Copyright � Tektronix, Inc. All rights reserved.
Tektronix products are covered by U.S. and foreign patents,
issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and
price change privileges reserved.
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
TEKTRONIX, TEK, TEKPROBE, and Tek Secure are
registered trademarks of Tektronix, Inc.
DPX, WaveAlert, and e*Scope are trademarks of
Tektronix, Inc.
Reference
TDS3000B Series
Digital Phosphor
Oscilloscopes
071-0923-00
������������
071092300
Tektronix warrants that the products that it manufactures and
sells will be free from defects in materials and workmanship
for a period of one (1) year from the date of shipment from an
authorized Tektronix distributor. If a product proves defective
within the respective period, Tektronix will provide repair or
replacement as described in the complete warranty statement.
To arrange for service or obtain a copy of the complete
warranty statement, please contact your nearest Tektronix sales
and service office.
EXCEPT AS PROVIDED IN THIS SUMMARY OR THE
APPLICABLE WARRANTY STATEMENT, TEKTRONIX
MAKES NO WARRANTY OF ANY KIND, EXPRESS OR
IMPLIED, INCLUDING WITHOUT LIMITATION THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT
SHALL TEKTRONIX BE LIABLE FOR INDIRECT,
SPECIAL OR CONSEQUENTIAL DAMAGES.
Contacting Tektronix
Phone
1-800-833-9200*
Address
Tektronix, Inc.
Department or name (if known)
14200 SW Karl Braun Drive
P.O. Box 500 Beaverton, OR 97077 USA
Web site
www.tektronix.com
Sales support
1-800-833-9200, select option 1*
Service support
1-800-833-9200, select option 2*
Technical
support
Email: [email protected]
1-800-833-9200, select option 3*
1-503-627-2400
6:00 a.m. – 5:00 p.m. Pacific time
*
This phone number is toll free in North America. After
office hours, please leave a voice mail message.
Outside North America, contact a Tektronix sales office
or distributor; see the Tektronix web site for a list of
offices.
To avoid potential hazards, use this product only as
specified in the TDS3000B Series Digital Phosphor
Oscilloscope User Manual which is located on the
TDS3000B Series Digital Phosphor Oscilloscope
Manuals and Application Notes CD. While using this
product, you may need to access other parts of the
system.
Read the General Safety Summary in the user manual
and in other system manuals for warnings and cautions
related to operating the system.
Preventing Electrostatic Damage
CAUTION. Electrostatic discharge (ESD) can
damage components in the oscilloscope and its
accessories. To prevent ESD, observe these
precautions when directed to do so.
Use a Ground Strap. Wear a grounded antistatic wrist strap
to discharge the static voltage from your body while
installing or removing sensitive components.
�
Contents
Safety Summary
2
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
TDS3000B Initial Setup
5
. . . . . . . . . . . . . . . . . . . . . . .
QuickMenus
9
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Menus
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Acquire Menu
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
XY Cursor Menu
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
YT Cursor Menu
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Display Menu
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Measure Menu
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Save/Recall Menu
. . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Trigger Menus
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Utility Menus
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Vertical Menus
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
e*ScopeE Web-based Remote Control
. . . . . . . . . . . . 41
Automatic Measurements
. . . . . . . . . . . . . . . . . . . . . 44
Use a Safe Work Area. Do not use any devices capable of
generating or holding a static charge in the work area
where you install or remove sensitive components.
Avoid handling sensitive components in areas that have a
floor or benchtop surface capable of generating a static
charge.
Handle Components Carefully. Do not slide sensitive
components over any surface. Do not touch exposed
connector pins. Handle sensitive components as little as
possible.
Transport and Store Carefully. Transport and store sensitive
components in a static-protected bag or container.
Manual Storage
The oscilloscope front cover has a convenient place to
store this manual.
4
More Information
Refer to the TDS3000B Series Digital Phosphor
Oscilloscope User Manual, which is located on the
TDS3000B Series Digital Phosphor Oscilloscope
Manuals and Application Notes CD, for more
information about using this product.
Connect the oscilloscope probe
to channel 1. Attach the probe
tip and reference lead to the
PROBE COMP connectors.
CH 1
PROBE COMP
4.
Push the AUTOSET button.
You should see a square wave
in the display (approximately
5 V at 1 kHz).
Probe Compensation
5.
Check the shape of the dis-
played waveform.
Overcompensated
Compensated correctly
Undercompensated
6.
Adjust your probe if necessary
to achieve correct compensa-
tion.
L
�
TDS3000B Initial Setup
The following procedures describe how to quickly verify
that the oscilloscope is powering up and functioning
correctly, compensate passive probes using the built-in
compensation signal, and run the signal path compensa-
tion (SPC) routine for maximum signal accuracy.
� You should perform all initial setup procedures the
first time you use the oscilloscope.
� You should perform the probe compensation
procedure whenever you attach a passive probe for
the first time to any input channel.
� You should run the signal path compensation routine
whenever the ambient temperature changes by 10� C
or more.
Functional Check
1.
Connect the oscilloscope power cable
2.
Turn on the oscilloscope. Wait
for the confirmation that the
oscilloscope has passed all
self-tests.
On/Standby
button
Signal Path Compensation (SPC)
The signal path compensation (SPC) routine optimizes
the oscilloscope signal path for maximum measurement
accuracy. You can run the routine anytime but you
should always run the routine if the ambient temperature
changes by 10� C or more.
To compensate the signal path, do the following steps:
1 Disconnect all probes or cables.
2 Push the UTILITY button.
3 Push the System screen button to select Cal.
4 Push the Signal Path screen button.
5 Push OK Compensate Signal Path.
Setting the User Interface Language
To set the user interface language, do the following
steps:
1 Push the UTILITY button.
2 Push the System screen button to select Config.
3 Push a side menu button to select a language.
�
Adjusting Time and Date
To set your oscilloscope to the current date and time, do
the following steps:
1 Push the UTILITY button.
2 Push the System menu button to select Config.
3 Push the Set Date & Time menu button.
4 Use the side menu buttons and general purpose knob
to adjust the date and time values.
5 Push the OK Enter Date/Time menu button to set
the instrument date and time.
User Manual
The TDS3000B Series DPO Oscilloscope Manuals and
Application Notes CD contains the TDS3000B Series
Digital Phosphor Oscilloscope User Manual, translated
into all supported languages. The user manual contains
complete information on how to operate your
TDS3000B Series oscilloscope.
The CD also contains manuals for all TDS3000B Series
accessories, application notes, oscilloscope and probe
fundamentals, and software to install the Adobe Acrobat
Reader, which is required to view the online documents.
To use a QuickMenu, do these steps
1 Push the QUICKMENU button on the front panel.
2 Push the screen button that corresponds to the control
you need to set. Push the screen button repeatedly to
choose one of the settings. The small arrow icon
indicates that there are additional settings available
that are not shown.
You can use most of the front-panel controls at the same
time you are using a QuickMenu. For example, if you
push a channel button to select a different channel, the
QuickMenu changes to show information about that
channel.
You can still use the regular menus. For example, if you
push the MEASURE button, you can set up and take
automatic waveform measurements in the usual way. If
you return to the QuickMenu, the measurement is still
displayed on the screen.
You may have optional application modules installed
that also have QuickMenu displays. To select the Quick-
Menu you want to use, push the MENU screen button.
This menu item is shown only if application modules
that contain a QuickMenu are installed.
�
QuickMenus
The QuickMenu feature simplifies the use of the oscillo-
scope. When you push the QUICKMENU button, a set
of frequently used menu functions show on the display.
Scope is one type of QuickMenu that you can use to
control the basic oscilloscope functions. You can per-
form many tasks without using the regular menu system.
The following sections briefly describe each oscilloscope
menu item.
Acquire Menu
Bottom
Side
Description
Mode
Sample
Use for normal acquisition.
Peak Detect
Detects glitches and reduces
the possibility of aliasing.
Envelope N
Captures variations of a signal
over a period of time. (Adjust N
with the general purpose knob.)
Average N
Reduces random or uncorre-
lated noise in the signal display.
(Adjust N with the general
purpose knob.)
Horizontal
Resolution
Fast Trigger
(500 points)
Acquires 500 point waveforms
at a fast repetition rate.
Normal
(10k points)
Acquires 10,000 point wave-
forms with more horizontal
detail.
Reset Hori-
zontal Delay
Set to 0 s
Resets the horizontal delay to
zero.
��
Bottom
Description
Side
Autoset
Normal
Autoset
Executes the autoset function.
(Optional application modules
may add choices to execute
specialized autoset functions.)
Undo Autoset
Reverts to the settings before
the last autoset.
WaveAlert
Wfm Anomaly
Detection
On Off
Enables or disables the
WaveAlert Waveform Anomaly
Detection function. See page
13.
Sensitivity
nn.n%
Sets the comparison sensitivity.
Use the general purpose knob
to set sensitivity from 0% (least
sensitive) to 100% (most sensi-
tive).
Beep On
Anomaly
On Off
When On, the oscilloscope
beeps if it detects a waveform
anomaly on any active channel.
Stop On
Anomaly
On Off
When On, the oscilloscope
stops waveform acquisition if
there is a waveform anomaly on
any channel. The input wave-
forms and anomalies remain
displayed on the screen.
XY Cursor Menu
The following cursor menu items are available in the XY
display mode (DISPLAY > XY Display > Triggered
XY). Push the CURSOR button to show the cursor
menu.
Bottom
Side
Description
Function
Off
Turns cursors off.
Waveform
Turns waveform cursors on and
displays measurements in
rectangular format (X and Y
values). Use the front-panel
SELECT button to select a
cursor. Use the general purpose
knob to move the active cursor.
Mode
Independent
Sets cursors to move indepen-
dently.
Tracking
Sets cursors to move together
when the reference cursor is
selected.
NOTE. The Advanced Analysis application
module (TDS3AAM) adds more XY cursor
features, including graticule cursors and polar
readouts.
��
Bottom
Description
Side
WaveAlert
(cont.)
Hardcopy
On Anomaly
On Off
When On, the oscilloscope
sends an image of the screen to
the hard copy device or floppy
file if there is a waveform anom-
aly on any channel.
Wfms To Disk
On Anomaly
On Off
When On, the oscilloscope
saves the anomalous wave-
form(s) data to the floppy disk
drive.
Highlight
Entire Wfm
Highlights the entire anomalous
waveform.
Highlight
Anomaly
Highlights just the anomalous
data in a waveform.
WaveAlert� Waveform Anomaly Detection. WaveAlert
provides a way to detect when a waveform deviates from
a steady-state condition. WaveAlert monitors the current
waveform acquisition and compares it to the previous
waveform acquisition. If the current acquisition exceeds
the comparison tolerance, the oscilloscope considers the
current acquisition as an anomaly and responds by
stopping acquisitions, emitting a beep, saving the
anomalous waveform to a disk file, printing the screen
image to a hard copy device, or any combination of the
above. Refer to the TDS3000B Series Digital Phosphor
Oscilloscope User Manual for more information.
YT Cursor Menu
The following YT cursor menu items are available in the
YT display mode (DISPLAY > XY Display > Off
(YT)). Push the CURSOR button to show the cursor
menu.
Bottom
Side
Description
Function
Off
Turns cursors off.
H Bars
Use to take vertical measure-
ments.
V Bars
Use to take both vertical and
horizontal measurements.
Bring Selected
Cursor to
Center Screen
Moves the active cursor to center
screen.
Bring Both Cur-
sors On Screen
Moves any off-screen cursor onto
the screen.
Mode
Independent
Sets cursors to move indepen-
dently.
Tracking
Sets cursors to move together
when cursor 1 is selected.
��
Bottom
Side
Description
V Bar Units
Sec (s) /
1/sec (Hz)
Sets horizontal units to se-
conds or frequency (Hz).
Ratio (%)
Sets V Bar measurement
units to percent.
Phase (degrees)
Sets V Bar measurement
units to degrees.
Use cursor posi-
tions as %/°
Sets V Bar measurement
scale so that 0% or 0° is the
current position of left V Bar
cursor and 100% or 360° is
the current position of right V
Bar cursor.
Use 5 divs as
%/°
Sets V Bar measurement
scale so that 5 screen major
divisions is 100% or 360°,
where 0% or 0° is –2.5 divi-
sions and 100% or 360° is
+2.5 divisions from center
graticule.
Set to on to see dots only. Set
to off to see dots and vectors.
Persist Time
Sets persist time.
Set to Auto
Sets the WAVEFORM
INTENSITY knob to control the
persistence time.
Clear
Persistence
Clears any displayed persis-
tence.
Backlight In-
tensity
High
Use for bright ambient condi-
tions.
Medium
Use for dim ambient conditions.
Low
Use to extend battery-operation
time.
Graticule
Full, Grid,
Cross Hair,
Frame
Selects the graticule type.
��
Bottom
Side
Description
H Bar Units
Base
Sets the H Bar units to be the
same as the selected wave-
form’s vertical measurement
units (volts, IRE, dB, and so
on).
Ratio (%)
Sets H Bar measurement units
to percent.
Use cursor
positions as
100%
Sets the H Bar measurement
scale so that 0% is the current
position of lowest H Bar cursor
and 100% is the current posi-
tion of highest H Bar cursor.
Use 5 divs as
100%
Sets H Bar measurement scale
so that 5 screen major divisions
is 100%, where 0% is –2.5
divisions and 100% is +2.5
divisions from center graticule.
Cursor Movement. Use the general purpose knob to move
the active cursor. Push the SELECT button to change
which cursor is active.
Faster Cursor Movement. Push COARSE to set the general
purpose knob to make faster cursor movements.
Turns off an XY display.
Triggered XY
Turns on triggered XY display.
Gated XYZ
Displays XY signals when
Z-channel signal is above a set
level. Only available on 4-chan-
nel instruments.
Ch1 (X)
Versus
Sets Ch2, Ch3, or Ch4, as Y
versus Ch1 as X.
Ref1 (X)
Versus
Sets Ref2, Ref3, or Ref4 as Y
versus Ref1 as X.
Gated By
Sets Ch2, Ch3, or Ch4 as Z
channel (gate) source, and
gating channel threshold level.
Color Palette
Normal
Selects the color display.
Monochrome
Sets all waveforms to high-con-
trast black and white.
XY Waveforms. For XY waveforms, channel 1 or Ref 1 is
displayed in the horizontal axis. Use the Vertical POSI-
TION and SCALE controls to adjust the horizontal
position and size of the XY waveform.
��
Another channel or reference waveform is displayed in
the vertical axis. Use the Vertical POSITION and
SCALE controls for that channel to adjust the vertical
position and size of the XY waveform.
XY Waveform Triggering. The XY waveform is triggered
so you can synchronize periodic input signals to the XY
waveform. This feature is useful when only one part of
the period contains valid information that you want to
see in the XY format. Set the time base and trigger
location to acquire just that part of the period.
When you want to see the complete period of the signals
regardless of the time base setting, set the trigger source
to an unused channel and the trigger mode to Auto.
Gated XYZ. Displays XY signals only when the Z (gating)
channel is true. Gated XYZ is similar to analog oscillo-
scope modulated XYZ mode except that the displayed
XY signal is either on or off; there is no intensity modu-
lation. Gated XYZ is useful for displaying constellation
diagrams.
The general purpose knob sets the Z-channel threshold
level.
Moves the active cursor to
center screen.
Bring Both Cur-
sors On Screen
Moves any off-screen cursor
onto the screen.
High-Low
Setup
Auto Select
Automatically uses the best
measurement method depend-
ing on the measurement type.
Histogram
Use to measure pulses.
Min-Max
Use to measure other wave
shapes.
��
Measure Menu
Bottom
Side
Description
Select
Measurement
Select an automatic measure-
ment for the selected (active)
channel. You can display up to
four measurements at one time.
See page 44.
Remove
Measurement
Measurement
1-4
Removes a specific measure-
ment.
All
Measurements
Removes all measurements.
Gating
Off
Use to take measurements on
the full waveform record.
Screen
Use to take measurements on
the portion of the waveform on
screen.
Cursors
Use to take measurements on
the portion of the waveform
between the V Bar cursors.
Set Levels in %
or units
Use to choose custom refer-
ence levels in relative or abso-
lute units.
High
Reference
Sets custom high reference
level.
Mid
Reference
Sets custom mid reference
level.
Mid 2
Sets the custom mid reference
level for the Delay and Phase
measurement second wave-
form.
Low
Reference
Sets custom low reference
level.
Set to
Defaults
Sets reference levels to default
values.
Indicators
Measurement
1-4
Selects the measurement for
which to display markers that
show the portion of the wave-
form used to calculate the
measurement value.
Off
Turns measurement indicators
off.
��
Save/Recall Menu
Bottom
Side
Description
Save Current
To File
Saves a setup to disk.
Setup
To Setup 1
...
To Setup 10
Saves a setup to nonvolatile
memory.
Recall Saved
From File
Recalls a setup from disk.
Setup
Recall Setup 1
...
Recall Setup
10
Recalls a setup from nonvolatile
memory.
Recall Factory
Setup
OK Confirm
Factory Init
Initializes the setup to a known
setup.
Save Wfm
To File
Shows file-format choices:
Internal, Spreadsheet, or Math-
cad. Saves active waveforms to
disk.
To Ref1
...
To Ref4
Saves the selected waveform to
nonvolatile memory.
Recall Wfm
From File
Recall Ref1
...
Recall Ref4
Recalls a waveform from disk
and displays it as a reference
waveform.
Edge Trigger Menu
Bottom
Side
Description
Source
Ch1
...
Ch4
Sets the trigger source to a
specific channel.
AC Line
Selects the AC line trigger source
(for AC line operation only)
Ext
Sets the oscilloscope to trigger
on the external trigger source.
Ext/10
Ext/10 attenuates the external
trigger signal by a factor of 10.
Ext Probe
nnX
Voltage / Current
(4-channel only)
Set this value to match the
attenuation factor and the type of
probe (voltage or current) that is
attached to the external trigger
connector. Push the menu button
to select the probe type. Use the
general purpose knob to set the
attenuation factor. The default
values are 1x and voltage.
��
File Utilities Menu
Bottom
Side
Description
File
Delete
Deletes a file.
Utilities
Rename
Names or renames a file.
Copy
Copies a file to another directo-
ry.
Print
Prints a file to a printer con-
nected to one of the hard copy
ports.
Create
Directory
Creates a new directory.
Confirm
Delete
Turns On or Off a confirmation
message before files are de-
leted.
Overwrite
Lock
Sets file overwrite protection to
On or Off.
Format
Formats a disk (erases all files).
Labels
Create/edit labels for reference waveforms and
instrument setups stored in nonvolatile memory.
Sets the trigger source to the
lowest-numbered active channel
in the display.
Alternating
(all active chan-
nels)
Sequentially uses each active
channel as a trigger source, from
the lowest-numbered active
channel to the highest-numbered
active channel.
Coupling
DC
Selects DC coupling.
HF Reject
Rejects frequencies above
30 kHz in the trigger signal.
LF Reject
Rejects frequencies below
80 kHz in the trigger signal.
Noise Reject
DC coupling with low sensitivity
to reject noise in the trigger
signal.
Slope
/ (rising edge)
Triggers on the rising edge of a
signal.
\ (falling edge)
Triggers on the falling edge of a
signal.
��
Bottom
Description
Side
Level
Level
Use to set the trigger level with
the general purpose knob.
Set to TTL
Sets the trigger level to +1.4 V
for TTL logic.
Set to ECL
Sets the trigger level to –1.3 V for
ECL logic (Vee = –5.2 V).
Set to 50%
Sets the trigger level to the 50%
amplitude level of the signal.
Mode &
Holdoff
Auto (untriggered
roll)
Enables free-running and roll-
mode acquisitions.
Normal
Triggers only on valid trigger
events.
Holdoff (time)
Sets holdoff to a specific time.
Holdoff
(% of record)
Sets holdoff to a percent of the
record duration.
Set to Min
Sets holdoff to the minimum
value.
B Trigger Menu
Bottom
Side
Description
B Trigger
After A
B Trigger After
A Time
Sets the oscilloscope to trigger
on the next B-trigger event that
occurs after a specified period
of time from the A trigger. Use
the general purpose knob to set
the time value.
Set to Delay
Time (B�M)
then Set
(B�M) to 0s
Sets the B Trigger After A Time
value to the horizontal B�M
value, then sets B�M to zero
seconds. B�M is the delay
time from the B trigger point to
the expansion point (center
screen).
Set to Min
Sets B Trigger After A time to
26.4 ns.
B Events
Sets the oscilloscope to trigger
on the nth B-trigger event after
the A trigger. Use the general
purpose knob to set the event
value.
Set to Min
Sets the B events count to 1.
��
Video Trigger Menu
Bottom
Side
Description
Standard
525/NTSC
Triggers on a NTSC signal.
625/PAL
Triggers on a PAL signal.
SECAM
Triggers on a SECAM signal.
Source
Selects the trigger source. See
Edge Trigger Menu for descrip-
tions of these menu items.
Trigger On
Odd
Triggers on odd or even fields in
Even
an interlaced signal.
All Fields
Triggers on any field in an
interlaced or noninterlaced
signal.
All Lines
Triggers on all lines.
NOTE. The Extended Video application module
(TDS3VID) adds more video features including
custom video triggering, video picture,
vectorscope, and analog HDTV triggering.
Sets Source, coupling, slope,
Coupling
and level for the B trigger.
These settings are independent
Slope
of similar settings for the A
trigger. See Edge Trigger Menu
Level
for descriptions of these menu
items.
Utility Menus
Configuration System Menu
Bottom
Side
Description
Language
English
Use to choose your native
Français
language. Most on-screen text
appears in the language you
Deutsch
appears in the language you
choose.
Italiano
Español
Português
(Russian)
(Japanese)
(Korean)
(Simplified
Chinese)
(Traditional
Chinese)
��
Bottom
Description
Side
Set Date &
Time
Display Date/
Time
Use to turn the date/time
display On or Off.
Hour Min
Use to set the internal clock
with the current hour and min-
ute.
Month Day
Use to set the internal clock
with the current month and day.
Year
Use to set the internal clock
with the current year.
OK Enter
Date/Time
Sets the date and time of the
internal clock.
Battery
Time-Outs
Power Off
Time-Out
Use to set the time before an
automatic shut down.
Backlight
Time-Out
Use to set the time before the
backlight automatically turns off.
Tek Secure
Erase Memory
Erases all nonvolatile waveform
and setup memory.
Version
Use to see the firmware ver-
sion.
Sets oscilloscope Ethernet
parameters such as the IP
address, instrument name, and
domain name.
DHCP/
BOOTP
Enables or disables DHCP/
BOOTP support.
Debug
Enables or disables a debug
message window.
Test
Connection
Tests the connection to your
network.
Ethernet Print-
Add Printer
Adds, renames, or deletes an
er Settings
Rename
Printer
Ethernet network printer from
the oscilloscope printer list.
Delete Printer
Confirm
Delete
Enables or disables displaying a
printer name delete confirmation
message.
More Information. Refer to the TDS3000 & TDS3000B
Series Digital Phosphor Oscilloscope Programmer
Manual for more information.
��
I/O System Menu
Bottom
Side
Description
GPIB
(TDS3GV)
Talk/Listen Ad-
dress
Sets the GPIB address.
Hard Copy
(Talk Only)
Sets the GPIB port to talk only
for making hard copies.
Off Bus
Disables the GPIB port.
Debug
Enables and disables a mes-
sage window to help you debug
GPIB problems.
RS-232
(TDS3GV)
Baud Rate
Sets the baud rate in steps from
1200 to 38400.
Flagging
Use to enable hard flagging
(RTS/CTS) or turn flagging off.
EOL
Selects end-of-line terminator.
Debug
Enables and disables a mes-
sage window to help you debug
RS-232 problems.
Set RS-232
Parameters to
Default Values
Sets baud rate = 9600, hard
flagging = on, and EOL = LF.
Hard Copy System Menu (Utility > System: Hard Copy )
Bottom
Side
Description
Format
Choose the correct hard copy
format for your printer.
Options
Portrait
Prints in a portrait orientation.
Landscape
Prints in a landscape orienta-
tion.
Compression
Turn on or off hard copy file .gz
compression format.
Ink Saver
Off
Prints the display colors.
On
Prints colors optimized for
printing on white paper.
Preview
Shows a preview of the hard
copy colors on the display.
Port
Centronics
Uses the parallel printer port.
GPIB
Uses the GPIB port.
RS-232
Uses the RS-232 port.
File
Saves the hard copy to the
floppy disk.
Ethernet
Uses the Ethernet LAN port.
Clear Spool
Stops a hard copy in progress.
Clears the print spooler.
��
Calibration System Menu
Bottom
Side
Description
System Cal
Signal Path
Compensates the signal paths
to obtain best measurement
accuracy.
Factory Cal
Used to calibrate the oscillo-
scope. This is a service function
only.
Cal Due
Control
Notify After
Hours of Op-
eration
Sets the number of hours of
operation before notifying you
that a calibration is due.
Notify After
Years of
Elapsed Time
Sets the number of years before
notifying you that a calibration is
due.
Signal Path Compensation. You can run the Signal Path
Compensation routine anytime but you should always
run the routine if the ambient temperature changes by
10 °C or more.
Sets input coupling to DC.
AC
Sets input coupling to AC.
GND
Provides a 0 V signal reference.
The input BNC disconnects
from internal circuits.
�
Sets input resistance to 50 �
or 1 M�.
Invert
Invert Off
Use for normal operation.
Invert On
Inverts the polarity of the signal
in the display.
Bandwidth
Full Bandwidth
Sets bandwidth to the full
oscilloscope bandwidth.
150 MHz
Sets the bandwidth to 150 MHz
(not available on some models).
20 MHz
Sets the bandwidth to 20 MHz.
Fine Scale
Enables fine scale adjustment
with the general purpose knob.
��
Diagnostic System Menu
Bottom
Side
Description
System Diags
Execute
Starts diagnostics.
Loop
Once
Executes the diagnostic loop
once.
Always
Executes the diagnostic loop
continuously.
Until Fail
Executes until a failure occurs.
Error Log
Page Up
Use to see the previous error
log page.
Page Down
Use to see the next error log
page.
Enables numerical vertical
position adjustment.
Set to 0 divs
Sets vertical position to center
screen.
Offset
Vertical Offset
Enables vertical offset adjust-
ment with the general purpose
knob.
Set to 0 V
Sets vertical offset to 0 V.
Probe Setup
Voltage Probe
Use to set probe gain or attenu-
Current Probe
ation for probes that do not
have the TekProbe II interface.
Deskew
Use to adjust the time skew
correction for each probe.
Set to 0
Use to set probe skew correc-
tion to zero.
��
Math Menu
Bottom
Side
Description
Dual Wfm
Math
Set 1st Source
to
Selects the first source wave-
form.
Set Operator
to
Selects the math operator: +, –,
�, or �
Set 2nd
Source to
Selects the second source
waveform.
Ref Menu
Bottom
Side
Description
Ref 1
Save Ch1 to
Ref1
Saves channel 1 to reference
waveform Ref 1.
Save Ch2 to
Ref1
Saves channel 2 to reference
waveform Ref 1.
Save Ch3 to
Ref1
Saves channel 3 to reference
waveform Ref 1.
Save Ch4 to
Ref1
Saves channel 4 to reference
waveform Ref 1.
Save Math to
Ref1
Saves the math waveform to
reference waveform Ref 1.
Ref 2
Ref 3
Ref 4
Identical settings for reference waveforms Ref 2,
Ref 3, and Ref 4.
Configuring Your Oscilloscope Ethernet Settings
Before using the e*Scope feature, you must set your
oscilloscope Ethernet network settings. The TDS3000B
Series Digital Phosphor Oscilloscope User Manual,
Appendix G, Ethernet Setup, describes how to enter the
Ethernet network parameters for your oscilloscope.
Accessing e*Scope
After your instrument is set up with the correct ethernet
settings, you are ready to access that instrument over the
Internet. To access you instrument’s e*Scope server, do
the following steps:
1 On a PC or work station, open a network browser
program.
2 In the Location or Address field (where you normally
enter a URL), type in the IP address of the
TDS3000B instrument to which you want to connect.
For example, 188.121.212.107. Do not precede the IP
address with any characters. Then press the Return
key.
3 The browser program loads the instrument’s e*Scope
Web-based Remote Control home page, which is an
image of the screen contents when your browser
accessed the instrument.
��
e*Scope Web-based Remote Control
e*Scope introduces the next generation of oscilloscope
functionality. With e*Scope you can access any Internet-
connected TDS3000B Series oscilloscope from a
browser on your work station, PC, or laptop computer.
No matter where you are, your TDS3000B is as close as
the nearest browser!
There are two levels to e*Scope: basic and advanced.
The basic level, which is hosted on the oscilloscope, lets
you view a screen image of the current acquisition, save
or load waveform and setup files, and remotely control
the oscilloscope using the oscilloscope programming
commands.
The advanced level, which is hosted on your system,
provides a graphical user interface to remotely control
the oscilloscope. To obtain the advanced level software,
go to the Tektronix TDS3000B website and download
the web-based remote control software.
NOTE. You can also use e*Scope on TDS3000
Series oscilloscopes that are equipped with a
TDS3EM Ethernet communication module and
are running firmware version 3 or above.
The menu at the top of the page provides the following
functions:
� Home displays the oscilloscope screen.
� Applications takes you to the application URL
specified in the Configure tab. The default URL is the
application notes site at www.tektronix.com.
� Configure lets you specify URLs for the advanced
Web-based Control HTML pages (Control),
application notes (Applications), and help files
(Help). The default values are for the Tektronix web
site. However, you can host any of these on your
local Intranet for faster access and better system
security.
� Data lets you upload or download waveform data
files and oscilloscope settings, as well as remotely
control the oscilloscope using oscilloscope
programing commands.
� Status displays the version screen, which shows the
firmware version and lists installed application
modules.
� Control displays the advanced Web-based Control
page that lets you remotely control the oscilloscope.
��
Automatic Measurements
Measure-
ment
Definition
Amplitude
Measured over the entire waveform.
Amplitude = High (100%) – Low (0%)
Burst
Width
The duration of a burst. Measured over the entire
waveform.
Cycle
Mean
The arithmetic mean over the first cycle in the
waveform.
Cycle
RMS
The true Root Mean Square voltage over the first
cycle in the waveform.
Delay
The time between the Mid Reference crossings of
two different waveforms.
Fall Time
Time that the falling edge of the first pulse in the
waveform takes to fall from 90% to 10% of its
amplitude.
Frequency
Reciprocal of the period of the first cycle in the
waveform. Measured in Hertz (Hz).
High
The value used as 100%. Calculated using either the
min/max or the histogram method. Measured over
the entire waveform.
Low
The value used as 0%. Calculated using either the
min/max or the histogram method. Measured over
the entire waveform.
Positive Duty Cycle � Positive Width
Period
� 100%
Measurement of the first cycle in the waveform.
Positive
Overshoot
Measured over the entire waveform.
Positive Overshoot � Max–High
Amplitude � 100%
Positive
Width
Measurement of the first positive pulse in the
waveform. The time between the 50% amplitude
points.
Rise Time
Time that the leading edge of the first pulse in the
waveform takes to rise from 10% to 90% of its
amplitude.
RMS
The true Root Mean Square voltage over the entire
waveform.
Snapshot
All Mea-
surements
Displays all measurement values (except two-chan-
nel measurements) at the time you pushed the
Snapshot All Measurements button, for the
selected waveform. Push the Snapshot All
Measurements button again to update the list
values. Push the MENU OFF button to clear the list
from the screen.
��
Measure-
ment
Definition
Max
The maximum amplitude. The most positive peak
voltage measured over the entire waveform.
Mean
The arithmetic mean over the entire waveform.
Min
The minimum amplitude. The most negative peak
voltage measured over the entire waveform.
Negative
Duty Cycle
Negative Duty Cycle � Negative Width
Period
� 100%
Measurement of the first cycle in the waveform.
Negative
Overshoot
Measured over the entire waveform.
Negative Overshoot � Low–Min
Amplitude � 100%
Negative
Pulse
Width
Measurement of the first negative pulse in the
waveform. The time between the 50% amplitude
points.
Phase
The amount one waveform leads or lags another in
time. Expressed in degrees, where 360� comprise
one waveform cycle.
Pk–Pk
Measured over the entire waveform.
Peak-to-peak = Max – Min
Period
Time it takes for the first complete signal cycle to
complete in the waveform. Measured in seconds.