Sri TCD Quick Start Manual

Text of Sri TCD Quick Start Manual

Thermal Conductivity Detector - TCD
Overview The Thermal Conductivity Detector (TCD) is the most universal detector available. Depending on the compound, the TCD responds with a detection range of 0.01% to 100% (100-1,000,000ppm). The SRI TCD consists of four filaments housed in a stainless steel detector block. The TCD detector block is instal
One of four TCD filaments
TCD filament bridge red/green black/green red/blue black/blue R S S R DETECTORS Thermal Conductivity Detector - TCD Theory of Operation The TCD detector measures the difference in thermal conductivity in the carrier gas flow and the analyte peaks. Every compound possesses some degree of thermal conductivity, and may therefore be me
Carrier: Helium @ 10mL/min
TCD gain = LOW TCD temp = 100oC Temperature program: Initial Hold Ramp Final 80oC 15.00 0.00 80oC Results: Component Retention Area Oxygen 1.633 19.7500 N2 2.150 121.0880 Methane 3.033 563.6130 Ethane 7.550 128.2185 CO2 9.983 11.9860 Propane 13.683 113.9220 Iso-Butane 18.150 69.4960 N-Butane 18.766 67
Thermal Conductivity Detector - TCD
TCD Room Air Analysis TCD Breath Analysis Column: 3’ Silica Gel Carrier: Helium at 10mL/min Sample: 0.5cc room air, direct injection TCD current: LOW TCD temperature: 100oC Temperature Program: Initial Hold Ramp Final 80oC 4.00 0.00 80oC The CO2 content of the room air analyzed is approxima
General Operating Procedure
1. Check to make sure that the TCD filament current is switched OFF. Plug in and turn on your GC. Allow the TCD detector oven to reach temperature (100oC) and stabilize. With the “Display Select” switch in the UP position, press on the TCD Temperature Actual button on the front control panel to read the TCD cell temperatur
All TCD detectors are susceptible to filament damage or destruction if operated at high current in the
absence of carrier and/or reference gas flow. The filaments will incandesce and burn out if the carrier or reference gas flow is interrupted due to a variety of possible factors such as a column break, inadvertent column disconnection during column changes
Standard TCD filament with rubber O-ring gasket
8670-9120 High temperature TCD filament with copper gasket 8690-9123 TCD Troubleshooting When the TCD fails to perform normally, review operating conditions to ensure that carrier gas flow to the detector is unimpeded, and that the column oven temperature, carrier gas flow rate, and carrier gas EPC pressure are
Replacing the TCD Filaments
A. Standard TCD detector block access 1. With a small blade screwdriver, free the filament leads from the bridge terminal by loosening the setscrews. 2. Remove the detector assembly cover by unscrewing the thumbscrew then sliding the cover off toward the right-hand edge of the GC; gently remove the white insulation to reveal t
(Standard TCD detector block access continued)
4. Cut the fiberglass tape wrapped around the detector block and peel it off. Unwrap and remove the heater rope from the detector block (it is probably affixed to the thermocouple wires with more fiberglass tape). 5. Disconnect the thermocouple by loosening the small philips head screw which holds it on the d
Replacing the TCD Filaments continued
B. High temperature TCD detector block access The high temperature TCD assembly is the same as the standard: outer housing around an inner clamshell case. The high temp detector block uses gland nuts and copper gaskets to secure the four filaments in its two cells. Instead of the heater rope, it employs a heating cart

Sri Gas Detectors Owners Instructions: