Prostat PAR-809C Operation & User’s Manual
Text of the manual
PROSTAT® PAR�809C VARIABLE RESISTANCE REFERENCE
Sec� on
Topic
Page
I.
Introduc� on
4
II.
Cau� ons & Warnings
4
III.
Inspec� on & Prepara� on for use
4
IV.
Applica� ons
5
V
General Opera� ons
5
VI.
Maintenance
9
General Specifi ca� ons
10
Table of Contents
Copyright © 2011 by Prostat® Corpora� on. All rights reserved. Printed in the U
I. Introduc� on
A. The Prostat PAR-809C Variable Resistance Refer-
ence is a precision wide range resistance instru-
ment used to measure the resistance range per-
formance of wrist strap testers, footwear testers
and other resistance indica� ng meters, fi xtures
and instruments.
1. The device under test’s resistance perfor-
mance range is indicated b
1. Slide ON/OFF switch to OFF posi� on un� l unit is intended for use.
2. NOTE that the ba� ery provides power to indica� ng lights only.
a. The ba� ery does not aff ect the func� on of the instrument as a resistance reference.
b. The posi� on lights are provided as a convenience for the opera� on of the unit during use.
IV. Applica� ons
A
8. The resistance range posi� ons where the device under test indicates PASS are those resistance
values accepted by the wrist strap tester.
a. Resistance posi� ons below the PASS range should all be rejected by the device under test
as too low.
b. Resistance posi� ons above the PASS range should all be rejected by the device under test
as too high.
B.
3. Rotate the PAR-809C’s Posi� on Selector knob to Posi� on 1, the lowest resistance reference.
4. Turn ON the PAR-809C’S posi� on indica� ng lights by sliding the ON/OFF switch located on the
side of the instrument to the ON posi� on.
5. Insure the Foot Wear device under test is ON and func� onal.
6. Using the second PAR-809C test lead with
4. Plug the two meter test lead banana plugs into the PAR-809C receptacles.
5. Rotate the PAR-809C’s Posi� on Selector knob to Posi� on 1, the lowest resistance reference.
6. Turn ON the PAR-809C’s posi� on indica� ng lights by sliding the ON/OFF switch located on the
side of the instrument to the ON posi� on.
7. Turn the Meter switch to the co