Chroma 63200 Series Operation & Programming Manual
Read Online
Read the Chroma 63200 Series Operation & Programming Manual online using the built-in PDF viewer.
Text of the manual
DC Electronic Load
63200 Series
Operation & Programming
Manual
Version 1.9
June 2006
P/N A11 000599
The information in this document is subject to change without notice.
Chroma ATE INC. makes no warranty of any kind with regard to this manual, including, but
not limited to, the implied warranties of merchantability and fitness for a particular purpose.
Chroma ATE INC. shall not be held liable for errors contained herein or direct, indirect,
special, incidental or consequential damages in connection with the furnishing, performance,
or use of this material.
CHROMA ATE INC.
66 Hwa-Ya 1st Rd., Hwa-Ya Technical Park, Kuei-Shan Hsiang, Taoyuan Hsien, Taiwan
Copyright Notices. Copyright 2003-2006 Chroma ATE INC., all rights reserved.
Reproduction, adaptation, or translation of this document without prior written permission is
prohibited, except as allowed under the copyright laws.
ii
All Chroma instruments are warranted against defects in material and workmanship for a
period of one year after date of shipment. Chroma agrees to repair or replace any assembly
or component found to be defective, under normal use during this period. Chroma’s
obligation under this warranty is limited solely to repairing any such instrument, which in
Chroma’s sole opinion proves to be defective within the scope of the warranty when returned
to the factory or to an authorized service center. Transportation to the factory or service
center is to be prepaid by purchaser. Shipment should not be made without prior authorization
by Chroma.
This warranty does not apply to any products repaired or altered by persons not authorized by
Chroma, or not in accordance with instructions furnished by Chroma. If the instrument is
defective as a result of misuse, improper repair, or abnormal conditions or operations, repairs
will be billed at cost.
Chroma assumes no responsibility for its product being used in a hazardous or dangerous
manner either alone or in conjunction with other equipment. High voltage used in some
instruments may be dangerous if misused. Special disclaimers apply to these instruments.
Chroma assumes no liability for secondary charges or consequential damages and in any
event, Chroma’s liability for breach of warranty under any contract or otherwise, shall not
exceed the purchase price of the specific instrument shipped and against which a claim is
made.
Any recommendations made by Chroma for use of its products are based upon tests believed
to be reliable, but Chroma makes no warranty of the results to be obtained. This warranty is
in lieu of all other warranties, expressed or implied, and no representative or person is
authorized to represent or assume for Chroma any liability in connection with the sale of our
products other than set forth herein.
CHROMA ATE INC.
66 Hwa-Ya 1st Rd., Hwa-Ya Technical Park,
Kuei-Shan Hsiang, Taoyuan Hsien, Taiwan
Tel: 886-3-327-9999
Fax: 886-3-327-2886
http://www.chromaate.com
iii
For the following equipment:
Product Name: DC Electronic Load
Model Name:63201, 63202, 63203, 63204, 63205, 63206, 63207, 63208, 63209 ,63210
Manufacturer’s Name: Chroma ATE Inc.
Manufacturer’s Address:
66 Hwa-Ya 1st Rd., Hwa-Ya Technical Park,
Kuei-Shan Hsiang, Taoyuan Hsien, Taiwan
is herewith confirmed to comply with the requirements set out in the Council
Directive on the Approximation of the Laws of the Member States Relating to
Electromagnetic Compatibility (89/336/EEC) and electrical equipment designed
for use within certain voltage limits (73/23/EEC;93/68/EEC)
For electromagnetic compatibility, the following standards were applied:
EMC:
EN61326-1(1997):CISPR22:1994+A1 Class A
IEC 1000-3-2:1995 Harmonics Current
IEC 1000-3-3:1995 Voltage Fluctuations
IEC 1000-4-2:1995 Electrostatic Discharge
IEC 1000-4-3:1995 Radio-Frequency Electromagnetic Field
IEC 1000-4-4:1995 Fast Transient Burst
IEC 1000-4-5:1995 Surge Immunity test
IEC 1000-4-6:1995 Immunity To Conducted Disturbances, Induced
By Radio Frequency Fields
IEC 1000-4-11:1994 Voltage Dips, Short Interruptions and Voltage
Variations Immunity Test
For safety requirement, the following standard was applied:
Safety:
EN61010-1(1993)+A2(1995)
Taiwan May 2005
Place
Date
Vice President, Engineering
Warning:
This is a class A product. In a domestic environment this product may cause radio
interference in which case the user may be required to take adequate measures.
iv
The following general safety precautions must be observed during all phases of operation,
service, and repair of this instrument. Failure to comply with these precautions or specific
WARNINGS given elsewhere in this manual will violate safety standards of design,
manufacture, and intended use of the instrument. Chroma assumes no liability for the
customer’s failure to comply with these requirements.
BEFORE APPLYING POWER
Verify that the product is set to match the available line voltage and the correct fuse is
installed.
PROTECTIVE GROUNDING
Make sure to connect the protective grounding to prevent an electric shock before turning on
the power.
NECESSITY OF PROTECTIVE GROUNDING
Never cut off the internal or external protective grounding wire, or disconnect the wiring of
protective grounding terminal. Doing so will cause a potential shock hazard that may bring
injury to a person.
FUSES
Only fuses with the required rated current, voltage, and specified type (normal blow, time
delay, etc.) should be used. Do not use repaired fuses or short-circuited fuse holders. To
do so could cause a shock or fire hazard. Caution: Be sure to remove the power cord before
changing the fuses.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes.
DO NOT REMOVE THE COVER OF THE INSTRUMENT
Operating personnel must not remove the cover of the instrument. Component replacement
and internal adjustment can be done only by qualified service personnel.
CLEANING THE INSTRUMENT
Please use a clean and dry cloth to clean the instrument surface.
v
DANGER – High voltage.
Explanation: To avoid injury, death of personnel, or damage to the
instrument, the operator must refer to an explanation in the instruction
manual.
Protective grounding terminal: To protect against electrical shock in
case of a fault. This symbol indicates that the terminal must be
connected to ground before operation of equipment.
WARNING
The WARNING sign denotes a hazard. It calls attention to a procedure,
practice, or the like, which, if not correctly performed or adhered to, could
result in personal injury. Do not proceed beyond a WARNING sign
until the indicated conditions are fully understood and met.
CAUTION
The CAUTION sign denotes a hazard. It calls attention to an operating
procedure, or the like, which, if not correctly performed or adhered to,
could result in damage to or destruction of part or all of the products. Do
not proceed beyond a CAUTION sign until the indicated conditions are
fully understood and met.
vi
The following lists the additions, deletions and modifications in this manual at each revision.
Date
Version Revised Sections
Mar. 2003
1.3
Update “CE Conformity Declaration” to add model
63205/63207/63209
Update “Specifications” for changes and addition of model
63205/63207/ 63209
Add the rear panel of Model 63205/630206/63207/63208/63209 in
“Rear Panel Description”
May 2003
1.4
Add a note to the “Specifications” for the fuse of 63205
Jan. 2004
1.5
Modify the “Specifications” in Chapter 1 for power rating changes for
all models.
Add a section of “Dimension Outline for 63200 Series” in Chapter 1.
Jan. 2005
1.6
Add the following sections:
− “Diagram of RS485 Parallel Connections” in Chapter 2.
− “Load Surge Capability” in Chapter 3.
− “Timer Function for Battery Discharge Testing” in Chapter 3.
Modify the following sections:
− “Specifications” in Chapter 1.
− “Protection Features” in Chapter 3.
Jun. 2005
1.7
Change the address of Chroma in CE-Conformity Declaration
Nov. 2005
1.8
Modify the “Specifications” in Chapter 1.
Jun. 2006
1.9
Modify the followings:
− “Specifications” to add a new model 63210.
− Description in the section of “Parallel Connections”.
− “Load Connections” in the section of “Application Connection”.
− Commands in the chapter of “Language Dictionary”.
− “Questionable Status” in the chapter of “Status Reporting”
− “Command Summary” in the chapter of “Index”.
Add “Installing the 63200 Protective Cover” in the section of
“Dimension Outline for 63200 Series”.
vii
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Table of Contents
1.
General Information................................................................................................ 1-1
1.1
Introduction........................................................................................................ 1-1
1.2
Description ......................................................................................................... 1-1
1.3
Overview of Key Features.................................................................................. 1-2
1.4
Specifications ..................................................................................................... 1-2
1.5
Dimension Outline for 63200 Series................................................................ 1-12
2.
Installation ................................................................................................................ 2-1
2.1
Introduction........................................................................................................ 2-1
2.2
Inspection ........................................................................................................... 2-1
2.3
Installing............................................................................................................. 2-1
2.3.1
Changing Line Voltage .............................................................................. 2-1
2.3.2
Turn-On Self-Test ...................................................................................... 2-2
2.4
Application Connection ..................................................................................... 2-3
2.4.1
Load Connections....................................................................................... 2-3
2.4.2
Remote Sensing Connections..................................................................... 2-5
2.4.3
Parallel Connections................................................................................... 2-5
2.4.4
Diagram of RS485 Parallel Connections ................................................... 2-6
2.5
Remote Control Connection............................................................................... 2-7
3.
Operation Overview................................................................................................. 3-1
3.1
Introduction........................................................................................................ 3-1
3.2
Front Panel Description...................................................................................... 3-1
3.3
Rear Panel Description....................................................................................... 3-2
3.4
Local/Remote Control........................................................................................ 3-4
3.5
Modes of Operation............................................................................................ 3-5
3.5.1
Constant Current Mode.............................................................................. 3-5
3.5.2
Constant Resistance Mode ......................................................................... 3-7
3.5.3
Constant Voltage Mode.............................................................................. 3-8
3.5.4
Constant Power Mode................................................................................ 3-8
3.5.5
Load Surge Capability................................................................................ 3-9
3.5.6
Timer Function for Battery Discharge Testing ........................................ 3-10
3.6
Measurements................................................................................................... 3-11
3.7
Slew Rate & Minimum Transient Time........................................................... 3-11
3.8
Start/Stop Sink Current .................................................................................... 3-12
3.9
Short On/Off..................................................................................................... 3-13
3.10
Load On/Off ..................................................................................................... 3-14
3.11
Protection Features........................................................................................... 3-14
3.12
Save/Recall Setting .......................................................................................... 3-16
3.13
Program............................................................................................................ 3-16
3.14
External Waveform Control............................................................................. 3-16
3.15
Voltage & Current Monitor.............................................................................. 3-16
4.
Local Operation........................................................................................................ 4-1
4.1
Introduction........................................................................................................ 4-1
4.2
Local Operation.................................................................................................. 4-1
4.2.1
Setting the Operation Mode ....................................................................... 4-4
4.2.2
Setting the Program.................................................................................. 4-10
ix
High Power DC Electronic Load 63200 Series Operation & Programming Manual
4.2.3
Running the Program ............................................................................... 4-12
4.2.4
Setting the Specification .......................................................................... 4-13
4.2.5
Setting the Configuration ......................................................................... 4-13
4.2.6
Recalling Files.......................................................................................... 4-17
4.2.7
Saving File/Default/Program.................................................................... 4-17
4.2.8
Going To Local ........................................................................................ 4-18
4.2.9
Setting System and RS-232C Connection................................................ 4-18
4.2.10
Online Change Level................................................................................ 4-20
5.
Basic Information for Programming...................................................................... 5-1
5.1
Introduction........................................................................................................ 5-1
5.1.1
Setting the GPIB Address and RS-232C Parameters................................. 5-1
5.1.2
Wire Connection of RS-232C .................................................................... 5-1
5.2
GPIB Capability of the Electronic Load ............................................................ 5-2
5.3
RS-232C in Remote Control .............................................................................. 5-3
6.
Introduction to Programming................................................................................. 6-1
6.1
Basic Definition.................................................................................................. 6-1
6.2
Numerical Data Formats .................................................................................... 6-2
6.3
Character Data Formats...................................................................................... 6-3
6.4
Separators and Terminators................................................................................ 6-3
7.
Language Dictionary................................................................................................ 7-1
7.1.1
Common Commands.................................................................................. 7-1
7.1.2
Common Command Dictionary ................................................................. 7-1
7.2
Specific Commands............................................................................................ 7-4
7.2.1
CONFigure Sub-system ............................................................................. 7-5
7.2.2
COMMunicate Sub-system........................................................................ 7-8
7.2.3
CURRENT Sub-system............................................................................ 7-10
7.2.4
FETch Sub-system ................................................................................... 7-13
7.2.5
LOAD Sub-system................................................................................... 7-15
7.2.6
MEASure Sub-system.............................................................................. 7-17
7.2.7
MODE Sub-system .................................................................................. 7-19
7.2.8
POWer Sub-system .................................................................................. 7-20
7.2.9
PROGram Sub-system ............................................................................. 7-21
7.2.10
RESistance Sub-system............................................................................ 7-25
7.2.11
SPECification Sub-system ....................................................................... 7-27
7.2.12
STATus Sub-system................................................................................. 7-29
7.2.13
VOLTage Sub-system.............................................................................. 7-31
7.2.14
SYSTem Sub-system................................................................................ 7-33
8.
Status Reporting....................................................................................................... 8-1
8.1
Introduction........................................................................................................ 8-1
8.2
Register Information in Common ...................................................................... 8-1
8.3
Questionable Status............................................................................................ 8-2
8.4
Output Queue ..................................................................................................... 8-3
8.5
Standard Event Status ........................................................................................ 8-3
8.6
Status Byte Register ........................................................................................... 8-4
8.7
Service Request Enable Register ....................................................................... 8-5
9.
INDEX....................................................................................................................... 9-1
x
High Power DC Electronic Load 63200 Series Operation & Programming Manual
9.1
Command Summary........................................................................................... 9-1
9.2
Parallel Operation............................................................................................... 9-4
xi
This manual contains specifications, installation, operation, and programming instructions of
63200 series high power electronic loads. All the machines are tested according to safety
standard EN61010-1: TYPE POLLUTION II and INSTALLTION CATEGORY II.
1.2 Description
The functions of the 63200 series loads are the same except the variations on input voltage,
load current, and power ratings. They can be operated independently in constant current
(CC) mode, constant resistance (CR) mode, constant voltage (CV) mode or constant power
(CP) mode.
VOLTS
AMPS
WATTS/OHMS
CP
CV
CR
CC
CURSOR
GO/NG
7
4
1
0
.
2
5
8
9
6
3
EDIT
16V/80V
MODEL
60A/600A
DC ELECTRONIC LOAD
63203
4KW
5.2KW
Figure 1-1 The Front Panel of the Electronic Load
There are three groups of keypads on the electronic load front panel shown in Figure 1-1.
They are System keypad, Function keypad and Entry keypad from left to right.
1-1
High Power DC Electronic Load 63200 Series Operation & Programming Manual
1.3
Overview of Key Features
� Local operation on front panel keypad.
� Remote control via GPIB, RS-232C interface or remote controller (optional).
� Photo-couple isolation supplies true floating Load.
� Automatic fan speed control to reduce audio noise.
� Constant current (CC), constant resistance (CR), constant voltage (CV) and constant
power (CP) operation modes.
� Programmable slew rate, load level, load period and conduct voltage (Von).
� Programmable dynamic loading with speed up to 20KHz.
� Minimum input resistance allows load to sink high current even with low input voltage.
� Selective voltage and current ranges.
� Remote sensing capability.
� 100 sets of memories to save/recall user-definable setups.
� 10 sets of programs to link files for automatic test.
� 15-bit A/D converter with precision measurement.
� Short circuit simulation.
� Automatic GO/NG inspection to confirm UUT within spec.
� Isolated voltage and current monitoring waveforms output.
� Isolated external Vdc reference input to control the load current.
1.4
Specifications
AC input:
115/230 interchangeable Vac line
Fuse:
3.15A, 250V *
Amplitude:
±10%
Frequency:
47 to 63 Hz
Maximum VA:
300VA
*
The fuses in 63201, 63202 are specified as 2A, 250V. The fuses in 63203, 63204,
63205, 63208, 63209 and 63210 are specified as 2.5A, 250V. The specifications of Load
are listed below.
� Notice
1.
The equipment is for indoor use only.
2.
The altitude up to 2000 meters is allowed to use the equipment.
3.
All specifications are tested under 20°C ∼ 30°C except otherwise stated.
4.
The range for operation temperature is 0°C ∼ 40°C.
5.
The relative humidity is from 10% to 90%.
6.
The specifications of DC current accuracy are tested after the input is applied for 30
seconds.
7.
The typical temperature coefficient is 100ppm.
1-2
This equipment is not intended for performing measurements on CAT I, II, III or IV.
CAT IV – is for measurements performed at the source of the low-voltage installation.
CAT III – is for measurements performed in the building installation.
CAT II – is for measurements performed on circuits directly connected to the low-voltage
installation.
CAT I – is for measurements performed on circuits directly connected to mains.
1-3
High Power DC Electronic Load 63200 Series Operation & Programming Manual
SPECIFICATION
Model
63201
63202
63203
Power*1
260W
2600W
260W
2600W
520W
5200W
Current
0-30A
0-300A
0-5A
0-50A
0-60A
0-600A
Voltage
1-80V
2.5-500V
1-80V
Min. operating voltage
1V @ 30A
1V @ 300A
2.5V @ 5A
2.5V @ 50A
1V @ 60A
1V @ 600A
Constant Current mode
Range
0-30A
0-300A
0-5A
0-50A
0-60A
0-600A
Resolution
7.5mA
75mA
1.25mA
12.5mA
15mA
150mA
Accuracy
0.1%+0.1%FS
0.2%+0.1%FS
0.1%+0.1%FS
0.2%+0.1%FS
0.1%+0.1%FS
0.2%+0.1%FS
Constant Resistance mode
Range
0.005-20ohm
0.25-1000ohm
0.25-1000ohm
10-40000ohm
0.0025-10ohm
0.125-500ohm
Resolution
12 bits
12 bits
12 bits
12 bits
12 bits
12 bits
Accuracy*2
0.6mho+0.35%
0.9mho+0.1%
0.012mho+0.35%
0.04mho+0.1%
1.2mho+0.35%
*4
1.2mho+0.1%
Accuracy*3 (Vin>7V) 0.6mho+0.35% 0.012mho+0.35% 0.012mho+0.35% 112.5u mho+0.35% 1.2mho+0.35% 0.024mho+0.35%
Constant Voltage mode
Range
1-16V
1-80V
2.5-125V
2.5-500V
1-16V
1-80V
Resolution
4mV
20mV
31mV
125mV
4mV
20mV
Accuracy
0.05%+0.1%FS
0.05%+0.1%FS
0.05%+0.1%FS
0.05%+0.1%FS
0.05%+0.1%FS 0.05%+0.1%FS
Constant Power mode
Range
0.6-260W
6-2600W
0.625-260W
6.25-2600W
1.2-520W
12-5200W
Resolution
7.5mW
75mW
3.125mW
31.25mW
22.5mW
225mW
Accuracy
0.5%+0.5%FS
0.5%+0.5%FS
0.5%+0.5%FS
0.5%+0.5%FS
0.5%+0.5%FS
0.5%+0.5%FS
Dynamic mode
Timing
T1&T2
0.025-10mS
1mS-30S
0.025-10mS
1mS-30S
0.025-10mS
1mS-30S
Resolution
1uS
1mS
1uS
1mS
1uS
1mS
Accuracy
1uS+100ppm
1mS+100ppm
1uS+100ppm
1mS+100ppm
1uS+100ppm
1mS+100ppm
Slew rate
5mA-1.25A/uS
50mA-12.5A/uS
0.8mA-0.2A/uS
8mA-2A/uS
10mA-2.5A/uS 100mA-25A/uS
Resolution
5mA/uS
50mA/uS
0.8mA/uS
8mA/uS
10mA/uS
100mA/uS
Current
Range
0-30A
0-300A
0-5A
0-50A
0-60A
0-600A
Resolution
7.5mA
75mA
1.25mA
12.5mA
15mA
150mA
Accuracy
0.4%FS
0.4%FS
0.4%FS
Measurement
Voltage read back
Range
0-16V
0-80V
0-125V
0-500V
0-16V
0-80V
Resolution
15 bits
15 bits
15 bits
15 bits
15 bits
15 bits
Accuracy
0.05%+0.05%FS
0.05%+0.05%FS
0.05%+0.05%FS
1-4
Current read back
Range
0-30A
0-300A
0-5A
0-50A
0-60A
0-600A
Resolution
15 bits
15 bits
15 bits
15 bits
15 bits
15 bits
Accuracy
0.1%+0.1%FS
0.1%+0.1%FS
0.1%+0.1%FS
Power read back
Range
0-260W
0-2600W
0-260W
0-2600W
0-520W
0-5200W
Resolution
15 bits
15 bits
15 bits
15 bits
15 bits
15 bits
Accuracy
0.3%+0.3%FS
0.3%+0.3%FS
0.3%+0.3%FS
General
Short circuit
Current
30A
300A
5A
50A
60A
600A
SIZE (mm)
440(W) X 177(H) X 644(D)
440(W) X 177(H) X 644(D)
440(W) X 353(H) X 644(D)
WEIGHT
35Kg
35Kg
58Kg
EMC & Safety
CE
CE
CE
*Specifications are subject to change without notice.
1. The power rating specifications at ambient temperature = 25°C and see the diagram below for power derating.
Power Rating
100%
77%
°C
Ambient Temperature
10°C 20°C 30°C 40°C
2. The Vin must be greater than min. operating voltage of each model.
3. The Vin must be greater than 7V of each model.
All CR mode accuracy should be referring to the CC mode accuracy.
4. Setting error will be 1% for R < 0.005Ω at CRL range.
1-5
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Model
63204
63205
63206
Power*1
520W
5200W
650W
6500W
1040W
10400W
Current
0-10A
0-100A
0-18A
0-180A
0-60A
0-600A
Voltage
2.5-500V
1-80V
1-80V
Min. operating voltage
2.5V @ 10A
2.5V @ 100A
1V @ 18A
1V @ 180A
1V @ 60A
1V @ 600A
Constant Current mode
Range
0-10A
0-100A
0-18A
0-180A
0-60A
0-600A
Resolution
2.5mA
25mA
4.5mA
45mA
15mA
150mA
Accuracy
0.1%+0.1%FS
0.2%+0.1%FS
0.1%+0.2%FS
0.1%+0.2%FS
0.1%+0.2%FS 0.1%+0.2%FS
Constant Resistance mode
Range
0.125-500ohm
5-20000ohm
0.008-32ohm
0.4-1600ohm
0.0025-10ohm 0.125-500ohm
Resolution
12 bits
12 bits
12 bits
12 bits
12 bits
12 bits
Accuracy*2
0.024mho+0.35% 0.08mho+0.1% 0.375mho+0.35%
0.75mho+0.1%
1.2mho+0.35%
*4
1.2mho+0.1%
Accuracy*3 (Vin>7V) 0.024mho+0.35% 225u mho+0.35% 0.375mho+0.35% 0.075mho+0.35% 1.2mho+0.35% 0.024mho+0.35%
Constant Voltage mode
Range
2.5-125V
2.5-500V
1-16V
1-80V
1-16V
1-80V
Resolution
31mV
125mV
4mV
20mV
4mV
20mV
Accuracy
0.05%+0.1%FS 0.05%+0.1%FS 0.05%+0.1%FS
0.05%+0.1%FS
0.05%+0.1%FS 0.05%+0.1%FS
Constant Power mode
Range
1.25-520W
12.5-5200W
0.36-650W
3.6-6500W
1.2-1040W
12-10400W
Resolution
6.25mW
62.5mW
4.6mW
46mW
22.5mW
225mW
Accuracy
0.5%+0.5%FS
0.5%+0.5%FS
0.5%+0.5%FS
0.5%+0.5%FS
0.5%+0.5%FS 0.5%+0.5%FS
Dynamic mode
Timing
T1&T2
0.025-10mS
1mS-30S
0.025-10mS
1mS-30S
0.025-10mS
1mS-30S
Resolution
1uS
1mS
1uS
1mS
1uS
1mS
Accuracy
1uS+100ppm
1mS+100ppm
1uS+100ppm
1mS+100ppm
1uS+100ppm
1mS+100ppm
Slew rate
1.6mA-0.4A/uS
16mA-4A/uS
3mA-0.75A/uS
30mA-7.5A/uS
12mA-3A/uS
100mA-25A/uS
Resolution
1.6mA/uS
16mA/uS
3mA/uS
30mA/uS
12mA/uS
100mA/uS
Current
Range
0-10A
0-100A
0-18A
0-180A
0-60A
0-600A
Resolution
2.5mA
25mA
4.68mA
46.8mA
15mA
150mA
Accuracy
0.4%FS
0.4%FS
0.4%FS
Measurement
Voltage read back
Range
0-125V
0-500V
0-16V
0-80V
0-16V
0-80V
Resolution
15 bits
15 bits
15 bits
15 bits
15 bits
15 bits
Accuracy
0.05%+0.05%FS
0.05%+0.05%FS
0.05%+0.05%FS
1-6
Current read back
Range
0-10A
0-100A
0-18A
0-180A
0-60A
0-600A
Resolution
15 bits
15 bits
15 bits
15 bits
15 bits
15 bits
Accuracy
0.1%+0.1%FS
0.1%+0.1%FS
0.1%+0.1%FS
Power read back
Range
0-520W
0-5200W
0-650W
0-6500W
0-1040W
0-10400W
Resolution
15 bits
15 bits
15 bits
15 bits
15 bits
15 bits
Accuracy
0.3%+0.3%FS
0.3%+0.3%FS
0.3%+0.3%FS
General
Short circuit
Current
10A
100A
18A
180A
60A
600A
SIZE (mm)
440(W) X 353(H) X 644(D)
440(W) X 310(H) X 644(D)
440(W) X 443.7(H) X 644(D)
WEIGHT
58Kg
64Kg
90Kg
EMC & Safety
CE
CE
CE
*Specifications are subject to change without notice.
1. The power rating specifications at ambient temperature = 25°C and see the diagram below for power derating.
Power Rating
100%
77%
°C
Ambient Temperature
10°C 20°C 30°C 40°C
2. The Vin must be greater than min. operating voltage of each model.
3. The Vin must be greater than 7V of each model.
All CR mode accuracy should be referring to the CC mode accuracy.
4. Setting error will be 1% for R < 0.005Ω at CRL range.
1-7
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Model
63207
63208
63209
Power*1
1040W
10400W
1560W
15600W
1560W
15600W
Current
0-30A
0-300A
0-60A
0-600A
0-100A
0-1000A
Voltage
1-80V
1-80V
1-80V
Min. operating voltage
1V @ 30A
1V @ 300A
1V @ 60A
1V @ 600A
1V @ 100A
1V @ 1000A
Constant Current mode
Range
0-30A
0-300A
0-60A
0-600A
0-100A
0-1000A
Resolution
9.3mA
75mA
15mA
150mA
31.25mA
250mA
Accuracy
0.1%+0.2%FS 0.1%+0.2%FS
0.1%+0.2%FS
0.1%+0.2%FS
0.1%+0.2%FS
0.1%+0.2%FS
Constant Resistance mode
Range
0.005-20ohm 0.25-1000ohm
0.0025-10ohm
0.125-500ohm
0.0015-6ohm
0.075-300ohm
Resolution
12 bits
12 bits
12 bits
12 bits
12 bits
12 bits
Accuracy*2
0.6mho+0.35% 0.9mho+0.1%
1.2mho+0.35% *4
1.2mho+0.1%
1.2mho+0.35% *4
1.2mho+0.1%
Accuracy*3 (Vin>7V) 0.6mho+0.35% 0.012mho+0.35% 1.2mho+0.35%
0.024mho+0.35%
1.2mho+0.35%
0.024mho+0.35%
Constant Voltage mode
Range
1-16V
1-80V
1-16V
1-80V
1-16V
1-80V
Resolution
4mV
20mV
4mV
20mV
4mV
20mV
Accuracy
0.05%+0.1%FS 0.05%+0.1%FS
0.05%+0.1%FS
0.05%+0.1%FS
0.05%+0.1%FS
0.05%+0.1%FS
Constant Power mode
Range
0.744-1040W
6-10400W
1.2-1560W
12-15600W
2.5-1560W
20-15600W
Resolution
9.3mW
75mW
22.5mW
225mW
31.255mW
250mW
Accuracy
0.5%+0.5%FS 0.5%+0.5%FS
0.5%+0.5%FS
0.5%+0.5%FS
0.5%+0.5%FS
0.5%+0.5%FS
Dynamic mode
Timing
T1&T2
0.025-10mS
1mS-30S
0.025-10mS
1mS-30S
0.025-10mS
1mS-30S
Resolution
1uS
1mS
1uS
1mS
1uS
1mS
Accuracy
1uS+100ppm 1mS+100ppm
1uS+100ppm
1mS+100ppm
1uS+100ppm
1mS+100ppm
Slew rate
6mA-1.5A/uS 50mA-12.5A/uS
12mA-3A/uS
100mA-25A/uS
20mA-5A/uS
166mA-41.6A/uS
Resolution
6mA/uS
50mA/uS
12mA/uS
100mA/uS
20mA/uS
166mA/uS
Current
Range
0-30A
0-300A
0-60A
0-600A
0-100A
0-1000A
Resolution
9.37mA
75mA
15mA
150mA
31.25mA
250mA
Accuracy
0.4%FS
0.4%FS
0.4%FS
Measurement
Voltage read back
Range
0-16V
0-80V
0-16V
0-80V
0-16V
0-80V
Resolution
15 bits
15 bits
15 bits
15 bits
15 bits
15 bits
Accuracy
0.05%+0.05%FS
0.05%+0.05%FS
0.05%+0.05%FS
1-8
Current read back
Range
0-30A
0-300A
0-60A
0-600A
0-100A
0-1000A
Resolution
15 bits
15 bits
15 bits
15 bits
15 bits
15 bits
Accuracy
0.1%+0.1%FS
0.1%+0.1%FS
0.1%+0.1%FS
Power read back
Range
0-1040W
0-10400W
0-1560W
0-15600W
0-1560W
0-15600W
Resolution
15 bits
15 bits
15 bits
15 bits
15 bits
15 bits
Accuracy
0.3%+0.3%FS
0.3%+0.3%FS
0.3%+0.3%FS
General
Short circuit
Current
30A
300A
60A
600A
100A
1000A
SIZE (mm)
440(W) X 443.7(H) X 644(D) 546(W) X 843.7(H) X 700(D) (cabinet) 546(W) X 843.7(H) X 700(D) (cabinet)
WEIGHT
90Kg
150Kg
150Kg
EMC & Safety
CE
CE
CE
*Specifications are subject to change without notice.
1. The power rating specifications at ambient temperature = 25°C and see the diagram below for power derating.
Power Rating
100%
77%
°C
Ambient Temperature
10°C 20°C 30°C 40°C
2. The Vin must be greater than min. operating voltage of each model.
3. The Vin must be greater than 7V of each model.
All CR mode accuracy should be referring to the CC mode accuracy.
4. Setting error will be 1% for R < 0.005Ω at CRL range.
1-9
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Model
63210
Power*1
1450W
14500W
Current
0-15A
0-150A
Voltage
2.5-500V
Min. operating voltage
2.5V @ 15A
2.5V @ 150A
Constant Current mode
Range
0-15A
0-150A
Resolution
3.75mA
37.5mA
Accuracy
0.1%+0.1%FS
0.2%+0.1%FS
Constant Resistance mode
Range
0.083-333ohm
3.3-13200ohm
Resolution
12 bits
12 bits
Accuracy*2
0.036mho+0.35%
0.092mho+0.1%
Accuracy*3 (Vin>7V)
0.036mho+0.35%
337.5u mho+0.35%
Constant Voltage mode
Range
2.5-125V
2.5-500V
Resolution
31mV
125mV
Accuracy
0.05%+0.1%FS
0.05%+0.1%FS
Constant Power mode
Range
5-1450W
50-14500W
Resolution
25mW
250mW
Accuracy
0.5%+0.5%FS
0.5%+0.5%FS
Dynamic mode
Timing
T1&T2
0.025-10mS
1mS-30S
Resolution
1uS
1mS
Accuracy
1uS+100ppm
1mS+100ppm
Slew rate
3mA-0.75A/uS
25mA-6A/uS
Resolution
3mA/uS
25mA/uS
Current
Range
0-15A
0-150A
Resolution
3.75mA
37.5mA
Accuracy
0.4%FS
Measurement
Voltage read back
Range
0-125V
0-500V
Resolution
15 bits
15 bits
Accuracy
0.05%+0.05%FS
1-10
Current read back
Range
0-15A
0-150A
Resolution
15 bits
15 bits
Accuracy
0.1%+0.1%FS
Power read back
Range
0-1450W
0-14500W
Resolution
15 bits
15 bits
Accuracy
0.3%+0.3%FS
General
Short circuit
Current
15A
150A
SIZE (mm)
546(W) X 843.7(H) X 700(D) (Cabinet)
WEIGHT
150Kg
EMC & Safety
CE
*Specifications are subject to change without notice.
1. The power rating specifications at ambient temperature = 25°C and see the diagram below for power derating.
Power Rating
100%
77%
°C
Ambient Temperature
10°C 20°C 30°C 40°C
2. The Vin must be greater than min. operating voltage of each model.
3. The Vin must be greater than 7V of each model.
All CR mode accuracy should be referring to the CC mode accuracy.
4. Setting error will be 1% for R < 0.005Ω at CRL range.
1-11
High Power DC Electronic Load 63200 Series Operation & Programming Manual
1.5
Dimension Outline for 63200 Series
� Model 63201 and 63202
� Model 63203 and 63204
1-12
High Power DC Electronic Load 63200 Series Operation & Programming Manual
� Model 63208 and 63209 and 63210
(Top View)
Top Side
Common Difference ± 2.0 mm
Unit: mm
Rack mount note for model 63207:
Notices for Rack Mount Installation
1 The four wheels must be removed.
2 The supporter must be removed.
3 Use the original banner screws to secure the rack
frame.
4 For safety, place it at the bottom of the rack.
1-14
Installing the 63200 Protective Cover:
Step 1: Insert the cable into protective cover.
Step 2: Socket the Bakelite into the positive/negative copper plate until it touches the chassis
and then secure the M6x20L screw and M6 screw nut to block the Bakelite.
Step 3: Place a washer on the M6x20L screw and let it go through the positive/negative
copper plate. Connect the two cables with an insulator to secure the cables on the
positive/negative copper plate.
Step 4: Cover the positive/negative copper plate with protective cover on the rear panel.
Step 5: Use the attached M6x10L screw to secure the protective cover.
Step 6: Complete installation.
1-15
This chapter describes how to install the 63200 series Load as well as a turn-on check
procedure and application considerations.
2.2
Inspection
As soon as the instrument is unpacked, inspect any damage that might have occurred in
shipping. Keep all packing materials in case that the instrument has to be returned. If any
damage is found, please file a claim to the carrier immediately. Do not return the instrument
to Chroma without prior approval.
In addition to this manual, be sure that the following items are also received – a power cord,
this manual, a pair of load terminal screws kit, a V-sense cable and a GPIB cable.
2.3
Installing
The Electric Loads can operate well within the temperature range from 0 to 40 ºC. However,
you must install the Electric Load in an area that has enough space around the unit for
adequate air flowing through and escaping from the back.
2.3.1
Changing Line Voltage
The Electronic Load can operate with a 115/230 Vac input as indicated on the left hand side
of the unit (facing the unit). If the factory set voltage does not correspond to your nominal
line voltage, set the switch to the correct line voltage as shown in Figure 2-1 before plugging
in the power cord and turning on the power.
� Notice
Line fuses do not need to be changed when the line voltage is changed. The line fuses will
protect the Electronic Load for incorrect voltage setting.
2-1
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Figure 2-1 Line Voltage Switch
2.3.2
Turn-On Self-Test
Check the following before turning on the Load.
1.
The unit has been set to the correct line voltage by factory. Refer to the line voltage
indicated on the left hand side of the panel.
2.
The power cord is connected to the AC input socket.
� WARNING
The power supplies a chassis ground through a third connector. Be sure that your outlet is of
three-conductor type with the correct pin connected to ground.
Turn on the Load by the power switch on the front panel and observe the display.
Immediately after turning on, the Electronic Load executes a self-test to check the RS232,
GPIB interface board and the system.
The LCD displays
FRONT PANEL TESTING
[ RS232 ] PASS
FRONT PANEL TESTING
CHECKSUM PASS
GPIB ADDRESS X
CHROMA ATE INC.
MODEL 632XX
and then it goes to the output mode setting display.
All of the 7-segment LEDs which display the measuring V, I, P (or R) will be activated when
2-2
the self-test completes. If any error is found in self-test, the display will stop here.
In case of failure, return the Load to Chroma sales or service office for repair.
2.4
Application Connection
2.4.1
Load Connections
� WARNING
To satisfy safety requirements, load wires must be heavy enough not to overheat while
carrying the short-circuit output current of the device connected to the Electronic Load.
Input connections are made to the + and − terminal block on the rear of the Load. The major
considerations in making input connections are the wire size, length and polarity. The
minimum wire size required to avoid overheating may not be enough to maintain good
regulation. The wires should be large enough to limit the voltage drop to less than 0.5V per
lead. The wires should be as short as possible, and bundled or tied together to minimize
inductance and noise. Connect the wire from the PLUS (+) terminal to the HIGH potential
output terminal of the power supply (UUT). Connect the wire from the MINUS (−) terminal
to the LOW potential output terminal of the power supply (UUT). Figure 2-2 illustrates the
typical setup of the Load to the UUT. When using Model 63208,63209 and 63210 the
distance should be 1 meter for chassis front and 2.5 meters for rear without any foreign object
blocking the vents or the hardware temperature may rise and cause OTP.
+
_
UUT
Figure 2-2 Load & Remote Sensing Connection
� WARNING
2-3
High Power DC Electronic Load 63200 Series Operation & Programming Manual
When in the test conditions that hardware protection is not available such as discharging
battery test, etc., it is necessary to do external protection for hardware to prevent any hazard
from occurring. The configuration and operation for safety are described below.
To avoid unclear explosion caused by operating the hardware, the factory suggests
configuring the battery as Figure 2-3 shows to protect the LOAD for usage when doing
battery discharge test.
LOAD
Fuse
R
No.2 NFB
No.1 NFB
BATTERY
Figure 2-3 Configuring Battery for Discharge
(1)
(2)
(3)
(1)
(2)
(3)
(1)
(2)
(3)
NFB: Use No Fuse Breaker, the capacity (current) needs to be smaller than the battery
maximum current so that it can cutoff when the MOSFET shorts due to aging.
R: It is recommended to install the resistance above 100kΩ1W to avoid giving LOAD
extreme voltage in a sudden.
Fuse: It is suggested to calculate the KW to be discharged for using appropriate Fuse. It
is recommended to use 20% higher.
� If more than two LOADs are used for parallel discharge test, each LOAD has to add a
fuse for protection at the front.
Operating Procedure:
Before inputting voltage to LOAD, first switch on No.1 NFB to avoid giving sudden
high voltage to the MOSFET inside LOAD to cause it aging.
Next switch on No.2 NFB to begin Battery discharge test.
To stop discharge test, first press OFF on the LOAD, next switch off No.2 NFB and then
No.1 NFB to shut down the discharge test completely.
EX: When battery discharge 300V (maximum current is 100A) is in use, how to configure the
wire if to discharge 2KW?
Ans: I = P / V = 2000 / 300 = 6.6A
Since the maximum current of battery is 100A, the NFB used must be smaller than 100A.
It is suggested to use a 20A NFB.
For R, it is suggested to use 100kΩ/ 1W.
Fuse must be larger than loading discharge current. It is 6.6A so use a 10A Fuse is fine.
� If there are any questions regarding operation, please contact the FAE of
Chroma.
2-4
Remote Sensing Connections
There are two sensing points for the Electronic Load. One is measurement at Load
terminal, and another is measurement at Vsense. The Load will automatically switch to
Vsense when Vsense terminals are connected to UUT, otherwise it will measure at Load
terminals. Remote sensing compensates for voltage drop in applications that require long
lead lengths. It is useful when the load is operating in CV or CR mode, or when it needs
precise measurement. Figure 2-2 also illustrates a typical setup for remote sensing
operation.
� Notice
The potential of V-sense red connector must be higher than that of V-sense black connector.
2.4.3
Parallel Connections
63200 series Loads can be paralleled to increase power dissipation. The Loads can be
directly paralleled in CC mode for static or dynamic operation while in CR and CP modes the
Loads can be paralleled for static operation only. It is not suggested to parallel Loads in CV
mode.
The Loads is linked via RS-485 cable to RS485 port. RS485 address represents the ID of
each Load in the parallel group.
There are two operation modes in this parallel function:
The first is MASTER mode. The master gets the total setting commands from the remote
control or front panel and then informs the slaves what setting current they should do
according to their model. (The slave model has to be ready first.) So the only one you need to
program is the master.
The second is MASTER SYNC mode. Each Load gets the current setting from its own remote
control or front panel, the master controls the loads H/W synchronization action so that all
loads can be operated with same action at the same time no matter in static or dynamic mode.
Referred to Section 4.2.9 for detail information of setting the Loads parallel function
parameters.
2-5
High Power DC Electronic Load 63200 Series Operation & Programming Manual
2.4.4
Diagram of RS485 Parallel Connections
2.4.4.1 Pin Assignment
2.4.4.2 Connecting Master & Salve
2-6
Remote Control Connection
The remote operation of Load can be done through GPIB, RS-232C or remote controller.
These connectors on the rear panel connect the Load to the controller or computer. The
GPIB interface of the electronic load is standard. The remote controller is optional. The
63200 series Remote Controller can control the load via RS-232C port. Connect the Remote
Controller to the Electronic Load before powering it on.
2-7
Chroma 63200 series electronic loads are suitable for design, manufacturing, testing and
quality assurance. It contains a processor, GPIB and RS-232C connectors, front panel
keypad, display, and power stage. Its built-in remote control function allows you to control,
read back current, voltage and status. The Save/Recall feature allows you to save up to 100
files, 10 programs, and one default setting. All of them can be saved in load EEPROM for
future use.
The load contains four cooling fans. The fan speed increases or decreases automatically when
the load power rises or falls. This feature reduces the overall noise level as the fans do not
always run at the maximum speed.
Each load can operate in constant current (CC), constant resistance (CR), constant voltage
(CV), and constant power (CP) modes. If the application is larger than one of DC load can
provide in power or current rating, the DC loads operation in parallel is a suggested way.
However, it can be applied in CC mode, CP mode and CR mode.
3.2
Front Panel Description
The front panel of the load includes a 20 × 2 characters LCD display, three segment LED, 9
led status indicators, and keypads. The LCD display will show which function is being
performed when you use the keypads. One of the keys has two functions. Figure 3-1
shows the front panel of the loads.
3-1
High Power DC Electronic Load 63200 Series Operation & Programming Manual
VOLTS
AMPS
WATTS/OHMS
CP
CV
CR
CC
CURSOR
GO/NG
7
4
1
0
.
2
5
8
9
6
3
EDIT
16V/80V
MODEL
60A/600A
DC ELECTRONIC LOAD
63203
4KW
5.2KW
Figure 3-1 The Front Panel
3.3
Rear Panel Description
The Mainframe rear panel includes ports of RS-232C, RS485, GPIB, two remote sense, three
BNC connectors, a pair of input terminals, an AC LINE socket, and a fuse holder. Figure
3-2 shows the rear panel of Mainframe 63200.
3-2
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Figure 3-2B The Rear Panel of 63205/63206/63207/63208/63209/63210
3.4
Local/Remote Control
Local (front panel) control is in effect immediately after the power is applied. The front
panel keypad and display allow manual control when Load is used in bench test applications.
Remote control goes into effect as soon as the Load receives a command via GPIB, RS-232C
or remote controller. When the remote control is in effect, only the computer can control the
Load. The front panel keypad has no effect except the LOCAL key. You can return to
local control by pressing LOCAL key.
Most of the functions that perform remotely can be done locally too at the Load front panel.
Details of local operation are given in Chapter 4 Local Operation. Fundamentals of remote
programming are described in the latter part of this manual.
3-4
Modes of Operation
There are four modes of operation: Constant Current (CC), Constant Resistance (CR),
Constant Voltage (CV), and Constant Power (CP).
You can select the mode by pressing CC, CR, CV, or CP keys under the FUNCTION keypad.
The parameters in current, resistance, voltage or power mode can be programmed easily when
the mode is selected.
All data set in CC/CR/CV/CP mode will be rescaled to fit the resolution of current/voltage
levels or slew rate. In local mode any value can be set to the Load from the keypad. There
is no upper and lower limit that would cause an error. The Load automatically selects data,
which is rescaled from the programmed value, truncates and checks high, low boundary
before fitting it into the memory. When the programmed data is over the boundary, the Load
will set the maximum or minimum level for it. In remote mode the programmed value
cannot be over boundary. An error will occur when data is over the maximum or minimum
value.
3.5.1
Constant Current Mode
I current setting
Load
Current
V
Input Voltage
Figure 3-3 Constant Current Mode
In CC mode, the Load will sink a current in accordance with the programmed value
regardless of the input voltage. The CC mode can be set by CC key in the front panel.
Press A/B key to select the current waveform in Static mode. Static function checks the
stability of output voltage from a power supply.
Press DYNA key to select the current waveform in Dynamic mode. This key is only valid
under CC mode, and Dynamic function checks the transient response.
Press RANGE key to select the current range suitable for your application no matter in static
or dynamic current mode. Low range can supply better current resolution
3-5
High Power DC Electronic Load 63200 Series Operation & Programming Manual
You can program two different static settings, A and B under all static CC, CR, CV, and CP
modes. Both A and B states use the same range. You can select A or B through the A/B
key. Slew rate determines the rate that Load level changes from one state to another.
Figure 3-4 shows the load current level after pressing A/B key.
CCLA: 4A, CCLB:2A, CCL : 0.2A/µS, CCL : 0.08A/µS
Time
Current
rise slew rate
fall slew rate
State A
State B
4A
2A
10uS
25uS
press key "A/B"
Figure 3-4 Load Level after Pressing A/B Key
Dynamic load operation enables you to program two load levels (CCDL1, CCDL2), load
duration (CCDLT1, CCDLT2), and slew rate (CCDL , CCDL ). The loading level is
switched between those two load levels according to your specific setting during operation.
The dynamic load is commonly used to test the UUT’s performance under transient loading
condition. Figure 3-5 shows current waveform of dynamic function.
CCDL1:4A, CCDL2:2A, CCDL :1A/µS, CCDL : 1A/µS,
CCDLT1:10mS, CCDLT2:10mS
Time
Current
Load1
load2
10mS
10mS
10mS
10mS
4A
2A
T1
T2
T1
T2
Figure 3-5 Dynamic Current Waveform
3-6
Slew Rate (Rise, Fall A/µS)
Slew rate determines the rate at which the current input of changes to a newly programmed
value. There are two slew rate values, which are rise rate and fall rate.
CC mode voltage specification setting
The Load can do GO/NG test by pressing the GO/NG key under loading condition. It will
check if the measured data is within the specification that set at the end of each mode setting.
We can select are two types of parameters, percent or value, under the configure setting.
This operation will be described in detail in section 4.2.4 Setting the Specification.
3.5.2
Constant Resistance Mode
V
Input
Voltage Slope
(R setting)
I
Load Current
Figure 3-6 Constant Resistance Mode
In CR mode, the Load will sink a current linearly proportional to the input voltage in
accordance with the programmed resistance. There is a double pole RC filter of input
voltage, so high frequency parts will be removed. The load sink current of CR mode is
proportioned to the input voltage through a double pole RC filter. To prevent the load
current change caused by the input voltage variation, the power source impedance should be
as low as possible, and remote sensing cable must be used to sense load input voltage when
high sink current (low setting resistance) is programmed.
Resistance can be programmed in either of low range or high range by the RANGE key. The
low range is used for input voltage in low voltage range while the high range is for the input
voltage over low voltage range. The current range of CR mode is high range.
There are two resistance levels (A or B) for CR function as static CC mode. Both A and B
states use the same range. You can select CRLA or CRLB using A/B key. Slew rate
determines the rate that the load level changes from one state to another.
3-7
High Power DC Electronic Load 63200 Series Operation & Programming Manual
3.5.3
Constant Voltage Mode
V Voltage setting
Input
Voltage
I
Load Current
Figure 3-7 Constant Voltage Mode
In CV mode the Load will sink current to control the voltage source in programmed value.
There are two response speeds of CV modes: fast and slow. The fast/slow respond speed is
the slew rate of current change.
Voltage can be programmed in either of low range or high range by the RANGE key. The
low range is used for input voltage in low voltage range while the high range is for the input
voltage over low voltage range.
There are two voltage levels (A or B) for CV function. You can select CVHA or CVHB
using A/B key.
3.5.4
Constant Power Mode
I
Load Power setting
Current P = V * I
V
Input Voltage
Figure 3-8 Constant Power Mode
In CP mode, the Load will sink a current according to the programmed power. This mode is
operated under the F/W calculation. That is, take the measured V data, divide the Power
setting and get the I setting value. High frequency parts will be removed as there is a lower
3-8
pass filter for the measuring data.
Power can be programmed in either of low range or high range by the RANGE key. The low
power range is operated under low current range mode while the high power range is under
high current range mode.
There are two power levels (A or B) for CP function as other modes. Both A and B states
use the same range. You can select CPLA or CPLB using A/B key. Slew rate determines
the rate that the load level changes from one state to another.
3.5.5
Load Surge Capability
Chroma’s 63200 Series DC Loads provide a unique load surge simulation capability, which
allows users to overdrive the loads up to 2.7 times their rated power for short periods. This
feature is ideal when the average power require by the UUT is low compared to short-term
peak power demands. Plasma Display Panel (PDPs) testing is one typical application, others
include battery 3C discharge, breaker & fuse over rating (300% to 1000%) tests, car engine
startup simulation and DC motor startup simulation.
The amount of surge loading available using the 63200 loads is related to the initial loading
conditions. Figures 3-9 and 3-10 show the relationship of initial state (Load_Low under
Dynamic mode) and the maximum acceptable overdrive power.
Under this operation, the load will display an Over Power Protection Alarm (OPP) and will
disable the load current if the user violates the maximum load surge capability showed in the
figures below.
Figure 3-9 Load Surge Capability (Static Mode)
3-9
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Figure 3-10 Load Surge Capability (Dynamic Mode)
Note 1: The Initial state under Static Mode should last at least 1 second.
Note 2: This load surge capability will be regulated by the temperature de-rating
characteristics. (Refer to Note 1 in Specifications)
Note 3: Examples below assume the use of the Model 63201 load with a continuous rating
of 2600W/300A/1-80VDC.
Example 1: STATIC LOADING
The Model 63201 can be overdriven to approximately 5200W (200% of its rated continuous
power rating) for 0.06 seconds when the starting power is 650W (25% of its rated power).
This is represented by DOT on the blue curve in Figure 3-9.
Example 2: DYNAMIC LOADING
The Model 63201 is capable of a zero – to- 6500W (250%) pulse at a duty cycle of 5%.
This is represented by the DOT on the purple curve in Figure 3-10.
3.5.6
Timer Function for Battery Discharge Testing
The 63200 Loads include unique timing & measurement function allowing for precision time
settings and measurements in the range of 1s to 99999s. This feature allows users to set a
final voltage & timeout value for battery discharge testing and similar applications.
For example, Figure 3-11 below shows that the 63200’s internal timer can be initiated
automatically when the battery voltage falls below a preset value. The timer will continue
counting until the second preset voltage value is reached.
3-10
The Load measures current, voltage, and power of the UUT and resistance of the Loading.
The sampling rate is about 8 mS. Voltage and current measurements are performed with a
15-bit resolution of full scale ratings.
There are three sets of 7-segment LEDs for the measuring data. One is for voltage, another is
for current, and the other is for Power or Resistance that you can select under the
configuration setting. The OHMS led will be on when you select the resistance
measurement.
3.7
Slew Rate & Minimum Transient Time
Slew rate is defined as the change in current over time. A programmable slew rate allows a
controlled transition from one load setting to another to minimize induced voltage drops on
inductive power wiring, or to control induced transients on a test device. If the transient
from one setting to another is large, the actual transient time can be calculated by dividing the
current transition by the slew rate. The actual transition time is defined as the time required
for the change of input from 10% to 90% or from 90% to 10% of the programmed excursion.
If the transition from one setting to another is small, the small signal bandwidth of Load will
limit the minimum transition time for all programmable slew rates. Because of the limit, the
actual transition time is longer than the expected time based on the slew rate. Therefore,
both minimum transition time and slew rate must be considered in the determination of actual
transition time. The minimum transition time is from 24 µS to 6 mS depending on the slew
rate setting.
3-11
High Power DC Electronic Load 63200 Series Operation & Programming Manual
3.8
Start/Stop Sink Current
To simulate the transient characteristics of load to UUT, the critical problems are when and
how the Load starts sinking current to UUT. You may set the conducting voltage Von to
solve the problems. The Load will start or stop sinking current when the UUT output
voltage reaches the Von voltage. You can start sinking current when the load is ON, and the
input voltage is over Von voltage, but stop sinking when the load is OFF, or the input voltage
is below Von voltage. See Figure 3-9 and 3-10 for start and stop sinking current.
There are two operation modes for Von control, latch and non-latch. Latch means that when
voltage is over Von voltage, Load will start sinking current continuously in spite of the input
voltage drop is below Von voltage. Non-latch means that when the input voltage is below
Von voltage, the Load will stop sinking current. The Von voltage and operation mode are
set in configuration.
Figure 3-12 Start Sinking Current (Von Non-Latch)
3-12
Figure 3-13 Stop Sinking Current (Von Non-Latch)
3.9
Short On/Off
The Load can simulate a short circuit at input by setting it on with full-scale current. The
short circuit can be on/off from the front panel or via remote control. There are two
operations for SHORT key on the front panel. One is toggled on/off, and the other is
controlled by key. They are selected in configuration. The SHORT key will be enabled
only when the Load is ON.
Toggled on/off means pressing SHORT once to enable short circuit, and again to disable it.
Control by Key means pressing SHORT and holding it to enable short circuit, and releasing it
to return to normal operation.
The actual value of electronic short depends on the mode and range that are active when the
short is turned on. Its basic limit is the maximum power range the Load can supply. In CC
mode it is equivalent to the programming of full-scale current. In CR mode it is equivalent
to the programming of the minimum resistance for the present resistance range. In CV mode
it is equivalent to the programming of zero voltage. In CP mode it is equivalent to the
programming of the maximum power for the present power range. Turning on the short
circuit does not affect the programmed setting, and Load input will return to the previous
programmed values when the short circuit is turned off.
Note that turning on the short circuit may cause the Load to sink so much current to trig
protection circuit, and will turn off the Load.
� Notice
The 63200 series have a TTL signal from RS-485 pin 6 for your application to control the
external short relay. TTL indicates high/low when the short key is set on/off.
3-13
High Power DC Electronic Load 63200 Series Operation & Programming Manual
3.10
Load On/Off
A module’s input can be toggled on/off through the blue LOAD ON/OFF key on the front
panel, or the remote control. The on/off change for input is done according to the slew
rate.
Turning off the load does not affect the programmed setting. The load will return to the
previous programmed values when the Load is turned on again.
3.11
Protection Features
The load includes the protection features: Over power, Over temperature, Fan fail, and
Abnormal Alarms for Reverse Voltage and Over Voltage.
The appropriate bits in the Load status registers are set when any of the protection features or
alarm listed above is active. The Load’s buzzer will beep sound to inform you till protection
or alarm status is reset. When any of the protection or alarm occurs, the Load input will turn
off.
�
Over voltage
The over voltage alarm circuit is set at a level slightly above the voltage range specified
in the Load specification. The over voltage status register bit is set when the OV
condition occurs and will remain set till it is reset. The Load will appear OVP as below
when over voltage alarm occurs.
PROTECTION
OVP
� Over current
When Load is operating in CR or CV mode, it is possible to attempt to sink current more
than it is rated for. The limit level of current is set at a level slightly above the current
of the Load. The over current status register bit is set when the OC condition occurs,
and will remain set till it is reset. The Load will appear as below when over current
protection occurs.
PROTECTION
OCP
� Over power
The over power protection circuit is set at a level slightly above the power range
specified in the Load specifications. The over power status register bit is set when the
OP condition occurs, and will remain set till it is reset. The Load will appear as below
when over power protection occurs. This protection will be activated also if the power
3-14
exceeds the maximum surge load capability mentioned in section 3.5.5.
PROTECTION
OPP
� Over temperature
The Load has an over temperature protection circuit, which will turn off the load if
internal temperature exceeds the safe limit. The over temperature status register bit is
set when the OT condition occurs, and will remain set till it is reset. Load will appear
as below when over temperature protection occurs.
PROTECTION
OTP
� Reverse Voltage
The Load can conduct a reverse current when the polarity of UUT connection is not
correct. The maximum safe reverse current is same as the Load rated current. If the
UUT reverse current is over the Load rated current, the Load may be damaged. If a
reverse voltage condition is detected, you must turn off the power to UUT immediately
and correct the connection. The reverse voltage status register bit is set when the RV
condition occurs, and will remain set till it is reset. The Load will appear as below
when reverse voltage alarm occurs.
PROTECTION
REV
� FAN FAIL
The Load has a fan fail protection circuit, which will turn off the load if any of the four
fans is out of order. The fan fail status register bit is set when the condition occurs, and
will remain set till it is reset. The Load will appear as below when fan fail protection
occurs.
FAN FAIL
All of the protections or alarms will latch when they are tripped. When any protection or
alarm occurs, the load will turn off the input, and beep till you remove the condition and reset
the protection by pressing ENTER.
�CAUTION
To protect the Electronic Load from possible damage, the input voltage must not exceed the
maximum input voltage rating specification. In addition, the Load + terminal potential must
be more than the − terminal potential.
3-15
High Power DC Electronic Load 63200 Series Operation & Programming Manual
3.12
Save/Recall Setting
The Electronic Load setting can be saved and recalled for various test setups use. This
simplifies the repetitive programming for different things. The present setting of mode
parameters (CC, CR, CV, CP), programs and power on status (DEFAULT) can be saved in
EEPROM using the SAVE key. Later you can recall the settings from the specified file via
RECALL key.
3.13
Program
The program feature is very powerful. It allows you to simulate various test conditions.
There are ten programs in the Electronic Load. Each program has ten sequences. The
setting mapped to the program sequence in file is one on one. It means that program 1,
sequence 1 maps to file 1, and program 3, sequence 4 maps to file 24. Please see section
4.2.2 and 4.2.3 for setting and running the program.
3.14
External Waveform Control
The external dynamic test, operated in the CC mode, is similar to that under the Dynamic test,
but the load level switching is controlled by the duty cycle of an External TTL signal. It
works the same way as the dynamic test except that the Period control signals are not
generated internally, but are inputted from V EXT. Connectors are on the rear panel. A
0-to-10V external signal corresponds to the 0-to-full scale input range, so that users should
apply DC offset for the external signal in the range from 0 to 10V. For the configuration of
external waveform control usage, refer to section 4.2.5 for details.
3.15
Voltage & Current Monitor
The 63200 series have two isolated BNC connectors to monitor load voltage and Current, the
output signal from I MON. and V MON. Connectors are on the rear panel. A 0-to-10V
output signal corresponds to the 0-to-full scale load V&I range.
3-16
This chapter describes how to operate the electronic load from the local panel in details.
The descriptions include panel control and indicators.
4.2
Local Operation
In order to use the front panel keys to control the electronic load, local operation must be in
effect. Immediately after the power is applied, local operation will be in effect. When
local operation is in effect, you can use the display with keypad on the front panel to control
the Load. The display can be used to view the programmed setting. The input voltage,
current, power or resistance is displayed on the three 7-segment Leds.
� Notice
When you edit setting, the display will blink to let you know which setting is to be edited or
has been selected.
In the remote state, the keys on the front panel have no effect. Only remote controller can
program the Load. The 7-segment will show the present input voltage and current readings.
The Load LCD display will show REMOTE message.
� Notice
When setting the load level, the resolution of current, voltage, power, resistance and slew rate
will be different from the entered values. The displayed value will remain as the one entered.
But the output will be the actual value D/A programmed in the load. These parameters
except resistance will be degraded to low values when entered.
4-1
High Power DC Electronic Load 63200 Series Operation & Programming Manual
VOLTS
AMPS
WATTS/OHMS
CP
CV
CR
CC
CURSOR
GO/NG
7
4
1
0
.
2
5
8
9
6
3
EDIT
16V/80V
MODEL
60A/600A
DC ELECTRONIC LOAD
63203
4KW
4
5
6
1
3
2
5.2KW
Figure 4-1 Front Panel of the Load
1.
Line switch
Turn the ac power on/off.
2.
LCD display
Display setting information normally.
3.
7-segment LEDs
Show the V, I, P, R measured data.
4.
System keys
SYS
To set the remote control status, ext. wave enable and
Parallel function parameters.
RECALL
To recall the saved settings from EEPROM, and all settings
from specified files (1 to 101).
SAVE
To save all of the present mode settings in the specified files (1
to 100). The save program is from 1 to 10. Saving
DEFAULT is to save the status for the next time the Electronic
Load is turned on. All saved settings are stored in EEPROM,
and will not be lost when ac power is cycled.
CONF
To select configuration data for editing.
5.
Function keys
CC
To select the constant current mode for editing and the CC led
aside this key will be on.
CR
To select the constant resistance mode for editing and the CR
led aside this key will be on.
CV
To select the constant voltage mode for editing and the CV led
4-2
aside this key will be on.
CP
To select the constant power mode for editing and the CP led
aside this key will be on.
RANGE
To select the setting level range for each mode.
CURRSOR EDIT
Enable to edit the digit by rotary knob under loading condition.
This key is valid only when the Load is in on status.
PROG.
To select a program for settings or running.
GO/NG
To enable the SPEC. checking test.
The GO/NG LED will turn red when any of the SPEC fails. It
will turn green when the check is all right.
A/B
To select one of the two static settings.
The LED aside this key will be on when you choose A setting,
and off when you choose B setting.
DYNA.
Choose the dynamic current waveform and run the setting under
Constant current mode. This key is only valid in CC mode and
the LED aside this key will be on when in this status.
SHORT
The Load can simulate a short circuit across the input. The
short circuit can be enabled when this key is pressed. When
the input is shorted, the LED aside this key will be active. The
SHORT key can be set in configuration to toggled on/off, or
active by pressing.
LOAD ON/OFF
The input can be toggled on or off by pressing LOAD. When
the input is turned on, the LED aside this key will be active.
6.
Entry keys
▲, ▼
They let you scroll through the choices in a parameter list for a
specific command. Parameter lists are circular. You can return
to the starting position by pressing either key continuously.
CLEAR
To clear the digit entered from keypad. This key lets you
correct wrong digits before they are entered.
ENTER
It executes the entered value or the parameter for the command
accessed presently. The parameters you have entered with
other keys are displayed but not entered into the Load until you
press this key. Before pressing ENTER you can change or
abort anything entered into the display previously.
0 - 9
They are used for entering numeric values.
4-3
High Power DC Electronic Load 63200 Series Operation & Programming Manual
.
It is a decimal point.
Rotary knob allows you to change the setting value
continuously by turning this knob.
4.2.1
Setting the Operation Mode
The CC, CR, CV, and CP keys are used to select the Load modes for local control.
The load levels and slew rates are common to CC, CR, and CP modes. CV mode sets
voltage level and response speed. There are two level settings for CC, CR, CV, and CP
modes. They can be switched by the A/B key.
Setting CC Values
There are four modes for CC operation: CCL, CCH, CCDL, CCDH. The current levels are
programmed in Amps. The slew rate levels are programmed in Amps/µS. The timings are
programmed in millisecond. The setting buffers of four CC modes are independent.
Changing the operation range doesn’t affect the settings of other ranges. The following
examples show how to set the CC values for the Load model number 63203 channel.
1.
Select mode
Press CC key to go into the CC static mode.
CCLA : 0.000A
CCLB : 0.000A
2. Select range
Press RANGE key to choose the high range constant current mode.
CCHA : 0.00A
CCHB : 0.00A
If press RANGE again, it will go back to the low range constant current mode.
3. Set Current Level
There are 4000 discrete steps from 0 to full scale in each range. Set level A current
level to 200 amps by pressing 2, 0, 0, ENTER. Set level B current level to 5 amp by
pressing 5, ENTER.
CCHA : 200.00A
CCHB : 5.00A
4-4
4. Set Slew Rate
There are 250 discrete steps in each range. Set the rise 20 A/µS and fall slew rates to
0.5A/µS by pressing 2, 0, ENTER for rise and 0, . , 5 ENTER for fall slew rate.
CCH
: 20.0A/uS
CCH
: 0.5A/uS
5. Set Voltage spec
Set the high voltage spec. to 6 V by pressing 6 , ENTER.
CCH VOLTAGE SPEC.
HIGH : 6.000V
Set the low voltage spec. to 4 V by pressing 4 , ENTER.
CCH VOLTAGE SPEC.
LOW : 4.000V
Then the display will go to the first editing page again.
� Notice
Spec. value can also be programmed by PERCENT under Configure setting. You have to
input the CENTER Voltage, and then enter the HIGH and LOW percent to the center voltage.
6. Set dynamic current mode.
Press DYNA key under CC mode, and select current range by RANGE key. Program
the high current level to 50A by pressing 5, 0, ENTER keys. Program the low current
level to 5A by pressing 5, ENTER keys.
CCDL1: 50.000A
CCDL2: 5.000A
Then set the dynamic function period T1 to 0.1 mS, period T2 to 0.2 mS by pressing
0, . , 1, ENTER and 0, . , 2 , ENTER . The range of dynamic period is from 0.025
mS to 30 Sec.
4-5
High Power DC Electronic Load 63200 Series Operation & Programming Manual
CCDLT1: 0.100mS
CCDLT2: 0.200mS
The slew rate and Voltage Spec settings are same as those of static current mode.
� Notice
If you press ENTER key, and the blinked data do not go to next, change the configuration
setting Enter Data Next to YES.
Setting CR Values
The Load CR mode is programmed by pressing CR. The resistance values can be
programmed in low voltage (CRL) or high voltage (CRH) range. The current is always in
high range. All resistance levels are programmed in ohm. The slew rate is in A/µS.
Following examples illustrate how to set the CR values for Load model number 63203.
1.
Select range
Press RANGE key to choose the high range CR mode.
CRHA : 500.000Ω
CRHB : 500.000Ω
If press RANGE again, it will go back to the low range CR mode
2. Set Resistance Level
Set level A resistance level to 200 Ohm by pressing 2, 0, 0, ENTER. Set level B
resistance level to 1 Ohm by pressing 1, ENTER.
CRHA : 200.000Ω
CRHB : 1.000Ω
3.
Set Slew Rate
Set the rise 20 A/µS and fall slew rates to 0.5A/µS by pressing 2, 0, ENTER for rise and
0, . , 5, ENTER for fall slew rate.
CRH
: 20.0A/uS
CRH
: 0.5A/uS
4-6
Set Voltage spec
Set the high voltage spec. to 6 V by pressing 6 , ENTER.
CRH VOLTAGE SPEC.
HIGH : 6.000V
Set the low voltage spec. to 4 V by pressing 4 , ENTER.
CRH VOLTAGE SPEC.
LOW : 4.000V
Then the display will go to the first editing page again.
Setting CV Values
The CV mode for the Load is programmed by pressing CV. The voltage values can be
programmed in low voltage (CVL) or high voltage (CVH) range. The current is always in
high range. All voltage levels are programmed in Volt.
The following examples illustrate how to set CV values for Load model number 63203.
1.
Select range
Press RANGE key to choose the low range CV mode.
CVLA : 16.000V
CVLB : 16.000V
If press RANGE again, it will go back to the high range CV mode.
2.
Set voltage Level
Set level A voltage level to 10 V by pressing 1, 0, ENTER.
Set level B voltage level to 8 V by pressing 8, ENTER.
CVLA : 10.000V
CVLB : 8.000V
3.
Set Response Speed
There are two response speeds for CV mode, fast and slow for different UUT testing.
Refer to Figure 4-2 and 4-3 for transfer functions.
4-7
High Power DC Electronic Load 63200 Series Operation & Programming Manual
CVL RESPONSE
1: FAST 2. SLOW
Press 2, ENTER to choose the slow response.
Figure 4-2 CV Response Transfer Function (FAST)
Figure 4-3 CV Response Transfer Function (SLOW)
4.
Set Current spec
Set the high current spec. to 8 A by pressing 8, ENTER.
CVL CURRENT SPEC.
HIGH : 8.000A
Set the low current spec. to 4 A by pressing 4, ENTER.
CVL CURRENT SPEC.
LOW : 4.000A
Then the display will go to the first editing page again.
4-8
Setting CP Values
The CP mode for the Load is programmed by pressing CP. The power values can be
programmed in low current (CPL) or high current (CPH) range. All power levels are
programmed in Walt. The slew rate is in A/µS.
The following examples illustrate how to set CP values for Load model number 63203.
1.
Select range
Press RANGE key to choose the high range CP mode.
CPHA : 12.000W
CPHB : 12.000W
If press RANGE again, it will go back to the low range CP mode
2.
Set Power Level
Set level A power level to 200 W by pressing 2, 0, 0, ENTER.
Set level B power level to 4000 W by pressing 4, 0, 0, 0, ENTER.
CPHA : 200.000W
CPHB : 4000.000W
3.
Set Slew Rate
Set the rise 20 A/µS and fall slew rates to 0.5A/µS by pressing 2, 0, ENTER for rise and
0, . , 5, ENTER for fall slew rate.
CPH
: 20.0A/uS
CPH
: 0.5A/uS
4.
Set Voltage spec
Set the high voltage spec. to 6 V by pressing 6, ENTER.
CPH VOLTAGE SPEC.
HIGH : 6.000V
Set the low voltage spec. to 4 V by pressing 4, ENTER.
CPH VOLTAGE SPEC.
LOW : 4.000V
4-9
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Then the display will go to the first editing page again.
4.2.2 Setting the Program
The Electronic Load is able to select customized basic tests, and link them to a program test
for automatic execution.
The PROG. key is used to select program or recall program for local control. There are ten
programs (1-10). Each program has ten sequences to map files from 1 to 100. Program 1
maps files from 1 to 10. Table 4-1 shows the relationship between the program sequence
and the corresponding file.
Program 1 Sequence No.
1
2
3
4
5
6
7
8
9
10
Corresponding File No.
1
2
3
4
5
6
7
8
9
10
Program 2 Sequence No.
1
2
3
4
5
6
7
8
9
10
Corresponding File No.
11
12
13
14
15
16
17
18
19
20
:
:
Program 10 Sequence No.
1
2
3
4
5
6
7
8
9
10
Corresponding File No.
91
92
93
94
95
96
97
98
99
100
Table 4-1 The Relationship of the Program Sequence and the Corresponding File.
When running a program you must set its corresponding file parameters first. If one
program sequence is not enough for you to test the UUT, you can use program chain function
to get more sequences.
Press PROG. key, and the LCD will display as follows. Press number 1 -10 followed by
ENTER to recall program from EEPROM, or use ▲, ▼ keys to edit the program.
PROGRAM SELECT
No: 1
1.
Setting the Program Chain
The chain function of program enables you to chain program so as to get more sequences
for testing. Set program chain number to 0 means no program chained. Program
chain function can chain itself for loop test, or other programs.
Press 1, ENTER to chain itself for loop test. The default setting is 0.
PROGRAM CHAIN
No: 1
4-10
Setting the Sequence Mode
There are three modes to control the method of sequence execution.
SKIP:
Skip the sequence. Load will not change the input status.
AUTO:
Use ON/OFF time to control Load input on/off. When ON/OFF time
passes, the Load will get to the next sequence automatically.
MANUAL:
Use ▲, ▼ or number 0 to 9 to control the execution sequence. Pressing
number key lets you select a random sequence number to execute.
Pressing 0 means to go to sequence 10.
Press 2, ENTER keys to set sequence 1 to manual mode. You need to set ten sequence
settings for one program. The default setting is SKIP.
[ SEQ 1 ] 1: AUTO
0: SKIP 2: MANUAL
3.
Setting the Sequence Load ON/OFF Time
The sequence ON/OFF time controls the Load input ON/OFF when the program’s
sequence is executed. The ON/OFF time range is 0 to 60 seconds. (30S each)
Press 8, ENTER keys to set OFF time for 8 seconds. The default setting is 0 second for
OFF time.
4.
Setting the Short Delay/ON Time
The range of short delay and on time is from 0 to 60 Sec. The short delay time starts
from the beginning of the Load on time. If the short time is set to 0 Sec., the Load will
not short. The default setting is 0 Sec.
Press 8, ENTER to set SHORT on time for 8 seconds.
SEQUENCE 1
SHORT DELAY : 0.0 S
SEQUENCE 1
LOAD OFF TIME : 8.0 S
SEQUENCE 1
LOAD ON TIME : 20.0 S
4-11
High Power DC Electronic Load 63200 Series Operation & Programming Manual
SEQUENCE 1
SHORT ON TIME : 8.0 S
5.
Setting the Sequence P/F Delay/ON Time
The sequence Pass/Failure Delay/ON time let you set the delay time for P/F checking
and on time for how long it checks when load condition changes. The failure status of
the sequence will latch when a program is executed. It means that any failure will be
memorized even when the UUT becomes stable within the specifications later. The
range of P/F delay and ON time is from 0 to 60 seconds. Press 1, ENTER to set the
sequence P/F delay time for 1 second and then 5, ENTER for P/F on time for 5 seconds.
Both default setting are 0 second.
SEQUENCE 1
P / F DELAY : 1.0 S
SEQUENCE 1
P / F ON TIME : 8.0 S
4.2.3
Running the Program
Press LOAD ON/OFF to run program when the program function is selected. The display
shows as below.
PROG. 1 SEQ. 1
[ON ][ KEY ] [ PASS ] [ S ]
The upper line shows the executed program and sequence number while the lower line shows
the Load, key, test result and short status.
ON/OFF:
It shows the Load input status.
KEY:
It displays when MANUAL mode is active and waiting for key input.
PASS/FAIL:
It shows the test result compared with SPEC setting.
S:
It display if Load is Short now.
When executing the program, the setting of sequence will recall files from EEPROM, and the
SPEC function is always ON. All function keys are disabled until LOAD ON/OFF is pressed
to stop the program execution, or the program run finishes. When the program run stops or
finishes, the LCD will display as follows.
PROGRAM OFF
RESULT : PASS
4-12
It means that all sequences have passed in the program tests. If the test fails, LCD will show
as follows.
PROG. XX Err.
REPORT 1, 2, 3,…….10
PROG. XX indicates the file number of the program failed, 1 to 10. In addition, 1, 2, 3...10
shown by LCD stand for the failed sequence numbers. During program chain testing, if the
failed program files are more than one set, you can use ▲, ▼ to read the contents of failed
programs.
4.2.4
Setting the Specification
The GO/NG key is to enable/disable specification checking function. The Load will
compare measurement data with the set specifications for HIGH and LOW boundary when
the SPEC TEST is ON, and the LED, GO/NG, is lighted on the panel. To set specifications
for the load; please refer to the section 3.5 at the end of each mode for parameters setting.
The specification unit for CC, CR, and CP modes is volt while CV mode is current. There
are three levels for each mode: CENTER, HIGH and LOW. The CENTER level must be set
by the value of channel input reference level. The HIGH and LOW levels can be set by the
value or percentage selected in configuration SPEC.ENTRY MODE. The HIGH/LOW
percentage range is from 0 to 100%.
The default setting of HIGH and LOW is 100%. The CENTER value is half of the range.
Please refer to section 4.2.5 for the specification selection set by Value or Percentage.
4.2.5
Setting the Configuration
The Electronic Load provides useful features such as Von point, Current limit, Auto Load On,
etc. To use these powerful features, you must set relevant parameters in accordance with
application needs by the configuration setup. This procedure is only needed for initial setup
of a test operation. The configuration of the Load is stored in EEPROM. To set
configuration you must press the CONF key.
Set the voltage range for CC, CP mode. There are two voltage ranges for CC or CP mode.
High range is for high voltage and low range is for low voltage so as to get better voltage
resolution. The default setting of V range is HIGH.
CC, CP V RANGE SELECT
1: HIGH 2: LOW
4-13
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Set Von point. Von is the conduction voltage level when the Electronic Load starts to sink
current and the UUT output reaches the Von voltage. The default setting of Von voltage is
1V.
Von POINT
VOLTAGE = 1.000V
Set Von latch. There are two operation modes for Von control. Von latch ON means that
Load will sink current continuously when it reaches Von voltage. Von latch OFF means that
Load will stop sinking current when UUT voltage is under Von voltage. The default setting
of Von latch is OFF. Figure 4-4 and 4-5 show the Von LATCH ON and OFF current
waveform respectively.
Von LATCH ENABLE
1: YES 2: NO
� CAUTION
If Von is set to 0V, the Load circuit will be ON in spite of no UUT. This will get overshoot
spike. If a UUT is applied, the overshoot may damage the UUT even though the Load
current sets small. So, do not set Von to 0V.
Figure 4-4 Von LATCH ON Current Waveform
4-14
Figure 4-5 Von LATCH OFF Current Waveform
Set CV mode CURR_LIMIT. This function will limit the Load current sinking to protect
the UUT in CV mode. The default setting of current limit is the maximum Load current.
CV CURRENT LIMIT
CURRENT = 200.000A
Set sign of voltage for display. The Electronic Load will show minus sign of the voltage if
MINUS is selected. No sign will show if select PLUS. The default setting is PLUS.
Selecting MINUS of SIGN OF VOLT. will occupy one digit. The displayed digits are five.
SIGN OF VOLTAGE
1. PLUS 2. MINUS
Set POWER/OHM for display. The right one of the three 7-segment LEDs can be selected
to display POWER or OHM measurement data. The default setting is POWER. The LED
“OHMS” will be lighted if you select OHM by pressing 2, ENTER.
POWER / OHM SELECT
1: POWER 2: OHM
Set the specifications of entry mode. The specifications of Load can be set by VALUE or
Percentage for HIGH and LOW data. The percentage values refer to CENTER value of
specification. The default setting of SPEC entry mode is percentage.
4-15
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Select data entry mode by ENTER. If YES is selected for data entry, the setting will go to
the next one after pressing ENTER. If NO is selected for data entry, the setting will remain
at the same line for you to change it again and again. The default setting is NO.
Select SOUND ENABLE. When you press the key on the module, it will produce a sound if
sound = YES. The default setting for sound is YES.
Select Load input status when it is powered ON. If YES is selected, the module will be
active according to AUTO LOAD ON mode setting. The default setting of AUTO LOAD
ON is NO.
Select the load on mode if AUTO LOAD ON is ON. If LOAD is selected, the Load will be
active as DEFAULT setting. If PROG is selected, the Load will be active as the program
saved last time. The default setting of AUTO LOAD ON MODE is LOAD.
Select short key mode. Set SHORT key mode for the Load. The default setting of SHORT
mode is TOGGLE.
Display the versions of host, load, and Panel.
Press ▼ key and the display will show:
WRTR: 00.85
BOOT: 00.85 MAIN: 00.85
SHORT KEY MODE
1: TOGGLE 2: HOLD
AUTO LOAD ON MODE
1: PROG 2: LOAD
AUTO LOAD ON
1: YES 2: NO
SOUND ENABLE
1: YES 2: NO
ENTER DATA NEXT
1: YES 2: NO
SPEC. ENTRY MODE
1: VALUE 2: PCet
4-16
Press ▼ key and the display will show:
4.2.6
Recalling Files
Press RECALL to recall files from 1 to 100. Files 1 to 100 are user data. File 101 is factor
set state. After a file is recalled, the display will go to mode editor for you to edit or view
the file. Pressing RECALL the display will show the file no. last recalled. The default file
no. is 1 when the Load is powered on.
Press RECALL, 3, ENTER to recall the number 3.
The data of the Load will be recalled when you execute file recall.
4.2.7
Saving File/Default/Program
There are 100 file locations (1 to 100) for you to save files. Press SAVE, 2, 0, ENTER to save
a file to location 20.
Press SAVE, ▲, ▼ until the display shows as follows. The DEFAULT states are used for
Electronic Load after power-on. Press 1, ENTER to save DEFAULT to EEPROM.
Press SAVE, ▲, ▼ until the display shows as follows.
SAVE DEFAULT
1: YES 2: NO
SAVE FILE
FILE NO : 20
RECALL FILE
FILE NO : 3
WRTR: 00.85
BOOT: 00.85 MAIN: 00.85
WRTR: 00.85
BOOT: 00.85 MAIN: 00.85
4-17
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Press 1, ENTER to save program.
SAVE PROGRAM
1: YES 2: NO
4.2.8
Going To Local
The CONF./LOCAL key operates as local key when Electronic Load is in remote mode.
You can press CONF./LOCAL key to go to local operation when Load is in remote state. In
local operation, CONF./LOCAL key operates as CONF key.
4.2.9
Setting System and RS-232C Connection
GPIB Address setting. Press 8, ENTER to set the Load GPIB address to 8.
GPIB ADDRESS
8
RS485 Address setting. Press 3, ENTER to set the Load RS485 address to 3. This is for
setting the models ID when used in parallel mode.
RS485 ADDRESS
3
RS-232C parameters setting. There are four parameters for you to set: Baud Rate, Parity
Check and Data Bit number.
Baud Rate:
0:4800, 1:9600, 2:19200, 3:38400, 4:57600, 5: 115200 bits/second
Parity Check:
0: NONE, 1: EVEN, 2: ODD
Data Bit:
1: 7 bits, 2: 8 bits
Stop Bit:
1: 1 bit, 2: 2 bits
The RS-232C connector on the rear panel is a 9-pin connector (DB-9, male connector). The
RS-232C connector bus signal is defined as below.
4-18
Note: Pin 1 and pin 9 (+5V) are for 63200 series Remote Controller only.
External wave Enable. The Load current can be controlled by the external DC voltage
input via the Vext BNC connector on the rear panel if this function is selected 1: YES.
The default is 2: NO. The Ext. Wave LED will be lighted when operating in this mode.
EXT. WAVE ENABLE
1: YES 2: NO
Parallel mode function. The Loads can be used in parallel to increase the loading power
capacity. There is only one master in the parallel group, the rest are slaves. These Loads
can be operated synchronized in static or dynamic mode via the master control. There are
four different types for the Load to set in operation. There are maximum six models in
parallel.
PARALLEL = NONE
PRESS 0 → 3 TO SELECT
0: NONE:
Disable the parallel mode function.
1: MASTER:
Used as the master for the parallel group, and this is the only one
controlled by front panel or PC in this group. Not only do the
synchronized operation, but also tell the slaves how many current they
should sink.
Measurement Display:
Choose 1. SUM to show the total current on the master or
2. Sole to show the current the loads owned.
SLAVE X MODEL:
Press 0 → 6 to setup the slaves’ model to use in parallel. NONE means
not exist.
2: SLAVE:
Setup the Load as slave.
3: MASTER SYN: Same as 1: MASTER except this just do the SYNC parallel function, all
the loading data of the slaves is decided by the front panel or PC input.
4-19
High Power DC Electronic Load 63200 Series Operation & Programming Manual
4.2.10 Online Change Level
The Load provides you two ways for changing level online. They are convenient for you to
change load directly with the rotary knob when LOADON. These two operation modes are
described below.
By Rotary knob:
In LOADON, change load with the rotary knob in resolution unit.
When the rotary knob rotates clockwise, it means as follows.
CC mode: raise the current value.
CR mode: raise the resistance value.
CV mode: raise the voltage value.
CP mode: raise the power value.
When the rotary knob rotates counterclockwise, it means as follows.
CC mode: lower the current value.
CR mode: lower the resistance value.
CV mode: lower the voltage value.
CP mode: lower the power value.
By CURSOR EDIT: In LOAD ON, press the mode key CC, CR, CV, or CP to make the
display show numeric values for the setting. Press CURSOR EDIT to
see the cursor blinking on the display. Press ▲, ▼ keys to move the
cursor to the digit you want to edit. Change the value by rotary knob.
Press the CURSOR EDIT key again to disable the cursor.
4-20
Basic Information for Programming
5.
Basic Information for Programming
5.1
Introduction
The following sections describe how to program the 63200 series electronic load remotely
from a GPIB controller or RS232C. The command set introduced here can be applied to all
63200 series electronic loads including 63203, 63204, that equipped with GPIB or RS232C.
Either GPIB or RS232C can be used one at a time. They cannot be used simultaneously. If
GPIB is used first in remote control, RS-232C will be disabled unless the machine is reset,
and vice versa.
5.1.1 Setting the GPIB Address and RS-232C Parameters
The electronic load is shipped with the GPIB address set to 5. The address can be only
changed via the “SYS” functional list menu (please see section 4.2.9). This menu is also
used to select the RS-232C interface, and specify the parameters such as baud rate and parity.
5.1.2 Wire Connection of RS-232C
The Electronic Load is shipped with the baud rate set to 115200, parity set to is None, and
data length 8 bits, stop1 bit. For RS-232C interface, only TxD and RxD signals are used for
data transfer. The RS-232C connector is a 9-pin D male connector. Following table shows
you the RS-232C connector pins and signals.
Pin No.
Input/Output
Description
1
---
+5V
2
INPUT
RxD
3
OUTPUT
TxD
4
OUTPUT
DTR
5
---
GND
6
INPUT
DSR
7
OUTPUT
RTS
8
INPUT
CTS
9
---
+5V
The connection between the computer (IBM PC compatible) and the Electronic Load is
illustrated below:
5-1
High Power DC Electronic Load 63200 Series Operation & Programming Manual
PIN
1
2
3
4
5
6
7
9
8
IBM PC
DCD
RX
TX
DTR
GND
DSR
RTS
RI
CTS
Electronic Load
No Connection
RX
TX
No Connection
GND
No Connection
No Connection
No Connection
No Connection
5.2
GPIB Capability of the Electronic Load
GPIB Capability
Response
Interface
Functions
Talker/Listener
All electronic load functions are programmable over
the GPIB. The electronic load can send and receive
messages over the GPIB. Status information is sent
using a serial poll.
AH1, SH1,
T6, L4
Service Request
The electronic load will set the SRQ line true if there is
an enabled service request condition.
SR1
Remote/Local
In local mode, the electronic load is controlled from
the front panel but will also execute commands sent
over the GPIB. The electronic load powers up in
local mode and remains there until it receives a
command over the GPIB. Once the electronic load is
in remote mode, REMOTE will be shown on the front
panel LCD, all front panel keys except LOCAL are
disabled. Pressing LOCAL key on the front panel
will return to local mode. Local can be disabled
using local lockout, so only the controller or the power
switch can return to local mode.
RL1
5-2
Basic Information for Programming
5.3
RS-232C in Remote Control
When you use RS-232C in remote control, you have to send the remote command
CONFigure : REMote ON first to let control procedure enter into remote state, and then
execute other command set. When control comes to an end, you have to send the command
CONFigure : REMote OFF to let control procedure return to local operation mode.
The RS-232C control commands are same as those of GPIB. When the RS-232C command
string comes to an end for sending, should be added. Its ASCII code is 0A
hexadecimal (or 10 decimal).
5-3
Introduction to Programming
6.
Introduction to Programming
6.1
Basic Definition
GPIB statement includes instrument control and query commands. A command statement
sends an instruction to the electronic load and query command request information from the
electronic load.
Simple Command
A simple command statement consists of a command or keyword usually followed by a
parameter or data:
LOAD ON
or
LOAD:SHORT OFF
Compound Command
When two or more keywords are connected by colons (:), it creates a compound command
statement. The last keyword usually is followed by a parameter or data:
CURRent:STATic:L1 3
or
CONFigure:VOLTage:RANGe H
Query Command
A simple query command consists of a keyword followed by a question mark:
MEASure:VOLTage?
MEASure:CURRent?
or
CHAN?
Forms of Keywords
Every keyword has two forms:
Long-Form
In this form, the word is spelled out completely to identify its function.
For instance, CURRENT, VOLTAGE, and MEASURE are long-form
keywords.
Short-Form
In this form, the word contains only the first three or four letters of the
long-form. For instance, CURR, VOLT, and MEAS are short-form
keywords.
In keyword definitions and diagrams, the short-form part of each keyword is emphasized in
UPPER CASE letters to help you remember it. However, the electronic load will accept
Volt, volt, voltage, VOLTAGE, volTAGE, etc. regardless to which form you apply. If the
keyword is incomplete, for example, “VOL” or “curre”, it will not be recognized.
6-1
High Power DC Electronic Load 63200 Series Operation & Programming Manual
6.2
Numerical Data Formats
Chroma 63200 electronic load accepts the numerical data type listed in Table 2-1. Numeric
data may be followed by a suffix to specify the dimension of the data. A suffix may be
preceded by a multiplier. Chroma 63200 makes use of the suffixes listed in Table 2-2 and
multipliers listed in Table 2-3.
Symbol
Description
Example
NR1
Digits with no decimal point. The decimal point is
assumed to be to the right of the least-significant digit.
123, 0123
NR2
Digits with a decimal point.
123., 12.3, 0.123, .123
NR3
Digit with a decimal point and an exponent.
1.23E+3, 1.23E-3
NRf
Flexible decimal form that includes NR1 or NR2 or
NR3.
123, 12.3, 1.23E+3
NRf+
Expanded decimal form that includes NRf and MIN,
MAX. MIN and MAX are the minimum and
maximum limit values for the parameter.
123, 12.3, 1.23E+3,
MIN, MAX
Table 2-1 Numerical Data Type
Mode
Class
Preferred Suffix Secondary Suffix
Referenced Unit
CC
Current
A
Ampere
CR
Resistance
OHM
Ohm
CV
Amplitude
V
Volt
CP
Power
W
Watt
ALL
Time
S
Second
ALL
Slew Rate
A/µS
Amperes/micro Second
Table 2-2 Suffix Elements
Multiplier
Mnemonic
Definition
1E6
MA
mega
1E3
K
kilo
1E-3
M
milli
1E-6
U
Micro
1E-9
N
Nano
Table 2-3 Suffix Multipliers
6-2
Introduction to Programming
6.3
Character Data Formats
For command statements, the data format permits entry of required characters. For
query statements, character strings may be returned in either of the forms shown in the
following table. It depends on the length of the returned string.
Symbol
Character Form
Crd
Character Response Data. They permit the return up to 12 characters.
aard
Arbitrary ASCII Response Data. They permits the return of undelimited
7-bit ASCII. This data type is an implied message terminator (refer to
Separators and Terminators).
6.4
Separators and Terminators
In addition to keywords and parameters, GPIB program statements require the following:
Data Separators:
Data must be separated from the previous command keyword by a space. This is shown in
examples as a space (CURR 3) and on diagrams by the letters SP inside a circle.
Keyword Separators:
Keywords (or headers) are separated by colon (:), semicolon (;), or both. For example:
� LOAD:SHOR ON
� MEAS:CURR?;VOLT?
� CURR:STAT:L1 3;:VOLT:L1 5
Program Line Separators:
A terminator to inform GPIB that it has reached the end of a statement. Normally, this is
sent automatically by your GPIB programming statements. Termination also occurs with
other terminator codes, such as EOI. In this manual, the terminator is assumed at the end of
each example line of code. If it needs to be indicated, it is shown by the symbol , which
indicates “new line” and represents the ASCII coded byte as 0A hexadecimal (or 10 decimal).
Traversing the Command Tree:
� The colon “:” separates keywords from each other which represents changes in branch
level to the next lower one. For example:
CONF:VOLT:ON 5
CONF is a root-level command, VOLT is the first branch, and ON is the second branch.
Each “:” moves the command interpretation downward to the next branch.
� The semicolon “;” allows you to combine command statements into one line. It returns
the command interpretation to the previous colon.
6-3
High Power DC Electronic Load 63200 Series Operation & Programming Manual
For example: Combine the following two command statements:
RES:RISE 100 and
RES:L1 400
which can be formed into one command line as follows:
RES:RISE 100;L1 400
5.
To return to the root-level form you can
1. Enter a new-line character. This is symbolized as “” and can be linefeed “LF”
or/and end-of-line “EOL”. Or else,
2. Enter a semicolon followed by a colon “;:”.
Please refer to the following figure.
1. (root):VOLT:L1: 30
Starting a New Line to return to the Root.
2.
(root):SPEC:VOLT:H 30;
:L 5;:
(root)RES:L1 400;
:RISE 1000;:
6-4
Commands for operating the 63200 Electronic Load remotely are grouped into subsystems.
Each command belonging to the same subsystem is arranged in alphabetic order. A syntax
chart of the subsystem that contains the commands in the same group is included. Sub-
systems are ordered alphabetically according to their names in the following sections.
7.1.1
Common Commands
Common commands defined by the IEEE488.2 standard are generic commands and queries.
The first part of the language dictionary covers the commands. Each of them has a leading
“*”.
Common Commands Syntax
*CLS
Clear status
*ESE
Standard event status enable
*ESE?
Return standard event status enable
*ESR?
Return event status enable
*IDN?
Return instrument identification
*OPC
Enable “operation complete” bit in ESR
*OPC?
Return a “1” when operation complete
*RCL
Recall instrument state
*RST
Reset
*SAV
Save instrument state
*SRE
Set service request enable register
*SRE?
Return service request enable register
*STB?
Return status byte
7.1.2
Common Command Dictionary
*CLS
Clear Status Command
Description This command clears the following actions:
1. Clear Questionable Status Event register
2. Clear Standard Event Status Event register
3. Clear Status Byte
4. Clear the Error Queue
Syntax *CLS
Parameters None
7-1
High Power DC Electronic Load 63200 Series Operation & Programming Manual
*ESE
Standard Event Status Enable Command/Query
Description This command sets the condition of the Standard Event Status
Enable register to determine which event the Standard Event Status
Event register (see *ESR?) is allowed to set the ESB (Event
Summary Bit) for the Status Byte register. A "1" in the bit position
enables the corresponding event. All the events enabled by
Standard Event Status Event register are logically ORed to cause the
ESB (bit 5) of the Status Byte register to be set. See the
descriptions of these registers in Chapter 8 Status Reporting.
Syntax *ESE
Parameters 0 to 255
Example *ESE 48
This command enables the CME and EXE events
for the Standard Event Status Event register.
Query Syntax *ESE?
Return Parameters
Query Example *ESE?
This query returns the current setting for "Standard
Event Status Enable".
*ESR?
Standard Event Status Register Query
Description This query reads the Standard Event Status register. Reading
the register clears it. See detailed explanation of this register in
Chapter 8 Status Reporting.
Bit Configuration of Standard Event Status Event Register
Bit Position
7
6
5
4
3
2
1
0
Bit Name
0
0
CME
EXE
DDE
QYE
0
OPC
Bit Weight
128
64
32
16
8
4
2
1
CME = Command error
DDE = Device-dependent error
EXE = Execution error
OPC = Operation complete
QYE = Query error
Query Syntax *ESR?
Return Parameters
Query Example *ESR?
Return the Standard Event Status Register readings.
Return Example 48
*IDN?
Identification Query
Description This query requests the Electronic Frame (632xx) to identify itself.
Query Syntax *IDN?
Return Parameters
Query Example *IDN?
String
Information
Chroma
Manufacture
632xx
Model name
12345678
Serial number
01.00
Firmware version
7-2
Return Example Chroma,63203,12345678,01.00
*OPC
Operation Complete Command
Description This command causes the interface to set the OPC bit (bit 0) of the
Standard Event Status register when the Electronic Frame (6320) has
completed all pending operations.
Syntax *OPC
Parameters None
*OPC?
Operation Complete Query
Description This query returns an ASCII “1” when all pending operations are
completed.
Query Syntax *OPC?
Return Parameters
Query Example 1
*RCL
Recall Instrument State Command
Description This command restores the electronic load to a state that was
previously stored in memory with *SAV command to the
specified location (see *SAV).
Syntax *RCL
Parameters 1 to 100
Example *RCL 50
*RST
Reset Command
Description This command forces an ABORt, *CLS, LOAD:PROT:CLE
command.
Syntax *RST
Parameters None
*SAV
Save Command
Description This command stores the present state of electronic load in a
specified location in memory.
Syntax *SAV
Parameters 1 to 100
Example *SAV 50
*SRE
Service Request Enable Command/Query
Description This command sets the condition of the Service Request Enable
register to determine which event of the Status Byte register (see
*STB) is allowed to set the MSS (Master Status Summary) bit. A
"1" in the bit position enabled by bits is logically ORed to cause Bit
6 (the Master Summary Status Bit) of the Status Byte register to be
set. See details regarding the Status Byte register in Chapter 8
Status Reporting.
Syntax *SRE
7-3
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Parameters 0 to 255
Example *SRE 24
Enable the QUES and MAV bit of the Service
Request Enable
Query Syntax *SRE?
Return Parameters
Query Example *SRE? Return the current setting for "Service Request Enable".
*STB?
Read Status Byte Query
Description This query reads the Status Byte register. Note that the MSS
(Master Summary Status) bit is returned in Bit 6 instead of RQS bit.
This bit indicates if the electronic load has at least one reason for
requesting service. *STB? does not clear the Status Byte register.
It is cleared only when subsequent action has cleared all bits set.
Refer to Chapter 8 Status Reporting for more information about this
register.
Bit Configuration of Standard Byte Register
Bit Position
7
6
5
4
3
2
1
0
Bit Name
0
MSS
RQS
ESB
MAV QUES
0
0
0
Bit Weight
128
64
32
16
8
4
2
1
MSS = master status summary
ESB = event status byte summary
RQS = request for service
QUES = questionable status summary
MAV = message available
Query Syntax *STB?
Return Parameters
Query Example *STB?
Return the contents of "Status Byte".
Return Example 24
7.2
Specific Commands
The 63200 series products are built-in with the following specific GPIB commands.
7-4
Description Set if the load module will do Auto Load On during power-on.
Syntax CONFigure:AUTO:LOAD
Parameters 0 | OFF, 1 | ON
Example CONF:AUTO:LOAD ON
Start Auto Load On during power on.
CONF:AUTO:LOAD OFF Close Auto Load On during power on.
Query Syntax CONFigure:AUTO:LOAD?
Return Parameters 0 | 1
Query Example CONF: AUTO: LOAD?
Return Example 0 (OFF)
Return the status of Auto Load On
CONFigure:AUTO:MODE
Description Set type of Auto Load On as LOAD ON or PROGRAM ON.
Syntax CONFigure:AUTO:MODE
Parameters 0 | PROGRAM, 1 | LOAD
Example CONF:AUTO:MODE LOAD
Set Auto Load On as general
Load On.
CONF:AUTO:MODE PROGRAM
Set Auto Load On as
Program On.
Query Syntax CONFigure:AUTO:MODE?
Return Parameters 0 | 1
Query Example CONF:AUTO:MODE? Return the execution type of
Auto Load On.
Return Example 0 ( PROGRAM)
CONFigure:DISPlay
Description Set measure display mode
Command Syntax CONFigure:DISPlay
7-5
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Parameters 0 | RESISTANCE, 1 | POWER
Example CONF:DISP POWER
Set the power display mode.
CONF:DISP 0
Set resistance display mode.
Query Syntax CONFigure:DISPlay?
Return Parameters 0 | 1
Query Example CONF:DISP?
Return the execution type of display mode.
Return Example 1 (POWER)
CONFigure:KEY
Description Set enter data key mode
Command Syntax CONFigure:KEY
Parameters 0 | NEXT, 1 | HOLD
Example CONF:KEY 0
Set enter data next mode.
CONF:KEY HOLD
Set enter data hold mode.
Query Syntax CONFigure:KEY?
Return Parameters 0 | 1
Query Example CONF:KEY?
Return the execution type to enter data mode.
Return Example 0 ( NEXT )
CONFigure:REMote
Description Set the status of remote control (only effective in RS232C).
Syntax CONFigure:REMote
Parameters 0 | OFF, 1 | ON
Example CONF:REM ON
Set to remote control.
CONFigure:SAVE
Description Store CONFigure data into EEPROM.
Syntax CONFigure:SAVE
Parameters None
Example CONF:SAVE
CONFigure:SOUNd
Description Set the buzzer sound ON/OFF.
Syntax CONFigure:SOUNd
Parameters 0 | OFF, 1 | ON
Example CONF:SOUN 0
CONF:SOUN OFF
Query Syntax CONFigure:SOUNd?
Return Parameters 0 | 1
Query Example CONF:SOUN?
Return the control status of the buzzer sound.
Return Example : 0 (OFF )
CONFigure:VOLTage:LATCh
Description Set the action type to Von.
Syntax CONFigure:VOLTage:LATch
Parameters 0 | OFF, 1 | ON
Example CONF:VOLT:LAT 1
Set the action type to Von as Latch.
7-6
Set the action type toVon as Non Latch
(Refer to the former part of this
manual for detailed).
Query Syntax CONFigure:VOLTage:LATCh?
Return Parameters 0 | 1
Query Example CONF:VOLT:LAT?
Return the action type to Von.
Return Example 0 (OFF)
CONFigure:VOLTage:LATCh:RESet
Description Reset Von signal.
Syntax CONFigure:VOLTage:LATCh:RESet
Parameters None
Example CONF:VOLT:LATC:RES
CONFigure:VOLTage:ON
Description Set the voltage of sink current to on.
Syntax CONFigure:VOLTage:ON [suffix]
Parameters For valid voltage range refer to respective specification.
Example CONF:VOLT:ON 1
Set Von=1V.
CONF:VOLT:ON 300mV
Set Von=300mV.
Query Syntax CONFigure:VOLTage:ON?
Return Parameters [Unit=Voltage]
Query Example CONF:VOLT:ON?
Return the setting of Von value.
Return Example 3.5
CONFigure:VOLTage:POLarity
Description Set the voltage display polarity.
Syntax CONFigure:POLarity
Parameters 0 | MINUS, 1 | PLUS,
Example CONF:POL 0
Set voltage display to minus.
CONF:POL PLUS
Set voltage display to plus.
Query Syntax CONFigure:POLarity?
Return Parameters 0 | 1
Query Example CONF:POL?
Return the display polarity.
Return Example 0 ( MINUS )
CONFigure:VOLTage:RANGe
Description Set voltage measurement range in CC mode.
Syntax CONFigure:VOLTage:RANGEe
Parameters 0 | L, 1 | H
Example CONF:VOLT:RANG 1
Set full-range to High.
CONF:VOLT:RANG L
Set full-range to Low.
Query Syntax CONFigure:VOLTage:RANGe?
Return Parameters 0 | 1
Query Example CONF:VOLT:RANG?
Return Voltage range.
Return Example 1 ( H )
7-7
High Power DC Electronic Load 63200 Series Operation & Programming Manual
7.2.2
COMMunicate Sub-system
COMMunicate
:ADDRess
:GPIB
:RS485
:SERial
:BAUD
:BITS
:PARity
:SBITs
COMMunicate:ADDRess:GPIB
Description Set the GPIB address
Syntax COMMunicate:ADDRess:GPIB
Parameters 1 – 30
Example COMM:ADDR:GPIB 16
Query Syntax COMMunicate:ADDRress:GPIB?
Return Parameters
Query Example COMM:ADDR:GPIB?
Return GPIB address.
Return Example 16
COMMunicate:ADDRess:RS485
Description Set the RS485 address
Syntax COMMunicate:ADDRess:RS485
Parameters 1 – 5
Example COMM:ADDR:RS485 5
Query Syntax COMMunicate:ADDRress:RS485?
Return Parameters
Query Example COMM:ADDR:RS485?
Return RS485 address.
Return Example 5
COMMunicate:SERial:BAUD
Description Set the baud rate
Syntax COMMunicate:SERial:BAUD
Parameters 0 | BAUD_48, 1 | BAUD_96, 2 | BAUD_192,
3 | BAUD_384, 4 | BAUD_576, 5 | BAUD_1152
Example COMM: SER:BAUD 5
COMM: SER:BAUD_1152
Query Syntax COMMunicate:SERial:BAUD?
Return Parameters 0 | 1 | 2 | 3 | 4 | 5
Query Example COMM:SER:BAUD?
Return RS232 baud rate.
Return Example 5 ( BAUD_1152 )
7-8
Description Set the number of the data bits
Syntax COMMunicate:SERial:BITS
Parameters 0 | BITS_8, 1 | BITS_7
Example COMM: SER:BITS 0
COMM: SER:BITS BITS_8
Query Syntax COMMunicate:SERial:BITS?
Return Parameters 0 | 1
Query Example COMM:SER:BITS?
Return RS232 data bits parameter.
Return Example 0 ( BITS_8)
COMMunicate:SERial:PARity
Description Set the parity scheme that is to be used for transmission
Syntax COMMunicate:SERial:PARity
Parameters 0 | NONE, 1 | EVEN, 2 | ODD
Example COMM: SER:PAR 0
COMM: SER:PAR NONE
Query Syntax COMMunicate:SERial:PARity?
Return Parameters 0 | 1 | 2
Query Example COMM:SER:PAR?
Return RS232 bits parameter.
Return Example 0 ( NONE )
COMMunicate:SERial:SBITs
Description Set the number of the stop bits
Syntax COMMunicate:SERial:SBITs
Parameters 0 | SBITS_2, 1 | SBITS_1
Example COMM: SER:SBITS 0
COMM: SER:SBIT SBITS_2
Query Syntax COMMunicate:SERial:SBITs?
Return Parameters 0 | 1
Query Example COMM:SER:SBIT?
Return RS232 stop bits parameter.
Return Example 0
7-9
High Power DC Electronic Load 63200 Series Operation & Programming Manual
7.2.3
CURRENT Sub-system
CURRent
:STATic
:L1
:L2
:RISE
:FALL
:DYNamic
:L1
:L2
:RISE
:FALL
:T1
:T2
CURRent:STATic:L1/L2
Description Set Static Load Current to constant current mode.
Syntax CURRent:STATic:L1 [suffix]
CURRent:STATic:L2 [suffix]
Parameters For valid value range refer to respective specification.
Example CURR:STAT:L1 20
Set Constant Current = 20A for Static
Load L1.
CURR:STAT:L2 10
Set Constant Current = 10A for Static
Load L2.
CURR:STAT:L1 MAX
Set Constant Current = maximum
value for Static Load L1.
CURR:STAT:L2 MIN
Set Constant Current = minimum value
for Static Load L2.
Query Syntax CURRent:STATic:L1?
CURRent:STATic:L2?
CURRent:STATic:L1? MAX
CURRent:STATic:L2? MIN
Return Parameters [Unit=Ampere]
Query Example CURR:STAT:L1?
Return the set current value of Static
Load L1.
Return Example 3.12
CURRent:STATic:RISE/FALL
Description Set the current slew rate for constant current static mode.
Syntax CURRent:STATic:RISE [suffix]
CURRent:STATic:FALL [suffix]
Parameters For valid value range refer to respective specification.
Example CURR:STAT:RISE 2.5
Set rise slew rate to 2.5A/µS for
static load.
7-10
Set fall slew rate to 1A/µS for
static load.
Query Syntax CURRent:STATic:RISE?
CURRent:STATic:FALL?
CURRent:STATic:RISE? MAX
CURRent:STATic:FALL? MIN
Return Parameters [Unit=A/µS]
Query Example CURR:STAT:RISE?
Return the rise slew rate of static
load.
Return Example 2.5
CURRent:DYNamic:L1/L2
Description Set Dynamic Load Current during constant current mode.
Syntax CURRent:DYNamic:L1 [suffix]
CURRent:DYNamic:L2 [suffix]
Parameters For valid value range refer to respective specification.
Example CURR:DYN:L1 20
Set dynamic load parameter L1 = 20A.
CURR:DYN:L2 10
Set dynamic load parameter L2 = 10A.
CURR:DYN:L1 MAX
Set dynamic load parameter L1 = max.
value
CURR:DYN:L2 MIN
Set dynamic load parameter L2 = min.
value.
Query Syntax CURRent:DYNamic:L1?
CURRent:DYNamic:L2?
CURRent:DYNamic:L1? MAX
CURRent:DYNamic:L2? MIN
Return Parameters [Unit=Ampere]
Query Example CURR:DYN:L1?
Return the setting current in dynamic
load L1.
Return Example 35.6
CURRent:DYNamic:RISE/FALL
Description Set current slew rate of constant current dynamic mode.
Syntax CURRent:DYNamic:RISE [suffix]
CURRent:DYNamic:FALL [suffix]
Parameters For valid value range refer to respective specification.
Example CURR:DYN:RISE 2.5
Set rise slew rate to 2.5A/µS.
CURR:DYN:FALL 1A/µS
Set fall slew rate to 1A/µS.
CURR:DYN:RISE MAX
Set rise slew rate to the maximum
value of dynamic load.
CURR:DYN:FALL MIN
Set fall slew rate to the minimum value
of dynamic load.
Query Syntax CURRent:DYNamic:RISE?
CURRent:DYNamic:FALL?
CURRent:DYNamic:RISE? MAX
CURRent:DYNamic:FALL? MIN
Return Parameters [Unit=A/µS]
Query Example CURR:DYN:RISE?
Return the rise slew rate of dynamic
7-11
High Power DC Electronic Load 63200 Series Operation & Programming Manual
load.
Return Example 2.5
CURRent:DYNamic:T1/T2
Description Set the duration parameter T1 or T2 for dynamic load.
Syntax CURRent:DYNamic:T1 [suffix]
CURRent:DYNamic:T2 [suffix]
Parameters For valid value range refer to respective specification.
Example CURR:DYN:T1 10mS
Set dynamic duration T1 = 10mS.
CURR:DYN:T2 2S
Set dynamic duration T2 = 2S.
CURR:DYN:T1 MAX
Set dynamic duration T1 to the
maximum value.
CURR:DYN:T2 MIN
Set dynamic duration T2 to the
minimum value.
Query Syntax CURRent:DYNamic:T1?
CURRent:DYNamic:T2?
CURRent:DYNamic:T1? MAX
CURRent:DYNamic:T2? MIN
Return Parameters [Unit=Sec]
Query Example CURR:DYN:T1?
Return the dynamic duration parameter
T1.
Return Example 0.15
7-12
Description Return the current measured at the input of the load.
Query Syntax FETCh:CURRent?
Return Parameters [Unit=Ampere]
Query Example FETC:CURR?
Return Example 3.15
FETCh:POWer?
Description Return the power measured at the input of the load.
Query Syntax FETCh:POWer?
Return Parameters [Unit=Watt]
Query Example FETC:POW?
Return Example 3.15
FETCh:RESistance?
Description Return the resistance measured at the input of the load
Query Syntax FETCh:RESistance?
Return Parameters [Unit=OHM]
Query Example FETC:RSE?
Return Example 10.22
FETCh:VOLTage?
Description Return the voltage measured at the input of the load.
Query Syntax FETCh:VOLTage?
Return Parameters [Unit=Voltage]
Query Example FETC:VOLT?
Return Example 8.12
FETCh:STATus?
Description Return the real time status of the load.
Query Syntax FETCh:STATus?
Return Parameters
7-13
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Bit Position
15
14
13
12
11
10
9
8
Condition
╳
PR
RS
FF
╳
╳
╳
SP
Bit Weight
╳
16384
8192
4096
╳
╳
╳
256
Bit Position
7
6
5
4
3
2
1
0
Condition
PF
ST
LD
OT
RV
OP
OV
OC
Bit Weight
128
64
32
16
8
4
2
1
Query Example FETC:STAT?
Read back the present status of the load.
Return Example 4
7-14
Description The LOAD command makes the electronic load active on or off.
Syntax LOAD:[STATe]
Parameters 0 | OFF, 1 | ON
Example LOAD ON
Activate the electronic load.
LOAD OFF
Inactivate the electronic load.
Query Syntax LOAD:[STATe]?
Return Parameters 0 | 1
Query Example LOAD?
Return if the electronic load is active.
Return Example 1 (ON)
LOAD:SHORt[:STATe]
Description Activate or inactivate short-circuited simulation.
Syntax LOAD:SHORt:[STATe ]
Example LOAD:SHOR 1
Activate short-circuited simulation.
LOAD:SHOR OFF
Inactivate short-circuited simulation.
Parameters 0 | OFF, 1 | ON
Query Syntax LOAD:SHORt:[STATe]?
Return Parameters 0 | 1
Query Example LOAD:SHOR?
Return the short-circuit simulation state.
Return Example 1 (ON)
LOAD:SHORt:KEY
Description Set the short key mode in the electronic load.
Syntax LOAD:SHORt:KEY TOGGLE
Parameters 0 | HOLD, 1 | TOGGLE
Example LOAD:SHOR:KEY TOGGLE
Set short key mode to Toggle.
LOAD:SHOR:KEY 0
Set short key mode to Hold.
Query Syntax LOAD:SHORt:KEY?
Return Parameters 0 | 1
Query Example LOAD:SHOR:KEY?
Return the short key mode in the
electronic load.
Return Example 1 (TOGGLE)
LOAD:PROTection?
7-15
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Description This command returns the status of the electronic load.
Query Syntax LOAD:PROTection:CLEar?
Return Parameters
Bit Position 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Condition
0 0 0 0 0 0 0 0 0 0 0 OT RV OP OV OC
Bit Weight
16 8 4 2 1
Query Example LOAD:PROT?
Return the status of the electronic load.
Return Example 0
LOAD:PROTection:CLEar
Description This command resets status of the electronic load.
Syntax LOAD:PROTection:CLEar
Parameters For valid value range refer to respective specification.
Example LOAD:PROT:CLE
LOAD:SAVE
Description This command saves the present state of electronic load as default.
Syntax LOAD:SAVE
Parameters None
Example LOAD:SAVE
7-16
Description Return the real time current measured at the input of the electronic
load.
Query Syntax MEASure:CURRent?
Return Parameters [Unit=Amper]
Query Example MEAS:CURR?
Return Example 3.15
MEASure:INPut
Description Select the input port of the electronic load to measure voltage.
Syntax MEASure:INPut
Parameters 0 | LOAD, 1 | UUT
Example MEAS:INP UUT
MEAS:INP LOAD
Query Syntax MEASure:INPut?
Return the input port that has been set.
Return Parameters 0 | 1
Query Example MEAS:INP?
Return Example 0 (LOAD)
MEASure:RESistance?
Description Return the real time resistance measured at the input of the electronic
load.
Query Syntax MEASure:RESistance?
Return Parameters [Unit=OHM]
Query Example MEAS:RES?
Return Example 8.12
MEASure:POWer?
Description Return the real time power measured at the input of the electronic
load.
Query Syntax MEASure:POWer?
Return Parameters [Unit=Watt]
Query Example MEAS:POW?
Return Example 8.12
MEASure:VOLTage?
7-17
High Power DC Electronic Load 63200 Series Operation & Programming Manual
Description Return the real time voltage measured at the input of the electronic
load.
Query Syntax MEASure:VOLTage?
Return Parameters [Unit=Voltage]
Query Example MEAS:VOLT?
Return Example 8.12
7-18