Weinmann MEDUMAT Standard2 Step-by-step Instructions
Text of the manual
• Start ventilation by height
• NIV therapy
• Resuscitation (CPR) with IPPV
• Resuscitation (CPR) with CCSV
• Anesthesia induction (RSI)
MEDUMAT Standard2
Step-by-step instructions
Important:
This document
does not replace the
instructions for use.
Complete information
can be found in
the instructions
for use.
Switch on ventilator
Select “New patient”
3
4
Select patient height and gender
Select “next”
5
6
Select the ventilation mode and check
the displayed ventilation parameters
Start ventilation
Use presets and work according to guidelines
By using the setting option of the Vt/kg body weight (BW) 4-10 ml/kg BW to be
applied, you will be working according to guidelines and determining the calculated
tidal volume for volume-controlled ventilation.
The ideal body weight and thus also the tidal volume to be applied are calculated
differently for male an
Switch on ventilator.
1
Select “New patient” in the start
menu.
2
Set the patient’s height and gender or
select the appropriate patient group:
Adult, Child, Infant.
3
Select the desired CPAP therapy using
the navigation button: PEEP, pMax,
ΔpASB. After adjusting the values,
begin the ventilation via “start”.
5
Now connect the patient to MEDUMAT
Standard². It is possible to adjust the
values using the navigation button
while the ventilation is running. You
can find other ventilation parameters in
the user menu on the
Modified by Prof. Dr. med. Thoralf Kerner
Logistical requirements
Oxygen supply: at least a 2-l bottle, filled ....................................................... Check
Emergency medical team familiar with NIV ..................................................... Check
Clinical requirements
Alert, cooperative (GCS > 12) .................................
In the event of imminent respiratory muscle fatigue, set pressure support (ASB).
Immediate intubation in the absence of clinical improvement or the occurrence of
contraindications!
Caution
• Continuous clinical observation and close patient contact
• No delay in pharmacological therapy or necessary intubation
• Prepared for intubation
The inspiratory trigger triggers a pressure support or a mechanical breath as soon
as inhalation effort is detected.
Trigger level
Corresponding unit value
Level 1 (sensitive)
Approx. 3 l/min
Level 2 (medium)
Approx. 6 l/min
Level 3 (insensitive)
Approx. 10 l/min
Setting the levels of the inspiration trigger:
(If “3 levels” has been
support and the
expiration trigger
Pressure support ΔpASB
The pressure support is always given as a value above PEEP. In addition to the set
PEEP, a patient receives this as soon as the inspiration trigger has been detected.
Example calculation:
PEEP = 5 mbar, ΔpASB = 10 mbar inspiration pressure in the inhalation phase = 15 mbar
Trigger level