! PROGRAMMABLE INPUTS

CUB4TC - Thermocouple types T, E, J, K, R, S, B, N, or mV CUB4RT - RTD types Pt385, Pt392, Ni672, Cu427 ! PROGRAMMABLE TEMPERATURE OFFSET ! SELECTABLE F or C WITH 1 or 0.1 RESOLUTION ! F OR C DISPLAY ANNUNCIATORS ! CONFORMS TO ITS-90 TEMPERATURE SCALE ! COLD JUNCTION COMPENSATION (Enable/Disable) CUB4TC ! 0.48 INCH (12.2 mm) HIGH DIGI ...

CUB4TC Thermocouple Inputs:

Input Impedance: 20 MΩ Lead Resistance Effect: 0.03µV/ohm Maximum Input Voltage: 30 VDC, TC+ to TC- Maximum Input Voltage TC-: 3 VDC max. with respect to common CUB4RT RTD Inputs: Lead Resistance: Cu427: 3 ohms/lead, 6 ohms total All others: 10 ohms/lead, 20 ohms total Balanced Lead Resistance: Autom ...

All conductors should meet voltage and current ratings of the unit. Also,

cabling should conform to appropriate standards of good installation, local codes and regulations. It is recommended that power supplied to the unit be protected by a fuse or circuit breaker. Unit should be mounted in a relatively stable temperature environment. Av ...

dISP - 0C, 0.0C, 0F, 0.0F

This program item sets the display for Fahrenheit or Celsius and whole number or tenths. Tenths is not available for thermocouple types R, S, and B. To change the display, use the “blank” button to scroll through the list of choices. Press the “SEL” button to enter the choice and return to the main menu. This menu item doe ...

Voltage Calibration Diagram

Resistance Calibration Diagram 1. Connect a precision Voltage Source with an output range of 0.000 to 105.000 mV and an accuracy of 0.03% or better to the SIG+ (#3) and SIG- (#5) terminals. Placement of jumper does not affect voltage calibration. 2. Enter the Program Menu and step through the Program Menu list by pressing the � ...

1. Connect a precision Voltage Source with an output range of -10.000 to 60.000

mV and an accuracy of 0.03% or better to the TC- (#5) and TC+ (#3) terminals. Cold junction calibration must be performed AFTER voltage calibration. 2. Enter the Program Menu and step through the Program Menu list by pressing the “SEL” button until CAL i ...

The meter meets NEMA 4X/IP65 requirements for indoor use,

when properly installed. The units are intended to be mounted on an enclosed panel. The viewing window and buttons are factory sealed for a wash down environment. A sponge rubber gasket and mounting clip are provided for sealing the unit in the panel cut-out. Installation Environment Th ...

The Company warrants the products it manufactures against defects in materials and workmanship

for a period limited to one year from the date of shipment, provided the products have been stored, handled, installed, and used under proper conditions. The Company’s liability under this limited warranty shall extend only to the repair or repla ...