Tektronix TDS3000B Series Reference

Specifications

Brand: Tektronix
Category: Test Equipment
Model: TDS3000B Series
File Type: PDF
File Size: —
Pages: 25

Read Online

Read the Tektronix TDS3000B Series Reference online using the built-in PDF viewer.

Text of the manual

Copyright � Tektronix, Inc. All rights reserved.

Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved. Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077 TEKTRONIX, TEK, TEKPROBE, and Tek Secure are registered trademarks of Tektronix, Inc. DPX, WaveAlert, and e*Scope are trademarks of Tektronix, Inc. Reference TDS3000B Series Digital Phosphor Oscilloscopes 071-0923-00 ������������ 071092300

Tektronix warrants that the products that it manufactures and

sells will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment from an authorized Tektronix distributor. If a product proves defective within the respective period, Tektronix will provide repair or replacement as described in the complete warranty statement. To arrange for service or obtain a copy of the complete warranty statement, please contact your nearest Tektronix sales and service office. EXCEPT AS PROVIDED IN THIS SUMMARY OR THE APPLICABLE WARRANTY STATEMENT, TEKTRONIX MAKES NO WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL TEKTRONIX BE LIABLE FOR INDIRECT, SPECIAL OR CONSEQUENTIAL DAMAGES. Contacting Tektronix Phone 1-800-833-9200* Address Tektronix, Inc. Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA Web site www.tektronix.com Sales support 1-800-833-9200, select option 1* Service support 1-800-833-9200, select option 2* Technical support Email: [email protected] 1-800-833-9200, select option 3* 1-503-627-2400 6:00 a.m. – 5:00 p.m. Pacific time * This phone number is toll free in North America. After office hours, please leave a voice mail message. Outside North America, contact a Tektronix sales office or distributor; see the Tektronix web site for a list of offices.

To avoid potential hazards, use this product only as

specified in the TDS3000B Series Digital Phosphor Oscilloscope User Manual which is located on the TDS3000B Series Digital Phosphor Oscilloscope Manuals and Application Notes CD. While using this product, you may need to access other parts of the system. Read the General Safety Summary in the user manual and in other system manuals for warnings and cautions related to operating the system. Preventing Electrostatic Damage CAUTION. Electrostatic discharge (ESD) can damage components in the oscilloscope and its accessories. To prevent ESD, observe these precautions when directed to do so. Use a Ground Strap. Wear a grounded antistatic wrist strap to discharge the static voltage from your body while installing or removing sensitive components. � Contents Safety Summary 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . TDS3000B Initial Setup 5 . . . . . . . . . . . . . . . . . . . . . . . QuickMenus 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Acquire Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 XY Cursor Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 YT Cursor Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Display Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Measure Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Save/Recall Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Trigger Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Utility Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Vertical Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 e*ScopeE Web-based Remote Control . . . . . . . . . . . . 41 Automatic Measurements . . . . . . . . . . . . . . . . . . . . . 44

Use a Safe Work Area. Do not use any devices capable of

generating or holding a static charge in the work area where you install or remove sensitive components. Avoid handling sensitive components in areas that have a floor or benchtop surface capable of generating a static charge. Handle Components Carefully. Do not slide sensitive components over any surface. Do not touch exposed connector pins. Handle sensitive components as little as possible. Transport and Store Carefully. Transport and store sensitive components in a static-protected bag or container. Manual Storage The oscilloscope front cover has a convenient place to store this manual. 4 More Information Refer to the TDS3000B Series Digital Phosphor Oscilloscope User Manual, which is located on the TDS3000B Series Digital Phosphor Oscilloscope Manuals and Application Notes CD, for more information about using this product.

Connect the oscilloscope probe

to channel 1. Attach the probe tip and reference lead to the PROBE COMP connectors. CH 1 PROBE COMP 4. Push the AUTOSET button. You should see a square wave in the display (approximately 5 V at 1 kHz). Probe Compensation 5. Check the shape of the dis- played waveform. Overcompensated Compensated correctly Undercompensated 6. Adjust your probe if necessary to achieve correct compensa- tion. L � TDS3000B Initial Setup The following procedures describe how to quickly verify that the oscilloscope is powering up and functioning correctly, compensate passive probes using the built-in compensation signal, and run the signal path compensa- tion (SPC) routine for maximum signal accuracy. � You should perform all initial setup procedures the first time you use the oscilloscope. � You should perform the probe compensation procedure whenever you attach a passive probe for the first time to any input channel. � You should run the signal path compensation routine whenever the ambient temperature changes by 10� C or more. Functional Check 1. Connect the oscilloscope power cable 2. Turn on the oscilloscope. Wait for the confirmation that the oscilloscope has passed all self-tests. On/Standby button

Signal Path Compensation (SPC)

The signal path compensation (SPC) routine optimizes the oscilloscope signal path for maximum measurement accuracy. You can run the routine anytime but you should always run the routine if the ambient temperature changes by 10� C or more. To compensate the signal path, do the following steps: 1 Disconnect all probes or cables. 2 Push the UTILITY button. 3 Push the System screen button to select Cal. 4 Push the Signal Path screen button. 5 Push OK Compensate Signal Path. Setting the User Interface Language To set the user interface language, do the following steps: 1 Push the UTILITY button. 2 Push the System screen button to select Config. 3 Push a side menu button to select a language. � Adjusting Time and Date To set your oscilloscope to the current date and time, do the following steps: 1 Push the UTILITY button. 2 Push the System menu button to select Config. 3 Push the Set Date & Time menu button. 4 Use the side menu buttons and general purpose knob to adjust the date and time values. 5 Push the OK Enter Date/Time menu button to set the instrument date and time. User Manual The TDS3000B Series DPO Oscilloscope Manuals and Application Notes CD contains the TDS3000B Series Digital Phosphor Oscilloscope User Manual, translated into all supported languages. The user manual contains complete information on how to operate your TDS3000B Series oscilloscope. The CD also contains manuals for all TDS3000B Series accessories, application notes, oscilloscope and probe fundamentals, and software to install the Adobe Acrobat Reader, which is required to view the online documents.

To use a QuickMenu, do these steps

1 Push the QUICKMENU button on the front panel. 2 Push the screen button that corresponds to the control you need to set. Push the screen button repeatedly to choose one of the settings. The small arrow icon indicates that there are additional settings available that are not shown. You can use most of the front-panel controls at the same time you are using a QuickMenu. For example, if you push a channel button to select a different channel, the QuickMenu changes to show information about that channel. You can still use the regular menus. For example, if you push the MEASURE button, you can set up and take automatic waveform measurements in the usual way. If you return to the QuickMenu, the measurement is still displayed on the screen. You may have optional application modules installed that also have QuickMenu displays. To select the Quick- Menu you want to use, push the MENU screen button. This menu item is shown only if application modules that contain a QuickMenu are installed. � QuickMenus The QuickMenu feature simplifies the use of the oscillo- scope. When you push the QUICKMENU button, a set of frequently used menu functions show on the display. Scope is one type of QuickMenu that you can use to control the basic oscilloscope functions. You can per- form many tasks without using the regular menu system.

The following sections briefly describe each oscilloscope

menu item. Acquire Menu Bottom Side Description Mode Sample Use for normal acquisition. Peak Detect Detects glitches and reduces the possibility of aliasing. Envelope N Captures variations of a signal over a period of time. (Adjust N with the general purpose knob.) Average N Reduces random or uncorre- lated noise in the signal display. (Adjust N with the general purpose knob.) Horizontal Resolution Fast Trigger (500 points) Acquires 500 point waveforms at a fast repetition rate. Normal (10k points) Acquires 10,000 point wave- forms with more horizontal detail. Reset Hori- zontal Delay Set to 0 s Resets the horizontal delay to zero. �� Bottom Description Side Autoset Normal Autoset Executes the autoset function. (Optional application modules may add choices to execute specialized autoset functions.) Undo Autoset Reverts to the settings before the last autoset. WaveAlert Wfm Anomaly Detection On Off Enables or disables the WaveAlert Waveform Anomaly Detection function. See page 13. Sensitivity nn.n% Sets the comparison sensitivity. Use the general purpose knob to set sensitivity from 0% (least sensitive) to 100% (most sensi- tive). Beep On Anomaly On Off When On, the oscilloscope beeps if it detects a waveform anomaly on any active channel. Stop On Anomaly On Off When On, the oscilloscope stops waveform acquisition if there is a waveform anomaly on any channel. The input wave- forms and anomalies remain displayed on the screen.

XY Cursor Menu

The following cursor menu items are available in the XY display mode (DISPLAY > XY Display > Triggered XY). Push the CURSOR button to show the cursor menu. Bottom Side Description Function Off Turns cursors off. Waveform Turns waveform cursors on and displays measurements in rectangular format (X and Y values). Use the front-panel SELECT button to select a cursor. Use the general purpose knob to move the active cursor. Mode Independent Sets cursors to move indepen- dently. Tracking Sets cursors to move together when the reference cursor is selected. NOTE. The Advanced Analysis application module (TDS3AAM) adds more XY cursor features, including graticule cursors and polar readouts. �� Bottom Description Side WaveAlert (cont.) Hardcopy On Anomaly On Off When On, the oscilloscope sends an image of the screen to the hard copy device or floppy file if there is a waveform anom- aly on any channel. Wfms To Disk On Anomaly On Off When On, the oscilloscope saves the anomalous wave- form(s) data to the floppy disk drive. Highlight Entire Wfm Highlights the entire anomalous waveform. Highlight Anomaly Highlights just the anomalous data in a waveform. WaveAlert� Waveform Anomaly Detection. WaveAlert provides a way to detect when a waveform deviates from a steady-state condition. WaveAlert monitors the current waveform acquisition and compares it to the previous waveform acquisition. If the current acquisition exceeds the comparison tolerance, the oscilloscope considers the current acquisition as an anomaly and responds by stopping acquisitions, emitting a beep, saving the anomalous waveform to a disk file, printing the screen image to a hard copy device, or any combination of the above. Refer to the TDS3000B Series Digital Phosphor Oscilloscope User Manual for more information.

YT Cursor Menu

The following YT cursor menu items are available in the YT display mode (DISPLAY > XY Display > Off (YT)). Push the CURSOR button to show the cursor menu. Bottom Side Description Function Off Turns cursors off. H Bars Use to take vertical measure- ments. V Bars Use to take both vertical and horizontal measurements. Bring Selected Cursor to Center Screen Moves the active cursor to center screen. Bring Both Cur- sors On Screen Moves any off-screen cursor onto the screen. Mode Independent Sets cursors to move indepen- dently. Tracking Sets cursors to move together when cursor 1 is selected. �� Bottom Side Description V Bar Units Sec (s) / 1/sec (Hz) Sets horizontal units to se- conds or frequency (Hz). Ratio (%) Sets V Bar measurement units to percent. Phase (degrees) Sets V Bar measurement units to degrees. Use cursor posi- tions as %/° Sets V Bar measurement scale so that 0% or 0° is the current position of left V Bar cursor and 100% or 360° is the current position of right V Bar cursor. Use 5 divs as %/° Sets V Bar measurement scale so that 5 screen major divisions is 100% or 360°, where 0% or 0° is –2.5 divi- sions and 100% or 360° is +2.5 divisions from center graticule.

Set to on to see dots only. Set

to off to see dots and vectors. Persist Time Sets persist time. Set to Auto Sets the WAVEFORM INTENSITY knob to control the persistence time. Clear Persistence Clears any displayed persis- tence. Backlight In- tensity High Use for bright ambient condi- tions. Medium Use for dim ambient conditions. Low Use to extend battery-operation time. Graticule Full, Grid, Cross Hair, Frame Selects the graticule type. �� Bottom Side Description H Bar Units Base Sets the H Bar units to be the same as the selected wave- form’s vertical measurement units (volts, IRE, dB, and so on). Ratio (%) Sets H Bar measurement units to percent. Use cursor positions as 100% Sets the H Bar measurement scale so that 0% is the current position of lowest H Bar cursor and 100% is the current posi- tion of highest H Bar cursor. Use 5 divs as 100% Sets H Bar measurement scale so that 5 screen major divisions is 100%, where 0% is –2.5 divisions and 100% is +2.5 divisions from center graticule. Cursor Movement. Use the general purpose knob to move the active cursor. Push the SELECT button to change which cursor is active. Faster Cursor Movement. Push COARSE to set the general purpose knob to make faster cursor movements.

Turns off an XY display.

Triggered XY Turns on triggered XY display. Gated XYZ Displays XY signals when Z-channel signal is above a set level. Only available on 4-chan- nel instruments. Ch1 (X) Versus Sets Ch2, Ch3, or Ch4, as Y versus Ch1 as X. Ref1 (X) Versus Sets Ref2, Ref3, or Ref4 as Y versus Ref1 as X. Gated By Sets Ch2, Ch3, or Ch4 as Z channel (gate) source, and gating channel threshold level. Color Palette Normal Selects the color display. Monochrome Sets all waveforms to high-con- trast black and white. XY Waveforms. For XY waveforms, channel 1 or Ref 1 is displayed in the horizontal axis. Use the Vertical POSI- TION and SCALE controls to adjust the horizontal position and size of the XY waveform. �� Another channel or reference waveform is displayed in the vertical axis. Use the Vertical POSITION and SCALE controls for that channel to adjust the vertical position and size of the XY waveform. XY Waveform Triggering. The XY waveform is triggered so you can synchronize periodic input signals to the XY waveform. This feature is useful when only one part of the period contains valid information that you want to see in the XY format. Set the time base and trigger location to acquire just that part of the period. When you want to see the complete period of the signals regardless of the time base setting, set the trigger source to an unused channel and the trigger mode to Auto. Gated XYZ. Displays XY signals only when the Z (gating) channel is true. Gated XYZ is similar to analog oscillo- scope modulated XYZ mode except that the displayed XY signal is either on or off; there is no intensity modu- lation. Gated XYZ is useful for displaying constellation diagrams. The general purpose knob sets the Z-channel threshold level.

Moves the active cursor to

center screen. Bring Both Cur- sors On Screen Moves any off-screen cursor onto the screen. High-Low Setup Auto Select Automatically uses the best measurement method depend- ing on the measurement type. Histogram Use to measure pulses. Min-Max Use to measure other wave shapes. �� Measure Menu Bottom Side Description Select Measurement Select an automatic measure- ment for the selected (active) channel. You can display up to four measurements at one time. See page 44. Remove Measurement Measurement 1-4 Removes a specific measure- ment. All Measurements Removes all measurements. Gating Off Use to take measurements on the full waveform record. Screen Use to take measurements on the portion of the waveform on screen. Cursors Use to take measurements on the portion of the waveform between the V Bar cursors.

Set Levels in %

or units Use to choose custom refer- ence levels in relative or abso- lute units. High Reference Sets custom high reference level. Mid Reference Sets custom mid reference level. Mid 2 Sets the custom mid reference level for the Delay and Phase measurement second wave- form. Low Reference Sets custom low reference level. Set to Defaults Sets reference levels to default values. Indicators Measurement 1-4 Selects the measurement for which to display markers that show the portion of the wave- form used to calculate the measurement value. Off Turns measurement indicators off. �� Save/Recall Menu Bottom Side Description Save Current To File Saves a setup to disk. Setup To Setup 1 ... To Setup 10 Saves a setup to nonvolatile memory. Recall Saved From File Recalls a setup from disk. Setup Recall Setup 1 ... Recall Setup 10 Recalls a setup from nonvolatile memory. Recall Factory Setup OK Confirm Factory Init Initializes the setup to a known setup. Save Wfm To File Shows file-format choices: Internal, Spreadsheet, or Math- cad. Saves active waveforms to disk. To Ref1 ... To Ref4 Saves the selected waveform to nonvolatile memory. Recall Wfm From File Recall Ref1 ... Recall Ref4 Recalls a waveform from disk and displays it as a reference waveform.

Edge Trigger Menu

Bottom Side Description Source Ch1 ... Ch4 Sets the trigger source to a specific channel. AC Line Selects the AC line trigger source (for AC line operation only) Ext Sets the oscilloscope to trigger on the external trigger source. Ext/10 Ext/10 attenuates the external trigger signal by a factor of 10. Ext Probe nnX Voltage / Current (4-channel only) Set this value to match the attenuation factor and the type of probe (voltage or current) that is attached to the external trigger connector. Push the menu button to select the probe type. Use the general purpose knob to set the attenuation factor. The default values are 1x and voltage. �� File Utilities Menu Bottom Side Description File Delete Deletes a file. Utilities Rename Names or renames a file. Copy Copies a file to another directo- ry. Print Prints a file to a printer con- nected to one of the hard copy ports. Create Directory Creates a new directory. Confirm Delete Turns On or Off a confirmation message before files are de- leted. Overwrite Lock Sets file overwrite protection to On or Off. Format Formats a disk (erases all files). Labels Create/edit labels for reference waveforms and instrument setups stored in nonvolatile memory.

Sets the trigger source to the

lowest-numbered active channel in the display. Alternating (all active chan- nels) Sequentially uses each active channel as a trigger source, from the lowest-numbered active channel to the highest-numbered active channel. Coupling DC Selects DC coupling. HF Reject Rejects frequencies above 30 kHz in the trigger signal. LF Reject Rejects frequencies below 80 kHz in the trigger signal. Noise Reject DC coupling with low sensitivity to reject noise in the trigger signal. Slope / (rising edge) Triggers on the rising edge of a signal. \ (falling edge) Triggers on the falling edge of a signal. �� Bottom Description Side Level Level Use to set the trigger level with the general purpose knob. Set to TTL Sets the trigger level to +1.4 V for TTL logic. Set to ECL Sets the trigger level to –1.3 V for ECL logic (Vee = –5.2 V). Set to 50% Sets the trigger level to the 50% amplitude level of the signal. Mode & Holdoff Auto (untriggered roll) Enables free-running and roll- mode acquisitions. Normal Triggers only on valid trigger events. Holdoff (time) Sets holdoff to a specific time. Holdoff (% of record) Sets holdoff to a percent of the record duration. Set to Min Sets holdoff to the minimum value.

B Trigger Menu

Bottom Side Description B Trigger After A B Trigger After A Time Sets the oscilloscope to trigger on the next B-trigger event that occurs after a specified period of time from the A trigger. Use the general purpose knob to set the time value. Set to Delay Time (B�M) then Set (B�M) to 0s Sets the B Trigger After A Time value to the horizontal B�M value, then sets B�M to zero seconds. B�M is the delay time from the B trigger point to the expansion point (center screen). Set to Min Sets B Trigger After A time to 26.4 ns. B Events Sets the oscilloscope to trigger on the nth B-trigger event after the A trigger. Use the general purpose knob to set the event value. Set to Min Sets the B events count to 1. �� Video Trigger Menu Bottom Side Description Standard 525/NTSC Triggers on a NTSC signal. 625/PAL Triggers on a PAL signal. SECAM Triggers on a SECAM signal. Source Selects the trigger source. See Edge Trigger Menu for descrip- tions of these menu items. Trigger On Odd Triggers on odd or even fields in Even an interlaced signal. All Fields Triggers on any field in an interlaced or noninterlaced signal. All Lines Triggers on all lines. NOTE. The Extended Video application module (TDS3VID) adds more video features including custom video triggering, video picture, vectorscope, and analog HDTV triggering.

Sets Source, coupling, slope,

Coupling and level for the B trigger. These settings are independent Slope of similar settings for the A trigger. See Edge Trigger Menu Level for descriptions of these menu items. Utility Menus Configuration System Menu Bottom Side Description Language English Use to choose your native Français language. Most on-screen text appears in the language you Deutsch appears in the language you choose. Italiano Español Português (Russian) (Japanese) (Korean) (Simplified Chinese) (Traditional Chinese) �� Bottom Description Side Set Date & Time Display Date/ Time Use to turn the date/time display On or Off. Hour Min Use to set the internal clock with the current hour and min- ute. Month Day Use to set the internal clock with the current month and day. Year Use to set the internal clock with the current year. OK Enter Date/Time Sets the date and time of the internal clock. Battery Time-Outs Power Off Time-Out Use to set the time before an automatic shut down. Backlight Time-Out Use to set the time before the backlight automatically turns off. Tek Secure Erase Memory Erases all nonvolatile waveform and setup memory. Version Use to see the firmware ver- sion.

Sets oscilloscope Ethernet

parameters such as the IP address, instrument name, and domain name. DHCP/ BOOTP Enables or disables DHCP/ BOOTP support. Debug Enables or disables a debug message window. Test Connection Tests the connection to your network. Ethernet Print- Add Printer Adds, renames, or deletes an er Settings Rename Printer Ethernet network printer from the oscilloscope printer list. Delete Printer Confirm Delete Enables or disables displaying a printer name delete confirmation message. More Information. Refer to the TDS3000 & TDS3000B Series Digital Phosphor Oscilloscope Programmer Manual for more information. �� I/O System Menu Bottom Side Description GPIB (TDS3GV) Talk/Listen Ad- dress Sets the GPIB address. Hard Copy (Talk Only) Sets the GPIB port to talk only for making hard copies. Off Bus Disables the GPIB port. Debug Enables and disables a mes- sage window to help you debug GPIB problems. RS-232 (TDS3GV) Baud Rate Sets the baud rate in steps from 1200 to 38400. Flagging Use to enable hard flagging (RTS/CTS) or turn flagging off. EOL Selects end-of-line terminator. Debug Enables and disables a mes- sage window to help you debug RS-232 problems. Set RS-232 Parameters to Default Values Sets baud rate = 9600, hard flagging = on, and EOL = LF.

Hard Copy System Menu (Utility > System: Hard Copy )

Bottom Side Description Format Choose the correct hard copy format for your printer. Options Portrait Prints in a portrait orientation. Landscape Prints in a landscape orienta- tion. Compression Turn on or off hard copy file .gz compression format. Ink Saver Off Prints the display colors. On Prints colors optimized for printing on white paper. Preview Shows a preview of the hard copy colors on the display. Port Centronics Uses the parallel printer port. GPIB Uses the GPIB port. RS-232 Uses the RS-232 port. File Saves the hard copy to the floppy disk. Ethernet Uses the Ethernet LAN port. Clear Spool Stops a hard copy in progress. Clears the print spooler. �� Calibration System Menu Bottom Side Description System Cal Signal Path Compensates the signal paths to obtain best measurement accuracy. Factory Cal Used to calibrate the oscillo- scope. This is a service function only. Cal Due Control Notify After Hours of Op- eration Sets the number of hours of operation before notifying you that a calibration is due. Notify After Years of Elapsed Time Sets the number of years before notifying you that a calibration is due. Signal Path Compensation. You can run the Signal Path Compensation routine anytime but you should always run the routine if the ambient temperature changes by 10 °C or more.

Sets input coupling to DC.

AC Sets input coupling to AC. GND Provides a 0 V signal reference. The input BNC disconnects from internal circuits. � Sets input resistance to 50 � or 1 M�. Invert Invert Off Use for normal operation. Invert On Inverts the polarity of the signal in the display. Bandwidth Full Bandwidth Sets bandwidth to the full oscilloscope bandwidth. 150 MHz Sets the bandwidth to 150 MHz (not available on some models). 20 MHz Sets the bandwidth to 20 MHz. Fine Scale Enables fine scale adjustment with the general purpose knob. �� Diagnostic System Menu Bottom Side Description System Diags Execute Starts diagnostics. Loop Once Executes the diagnostic loop once. Always Executes the diagnostic loop continuously. Until Fail Executes until a failure occurs. Error Log Page Up Use to see the previous error log page. Page Down Use to see the next error log page.

Enables numerical vertical

position adjustment. Set to 0 divs Sets vertical position to center screen. Offset Vertical Offset Enables vertical offset adjust- ment with the general purpose knob. Set to 0 V Sets vertical offset to 0 V. Probe Setup Voltage Probe Use to set probe gain or attenu- Current Probe ation for probes that do not have the TekProbe II interface. Deskew Use to adjust the time skew correction for each probe. Set to 0 Use to set probe skew correc- tion to zero. �� Math Menu Bottom Side Description Dual Wfm Math Set 1st Source to Selects the first source wave- form. Set Operator to Selects the math operator: +, –, �, or � Set 2nd Source to Selects the second source waveform. Ref Menu Bottom Side Description Ref 1 Save Ch1 to Ref1 Saves channel 1 to reference waveform Ref 1. Save Ch2 to Ref1 Saves channel 2 to reference waveform Ref 1. Save Ch3 to Ref1 Saves channel 3 to reference waveform Ref 1. Save Ch4 to Ref1 Saves channel 4 to reference waveform Ref 1. Save Math to Ref1 Saves the math waveform to reference waveform Ref 1. Ref 2 Ref 3 Ref 4 Identical settings for reference waveforms Ref 2, Ref 3, and Ref 4.

Configuring Your Oscilloscope Ethernet Settings

Before using the e*Scope feature, you must set your oscilloscope Ethernet network settings. The TDS3000B Series Digital Phosphor Oscilloscope User Manual, Appendix G, Ethernet Setup, describes how to enter the Ethernet network parameters for your oscilloscope. Accessing e*Scope After your instrument is set up with the correct ethernet settings, you are ready to access that instrument over the Internet. To access you instrument’s e*Scope server, do the following steps: 1 On a PC or work station, open a network browser program. 2 In the Location or Address field (where you normally enter a URL), type in the IP address of the TDS3000B instrument to which you want to connect. For example, 188.121.212.107. Do not precede the IP address with any characters. Then press the Return key. 3 The browser program loads the instrument’s e*Scope Web-based Remote Control home page, which is an image of the screen contents when your browser accessed the instrument. �� e*Scope Web-based Remote Control e*Scope introduces the next generation of oscilloscope functionality. With e*Scope you can access any Internet- connected TDS3000B Series oscilloscope from a browser on your work station, PC, or laptop computer. No matter where you are, your TDS3000B is as close as the nearest browser! There are two levels to e*Scope: basic and advanced. The basic level, which is hosted on the oscilloscope, lets you view a screen image of the current acquisition, save or load waveform and setup files, and remotely control the oscilloscope using the oscilloscope programming commands. The advanced level, which is hosted on your system, provides a graphical user interface to remotely control the oscilloscope. To obtain the advanced level software, go to the Tektronix TDS3000B website and download the web-based remote control software. NOTE. You can also use e*Scope on TDS3000 Series oscilloscopes that are equipped with a TDS3EM Ethernet communication module and are running firmware version 3 or above.

The menu at the top of the page provides the following

functions: � Home displays the oscilloscope screen. � Applications takes you to the application URL specified in the Configure tab. The default URL is the application notes site at www.tektronix.com. � Configure lets you specify URLs for the advanced Web-based Control HTML pages (Control), application notes (Applications), and help files (Help). The default values are for the Tektronix web site. However, you can host any of these on your local Intranet for faster access and better system security. � Data lets you upload or download waveform data files and oscilloscope settings, as well as remotely control the oscilloscope using oscilloscope programing commands. � Status displays the version screen, which shows the firmware version and lists installed application modules. � Control displays the advanced Web-based Control page that lets you remotely control the oscilloscope. �� Automatic Measurements Measure- ment Definition Amplitude Measured over the entire waveform. Amplitude = High (100%) – Low (0%) Burst Width The duration of a burst. Measured over the entire waveform. Cycle Mean The arithmetic mean over the first cycle in the waveform. Cycle RMS The true Root Mean Square voltage over the first cycle in the waveform. Delay The time between the Mid Reference crossings of two different waveforms. Fall Time Time that the falling edge of the first pulse in the waveform takes to fall from 90% to 10% of its amplitude. Frequency Reciprocal of the period of the first cycle in the waveform. Measured in Hertz (Hz). High The value used as 100%. Calculated using either the min/max or the histogram method. Measured over the entire waveform. Low The value used as 0%. Calculated using either the min/max or the histogram method. Measured over the entire waveform.

Positive Duty Cycle � Positive Width

Period � 100% Measurement of the first cycle in the waveform. Positive Overshoot Measured over the entire waveform. Positive Overshoot � Max–High Amplitude � 100% Positive Width Measurement of the first positive pulse in the waveform. The time between the 50% amplitude points. Rise Time Time that the leading edge of the first pulse in the waveform takes to rise from 10% to 90% of its amplitude. RMS The true Root Mean Square voltage over the entire waveform. Snapshot All Mea- surements Displays all measurement values (except two-chan- nel measurements) at the time you pushed the Snapshot All Measurements button, for the selected waveform. Push the Snapshot All Measurements button again to update the list values. Push the MENU OFF button to clear the list from the screen. �� Measure- ment Definition Max The maximum amplitude. The most positive peak voltage measured over the entire waveform. Mean The arithmetic mean over the entire waveform. Min The minimum amplitude. The most negative peak voltage measured over the entire waveform. Negative Duty Cycle Negative Duty Cycle � Negative Width Period � 100% Measurement of the first cycle in the waveform. Negative Overshoot Measured over the entire waveform. Negative Overshoot � Low–Min Amplitude � 100% Negative Pulse Width Measurement of the first negative pulse in the waveform. The time between the 50% amplitude points. Phase The amount one waveform leads or lags another in time. Expressed in degrees, where 360� comprise one waveform cycle. Pk–Pk Measured over the entire waveform. Peak-to-peak = Max – Min Period Time it takes for the first complete signal cycle to complete in the waveform. Measured in seconds.