Sri TCD Quick Start Manual
Text of Sri TCD Quick Start Manual
Thermal Conductivity Detector - TCD
Overview
The Thermal Conductivity Detector (TCD) is the most universal detector available. Depending on the
compound, the TCD responds with a detection range of 0.01% to 100% (100-1,000,000ppm). The SRI
TCD consists of four filaments housed in a stainless steel detector block. The TCD detector block is instal
One of four TCD filaments
TCD filament bridge
red/green
black/green
red/blue
black/blue
R
S
S
R
DETECTORS
Thermal Conductivity Detector - TCD
Theory of Operation
The TCD detector measures the difference in thermal conductivity in the carrier gas flow and the analyte
peaks. Every compound possesses some degree of thermal conductivity, and may therefore be me
Carrier: Helium @ 10mL/min
TCD gain = LOW
TCD temp = 100oC
Temperature program:
Initial
Hold
Ramp
Final
80oC
15.00
0.00
80oC
Results:
Component
Retention Area
Oxygen
1.633
19.7500
N2
2.150
121.0880
Methane
3.033
563.6130
Ethane
7.550
128.2185
CO2
9.983
11.9860
Propane
13.683
113.9220
Iso-Butane
18.150
69.4960
N-Butane
18.766
67
Thermal Conductivity Detector - TCD
TCD Room Air Analysis
TCD Breath Analysis
Column: 3’ Silica Gel
Carrier: Helium at 10mL/min
Sample: 0.5cc room air,
direct injection
TCD current: LOW
TCD temperature: 100oC
Temperature Program:
Initial
Hold
Ramp
Final
80oC
4.00
0.00
80oC
The CO2 content of the room air analyzed is approxima
General Operating Procedure
1. Check to make sure that the TCD filament current is switched OFF. Plug in and turn on your GC. Allow
the TCD detector oven to reach temperature (100oC) and stabilize. With the “Display Select” switch in the
UP position, press on the TCD Temperature Actual button on the front control panel to read the TCD cell
temperatur
All TCD detectors are susceptible to filament damage or destruction if operated at high current in the
absence of carrier and/or reference gas flow. The filaments will incandesce and burn out if the carrier or
reference gas flow is interrupted due to a variety of possible factors such as a column break, inadvertent
column disconnection during column changes
Standard TCD filament with rubber O-ring gasket
8670-9120
High temperature TCD filament with copper gasket
8690-9123
TCD Troubleshooting
When the TCD fails to perform normally, review operating conditions to ensure
that carrier gas flow to the detector is unimpeded, and that the column oven
temperature, carrier gas flow rate, and carrier gas EPC pressure are
Replacing the TCD Filaments
A. Standard TCD detector block access
1. With a small blade screwdriver, free the filament leads from the bridge terminal by loosening the
setscrews.
2. Remove the detector assembly cover by unscrewing the thumbscrew then sliding the cover off
toward the right-hand edge of the GC; gently remove the white insulation to reveal t
(Standard TCD detector block access continued)
4. Cut the fiberglass tape wrapped around the detector block and peel it off. Unwrap and remove the
heater rope from the detector block (it is probably affixed to the thermocouple wires with more
fiberglass tape).
5. Disconnect the thermocouple by loosening the small philips head screw which holds it on the
d
Replacing the TCD Filaments continued
B. High temperature TCD detector block access
The high temperature TCD assembly is the same as the standard: outer housing around an inner
clamshell case. The high temp detector block uses gland nuts and copper gaskets to secure the four
filaments in its two cells. Instead of the heater rope, it employs a heating cart