TESTO 510 Short Manual
Text of TESTO 510 Short Manual
Short manual testo 510 11
Short manual testo 510
1 Protection cap: Park position
2 Differential pressure sensor nipple
connection
3 Display
4 Control keys
5 Battery compartment, holding
magnets (on rear)
Basic settings
Instrument off >press and hold
2s > select with
(
),
confirm with
(
):
Unit of pressure/velocity: hPa, mbar, Pa, mmH2O, mmHg, psi, inH
Safety and the enviroment
12
Safety and the
environment
About this document
> Please read this documentation through carefully and
familiarise yourself with the product before putting it to use.
Keep this document to hand so that you can refer to it when
necessary. Hand this documentation on to any subsequent
users of the product.
> Pay particular attention
Functions and use
The testo 510 is a differential pressure measuring instrument. It
is normally used to measure small differences in pressure (e.g.
to check filter permeability), gas flow pressure measurement,
fine draught measurement as well as flow velocity measurement
with a Pitot tube.
Technical data
Measurement data
· S
At a glance
1 Protection cap: Park position
2 Differential pressure sensor nipple
connection
3 Display
4 Control keys
5 Battery compartment, holding
magnets (on rear)
First steps
- Inserting batteries:
1 To open the battery compartment, push the battery cover
down.
2 Insert batteries (2x 1.5 V type AAA). Observe t
Using the product 15
Using the product
For velocity to be measured, the unit of density and density
value must be set correctly, see chapter ‚First steps‘, section
‚Basic settings (configuration mode)‘.
- Switching the instrument on:
> Press
.
- Measuring mode is opened.
- Switching the display light on:
The instrument is switched on.
> Press
.
- Th
Using the product
16
- Timed mean value calculation for flow:
Function only available when unit m/s or fpm is set.
1 Press
several times, until Hold and Avg appear in the
display. The last result of mean value calculation is
displayed.
- If mean value calculation was carried out since the last
time the instrument was switched on, the last result is
displa