Instant screening for carbon monoxide poisoning.

Contents How to perform a bump test with a non-smoker 2 How to perform a bump test with verified gas 3 How to perform a calibration 4 How to perform a breath test 5 Infection control plays an important role in the medical device in ...

Attach the regulator to the ToxCO® monitor

If it’s been a while since you’ve used your ToxCO®, please check the batteries have charge. Replace them if necessary As routine maintenance, a bump test should always be performed after long periods of no use to check the device is reading correctly. You can perform a bump test using either a non-smoker or ce ...

As prompted, attach the regulator to the

ToxCO® monitor and follow the onscreen instructions to run the calibration Open the gas regulator to perform the test Close the regulator when the calibration is finished To perform a calibration test, firstly turn on the ToxCO® monitor Press the settings icon to open the settings menu Press the cylinder calibration ...

The ToxCO® should be wiped clean

with the alcohol-free antibacterial wipes provided with the monitor before and after each test A new SteriBreath™ mouthpiece should be used for each patient. It is best practice for the patient to insert their own mouthpiece Follow the instructions on the ToxCO® to complete a breath test The patient will be prompted to in ...

Bedfont® Scientific Ltd

Station Road, Harrietsham, Maidstone, Kent, ME17 1JA, England Tel: +44 (0)1622 851122, Fax: +44 (0)1622 854860 Email: [email protected], Web: www.bedfont.com Issue 1 - November 2020, Part No: MKT530 Bedfont® Scientific Limited reserve the right to change or update this literature without prior notice. Registered in: England and Wales ...