Planar IT-081 Operating Manual
Text of Planar IT-081 Operating Manual
TABLE OF CONTENTS
1. PRODUCT FAMILY GENERAL INFORMATION .............................................................. 4
2. GENERAL DESCRIPTION AND PRINCIPLE OF OPERATION....................................... 5
2.1. Introduction.................................................................................................................................................................................5
2.2. Safety Precautions.............................
1. PRODUCT FAMILY GENERAL INFORMATION
IT-08 is a multifunction handheld compact lightweight measurement instrument
incorporating two components. One of the components, MB-08 Base Unit, remains
invariable and provides color display, control and powering (from batteries or mains
adapter) features. The other component, MC-08x, is one of the removable modules, which
determine measurement functions of the whole instrument.
IT-08 is capable to satisfy all your r
2. GENERAL DESCRIPTION AND PRINCIPLE OF OPERATION
2.1. Introduction
This Operating Manual is intended for introducing the design, functions, and basic
instructions related to operation and servicing of the IT-081 TV Level Meter (Meter).
The IT-081 is designed for control and adjustment of television and broadcasting
distribution networks as well as of separate components of such networks, or other
electronic devices. The Meter allows measurement of channel
This Operating Manual is made in accordance with the hardware version 02.80.01 for
MB-08 Base Unit and software version 1.2.x.x/1.2.x.x for IT-081.
2.4. Environmental Conditions
Normal operating conditions:
a) ambient temperature (23±5) 0С;
b) relative air humidity (55±25)%;
c) atmospheric pressure 84-106 kPa (630-795 mm Hg);
d) voltage transients correspond to installation category CAT. II.
Rated operating conditions:
a) ambient
Operating frequency range ...............................................................................5 to 900 MHz
Resolution.................................................................................................................125 kHz
Channel allocation standard .................................... China, PAL B/G, PAL Ireland, PAL UK,
.................................................................................. Secam
2.7. Design and Operation Overview
2.7.1. Principle of Operation
The TV Level Meter is a super heterodyne receiver with triple frequency conversion
(auto or manual frequency tuning). Channel power is measured by means of analog-to-
digital converter (ADC) after signal peak detection at output of logarithmic detector of the
amplifier of the third IF. The principle of spectrum analyzer is based on sequential analysis
method with spectrum indication on the LC
Logarithmic Gain Amplifier Limiter (LGAL) performs logarithmation and signal
detection.
Peak Detector (PD) allows measuring the video carrier level by means of Analog-to-
Digital Converter (ADC). The digital code of the input signal peak level logarithm is defined
as real value and is corrected in accordance with the calibration matrix of the Control Unit
(CU) microcontroller.
The Control Unit receives the commands from the user entered via Keypad,
process