make sure there is no damage to the test leads before using the meter. l As soon as the battery indicator’s“ ” appear, replace the batt

Model UT30B/C/D/F: OPERATING MANUAL

l When measuring voltage higher than DC 60V or AC 30Vrms, pay extra attention to avoid electric shock. l Make sure to replace right type and right rating fuse. l Do not operate or store the Meter under high temperature or humid condition. l Do not change internal circuit to avoid damages to the meter and danger to the user. l Periodically wipe the case with damp cloth and mild detergent. Do not u

UNI-T UT30B D.International Electrical Symbols

Low Battery Earth Ground Warning Double Insulation AC current Diode DC current Buzzer Fuse Conforms to Standards of European Union. AC or DC 6 Model UT30B/C/D/F: OPERATING MANUAL
Model UT30B/C/D/F: OPERATING MANUAL E.General Specifications(1) 1. The maximum voltage, between any terminal and earth, is 600Vrms. A . The“COM”input terminal is always connected with the black test lead. B . The“VΩmA”input terminal is always connected with the red test lead and is used to measure voltage up to 500V,resistance and current up to 200mA. C . The“10A MAX”input termin

8 Relative humidity:Max.relative humidity 80% for temperature up to

31 oC decreasing linearly to 50% relative humidity at 40 oC. 9. Battery: 9V NEDA 1604 or 6F22 or 006P. 10. Low Battery Indication: Display “ ”.on LCD. 11. Dimension:75mmx130mmx36mm 12.
Model UT30B/C/D/F: OPERATING MANUAL F.Specification(1) Accuracy: (a% reading + b digit), which guarantee one year. Environmental Temperature:23 oC 5 oC Relative Humidity: <75% Input Impedance:10MΩ for all the ranges. Overload protection: At 200mV range, it’s protected at 230V(AC / DC Current), others are protected at 500V(AC or DC). Range Resolution Accuracy 200mV 2000mV(2V) 20V 200V 500V 100µV 1mV 10mV 100mV 1V UT30B UT30C

UNI-T UT30B 2. AC Voltage

Input Impedance: (approx. 5MΩ ) of UT30B\C\D, but all ranges of UT30F are 10MΩ . Frequency: 40-400Hz Display: RMS of Sine Wave Value (Average Value) Overload protection: At 200mV range, it is protected at 230V(AC / DC Current), others are protected at 500V(AC or DC ) Range Resolution Accuracy 200mV 2V 20V 200V 500V 100µV 1mV 10mV 100mV 1V UT30B UT30C UT30D UT30F (1.2%+10) (1.2%+

UNI-T Multimeter Owners Instructions: