Bante Instruments PHscan30S Instruction Manual
Text of the manual
Thank you for selecting the PHscan series pocket pH tester. This manual provides a step-by-step guide to help you operate the instrument, please
carefully read the following instructions before use.
Installing the Batteries
1.
Twist the electrode collar counter clockwise, pull the electrode away from the tester.
2.
Insert the two AAA bat
Switching the Tester On and Off
•
Press and hold the Meas key for about 5 seconds to switch on the tester.
•
Press and hold the Meas key to switch off the tester.
•
To enable the auto-power off function, refer to the Setup Menu section.
Setup Menu
The PHscan30 tester contains 7 menu items in the setup menu, the following table describes the
The PHscan30 tester allows up to 3 points pH calibration, we recommend that you perform at least 2 point calibration for accurate measurement,
the tester will automatically recognize and calibrate to following standard buffer values.
USA standard buffers
pH 4.01, 7.00, 10.01
NIST standard buffers
pH 4.01, 6.86, 9.18
Single point ca
3 Points Calibration
Make sure that you have selected 3 points calibration in the setup menu.
3.1
Repeat steps 1.1 through 1.3 above. When the first calibration point is completed, the display will show "pH 4.01/CAL 2", the tester prompts
you to continue with second point calibration.
3.2
Rinse the electrode with deionized water and place into th
Since pH electrode is susceptible to contamination, thoroughly clean with deionized water as necessary after each use.
•
If your samples contain the oil or grease, soak the electrode in mild detergent or electrode cleaning solution for at least 15 minutes.
•
If you do not use the tester for long periods, store the el
pH 4.01/7.00/10.01 buffer solutions, 480 ml
PHCS-OG
Electrode cleaning solution, removes oil and grease contaminants, 480 ml
PHCS-PR
Electrode cleaning solution, removes protein contamination, 480 ml
PHCS-ES
pH electrode storage solution, 480 ml
Specifications
pH
Model
PHscan30S/F
Range
-1.00 to 15.00 pH
Bante Instruments Inc.
Office: 4715 Castlewood St., Sugar land, TX 77479, USA
Tel: (+1) 346-762-7358
E-mail: [email protected]
Factory: A10, No.2066, Laifang Rd., Shanghai 201615, China
Tel: (+86) 21-6404-1598
E-mail: [email protected]
www.bante-china.com