Mettler Excellence HS153 Reference Manual
Text of the manual
Table of Contents
1
Introduction
5
2
Safety Information
6
2.1
Definition of signal warnings and symbols......................................................................
6
2.2
Product specific safety information.................................................................................
6
3
Design and Function
9
3.1
Overview.....................................................................................................................
9
3.1.1
Overview drying unit......................................................................................
9
3.1.2
Overview terminal .........................................................................................
10
3.1.3
Overview type plate .......................................................................................
10
3.2
Compliance Information................................................................................................
11
3.3
Function Description.....................................................................................................
11
4
Installation and Putting into Operation
13
4.1
Selecting the location ...................................................................................................
13
4.2
Unpacking ..................................................................................................................
13
4.3
Scope of delivery .........................................................................................................
13
4.4
Connecting instrument..................................................................................................
14
4.5
Setup wizard ...............................................................................................................
14
4.6
Setting up the instrument ..............................................................................................
15
4.7
Leveling the instrument.................................................................................................
16
4.8
Anti-theft device ...........................................................................................................
17
4.9
Setting reading angle of terminal....................................................................................
17
4.10
Setting date and time....................................................................................................
17
4.11
Adjusting after setup.....................................................................................................
18
4.12
Transporting, packaging and storing..............................................................................
18
4.12.1
Transporting the Moisture Analyzer..................................................................
18
4.12.2
Putting into operation after transport................................................................
18
4.12.3
Packaging and storing...................................................................................
19
4.13
Installing USB device driver...........................................................................................
19
5
Operating Elements and Interfaces
20
5.1
Operation keys ............................................................................................................
20
5.2
Electrical interfaces ......................................................................................................
20
6
Operation
22
6.1
Operation orientation....................................................................................................
22
6.2
User Interface ..............................................................................................................
22
6.2.1
Operation Keys .............................................................................................
22
6.2.2
Screen navigation .........................................................................................
22
6.2.3
Icons and buttons .........................................................................................
23
6.2.3.1
Status message icons..............................................................................
23
6.2.3.2
Title bar icons.........................................................................................
23
6.2.3.3
Action bar buttons ...................................................................................
23
6.2.4
Basic elements on the touch screen ................................................................
23
6.2.5
Home Screen................................................................................................
24
6.2.6
Work Screen.................................................................................................
25
6.2.7
Input dialogs ................................................................................................
25
6.2.7.1
Entering characters and numbers..............................................................
25
6.2.7.2
Entering numerical values ........................................................................
27
6.2.7.3
Changing date and time...........................................................................
27
6.2.8
Lists and tables ............................................................................................
28
6.3
Performing a Simple Measurement.................................................................................
29
Table of Contents
1
Moisture Analyzer
Method parameters .......................................................................................
33
7.1.1.1
Main measurement parameters.................................................................
33
7.1.1.2
Result and value handling........................................................................
42
7.1.1.3
Workflow handling .................................................................................
45
7.1.1.4
General method properties........................................................................
46
7.1.2
Creating a method.........................................................................................
47
7.1.2.1
Creating a method with the Method Wizard ................................................
47
7.1.2.2
Creating a method manually.....................................................................
48
7.1.2.3
Creating a method by consulting the method library ....................................
48
7.1.3
Editing a method...........................................................................................
49
7.1.4
Testing a method ..........................................................................................
50
7.1.4.1
Temperature assistant..............................................................................
50
7.1.4.2
Test measurement...................................................................................
51
7.1.4.3
Testing with AutoMet................................................................................
52
7.1.5
Preparing a method for QuickPredict ...............................................................
52
7.1.5.1
Prerequisites to use QuickPredict for a method............................................
53
7.1.5.2
Activating QuickPredict.............................................................................
54
7.1.5.3
Establishing the prediction knowledge .......................................................
56
7.2
Measurement ..............................................................................................................
59
7.2.1
Performing a measurement ............................................................................
59
7.2.2
Optimum sample preparation .........................................................................
61
7.2.3
Working with weighing-in aid.........................................................................
61
7.2.4
Working with shortcuts ..................................................................................
62
7.2.5
Working with QuickPredict .............................................................................
63
7.2.5.1
Performing measurements with QuickPredict ..............................................
63
7.2.5.2
Handling QuickPredict results ...................................................................
66
7.2.5.3
Working with uncertainty thresholds ..........................................................
67
7.3
Results .......................................................................................................................
67
7.3.1
Graphical evaluations of measurement results..................................................
67
7.3.2
Method versioning.........................................................................................
70
7.3.3
Exporting results ...........................................................................................
70
7.3.3.1
Exporting results to an external storage device............................................
71
7.3.3.2
Exporting results via FTP server.................................................................
71
7.3.3.3
Exporting results with EasyDirect Moisture..................................................
71
7.3.3.4
Single result export settings ......................................................................
72
7.3.3.5
Multiple result export settings....................................................................
72
7.4
Test/Adjust ..................................................................................................................
73
7.4.1
Tests ...........................................................................................................
73
7.4.1.1
Weight test - external...............................................................................
73
7.4.1.2
Temperature test......................................................................................
74
7.4.1.3
SmartCal test ..........................................................................................
76
7.4.1.4
Measures after failed test..........................................................................
77
7.4.2
Adjustments .................................................................................................
77
7.4.2.1
Weight adjustment - external ....................................................................
77
7.4.2.2
Temperature adjustment...........................................................................
78
7.4.3
History.........................................................................................................
80
7.5
Settings ......................................................................................................................
81
7.5.1
User preferences ...........................................................................................
82
7.5.1.1
Languages .............................................................................................
82
7.5.1.2
Screen ..................................................................................................
82
7.5.1.3
Sound....................................................................................................
83
7.5.1.4
Password...............................................................................................
83
7.5.2
Instrument settings........................................................................................
83
7.5.2.1
Regional settings.....................................................................................
83
7.5.2.2
Peripherals.............................................................................................
84
Table of Contents
2
Moisture Analyzer
Default settings .......................................................................................
86
7.5.2.4
Company information ..............................................................................
86
7.5.2.5
Touch screen adjustment .........................................................................
87
7.5.2.6
Network interfaces...................................................................................
87
7.5.3
Applications settings......................................................................................
91
7.5.3.1
Identifications management......................................................................
91
7.5.3.2
Print and export management...................................................................
92
7.5.3.3
Method selection.....................................................................................
95
7.5.4
Quality management .....................................................................................
95
7.5.4.1
Test / Adjustment settings.........................................................................
96
7.5.4.2
QuickPredict ...........................................................................................
99
7.5.5
User management.........................................................................................
99
7.5.5.1
Groups ..................................................................................................
99
7.5.5.2
Users..................................................................................................... 100
7.5.5.3
Account policies...................................................................................... 101
7.5.6
System and data management ....................................................................... 102
7.5.6.1
Result management................................................................................. 102
7.5.6.2
Export / Import........................................................................................ 103
7.5.6.3
Back up / Restore.................................................................................... 104
7.5.6.4
Export event logs..................................................................................... 105
7.5.6.5
License manager..................................................................................... 105
7.5.6.6
Reset ..................................................................................................... 106
7.5.6.7
Update................................................................................................... 106
7.5.7
Help and tutorials.......................................................................................... 106
7.5.7.1
Instrument tutorial ................................................................................... 106
8
Maintenance
107
8.1
Maintenance table........................................................................................................ 107
8.2
Cleaning..................................................................................................................... 107
8.2.1
Sample chamber........................................................................................... 108
8.2.2
Heating module ............................................................................................ 108
8.2.3
Fan grill ....................................................................................................... 111
8.2.4
Putting into operation after cleaning ................................................................ 111
8.3
Replacing the Dust filter................................................................................................ 111
8.4
Replacing the power line fuse........................................................................................ 111
8.5
Updating the software................................................................................................... 112
9
Troubleshooting
113
9.1
Malfunctions shown on the display................................................................................ 113
9.2
Malfunctions when operating the instrument.................................................................... 114
10
Accessories and Spare Parts
119
10.1
Accessories................................................................................................................. 119
10.2
Spare parts ................................................................................................................. 123
10.3
Spare parts ................................................................................................................. 124
11
Technical Data
125
11.1
General data ............................................................................................................... 125
11.2
Model-specific data...................................................................................................... 126
11.3
Dimensions................................................................................................................. 127
11.3.1
Drying unit ................................................................................................... 127
11.3.2
Terminal ...................................................................................................... 128
11.4
Interfaces specification ................................................................................................. 128
11.4.1
RS232C....................................................................................................... 128
11.4.2
USB host...................................................................................................... 129
11.4.3
USB device................................................................................................... 129
Table of Contents
3
Moisture Analyzer
Thank you for purchasing a Halogen Moisture Analyzer from METTLER TOLEDO.
Your Moisture Analyzer is fast, reliable and offers a high level of operating convenience and useful functions to
facilitate determination of the moisture content of your samples.
The Moisture Analyzer conforms with all common standards and directives. It supports stipulations, work
techniques and result records as demanded by all international quality assurance systems, e.g. GLP (Good
Laboratory Practice), GMP (Good Manufacturing Practice). The instrument has a CE declaration of conformity
and METTLER TOLEDO as the manufacturer has been awarded ISO 9001 and ISO 14001 certification. This
provides you with the assurance that your capital investment is protected in the long term by a high product
quality and a comprehensive service package (repairs, maintenance, servicing, adjustment service).
EULA
The software in this product is licensed under the METTLER TOLEDO End
User License Agreement (EULA) for Software.
When using this product you agree to the terms of the EULA.
u www.mt.com/EULA
Finding more information
u www.mt.com/moisture
u www.mt.com/hxhs
Search for software
u http://www.mt.com/moisture-software
Software version
This document is based on the software version V2.40 / V3.10.
Search for documents
u www.mt.com/library
For further questions, please contact your authorized METTLER TOLEDO dealer or service representative.
u www.mt.com/contact
Conventions and symbols
Key and button designations are indicated by a picture or bold text (e.g. Save).
Refers to an external document.
Note
For useful information about the product.
Elements of instructions
In this manual, step-by-step instructions are presented as follows. The action steps are numbered and can
contain prerequisites, intermediate results and results, as shown in the example. Sequences with less than two
steps are not numbered.
§
Prerequisites that must be fulfilled before the individual steps can be executed.
1
Step 1
ð Intermediate result
2
Step 2
ð Result
2 Safety Information
Two documents named "User Manual" and "Reference Manual" are available for this instrument.
•
The User Manual is printed and delivered with the instrument.
•
The electronic Reference Manual contains a full description of the instrument and its use.
•
Keep both documents for future reference.
•
Include both documents if you transfer the instrument to other parties.
Only use the instrument according to the User Manual and the Reference Manual. If you do not use the
instrument according to these documents or if the instrument is modified, the safety of the instrument may be
impaired and Mettler-Toledo GmbH assumes no liability.
2.1 Definition of signal warnings and symbols
Safety notes contain important information on safety issues. Ignoring the safety notes may lead to personal
injury, damage to the instrument, malfunctions and false results. Safety notes are marked with the following
signal words and warning symbols:
Signal words
WARNING
A hazardous situation with medium risk, possibly resulting in death or severe injury if
not avoided.
CAUTION
A hazardous situation with low risk, resulting in minor or moderate injury if not avoided.
NOTICE
A hazardous situation with low risk, resulting in damage to the instrument, other
material damage, malfunctions and erroneous results, or loss of data.
Warning symbols
General hazard: read the User Manual or the Reference Manual for information about the hazards
and the resulting measures.
Hot surface
Notice
2.2 Product specific safety information
Intended use
This instrument is designed to be used by trained staff. The instrument is intended for determining the weight
loss during drying of samples.
Any other type of use and operation beyond the limits of use stated by Mettler-Toledo GmbH without consent
from Mettler-Toledo GmbH is considered as not intended.
Moisture determination applications must be optimized and validated by the user according to local
regulations. Application-specific data provided by METTLER TOLEDO is intended for guidance only.
Responsibilities of the instrument owner
The instrument owner is the person holding the legal title to the instrument and who uses the instrument or
authorizes any person to use it, or the person who is deemed by law to be the operator of the instrument. The
instrument owner is responsible for the safety of all users of the instrument and third parties.
METTLER TOLEDO assumes that the instrument owner trains users to safely use the instrument in their
workplace and deal with potential hazards. METTLER TOLEDO assumes that the instrument owner provides the
necessary protective gear.
Personal protective equipment
Insulated gloves
Lab coat
Goggles
Safety notes
WARNING
Death or serious injury due to electric shock
Contact with parts that carry a live current can lead to injury and death. If the instrument
cannot be shut down in an emergency situation, people can be injured and the instrument
can be damaged.
1
Check that the voltage printed on the instrument is the same as your local power supply
voltage. If this is not the case, under no circumstances connect the instrument to the
power supply, but contact a METTLER TOLEDO representative.
2
Only use the three-core power cable with equipment grounding conductor supplied by
METTLER TOLEDO to connect your instrument.
3
Only connect it to a three-pin power socket with earthing contact.
4
Only use standardized extension cables with equipment grounding conductor for
operating the instrument.
5
Make sure that the power plug is accessible at all times.
6
Arrange the cables so that they cannot be damaged or interfere with the operation.
7
Keep all electrical cables and connections away from liquids.
WARNING
Injury or death due to toxic or corrosive substances
Heating up toxic or corrosive substances, e.g., acids, can result in toxic or corrosive vapors
that can cause injuries if they come in contact with the skin or the eyes or if they are
inhaled.
1
When using chemicals and solvents, comply with the instructions of the manufacturer
and the general laboratory safety rules.
2
Set up the instrument in a well-ventilated location.
3
When using dry substances that form toxic gases, place the instrument in a fume hood.
WARNING
Death or serious injuries due to flammable solvents
Flammable solvents in the vicinity of the instrument can ignite and lead to fire and
explosions.
1
Keep flammable solvents away from the instrument.
2
When using chemicals and solvents, comply with the instructions of the manufacturer
and the general laboratory safety rules.
Burns due to hot surfaces
During operation, parts of the instrument can reach temperatures that can cause burns if
touched.
1
Do not touch the area marked with the warning symbol.
2
Ensure sufficient free space around the instrument to avoid heat accumulation and
overheating (approx. 1 m free space above the heating module).
3
Never cover, tape or clog the vent over the sample chamber. Do not tamper with the vent
in any other way.
4
Exercise caution when removing a sample. The sample itself, the sample chamber, the
draft shield and the sample pan may be very hot.
5
Do not open the heating module during operation. Always let it cool down completely
before opening.
6
Do not modify the heating module in any way.
NOTICE
Damage to the instrument due to corrosive substances and vapors
Corrosive substances and corrosive vapors can damage the instrument.
1
When using chemicals and solvents, comply with the instructions of the manufacturer
and the general laboratory safety rules.
2
Ensure that the instrument parts touching your sample substance can not get altered by
it.
3
Wipe off any condensation of corrosive vapours after an operation.
4
Work with small samples.
NOTICE
Damage to the instrument due to the use of unsuitable parts
Using unsuitable parts with the instrument can damage the instrument or cause it to
malfunction.
−
Only use parts from METTLER TOLEDO that are intended to be used with your
instrument.
For related information concerning the QuickPredict function, see [Prerequisites to use QuickPredict for a
method } Page 53].
Design and Function
9
Moisture Analyzer
3 Design and Function
In order to work with your Moisture Analyzer, you need to familiarize yourself with the instrument layout, drying
unit, graphical user interface of the touch screen as well as the terminal.
3.1 Overview
3.1.1 Overview drying unit
Legend drying unit
1
Heating module (with thermal overload
protection)
2
Handles for opening the sample chamber
3
Power contacts for halogen lamp
4
Draft shield element
5
Sample chamber
6
Level indicator / Level sensor
7
Leveling screw
8
Sample pan handler
9
Sample pan holder
10
Contacts for temperature kit (optional)
11
Temperature sensor
12
Inspection window and vent
13
2 Aux ports (n.a.)
14
Socket for terminal connection cable
15
Fastening for terminal stand
16
Fan
17
RS232C (for production only)
18
Dust filter (optional)
19
Kensington slot for anti-theft purposes
20
Power supply socket
21
Power line fuse (plus replacement fuse)
Design and Function
10
Moisture Analyzer
3.1.2 Overview terminal
Legend terminal
1
Touch screen
2
USB Host port 1
3
Foot
4
USB Host port 2
5
USB Device port
6
RS232C port
7
Ethernet RJ45 port
8
Socket for terminal connection cable
9
SD Memory Card slot
3.1.3 Overview type plate
Drying unit type plate
The type plate of the drying unit is located on the side of the drying unit and contains the following information:
5
4
6
7
1
2 3
1
Model designation
2
Readability
3
Maximum capacity
4
Manufacturer
5
Power supply
6
Year of manufacture
7
Serial number (SNR)
Design and Function
11
Moisture Analyzer
Terminal type plate
The type plate of the terminal is located on the terminal and contains the following information:
1
2
1
Model designation
2
Serial number (SNR)
3.2 Compliance Information
European Union
The instrument complies with the directives and standards listed on the EU Declaration of Conformity.
United States of America
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to
part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful inter-
ference when the equipment is operated in a commercial environment. This equipment generates, uses, and
can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications. Operation of this equipment in a residential area is
likely to cause harmful interference in which case the user will be required to correct the interference at his own
expense.
Changes or modifications not expressly approved by the party responsible for compliance could void the user's
authority to operate the equipment.
The FCC Supplier Declaration of Conformity is available online.
u http://www.mt.com/ComplianceSearch
3.3 Function Description
METTLER TOLEDO Moisture Analyzers work on the thermogravimetric principle. At the start of the measurement
the Moisture Analyzer determines the weight of the sample, the sample is then quickly heated by the absorption
of the emitted infrared radiation. During the drying process the instrument continually measures the weight of
the sample and displays the weight reduction, up to the final result.
120 °C
With halogen heating technology, maximum heating power is quickly reached and allows use of high temper-
atures. Uniform heating of the sample material ensures repeatability of the drying results and makes it possible
to use small amounts of sample.
The METTLER TOLEDO Moisture portfolio comprises a range of moisture analyzers that differ from each other in
both hardware and software.
The following features are available for the HS153 model:
•
High safety and process control: management of multiple users with individual access rights. See User
management
Design and Function
12
Moisture Analyzer
•
Built-in testing methods to ensure performance: SmartCal and tests for the heating and weighing unit. See
[Test/Adjust } Page 73]
•
Integrated method development support. See Creating a method
•
Versatile data transfer and printing. See Exporting results
•
Quick and easy cleaning. See [Cleaning } Page 107]
•
ID management. See [Identifications management } Page 91]
•
Optional: Reduced measurement time and quick decision making with QuickPredictTM. See [Activating
QuickPredict } Page 54] and [Working with QuickPredict } Page 63]
Installation and Putting into Operation
13
Moisture Analyzer
4 Installation and Putting into Operation
4.1 Selecting the location
Requirements of the location
Place indoors on stable
table
Ensure sufficient spacing
Level the instrument
Provide adequate lighting
Avoid direct sunlight
Avoid vibrations
Avoid strong drafts
Avoid temperature fluctu-
ations
Sufficient distance for Moisture Analyzers: > 15 cm next to the instrument, > 1 m above the heating module.
4.2 Unpacking
Note
Store all parts of the packaging. This packaging guarantees the best possible protection for the transport of your
instrument.
After receiving the instrument, proceed as follows:
1
Open the package and remove drying unit, terminal and the accessories.
2
Remove the packaging from the instrument.
3
Check the Moisture Analyzer for transport damage and complete scope of delivery.
4
In the event of damage or missing accessories, immediately inform a METTLER TOLEDO representative.
4.3 Scope of delivery
Moisture Analyzer
Documentation
Accessories
•
1 Drying unit
•
1 Draft shield
•
1 Power cable
•
1 Terminal cable
•
1 Sample handler
•
1 Sample pan holder
•
1 User Manual
•
1 Application brochure «Guide
to Moisture Analysis»
•
1 Production certificate
•
1 Voucher for E-learning "Proper
Moisture Determination"
•
EU-countries: 1 CE declaration
of conformity
•
80 aluminium sample pans
•
3 Specimen samples (glass
fibre filter)
•
1 SmartCal sampler
•
Specific countries: 1 Wifi dongle
Installation and Putting into Operation
14
Moisture Analyzer
4.4 Connecting instrument
WARNING
Death or serious injury due to electric shock
Contact with parts that carry a live current can lead to injury and death. If the instrument
cannot be shut down in an emergency situation, people can be injured and the instrument
can be damaged.
1
Check that the voltage printed on the instrument is the same as your local power supply
voltage. If this is not the case, under no circumstances connect the instrument to the
power supply, but contact a METTLER TOLEDO representative.
2
Only use the three-core power cable with equipment grounding conductor supplied by
METTLER TOLEDO to connect your instrument.
3
Only connect it to a three-pin power socket with earthing contact.
4
Only use standardized extension cables with equipment grounding conductor for
operating the instrument.
5
Make sure that the power plug is accessible at all times.
6
Arrange the cables so that they cannot be damaged or interfere with the operation.
7
Keep all electrical cables and connections away from liquids.
Two different versions of drying units with country-specific power cable are available:
110 V AC or 230 V AC
Connecting the terminal to the drying unit
§
Drying unit and terminal are at the final location.
1
Connect one end of the supplied terminal cable to the system connection socket on the terminal.
2
Screw the connector firmly.
3
Connect the other end to the system connection socket on the drying unit.
4
Screw the connector firmly.
Connecting the drying unit to the power supply
§
The terminal is connected to the drying unit.
1
Connect one end of the power cable to the power supply socket on the drying unit.
2
Connect the other end to the power line outlet.
See also
2 Overview terminal } Page 10
2 Overview drying unit } Page 9
4.5 Setup wizard
When you switch on your Moisture Analyzer for the first time or when a factory reset was performed, the Setup
Wizard appears. The Setup Wizard is a step-by-step instruction, that helps to define the basic functions
(Language and Date format) and to install the sample chamber of your device. The buttons in the action bar
can be used for navigation.
It must be successfully finished. If it is not finished, the device is not operational and the function will reappear
when the device is switched on the next time.
Installation and Putting into Operation
15
Moisture Analyzer
After the application Setup Wizard is successfully finished, the settings can be changed via Settings of the user
home screen.
Note
The application Setup Wizard appears automatically when the instrument was switched on for the first time (or
when a factory reset was performed). It must be successfully finished. If it is not finished, the device is not
operational and the function will reappear when the device is switched on the next time.
After the application Setup Wizard is successfully finished, the settings can be changed via Settings of the user
home screen.
Note
The application Instrument tutorial starts automatically after the application Setup Wizard has succesfully
finished. This option can be skipped by inactivating the checkbox Proceed to the Instrument Tutorial. If you
want to redo the application Instrument tutorial, it can be found in the section Settings.
4.6 Setting up the instrument
§
Terminal and drying unit are connected.
1
To switch on the Instrument, press [
].
2
Open the sample chamber.
3
Place the draft shield element. Only one position is possible.
4
Insert the sample pan holder carefully. Check that the
sample pan holder is hooked correctly.
Installation and Putting into Operation
16
Moisture Analyzer
5
Insert the sample pan handler.
4.7 Leveling the instrument
Exact horizontal positioning and stable installation are prerequisites for repeatable and accurate results. To
compensate for small irregularities or inclinations (±2%) at the location, the instrument must be leveled.
Note
For leveling the drying unit use the function Leveling guide that can be found on Settings > Help and tutorials
> Instrument tutorial > 1. Leveling the instrument.
For exact horizontal positioning, the drying unit has a level indicator
(level) and 2 leveling screws. When the air bubble in the level
indicator is exactly in the center, the instrument is standing perfectly
horizontally. To level it, proceed as follows:
1
Position your Moisture Analyzer at the selected location.
2
Turn the two leveling feet until the air bubble is in the center of the
level indicator.
L = left foot
R = right foot
Air bubble at 12
o'clock:
turn both feet clockwise.
Air bubble at 3
o'clock:
turn left foot clockwise, right foot counter-
clockwise.
Air bubble at 6
o'clock:
turn both feet counterclockwise.
Air bubble at 9
o'clock:
turn left foot counterclockwise, right foot
clockwise.
Installation and Putting into Operation
17
Moisture Analyzer
Leveling the drying unit with optional stand for terminal mounted.
§
The drying unit is positioned at the selected location.
1
To loose the clamps for the safety feet of the stand, turn them
outwards.
2
Level the drying unit as previously described.
3
To lock the clamps for the safety feet, turn them inwards as far as
they will go.
4.8 Anti-theft device
For protection against theft, your drying unit is fitted
with a Kensington security slot for an anti-theft device.
To find a suitable anti-theft cable see [Accessories
and Spare Parts } Page 119].
4.9 Setting reading angle of terminal
−
To change the reading angle, fold out the two tilt feet.
4.10 Setting date and time
Navigation: Home > Settings > Instrument settings > Regional Settings
When you put your new instrument into operation for the first time, you should enter the current date and time.
These settings are retained even if you disconnect your instrument from the power supply.
Note
Various languages are available for the dialog with your instrument and can be selected in the menu:
Navigation: Settings > User preferences.
Set the current date
§
Regional Settings is selected.
1
Tap Date.
2
Set day, month and year.
3
Confirm with [Set date].
Installation and Putting into Operation
18
Moisture Analyzer
Set the current time
§
Regional Settings is selected.
1
Tap Time.
2
Set Hours and minutes.
3
Confirm with [Set time].
4.11 Adjusting after setup
To obtain accurate measuring results, it is necessary to adjust the integrated balance as well as the heating
module.
Adjusting is necessary:
•
before the instrument is used for the first time.
•
at regular intervals.
•
after a change of location.
The following adjustment options are possible:
•
Balance adjustment with external weight
•
Temperature adjustment with temperature kit
•
To verify the overall performance of the Moisture Analyzer after the aforementioned adjustments, perform a
SmartCal test.
For more information to perform these operations see [Test/Adjust } Page 73].
4.12 Transporting, packaging and storing
4.12.1 Transporting the Moisture Analyzer
Transporting the Moisture Analyzer over short distances
1
Switch off the instrument and unplug all interface cables.
2
Hold the instrument with both hands as shown.
3
Carefully lift the instrument and carry it in horizontal
position to the new location.
Transporting the Moisture Analyzer over long distances
For transporting the Moisture Analyzer over long distances, always use the original packaging or the designated
transport case (see "Accessories").
4.12.2 Putting into operation after transport
Putting into operation after transport:
1
Connect drying unit and terminal.
2
Check the level status. Level the Moisture Analyzer if necessary.
3
METTLER TOLEDO recommends to perform weight and temperature tests and, if necessary, adjustments
after transporting the Moisture Analyzer.
ð The Moisture Analyzer has been put into operation and is ready to use.
Installation and Putting into Operation
19
Moisture Analyzer
See also
2 Connecting instrument } Page 14
2 Test/Adjust } Page 73
2 Setting up the instrument } Page 15
4.12.3 Packaging and storing
Packaging
Store all parts of packaging in a safe place. The elements of the original packaging are developed specifically
for the Moisture Analyzer and its components to ensure optimal protection during transportation or storing.
Storing
Store the Moisture Analyzer under following conditions:
•
Indoor and in the original packaging.
•
According to the environmental condition, see "Technical data".
Note
When storing for longer than six months, the rechargeable battery may be down (date and time get lost).
4.13 Installing USB device driver
To transmit commands via USB device interface, a USB driver program has to be installed on the instrument
and PC. The driver program is automatically installed during setup of the instrument. However, in case of an
installation error, the software can be downloaded online and installed manually.
Find the download link on:
u www.mt.com/labweighing-software-download
Requirements
•
PC with one of the following Microsoft Windows® 32-bit/64-bit operating systems: Win 7 (SP1), Win 8 or
Win 10.
•
USB cable to connect the instrument to the PC.
Downloading and installing the USB driver
1
Open the download portal (see above) on your PC.
2
Download the driver software and extract the .zip file.
3
Right-click on the extracted .exe installation program and select Run as Administrator.
4
If a safety warning appears, allow windows to perform the installation.
5
Click Next and follow the installer's instructions.
Connecting the instrument to the PC
1
Connect the moisture analyzer to the preferred USB port on the PC.
2
Follow the instructions on the PC and install the software automatically (recommended).
3
Check that the instrument is listed as a device (COM Port) on your PC.
4
Set USB interface in the peripheral settings of the instrument.
Operating Elements and Interfaces
20
Moisture Analyzer
5 Operating Elements and Interfaces
5.1 Operation keys
Legend key functions
ON/OFF
Home
5.2 Electrical interfaces
1
2
2
3
3
4
5
6
7
Legend interfaces
1
System connection
Connects terminal and drying unit.
2
RS-232 interface (COM1)
Connects the instrument to a peripheral device
with a 9-pin male connector, e.g.:
•
Strip printer
•
PCs (for data exports)
•
EasyDirect moisture
Operating Elements and Interfaces
21
Moisture Analyzer
3
USB Host interface (Type A socket)
Connects the instrument to a peripheral USB-
device, e.g.:
•
Memory stick (up to 32 GB)
•
(A4 / Letter) Printer
4
USB Device connection (Type B socket,
COM4)
Connects the instrument to a peripheral USB-
host-device, e.g.
•
PCs (for data exports)
•
EasyDirect moisture
5
Ethernet interface
Connects the instrument to a TCP/IP network,
e.g.
•
Workstations (for data exports)
•
FTP-Servers
•
(A4 / Letter) Printer
DHCP and fix IP adresses are supported.
6
WLAN connection
Connects the instrument via WiFi dongle to a
wireless TCP/IP network e.g.
DHCP and fix IP adresses are supported.
•
Workstations (for data exports)
•
FTP-Servers
•
(A4 / Letter) Printer
Note
This function is not available in every country
7
SD Memory Card slot
Connects the instrument to following SD Memory
Card types:
•
SD (up to 32 GB)
•
SDHC (up to 32 GB)
Note
SDXC types are not supported.
6.1 Operation orientation
The Moisture Analyzer is mainly operated via the touchscreen and keys of the terminal.
Before a measurement can be made, a method must first be defined. Methods define the parameters of a
measurement, such as the temperature to which the sample is heated or the criterion after which the
measurement is finalized. The optimum parameters differ from sample to sample. The instrument and METTLER
TOLEDO support you in finding a suitable method for your sample.
Once a method has been defined, the actual measurement can be started: The sample is placed in the sample
chamber and prepared. Then the heating module is closed and the sample is heated. Depending on the sample
and the method, the measurement can take from a few minutes to a few hours.
When the measurement is finished, the end result is shown on the terminal screen. The result of the
measurement can also be exported or printed.
The Moisture Analyzer can be used immediately for another measurement.
6.2 User Interface
6.2.1 Operation Keys
Key
Designation
Description
ON/OFF
Switches the Moisture Analyzer on / into standby mode (to switch
it off completely, it has to be unplugged from the power supply).
Home
Directly returns to the home screen from any other menu level.
6.2.2 Screen navigation
The colored touch screen is a touch-sensitive WVGA monitor. It displays information and allows you to enter
commands by touching certain areas on its surface: You can choose the information displayed on the screen,
change terminal settings or perform certain operations on the instrument.
Only those elements which are available for the current dialog appear on the display.
NOTICE
Damage to the touch screen due to pointed or sharp objects
−
Operate the touch screen with your fingers.
Note
The sensitivity of the touch screen is set to a default value. The sensitivity of the touch screen can be defined
with the function Touch screen adjustment in the section Settings.
The surface navigation of the touch screen works in the same way as on most common touch screens:
Selecting a button or an icon
−
Tap on it.
Scrolling
−
Drag the button in the scroll bar up or down.
−
Or tap
or
6.2.3 Icons and buttons
6.2.3.1 Status message icons
Status messages are displayed by means of small icons in the status bar of the display. The status icons
indicate the following:
Icon
Status description
Diagnostic
Remedy
The built-in level sensor has
detected that the instrument is
not correctly leveled.
Instrument is out of level.
Immediately level the
instrument.
The battery must be replaced.
This battery ensures that the
date and time are retained when
the instrument is disconnected
from the power supply.
-
Contact your METTLER TOLEDO
support representative.
Service is due.
–
Contact your METTLER TOLEDO
support representative.
Quick Access icon (displayed
continuously)
Quick setting of brightness and
sound.
-
WLAN connected. Three bars
are indicating signal strength of
connection.
Settings can be modified.
•
Tap icon in the information
field.
•
Settings > Network
interfaces
WLAN not connected.
Check Wifi/Bluetooth dongle
(MTICWD-100).
Plug-in Wifi/Bluetooth dongle
(MTICWD-100).
Network not selected.
Settings can be modified.
•
Tap icon in the information
field.
•
Settings > Network
interfaces
6.2.3.2 Title bar icons
The title bar icons only appear when they are active. The icons which appear in the title bar, are listed with their
function below:
Parameter
Description
Create a shortcut and add it to your user home screen. Edit or remove an existing shortcut
see [Working with shortcuts } Page 62].
6.2.3.3 Action bar buttons
The action bar contains action buttons for triggering any of the actions which are required and available in the
current dialog (e.g. Back, ->0/T<-, Print, Save, Delete, OK). There are up to six action buttons available for a
work step.
6.2.4 Basic elements on the touch screen
Name
Explanation
1
Status bar
The status bar contains the user name, status icons as well as date and time.
See Status Message Icons.
2
Title bar
The title bar contains elements for the user's orientation and information, as well
as for special functions. See [Title bar icons } Page 23].
The content area is the main working area of the menus and the applications,
content depends on application or performed action. The content area also shows
graphic illustrations such as the curve diagram of a data set (e.g. drying curve).
4
Action bar
The action bar contains action buttons for triggering the actions which are
required and available in the current dialog (e.g. Back, ->0/T<-, Print, Save,
Delete, OK). See [Action bar buttons } Page 23].
6.2.5 Home Screen
The user Home screen is the main screen and appears after startup or login of the instrument. All sections of
the user interface can be accessed from the home screen. Returning to the Home screen is possible from all
other screens of the user interface by pressing the [
] key or tapping the [Home] button.
1
2
3
Name
Explanation
1
Main Menu
Measurement
Start measurement.
Prerequisite:
Method is defined and all necessary settings are made.
Results
Display, print and export results
Method Definition
Define, edit, test or delete a method
Test/Adjust
Adjust or test the integrated balance and the heating module and perform
SmartCal tests.
Settings
Define instrument-, user-, and data-management settings. Find help and tutorials
under this menu item.
2
User shortcuts
Shows user-specific shortcuts for frequently used methods. Shortcuts are saved in
the user profile.
3
Instrument info
Shows general info about the instrument and the software (e.g. serial number,
software version).
6.2.6 Work Screen
1
2
3
4
5
Name
Explanation
1
Shortcut button
Adding/editing a shortcut to the current method (to the home screen).
2
Value panel
Current measured (or predicted) values of the work process.
3
Parameter panel
Parameters of the current work process. A detailed overview of the method
parameters appears after tapping the parameter panel.
Identifications (IDs) appear after tapping the ID panel for entering or editing
values (comments). ID panel appears only if identification input is activated in
the menu.
4
Action buttons
Action buttons which are required and available for the current dialog (e.g. Back,
->0/T<-, Print, Save, Delete, OK).
5
Graphics panel
Graphical illustrations, e.g. of the drying curves, instructions for the user for
performing tasks, and weighing-in aid.
6.2.7 Input dialogs
6.2.7.1 Entering characters and numbers
The keyboard dialog allows you to enter characters including letters, numbers and a variety of special
characters. The country-specific layout of the keyboard is in accordance with the selected language, see
[Regional settings } Page 83].
1
2
3
Name
Explanation
1
Input field
Shows the data that has been entered.
2
Explanation field
Indicates the maximum number of characters allowed.
3
Shift
Switches the keyboard between standard and special characters.
Confirm with [OK] (To exit keyboard tap [Cancel]).
Function
Shifts between lower and upper case letters
Deletes last character
Cursor left
Cursor right
Note
It is also possible to place cursor directly in input field by tapping the respective position.
6.2.7.2 Entering numerical values
The numerical keypad dialog allows you to enter numerical values.
1
2
3
Name
Explanation
1
Switch
If a switch appears, located on the left side in the title bar; it must be switched on
for the entry of a value: O = OFF, I = ON
2
Input field
Shows the data that has been entered.
3
Explanation field
The valid input range is indicated.
Delete button
Deletes last character.
Cursor left
Jumps one character to the left.
Cursor right
Jumps one character to the right.
−
Confirm with [OK].
To exit keypad, tap [Cancel].
6.2.7.3 Changing date and time
This dialog (Picker view) allows you to set the date and time by picking from the built-in calendar/clock.
1
2
3
Name
Explanation
1
Pick button
Increases the value
2
Explanation field
Shows the current value
3
Pick button
Reduces the value
−
Confirm with [Set date]. (To exit without changing, tap [Cancel].)
6.2.8 Lists and tables
Simple list view
Basic elements of a simple list including a content title. If necessary, the content area can be scrolled vertically.
5
1
2
3
4
6
Name
Explanation
1
Content title
Shows the subject of the current content
2
Content area
Shows the relevant content
3
Arrow icon
[ ] Shows the list in ascending order.
[ ] Shows the list in descending order.
4
Scrollbar
[
] Scroll up
5
Scroller
6
[
] Scroll down
Further types of lists
Accordion panels are opened or closed by tapping [
].
Closable panels are opened or closed by tapping [ ].
6.3 Performing a Simple Measurement
To check the function of the instrument and to become familiar with the process, METTLER TOLEDO
recommends carrying out a simple example measurement, e.g. using a glass fibre filter to measure the
moisture content of water. During this first measurement, the instrument works with factory default settings.
§
The instrument is connected to the power supply.
−
To switch on, press
.
1
Tap Method Definition.
ð Method Definition menu appears.
2
To define a new method, tap New....
3
To define a new method manually, tap Define method manually.
ð The keypad appears.
4
Enter a name for your first method e.g. My first method.
5
Confirm with OK.
6
To store your new method with factory settings, tap Save.
7
Tap Home.
ð User home screen appears.
Select the method for measuring
1
Tap Measurement.
ð The methods list is displayed.
2
Tap My first method.
ð The work screen of the method My first method appears.
3
Open the sample chamber.
Placing the accessories
§
The display now prompts you to load the predefined accessories and tare the balance.
1
Place the empty accessories in the sample pan handler.
Place the sample pan handler in the sample chamber.
Ensure that the tongue of the pan handler fits exactly in
the slot of the draft shield element. The accessories
must lie flat in the pan holder.
Note
METTLER TOLEDO recommends to work with the sample
pan handler at all times during a measurement. The pan
handler is ergonomic, ensures correct positioning, is safe
and provides protection against possible burns due to the
hot accessories.
Taring
1
Close the sample chamber.
ð The instrument tares the balance (Starting mode: Automatic).
2
After taring, open the sample chamber.
Note
In this example, the glass fiber filter is part of the sample for the sake of simplicity. In routine liquid
measurements, the glass fiber filter is not part of the sample and must be tared together with the sample pan.
Start measuring
1
Place the specimen sample in the sample pan.
2
Wet the specimen sample with a few drops of water.
Tap on [->0/T<-] to tare.
4
Close the sample chamber.
ð The drying process starts automatically.
Drying process
During the measurement, the display continuously shows the real time results of the drying process in form of a
drying curve.
When the drying process is finished, the final measured moisture content of your sample is shown on the
display.
If control limits are defined for the method, the status (passed, warned or failed) is shown.
CAUTION
Burns due to hot surfaces!
Parts of the instrument can reach temperatures that cause injuries if touched.
Sample, sample pan and other parts inside the sample chamber may still be hot.
−
Do not touch the place marked with the warning symbol.
Remove Sample
§
The drying process is completed.
1
Open the sample chamber.
2
Carefully remove the sample pan handler from the sample chamber.
Note
To remove the sample pan from the handler, lift the pan slightly from below and pull it sideways out of the
handler.
3
• To perform a further measurement with the current
method, tap [Next Sample].
• To perform a measurement with a new method, tap
[Methods].
• To return to the home screen, press [
].
] to switch the instrument off into standby.
7 Software Description
7.1 Method Definition
Navigation: Home > Method definition
This function offers you to define new methods and to change, delete, export or import existing methods
depending on the user rights. It allows you to store up to 100 individual methods.
All parameters used in a drying method can be defined under this menu item.
Resources on methods and method definition:
u www.mt.com/moisture-guide
u www.mt.com/moisture-methods
7.1.1 Method parameters
7.1.1.1 Main measurement parameters
Navigation: Home > Method Definition > Method name > Main measurement parameters
These parameters are relevant for measuring and must be determined for all samples.
The following parameters can be defined:
Menu item
Explanation
Further Information
Drying program
Defines the drying program which is best suited for
the specific sample. (including temperature and
switch-off criteria)
see Setting the Drying Program
Display mode
Defines the unit for displaying and printing.
see [Display mode } Page 39]
Start weight
Defines a start weight so that the samples always
have approximately the same weight .
see [Start weight } Page 42]
7.1.1.1.1
Drying program
Navigation: Home > Method Definition > Method name > Main measurement parameters > Drying
program
This function offers various pre-defined drying programs for optimal adaptation of the drying properties to the
sample used. In addition, the drying temperature and the switch-off criterion can be set.
You can define the following parameters:
Parameter
Description
Values
Drying program
Defines the dynamics with which the drying temperature is
reached.
Standard* | Rapid |
Gentle | Steps
* Factory setting
Further information:
•
Standard drying see Settings of Standard drying
•
Rapid drying see Settings of Rapid drying
•
Gentle drying see Settings of Gentle drying
•
Steps drying see Settings of Step drying
See also
2 Test measurement } Page 51
Types of drying programs
You can define the following parameters:
This drying program is set in the factory and is suitable for most samples. The sample is
heated to the drying temperature with default dynamics.
Factory setting
Drying temperature = 105 °C, Switch-off criterion 3 = 1 mg / 50 s
The following parameters can be defined:
Parameter
Description
Values
Drying
temperature
Defines the drying temperature.
40…230 °C
(105 °C *)
Switch-off
criterion
Defines the criterion when the instrument should end the drying.
1 (1 mg / 10 s)
2 (1 mg / 20 s)
3 (1 mg / 50 s)*
4 (1 mg / 90 s)
5 (1 mg / 140 s)
Timed...
Free (mg / s)...
Free (% / s)...
SOC delay
Activates and defines the delay for the starting time of the switch-
off criterion.
OFF = no delay applied.
OFF* | ON |
0…99 min
Setting the drying temperature
Navigation: Method Definition > Name of method > Main measurement parameters > Drying temperature
In this menu item you can set the final drying temperature.
Note
The admissible input range for the drying temperature is indicated.
1 Start of drying
2 Final temperature
Setting switch-off criterion
see [Switch-off criterion (SOC) } Page 36]
Rapid drying
Rapid drying
This drying program is primarily suitable for samples with a moisture content over 30%.
Following the start, the selected temperature is exceeded by 40% for 3 minutes (the
maximum heating temperature is 230 °C) to compensate for the cooling due to vapor-
ization and to accelerate the drying process. The drying temperature is then lowered to the
set value and maintained.
Factory setting
Drying temperature = 105 °C, Switch-off criterion 3 = 1 mg / 50 s
The following parameters can be defined:
Defines the drying temperature.
40…230 °C
(105 °C *)
Switch-off
criterion
Defines the criterion when the instrument should end the drying.
1 (1 mg / 10 s)
2 (1 mg / 20 s)
3 (1 mg / 50 s)*
4 (1 mg / 90 s)
5 (1 mg / 140 s)
Timed...
Free (mg / s)...
Free (% / s)...
* Factory setting
Setting the drying temperature
Navigation: Method Definition > Name of method > Main measurement parameters > Drying temperature
In this menu item you can set the final drying temperature.
Note
The admissible input range for the drying temperature is indicated.
1 Start of drying
2 Final temperature
Setting switch-off criterion
see [Switch-off criterion (SOC) } Page 36]
Gentle drying
Gentle drying
This drying program is suitable for the gentle drying of substances which tend to form a
skin (e.g. substances containing sugar) or substances that contain highly volatile
components. With this program the temperature is continuously increased and attains the
selected drying temperature only after the elapse of the so-called ramp time. The «ramp»,
i.e. the time that should elapse between the start of drying and attainment of the final
temperature can be pre-set.
Factory setting
Drying temperature = 105 °C, Ramp time = 3:00 min,
Switch-off criterion 3 = 1 mg / 50 s
The following parameters can be defined:
Parameter
Description
Values
Drying
temperature
Defines the drying temperature.
40…230 °C
(105 °C *)
Ramp time
Defines the time in minutes that should elapse between the start
of drying and attainment of the final temperature for Gentle
drying.
0…480 min
Defines the criterion when the instrument should end the drying.
1 (1 mg / 10 s)
2 (1 mg / 20 s)
3 (1 mg / 50 s)*
4 (1 mg / 90 s)
5 (1 mg / 140 s)
Timed...
Free (mg / s)...
Free (% / s)...
* Factory setting
Setting the drying temperature
Navigation: Method Definition > Method name > Main measurement parameters > Drying program >
Gentle > Drying temperature
In this menu item you can set the drying temperature.
1 Start of drying
2 Final temperature
Setting the ramp time
Navigation: Method Definition > Method name > Main measurement parameters > Drying programGentle
> Ramp time
In this menu item you can define the Ramp time, i.e. the time that should elapse between the start of drying
and attainment of the final temperature. The ramp is started once the temperature reaches 50 °C.
1 Start of drying
2 Final temperature
3 Ramp time
Setting switch-off criterion
see [Switch-off criterion (SOC) } Page 36]
Switch-off criterion (SOC)
Navigation: Home > Method Definition > Method name > Main measurement parameters > Drying
program > Switch-off criterion
This function offers you different switch-off criteria. A switch-off criterion defines when the instrument should end
the drying or should start the next step ("Step drying"). Switch-off criteria (except for timed switch-off) ensure
that measurements are finished in the same condition (weight loss per unit of time) every time, ensuring
repeatable measurements.
Note
Repeatable results require same start weights, see [Start weight } Page 42].
The following parameters can be defined:
Defines the criterion when the instrument should end the drying.
1 (1 mg / 10 s)
2 (1 mg / 20 s)
3 (1 mg / 50 s)*
4 (1 mg / 90 s)
5 (1 mg / 140 s)
Timed...
Free (mg / s)...
Free (% / s)...
* Factory setting
•
Weight loss per unit of time (5 preprogrammed settings)
•
Free switch-off criterion (2 different settings)
•
Timed switch-off
Weight loss per unit of time
The switch-off defines in which drying condition, the measurement is completed. This switch-off is based on a
weight loss per unit of time. As soon as the mean weight loss is less than the preset value during a specified
time, the instrument considers drying as complete and automatically discontinues the measurement process.
Note
The switch-off criterion is inactive during the first 30 sec.
The following 5 preprogrammed settings are available
Switch-off
criterion
Δ g
Δ t
Description
1
1 mg
10 s
This setting is suitable for fast measurements to determine a trend.
2
1 mg
20 s
This setting is suitable for quick drying samples.
3
1 mg
50 s
This is the factory setting. It is suitable for most kind of samples.
4
1 mg
90 s
This setting is suitable for samples which dry moderately quickly or for
samples with higher accuracy requirements.
5
1 mg
140 s
This setting is suitable for samples which dry very slowly (trapped
moisture, skin formation) or for samples with very low moisture content
e.g. plastics. Not suitable for very temperature sensitive samples.
The graph below exemplifies the mode of operation of the switch-off operation (not to scale).
∆g
∆g
∆g
∆ g
(∆ t
)
∆t
(10 s)
∆t
(140 s)
∆t
(50 s)
t
1)
2)
1 (1 mg / 10 s)
3 (1 mg / 50 s)
5 (1 mg / 140 s)
W
W
=
=
t
=
time
1)
=
equal area
W
=
weight of sample
2)
=
mean weight loss per unit of time
1, 3, 5
=
switch-off criteria shown as an example
Free switch-off criterion
The free switch-off criterion is based on a user-defined mean weight loss per unit of time.
The following 2 settings are available:
•
Switch-off criterion > Free (mg / s)... (weight loss per unit of time)
•
Switch-off criterion > Free (% / s)... (weight loss in percent of the start weight per unit of time)
Timed switch-off
With this switch-off criterion the measurement lasts until the pre-set drying time has elapsed. The display
provides you with continuous information in the drying time.
Performing test measurements
For performing test measurements to determine a correct switch-off criterion, see [Test
measurement } Page 51].
SOC delay
Navigation: Home > Method Definition > Method name > Main measurement parameters > Drying
program (Standard) > SOC delay
SOC delay is only available for the Standard drying program.
With this function you can delay the starting time of the switch-off criterion. This can be useful for samples with
low moisture content which need extended time until they evaporate moisture e.g. plastics.
The following parameters can be defined:
Activates and defines the delay for the starting time of the switch-
off criterion.
OFF = no delay applied.
OFF* | ON |
0…99 min
* Factory setting
7.1.1.1.2
Display mode
Navigation: Home > Method Definition > Method name > Main measurement parameters > Display mode
With this function you can select the desired type of result display. You also define what type of values are
printed on the records.
The following parameters can be defined:
Parameter
Description
Values
Display mode
Defines the type of value for displaying and printing.
%MC* | %DC | %AM |
%AD | g| g/kg MC |
g/kg DC | –%MC
* Factory setting
•
%MC – Moisture Content (calculated value)
•
%DC – Dry Content (calculated value)
•
%AM – ATRO Moisture Content (calculated value)
•
%AD – ATRO Dry Content (Wet weight, calculated value)
•
g – Measured value in Grams
•
g/kg MC – Moisture Content (calculated value)
•
g/kg DC – Dry Content (calculated value)
•
–%MC – Moisture Content (negative value, calculated value)
Note
Calculated values are indicated with an asterisk in the display.
Detailed information:
%MC – Moisture Content
The moisture content of the sample is displayed (and printed out) as a
percentage of the wet weight (WW = initial weight = 100%). This is the
factory setting.
During the measurement the value is constantly displayed in percent and
as a graphical drying curve. The measured value is marked by
"%MC" (Moisture Content, e.g. 11.35 %MC) also for the printed results.
MC
WW
DW
WW
=
–
· 100 %
MC = Moisture Content [0…100%]
WW = wet weight
DW = dry weight
WW
DW
MC
Before drying
After drying
%DC – Dry Content
The dry content of the sample is displayed (and printed out) as a
percentage of the wet weight (WW = initial weight = 100%).
During the measurement the value is constantly displayed in percent and
as a graphical drying curve. The measured value is marked by
"%DC" (Dry Content, e.g. 88.65 %DC) also for the printed results.
DC
DW
WW
=
· 100 %
DC = dry content [100…0%]
WW = wet weight
DW = dry weight
Before drying
After drying
%AM – ATRO Moisture Content 1)
The moisture content of the sample is displayed (and printed out) as a
percentage of the dry weight (DW = final weight = 100%)
During the measurement the value is constantly displayed in percent and
as a graphical drying curve. The measured value is marked by
"%AM" (ATRO Moisture Content, e.g. 255.33 %AM) also for the printed
results.
AM
WW
DW
DW
=
–
· 100 %
AM = ATRO moisture content [0…1000%]
WW = wet weight
DW = dry weight
Before drying
After drying
%AD – ATRO Dry Content (Wet weight)1)
The wet weight of the sample is displayed (and printed out) as a
percentage of the dry weight (DW = final weight = 100%)
During the measurement the value is constantly displayed in percent and
as a graphical drying curve. The measured value is marked by
"%AD" (ATRO Dry Content, e.g. 312.56 %AD) also for the printed
results..
AD
WW
DW
=
· 100 %
AD = ATRO dry content [100…1000%]
WW = wet weight
DW = dry weight
Before drying
After drying
1) Comment on the ATRO display mode
If the current measured value in the ATRO display mode is greater or less than the predefined limit value (i.e.
greater than 999.99 %AD or less than –999.99 %AM), the ATRO result values are limited to 999.99%.
g – Weight in Grams
The weight of the sample is displayed (and printed out) in grams. With
this setting, the Moisture Analyzer is used as a precision balance.
During the measurement current weight is constantly displayed in
grams.
g/kg MC – Moisture Content
The moisture content of the sample is displayed (and printed out)in g/kg
of the wet weight (WW = initial weight = 1000 g/kg).
During the measuremnet the current measured value is constantly
displayed in percent and as a graphical drying curve. The measured
value is marked by "g/kg MC" (Moisture Content, e.g. 11.35 g/kg MC)
also for the printed results.
MC
WW
DW
WW
=
–
· 1000 g/kg
MC = Moisture Content [0…1000 g/kg]
WW = wet weight
DW = dry weight
WW
DW
MC
Before drying
After drying
g/kg DC – Dry Content
The dry content of the sample is displayed (and printed out) in g/kg of
the wet weight (WW = initial weight = 1000 g/kg).
During the measuremnet the current measured value is constantly
displayed in percent and as a graphical drying curve. The measured
value is marked by "g/kg DC" (Dry Content, e.g. 88.65 g/kg DC) also for
the printed results.
DC
DW
WW
=
· 1000 g/kg
DC = dry content [1000…0 g/kg]
WW = wet weight
DW = dry weight
Before drying
After drying
–%MC – Moisture Content
The moisture content of the sample is displayed (and printed out) as a
percentage of the wet weight (WW = initial weight = 100%).
During the measurement the value is constantly displayed in percent and
as a graphical drying curve. The measured value is marked by "-
%MC" (Moisture Content, e.g. -11.35 -%MC) also for the printed results
and shown as a negative value.
MC
WW
DW
WW
=
–
· 100 %
–
MC = Moisture Content [0…100%]
WW = wet weight
DW = dry weight
WW
DW
MC
Before drying
After drying
7.1.1.1.3
Start weight
Navigation: Home > Method Definition > Method name > Main measurement parameters > Start weight
The start weight affects the duration as well as the end result of a measurement for a number of reasons.
Defining a starting weight ensures that approximately the same amount of sample is used for each
measurement, which improves the repeatability of the measurements.
The weighing-in aid assists you in weighing-in the sample. For most of the samples the target weight is in the
range of 2–5 g (plastics 30 g). We recommend to cover the whole surface of the sample pan with a thin
uniform layer of sample.
The following parameters can be defined:
Parameter
Description
Values
Start weight
Defines a start weight so that the samples always have approxi-
mately the same weight .
OFF* | ON
(0.100…200.000 g)
Start weight
tolerance
Defines the tolerance of the Start weight.
10 %, recommended for sample weights up to 5 g.
1…25 %
(10 %)*
Weigh-in aid
Defines monitoring of target weight and tolerance.
Passive = tolerance is displayed.
Active = tolerance is monitored. If the startweight is out of
tolerance, the measurement cannot be started.
Passive* | Active
* Factory setting
7.1.1.2 Result and value handling
Navigation: Home > Method Definition > Method name > Result and value handling
The following parameters can be defined:
Defines the acceptable range for measuring results
in the unit of the selected display mode.
see [Control limits } Page 43]
QuickPredict
Activates the QuickPredict function for a method and
defines the specifics for its use.
see [QuickPredict } Page 44]
7.1.1.2.1
Control limits
Navigation: Home > Method Definition > Method name > Result and value handling > Control limits
With this function you can set warning and intervention limits for the quality and process monitoring. You can
also set limits for distinguish passed or failed.
The following parameters can be defined:
Parameter
Description
Values
Control limits
Defines the acceptable range for the measuring results in the unit
of the selected display mode.
OFF = no limits applied.
OFF* | ON
(0.00…100.00)
T1+ (Upper
warning limit)
Defines the upper warning limit in the selected unit of the display
mode. Measurement passed with warning (!) if above limit.
Value range in ATRO mode (%AD, %AM) up to 1000.
OFF* |
0.01…100.0 (1000)
T1- (Lower
warning limit)
Defines the lower warning limit in the unit of the selected display
mode. Measurement passed with warning (!) if below limit.
Value range in ATRO mode (%AD, %AM) up to 1000.
OFF* |
0.01…100.0 (1000)
T2+ (Upper inter-
vention limit)
Defines the upper intervention limit in the unit of the selected
display mode. Measurement failed (X) if above limit.
Value range in ATRO mode (%AD, %AM) up to 1000.
OFF* |
0.01…100.0 (1000)
T2- (Lower inter-
vention limit)
Defines the lower intervention limit in the unit of the selected
display mode. Measurement failed (X) if below limit.
Value range in ATRO mode (%AD, %AM) up to 1000.
OFF* |
0.01…100.0 (1000)
* Factory setting
Setting the limits
T2+ (Upper intervention limit)
(X)
Off
Off
T1+ (Upper warning limit)
(!)
Off
Off
Off
Off
T1- (Lower warning limit)
(!)
Off
Off
Off
Off
T2- (Lower intervention limit)
(X)
Off
Off
(✓)
Measurement is within the warning limits: passed (displayed in green)
(!)
Measurement is between the warning limits and the intervention limits: passed with warning (displayed
in yellow)
(X)
Measurement is outside the intervention limit: failed (displayed in red)
Control limits is off = no limits applied (displayed in blue)
Testing the limits
See [Test measurement } Page 51].
7.1.1.2.2
QuickPredict
Navigation: Home > Method Definition > Method name > Result and value handling > QuickPredict
The QuickPredict function predicts the result of a measurement shortly after its start, e.g. after two minutes. Find
more information on the optimum use of QuickPredict in chapter [Preparing a method for Quick-
Predict } Page 52].
The following parameters can be defined:
Parameter
Description
Values
QuickPredict
Activates or deactivates QuickPredict for a method.
OFF* | ON
QuickPredict
usage
Activates and deactivates QuickPredict for routine measurements
(only possible if the prediction knowledge is established).
Inactive* | Active
Reference Curves Lists the reference curves this method.
any
Measurement settings
Parameter
Description
Values
Allow saving of
prediction
Activates the Save and Continue button in the prediction screen.
(deactivated)* |
(activated)
Threshold to
allow saving
Activates and defines the threshold for the Save and Continue
function in the selected display mode.
OFF* | ON | 0.00 ...
100.00
Allow stopping of
measurement
Activates the Save and Stop button in the prediction screen.
(deactivated)* |
(activated)
Threshold to
allow stopping
Activates and defines the threshold for the Save and Stop function
in the selected display mode.
OFF* | ON | 0.00 ...
100.00
* Factory setting
See also
2 QuickPredict } Page 44
2 QuickPredict } Page 44
2 QuickPredict } Page 44
2 QuickPredict } Page 44
7.1.1.3 Workflow handling
Navigation: Home > Method Definition > Method name > Workflow handling
With this functions you can define the workflow for the measurement
You can define the following parameters:
Menu item
Explanation
Further Information
Starting mode
Defines how the sample chamber is operated.
see [Starting mode } Page 45]
7.1.1.3.1
Starting mode
Navigation: Home > Method Definition > Method name > Workflow handling > Starting mode
In this menu you can choose whether the sample chamber performs automatic or manual functions.
The following parameters can be defined:
Parameter
Description
Values
Starting mode
Defines how the sample chamber is operated.
Automatic* |
Manual
* Factory setting
Automatic
Your instrument is set in the factory to the automatic operating mode. This mode can be used for most kind of
samples. When closing the sample chamber, the weight of the sample is recorded and the measurement is
started.
Manual
We advise you to use the manual operating mode for samples which contain readily volatile substances. In the
manual operating mode, the process does not automatically start when the [Start drying] button is tapped.
However, the initial weight (wet weight) important for the determination of the moisture content is recorded. In
the manual operating mode you have time for further preparation of the sample (e.g. mixing with quartz sand
or even distribution of the sample) while weight losses due to evaporation during the preparation time are
measured from the outset. In the manual operating mode you can open the sample chamber during a drying
operation e.g. in standby mode. In contrast to the automatic mode, drying will not be stopped but simply inter-
rupted until the sample chamber is closed again.
7.1.1.3.2
Accessories
Navigation: Home > Method Definition > Method name > Workflow handling > Accessories
For a method, certain accessories can be specified that must be used for the measurements. During the
measurement, these accessories are then requested on the work screen.
The following parameters can be defined:
Parameter
Description
Values
Accessories
Defines the main accessory (pan) in which the sample is
prepared.
Sample pan* | HA-Cage
Additional acces-
sories (only
available if the
sample pan is the
main accessory:
•
Glass fibre
filter
•
Second glass
fibre filter
•
Downholder
Defines additional accessories to be used with the sample pan.
(deactivated)* |
(activated)
* Factory setting
7.1.1.4 General method properties
Navigation: Home > Method Definition > Method name > General method properties
7.1.1.4.1
Method name
Navigation: Home > Method Definition > Method name > General method properties > Method name
The following parameters can be defined:
Parameter
Description
Values
Method name
Renaming a method. Name should be unique and unambiguous. any
7.1.1.4.2
Sample preparation
Navigation: Home > Method Definition > Method name > General method properties > Sample prepa-
ration
If a method is created with the method wizard (see [Creating a method with the Method Wizard } Page 47]),
the wizard automatically fills the Sample preparation field with the recommended preparation steps. The
recommended preparation can be edited by tapping Sample preparation, but the number of characters in the
text is limited to 100.
The sample preparation recommendation is not visible in the workflow during routine measurements.
The following parameters can be defined:
Parameter
Description
Values
Sample prepa-
ration
Specifies the sample preparation for a method recommended by
the Method Wizard.
any
7.1.2 Creating a method
Prerequisite
The parameter menu appears only if a method already exists and is selected in the method definition dialog or
if a new method has been created.
Main menu
Submenu
Further information
Main measurement
parameters
Drying program
(including temperature and switch-off
criteria)
see Settings of Drying Program
Display mode
see [Display mode } Page 39]
Start weight
see [Start weight } Page 42]
Result and value handling
Control limits
see Control limits
Resolution
see Resolution
Free factor
see Free Factor
QuickPredict
see [QuickPredict } Page 44]
Workflow handling
Starting mode
see Starting Mode
Preheating
see Preheating.
Accessories
General method properties
Method name
see [Method name } Page 46]
Sample preparation
see [Sample preparation } Page 46]
7.1.2.1 Creating a method with the Method Wizard
Introduction
The Method Wizard is a feature that offers assistance in developing a suitable moisture method for specific
samples. Based on the characteristics of your sample as well as a test of the drying temperature and switch-off
criterion, the Method Wizard is able to formulate a suitable method for your sample.
The Method Wizard guides you through the development of your moisture method in three simple steps:
1.
Questionnaire: proposal of method parameters based on sample characteristics.
2.
Temperature assistant: determination of drying temperature.
3.
Test measurement: verification of temperature and selection of appropriate switch-off criterion
After each step of the Method Wizard you can decide whether to accept the current method settings or to
complete another step of the wizard. For steps 2 and 3, a reference moisture content value is needed.
Creating a new method with the Method Wizard
§
Reference value (for Steps 2 and 3) is available.
§
Method definition is selected.
1
Tap [New...].
2
Tap Define using Method Wizard.
ð The Method Wizard disclaimer text appears.
3
Read the disclaimer text and confirm by tapping [Next].
ð The keypad appears.
4
Enter a name for the new method. The name should be unique and unambiguous. 1 to 30 characters are
possible (including spaces).
5
Confirm with [OK].
ð A short introduction appears.
6
Read the instruction and tap Next.
ð The Method Wizard starts.
Step 1 (Questionnaire): Follow through the questions by tapping the answers and then tapping [Next].
ð After all questions are answered, you are proposed
a method and asked if you want to adopt this
method or continue with the next step.
8
• To continue, tap Continue with temperature
assistant (recommended).
• To leave the wizard, tap Accept method parameters
and exit wizard.
9
Confirm by tapping [Next].
10 Step 2 (Temperature assistant): Follow the
instructions of the wizard to determine a suitable drying
temperature. For questions about the temperature
assistant, see [Temperature assistant } Page 50]. This step may require several iterations.
ð After successful testing, you are proposed a drying temperature and asked if you want to complete the
next step or adopt the current method.
11 • To continue, tap Continue with test measurement (recommended).
• To exit the wizard, tap Accept method parameters and exit wizard.
12 Tap [Next].
13 Step 3 (Test measurement): Follow the instructions of the wizard to select an appropriate switch-off
criterion. For questions about the test measurement, see [Test measurement } Page 51].
ð After successful testing, you are recommended a switch-off criterion and are asked if you want to adopt
it or perform another test.
14 • To adopt the method, tap the desired switch-off criterion.
• To reject the proposed switch-off criterion and perform another test at a different drying temperature, tap
Start over again using a different drying temperature.
15 Confirm by tapping [Next].
16 To store the method, tap [Finish].
ð The method settings appear. The new parameters are already filled in.
17 NOTICE: Data loss: To definitively store your method settings, tap [Save] before you exit the method
settings. Otherwise all data is lost.
7.1.2.2 Creating a method manually
§
Method definition is selected.
1
Tap [New...].
2
Tap Define method manually.
ð The keypad appears.
3
Enter a name for the new method. The name should be unique (a method name can only exist once) and
unambiguous. 1 to 30 characters are possible (including spaces).
4
Confirm with [OK].
ð The parameter menu for the new method appears.
5
Set the desired parameters e.g. Main measurement parameters.
6
To store the method, tap [Save].
7.1.2.3 Creating a method by consulting the method library
The online METTLER TOLEDO moisture method library offers a wide range of tested and validated methods for
different industries and substances. Optimize your moisture analysis with the help of our moisture expertise and
profit from a wealth of know-how and widespread support.
Find the moisture library under:
u www.mt.com/moisture-methods
Downloading a method from the method library
§
PC with internet connection is available.
§
Storage device (e.g. USB Stick) with at least 500 MB memory space is available.
1
Open the method library (see above) on your PC.
2
Click Moisture Applications.
3
Fill out the form according to the instructions to find a moisture method for your sample.
4
Click the method you wish to download.
ð The access form opens.
5
Fill out the form and click Download to download the method file.
6
Transfer the file onto your storage device.
Importing the method to the istrument
§
Method Definition is opened.
1
Connect your storage device to the instrument.
2
Tap [Import].
ð Import methods screen opens.
3
Tap [Import from file].
ð The list of connected hard drives opens.
4
Tap the device and location where your method file is stored and select your file.
5
Confirm by tapping Import.
ð Your new method is imported and is now ready for routine measurements in the method list.
7.1.3 Editing a method
Editing an existing method
§
Method definition is selected.
1
Tap the method in the list you want to edit.
ð The parameter menu for the selected method appears.
2
Edit the desired parameters e.g. Main measurement parameters.
3
To store the method, tap [Save].
Copying an existing method
§
Method definition is selected.
1
Tap the method in the list you want to copy.
ð The parameter menu of the desired method appears.
2
Tap [Save as...].
ð The keypad appears.
3
Enter a new name for the copied method. The name should be unique and unambiguous. 1 to 30
characters are possible.
4
To store the method, tap [Save].
Removing an existing method
Note
All results of this method (incl. results of previous versions) as well as corresponding shortcuts are also
removed.
§
Method definition is selected.
1
Tap the method in the list you want to remove.
ð The parameter menu of the desired method appears.
2
Tap [Delete].
ð A message box appears.
Confirm with [Delete].
ð The desired method is removed.
See also
2 Working with shortcuts } Page 62
7.1.4 Testing a method
Navigation: Home > Method Definition > Method name > Test method
By testing a method before its application, its efficiency and the duration of a measurement can be optimised.
Both the temperature and the switch-off criterion can be adjusted to an ideal setting for the sample with a test,
shortening the test duration and increasing the reliability of the measurement at the same time.
Testing of a method is possible at any time after the method has been created.
7.1.4.1 Temperature assistant
Introduction
The temperature assistant is a test function that determines a suitable drying temperature for a specific sample
by drying the test sample while comparing the measured moisture content to a reference value. Based on this
procedure, the temperature assistant is able to specify a suitable temperature at the end of the test.
Testing procedure
§
A reference value for the test sample is available.
§
Test method is opened.
1
Tap Temperature assistant and confirm with Next.
2
Enter the reference value and confirm with Next.
ð A warning text appears, indicating the highest temperature possible in this test and asking if this peak
temperature is still acceptable for the sample or poses a danger.
3
• If the peak test temperature poses no danger to the sample, select Accept maximum drying temperature
of x °C [Value] and confirm with Next.
• If the peak test temperature poses a possible danger to the sample, select Enter another maximum
drying temperature, enter the new drying temperature and tap [OK].
ð The work screen appears.
4
Follow the instructions on the screen and carry out the test measurement like a normal measurement, see
Performing a measurement.
ð Once the procedure is finished, the instrument shows the result of the test:
• If the test was successful, the instrument proposes a suitable drying temperature. If this temperature differs
from the one previously defined, the instrument changes this parameter automatically in the method (but
does not yet save it).
• If the test fails (e.g. if the final moisture content was higher than the reference value), the instrument
proposes to reiterate the test (begin from step 1).
Saving or rejecting the proposed drying temperature
1
Below the screen showing the new proposed drying temperature, tap Next.
ð The method parameter screen appears. In the Main measurement parameters, the new temperature is
already entered.
2
To accept the new temperature, tap Save.
To reject the new temperature, tap Drying temperature, enter your desired temperature and confirm with OK
or press Cancel to discard all changes.
Note
The results of a temperature assistant test cannot be printed.
7.1.4.2 Test measurement
Introduction
A test measurement helps to define the best switch-off criterion (SOC) for a specific sample. In a test
measurement, the instrument records the points in the drying curve at which the different switch-off points (1 to
5 and, if available, customary SOC) have been reached. The measurement terminates after all SOC have been
reached (default setting) or after a custom set time period. After the test is finished, the user can evaluate the
drying curve to decide which criterion suits the method best.
0 1 2 3 4 5 6 7 8 9 10 11
t
MC (%)
0
5
10
15
20
1
2
4
5
3
The diagram illustrates the progress of a measurement. The points at which the individual switch-off criteria
were reached (1-5) are marked. If measurement duration has priority in future measurements, SOC 2 can be
selected for the method. However, if precision and reliability of results are more important, SOC 3 or SOC 4 are
the more appropriate choice.
By comparing the measured values with values of a reference method, e.g., drying oven and the standard
deviation, you can define the appropriate parameter settings. For more information see the application brochure
«Guide to Moisture Analysis».
For more information about switch-off criteria, see [Switch-off criterion (SOC) } Page 36].
Testing procedure
§
Test method is opened.
1
Tap Test measurement.
ð The test work screen appears.
2
If you wish for the test to stop after a certain time period, tap Test time and activate and set the test time.
3
Follow the instructions on the screen and carry out the test measurement like a normal measurement, see
Performing a measurement.
ð After the test, the instrument shows the results in the method journal.
Result view
In the graphical visualization of the test process, the details about the different SOC (time and % MC at the
point of reaching) can be accessed:
•
by clicking the points on the drying curve. The details appear below the title bar.
•
by navigating with
and
left and right on the screen. The details appear below the title bar.
•
by tapping More data. A detailed list about the test details opens.
Test measurements are recorded in the journal of a method and marked as test results.
Adjusting the switch-off criterion
The test measurement does not change the method parameters. To adjust the SOC of a method, the parameters
have to be changed manually. Before releasing the method, check if the determined switch-off criterion is set.
Printing results of the determined switch-off criteria
For a printout example , See Information on Printouts section «Special events».
Displaying results of the determined switch-off criteria
See Graphical Evaluations of Measurement Results
7.1.4.3 Testing with AutoMet
Introduction
In combination with the test measurement you can use the AutoMet function to enter the reference value and
the instrument to automatically establish an appropriate switch-off criterion for the defined method. This will be
the point at which the measurement most accurately matches the reference value. This is the free switch-off
criterion Free (mg / s).... The AutoMet test measurement only proposes the switch-off criterion Free (mg / s)...
if the drying trend means you can expect a precise moisture determination. The range for the switch-off criterion
Free (mg / s)... is therefore limited to between 1mg/20 seconds and 1mg/140 seconds.
Note
If the target value is reached but the switch-off criterion is still outside 1mg/20-140 seconds, you have not
selected the optimum temperature. If Free (mg / s)... is below 1mg/20 seconds, reduce the temperature. If
Free (mg / s)... is above 1mg/140 seconds, either switch-off criterion 5 will provide a good approximation or
it might be a good idea to increase the drying temperature. If the target value is not reached, increasing the
drying temperature may obtain the desired result.
Note
METTLER TOLEDO recommends usind the Method Wizard function to find an optimum switch-off criterion
instead of the AutoMet function.
AutoMet
1
Perform a test measurement with the same sample as for the reference measurement.
2
Tap [AutoMet].
3
Enter Measurement reference in display unit (target value) and tap [OK].
ð AutoMet suggests you an appropriate switch-off criterion for the temperature selected or gives further
instructions on the display.
4
Tap [Save] to assume this switch-off criterion into the method.
7.1.5 Preparing a method for QuickPredict
Introduction
The intended use of QuickPredict is to predict the final result of a measurement shortly after its start. To achieve
best prediction performance and take into account the different drying characteristics of samples, substance-
specific reference data is needed.
QuickPredict works on the basis of the so-called prediction knowledge, which consists of substance specific
data and which is stored as part of the method. To establish this prediction knowledge for a particular method,
six representative measurements must be made, which form the reference data set and from whose data the
prediction knowledge is then calculated.
Key factor for the performance of QuickPredict are the quality of the sample specific data in the prediction
knowledge as well as high repeatability of method and sample, which means that not every substance and
method is suitable.
The following chapter helps you in determining if your substance is suitable and explains how to set your
instrument and chosen method up for the use of QuickPredict.
7.1.5.1 Prerequisites to use QuickPredict for a method
QuickPredict cannot be used for every kind of substance and method. The function performance relies heavily
on repeatability and characteristics of the drying curve, which means that certain prerequisites must be met for
QuickPredict to function as intended.
Suitability of purpose (Disclaimer)
The prediction functionality intends to support the customer in getting an estimation on results by using a set of
algorithms to predict the end result of a measurement. By using the 'QuickPredict Module' you acknowledge to
be fully aware, that any predicted results come with an uncertainty and may not in every case represent the
final value. Although METTLER TOLEDO attempts to provide predictions with a certain degree of accuracy, it
makes no representation that the predicted result, including the prediction uncertainty, are accurate or suitable
for any particular purpose. Your use of the prediction functionality is at your own risk and you must validate the
predicted results according to internal guidelines and corresponding validating procedures for making your
decisions. [...]
Sample suitability
Substances that are well suited for results prediction are characterized by a drying curve with a characteristic
curving during the initial stage of the measurement.
A pronounced curvature of the drying curve during the early stages of the measurement improves prediction
precision and reduces prediction uncertainty. During the recording of the reference curves, the instrument
automatically analyzes each curve and indicates their suitability for usage with QuickPredict. The majority of
samples tested for moisture content fulfill the criteria stated above.
time
time
%MC
%MC
time
time
%MC
%MC
The compatibility of the substance with QuickPredict is verified during the calculation of the prediction
knowledge. As such, it is possible to test the suitability of a substance for QuickPredict without a license.
Method suitability
The reference data set for QuickPredict is only created for a specific combination of settings of a method. If
these (result influencing) settings are changed, the reference data set for this method is deleted. For this reason,
only use QuickPredict for stable and validated methods for which you are sure that they will not need any
further changes.
Settings irrelevant for the results, such as names, can still be changed after the reference curves have been
recorded.
The suitability of a method is determined when the QuickPredict function is activated for a method.
QuickPredict is only supported for methods with the following settings:
Parameter
Value
Drying program
•
Standard
•
Gentle
•
Rapid
Display mode
•
% MC
•
% DC
•
g
•
-% MC
Start weight
ON
Weigh-in aid
Active
7.1.5.2 Activating QuickPredict
7.1.5.2.1
License
QuickPredict is not available as a default function on all instruments.
For HX204-models with a serial number higher than S/N B922000000:
QuickPredict is available by default. No license is required.
For all HS153-models as well as HX204-instruments with a serial number lower than S/N
B922000000:
•
A software-update to V2.30 or higher is required.
•
A valid license is required, see Accessories
In the instrument info you can see if a license for the instrument is already activated.
Function of the QuickPredict License
Without a valid license, QuickPredict cannot be used in full functionality: The instrument can still record
reference curves and calculate the prediction knowledge. This can be useful to determine if a sample substance
is suitable for QuickPredict. However, to use QuickPredict in routine measurements, a valid license needs to be
ordered.
A license is only valid for a specific instrument, not for users or companies.
Activating a QuickPredict License
Activation Portal:
u activation.mt.com
§
No active license is registered in the instrument info.
§
License key has been ordered and received. See Accessories.
1
Open the activation portal (see above) on your PC.
2
Create a user profile under Register.
3
Once you are signed in on the portal, click Activation.
ð The Activation / Reactivation-page opens.
4
In the Add License Key-section, enter your license key in the input field.
ð Your license key is activated.
5
On your Moisture Analyzer, open Settings > System and data management > License manager.
6
Note the License Registration Information (LRI) that is shown under QuickPredict license (1).
7
In the portal, Enter the LRI from the instrument.
8
Generate a Short Activation Code (SAC).
On the instrument, tap QuickPredict license.
ð The input field for the SAC appears.
10 Enter the SAC from the portal and confirm by tapping OK.
1
ð The SAC appears in the License manager panel (2).
ð Your QuickPredict license is now active and Quick-
Predict is unlocked on your instrument.
7.1.5.2.2
Activating QuickPredict on the instrument
Enabling QuickPredict
If a license is activated for the instrument, QuickPredict is enabled as default. However, in case of an error or for
a re-activation, QuickPredict can be enabled manually with the following steps:
§
A valid QuickPredict license is active.
§
Settings is opened.
1
Tap Quality management.
ð The menu options open. QuickPredict is not active.
2
Tap QuickPredict.
ð The switch for QuickPredict appears.
3
Tap the switch.
ð A disclaimer text appears.
4
Scroll down, check the box for I accept the terms above and confirm with OK.
ð The switch turns green and the QuickPredict function is now enabled for the instrument. Note that the
function needs to be activated for every method separately.
Disabling QuickPredict
QuickPredict can be manually disabled for a instrument, e.g. if your process does not allow predicted results:
§
Settings is opened.
1
Tap Quality management.
ð The menu options open. QuickPredict is active.
2
Tap QuickPredict.
ð The switch for QuickPredict appears.
3
Tap the switch.
ð The switch turns red and the QuickPredict function is now disabled for the instrument. Note that this setting
applies to all methods.
7.1.5.2.3
Enable QuickPredict for a method
Before QuickPredict can be used for a specific method, the function must be enabled in the method settings.
§
Method Definition is opened.
1
Select the method for which you want to set up QuickPredict.
ð The method settings open.
2
Tap Result and value handling.
Tap QuickPredict (1).
ð The QuickPredict settings for the method appear.
4
Tap the switch in the upper left corner (2).
2
ð A disclaimer text appears.
5
Scroll down, check the box for I accept the terms
above and confirm with OK.
ð If any current settings are incompatible with Quick-
Predict, the corresponding settings are listed on the
screen. QuickPredict cannot be activated unless
these settings correspond to QuickPredict
requirements.
ð The switch turns green and the QuickPredict function is
now active for this method.
ð To use QuickPredict in routine measurements, the prediction knowledge must be established and Quick-
Predict usage must be set to Active. See [Establishing the prediction knowledge } Page 56]
7.1.5.3 Establishing the prediction knowledge
QuickPredict requires six measurements for the reference data set. To intercept environmental influences, three
measurements are started with a cold instrument, and three measurements are performed with a warmed-up
instrument. Thus, the prediction knowledge covers a wider range of values.
There are two ways to create the reference data set:
•
Creating the reference data set with help from the QuickPredict Wizard.
•
Creating the reference data set independently in expert mode (recommended only for experienced Quick-
Predict users).
7.1.5.3.1
Process schedule
The reference curves are divided into two groups:
•
Cold starts: Measurements/drying curves recorded with a cold instrument.
•
Warm starts: Measurements/drying curves recorded with a warmed-up instrument.
During the process, the instrument needs to be warmed up or cooled down accordingly, for several times. Any
heating up or cooling down times are indicated on the screen with a timer.
In QuickPredict Wizard: The wizard makes sure that the system is cold or hot before the reference curve is
recorded.
In expert mode: The user can decide on the schedule of the process but must ensure that the precondition of a
hot or cold instrument is fulfilled.
7.1.5.3.2
QuickPredict Wizard
The QuickPredict Wizard is the easiest and safest way to establish the prediction knowledge for a specific
method, as it guides you through all the necessary steps.
Starting the QuickPredict Wizard
§
QuickPredict is activated.
§
Method Definition is opened.
1
Select the method for which you want to set up QuickPredict.
1
ð The method settings open.
2
Tap Result and value handling.
3
Tap QuickPredict.
ð The QuickPredict settings for the method appear.
4
Tap [Wizard] (1).
ð An introductory text opens.
5
Read the text, tap Start recording of reference curves
and confirm by tapping [Next].
ð The QuickPredict Wizard starts directly. After starting,
the QuickPredict Wizard determines the state of the
instrument (cold or warm) and schedules the corresponding fastest recording sequence.
ð Depending on the state of your instrument, it warms up or cools down.
Recording procedure
§
The work screen for a reference measurement is opened. The upper right corner shows the number of
previously recorded reference curves, e.g. [1 / 6] means that you are currently recording the first reference
curve.
1
Follow the instructions on the screen to perform the reference measurement.
ð Upon completion of a measurement, the instrument evaluates the recorded curve for suitability for Quick-
Predict:
• If the recorded curve meets all quality criteria, the result is colored green for pass and the corre-
sponding reference curve is stored in the reference journal.
• If the recorded curve does not meet all quality criteria, the curve is rejected and the reason for the
rejection is stated. If your reference curves are rejected repeatedly, check your sample and method for
repeatability. Rejected reference curves are not stored in the reference database.
2
Repeat the procedure for the rest of the reference measurements. The samples must always be prepared and
presented in the same way.
ð Once all 6 reference curves have been successfully recorded, the instrument automatically calculates the
prediction knowledge. This might take several minutes. The progress of the calculation is shown on the
screen.
ð Upon completion, the screen shows if the prediction knowledge calculation has been successful or not.
Saving the QuickPredict settings
1
Tap [OK].
ð You are informed that QuickPredict has been added to your method.
2
Tap [Finish] to finish the procedure.
ð In the method settings, QuickPredict usage is now set to Active.
3
NOTICE: : Data loss: To definitively store your reference curves and prediction knowledge, tap [Save] before
you exit the method settings. Otherwise all recorded data is lost.
ð QuickPredict is now ready for routine measurements.
See also
2 Malfunctions shown on the display } Page 113
7.1.5.3.3
Expert mode
Recording the reference curves in expert mode is recommended only for experienced QuickPredict users. Expert
mode gives the user more leeway in the process, but also bears a higher risk of errors and data loss.
Starting expert mode
§
QuickPredict is activated.
§
Method Definition is opened.
1
Select the method for which you want to set up QuickPredict.
ð The method settings open.
2
Tap Result and value handling.
3
Tap QuickPredict.
ð The QuickPredict settings for the method appear.
4
Tap [Reference Curves].
ð The reference journal with the reference measurement list opens.
5
Tap one of the reference measurement bars for which you want to record a reference curve:
• For a reference measurement with a cold instrument, tap a
-bar.
• For a reference measurement with a warm instrument, tap a
-bar.
ð The recording procedure starts.
ð Depending on the state of your instrument, it is first warmed up or cooled down.
Recording procedure
§
The work screen for a reference measurement is opened. The upper right corner shows the number of
previously recorded reference curves, e.g. [1 / 6] means that you are currently recording the first reference
curve.
1
Follow the instructions on the screen to perform the reference measurement.
ð Upon completion of a measurement, the instrument evaluates the recorded curve for suitability for Quick-
Predict:
• If the recorded curve meets all quality criteria, the result is colored green for pass and the corre-
sponding reference curve is stored in the reference journal.
• If the recorded curve does not meet all quality criteria, the curve is rejected and the reason for the
rejection is stated. If your reference curves are rejected repeatedly, check your sample and method for
repeatability. Rejected reference curves are not stored in the reference database.
2
Open the next reference measurement tab in the reference journal.
3
Repeat the procedure for the rest of the reference measurements. The samples must always be prepared and
presented in the same way.
ð Once all 6 reference curves have been successfully recorded, a window appears: Ready for calculation
of prediction knowledge. Start now?
4
To start the calculation of the prediction knowledge, tap [Yes]
ð The calculation starts. This might take several minutes. The progress of the calculation is shown on the
screen.
ð Upon completion, the screen shows if the prediction knowledge calculation has been successful or not.
Saving the QuickPredict settings
1
Tap [OK].
ð You are informed that QuickPredict has been added to your method and that the usage has to be
activated.
2
If you would like to activate the QuickPredict usage now, check the box Activate 'QuickPredict usage'.
3
Tap [Finish] to finish the procedure.
ð In the method settings, QuickPredict usage is now set to Active.
4
NOTICE: : Data loss: To definitively store your reference curves and prediction knowledge, tap [Save] before
you exit the method settings. Otherwise all recorded data is lost.
ð QuickPredict is now active for routine measurements.
Calculating the prediction knowledge at a later time
In expert mode it is possible to start the prediction knowledge calculation at a later point in time than directly
after recording the reference curves. this option can also be used if, for example, a single reference curve is
newly recorded.
§
All reference curves are recorded and saved.
§
The measurement settings are opened.
1
Tap QuickPredict usage.
2
Set the switch to ON.
ð A window appears: Ready for calculation of prediction knowledge. Start now?
3
To start the calculation of the prediction knowledge, tap [Yes]
ð The calculation starts. This might take several minutes. The progress of the calculation is shown on the
screen.
ð Upon completion, the screen shows if the prediction knowledge calculation has been successful or not.
See also
2 Malfunctions shown on the display } Page 113
7.2 Measurement
Navigation: Home > Measurement
Introduction
Before a measurement can be performed, a method first needs to be defined. To define a suitable method for
your sample, see [Method Definition } Page 33].
The instrument guides you through the steps during a measurement and displays results in real time. Never-
theless, please note the instructions in this manual, especially at the beginning of your work with the moisture
analyzer, as a correct execution of the work steps and the handling of the samples can have a great influence
on the results.
7.2.1 Performing a measurement
Switching the instrument on
§
The instrument has been connected to the power supply for at least 60 minutes in order to reach operating
conditions.
1
To switch on the Instrument, press [
].
2
Log-in with your password, if needed.
Selecting a method
1
Tap on Measurement.
ð The methods list is displayed.
2
Select the method to determine your sample.
ð The work screen of the method appears.
3
Open the sample chamber.
Placing the accessories
§
The display now prompts you to load the predefined accessories and tare the balance.
1
Place the empty accessories in the sample pan handler.
Place the sample pan handler in the sample chamber.
Ensure that the tongue of the pan handler fits exactly in
the slot of the draft shield element. The accessories
must lie flat in the pan holder.
Note
METTLER TOLEDO recommends to work with the sample
pan handler at all times during a measurement. The pan
handler is ergonomic, ensures correct positioning, is safe
and provides protection against possible burns due to the
hot accessories.
Taring
1
Close the sample chamber.
ð The instrument tares the balance (Starting mode: Automatic).
2
After taring, open the sample chamber.
Starting the measurement
§
After taring, the display prompts you to add the sample to the accessory.
1
Add the sample to the accessory. If you have defined a start weight, weigh the sample using the weighing-
in aid.
2
Close the sample chamber.
3
Tap [Start drying].
ð The drying process starts.
ð During the measurement, the display continuously
shows the real time results of the drying process in
form of a drying curve.
ð When the drying process is finished, the final
measured moisture content of your sample is shown
on the display. If control limits are defined for the
method, the status (passed, warned or failed) is
shown.
Removing the sample
§
The drying process is completed.
§
The sample chamber is open.
1
CAUTION: Sample chamber and sample pan may still be hot. Carefully remove the sample pan
handler from the sample chamber.
Remove the accessories from the handler.
3
• To perform a further measurement with the current
method, tap [Next Sample].
• To perform a measurement with a new method, tap
[Measurement].
• To return to the home screen, press [
].
Stop drying
When you stop the measuring process by tapping [Stop drying], you can choose between two options:
•
Abort measurement
Already sampled data are stored and an entry will be made in the results. The result is marked as aborted.
•
Cancel
The process will be continued.
Adding a comment
At the end of a measurement you can add a comment to the result of the measurement. This comment is stored
in the measurement result and can be printed. A comment can only be entered before leaving the current
measurement.
1
To create a comment, tap Note.
ð The keyboard dialog appears.
2
Enter a comment.
3
Confirm with OK.
7.2.2 Optimum sample preparation
Preparation of the sample is decisive for the speed of the measurement process and the quality of the
measurement results.
Basic rules for the preparation of samples:
•
The amount of sample you select should be as small as possible but as large as necessary.
•
The more inhomogeneous the sample substance, the larger the amount of sample needed to obtain a
repeatable result.
•
Distribute the sample evenly over the sample pan. This increases the surface area of the sample and facil-
itates heat absorption.
•
With liquid, fat-containing, melting, highly reflecting samples or samples which form a skin on their surface
when heated, use the sample with the glass fiber filter available as optional equipment, see Accessories
and Spare Parts.
7.2.3 Working with weighing-in aid
The weighing-in aid can be defined for each method and facilitates weighing in of the sample to a target value.
This is particularly necessary if you require all the samples of a method you wish to process to have same
weight in order to improve the repeatability of the measurement results. In addition, the weighing-in aid can be
set as active so that the drying process cannot be started if the weight of sample is outside the set tolerance.
You are therefore compelled to weigh-in the correct quantity of sample. If all sample weighed are within
tolerances then this will improve the repeatability. The weighing-in aid is available only if start weight has been
activated. For further information see at section [Settings of Start Weight } Page 42].
Icon
Function
Lower weight limit (tolerance range)
Target weight
Upper weight limit (tolerance range)
7.2.4 Working with shortcuts
Shortcuts create a button on the home screen with which measurements can be started directly, which is most
useful for methods that are used very often.
Shortcuts are user-specific: every user can create shortcuts individually that only appear in the user profile. The
rights to create or edit shortcuts have to be defined by an administrator via the menu User management. If the
rights are denied, existing shortcuts can be used but it is not possible to create or to edit shortcuts. This enables
the administrator to preconfigure home-screens for operators to increase process security and efficiency.
Create a shortcut
Option 1: Creating a single shortcut directly from the work screen.
§
The method is opened in the work screen.
1
Tap [
] in the title bar.
ð A popup-window My shortcuts opens.
2
Tap Add a shortcut to my home for this method...
ð The window Shortcut properties appears.
3
If you want your shortcut to have a different name than the method, tap Shortcut name, enter the new name
and confirm with OK.
4
Tap Save.
ð The new shortcut appears on the home screen.
Option 2: Creating one or several shortcuts at once.
1
Tap [Measurement].
ð The list of the defined methods appears.
2
Tap the button Edit shortcuts... in the action bar.
ð A checklist of the defined method appears.
3
Check the methods from the list for which you want to
create a shortcut.
4
Tap [Save].
ð The new shortcuts appear on the home screen.
Edit a shortcut
1
Tap [Measurement].
ð The list of the defined methods appears.
2
Tap [Edit shortcuts...].
3
Tap the defined method from the list for which you want to edit the shortcut.
4
Define the shortcut name.
5
Tap [OK].
6
Tap [Save].
ð The shortcut has been edited.
Removing a shortcut
1
Tap [Measurement].
ð The list of the defined methods appears.
2
Tap [Edit shortcuts...].
3
Uncheck the shortcut that are to be removed from the home screen.
4
Tap [Save].
ð The shortcut is removed from the home screen.
7.2.5 Working with QuickPredict
Introduction
QuickPredict is an embedded function that predicts the result of a moisture measurement shortly after its start.
Based on the prediction knowledge of the method, QuickPredict analyzes the drying curve of a running
measurement and predicts the end result.
Note
Before QuickPredict can be used in routine operations, the function must be enabled for the method and the
prediction knowledge must be established. For the necessary settings and preparations, see [Preparing a
method for QuickPredict } Page 52].
7.2.5.1 Performing measurements with QuickPredict
If the Prediction Knowledge for a certain method has been established, the function can be activated and used
during measurements.
Performing measurements with QuickPredict
§
The prediction knowledge for the method is established.
§
QuickPredict is activated for the method.
1
Prepare the sample according to method conditions.
2
Start the measurement.
ð QuickPredict will try to release a prediction earliest after two minutes. Up to this point, the message
Prediction pending (1) is shown.
ð As soon as a prediction is possible, the screen shows the predicted result. Depending on the method
settings, Save and Stop and/or Save and Continue is now active.
ð If not stopped, the measurement will continue running until the final result is determined.
Elements on the prediction screen
1
2
3
4
5
6
7
8
10
9
Name
Explanation
1
QuickPredict icon
Appears during measurements for which QuickPredict Usage is activated.
2
Current prediction
Shows the current prediction of the end result value. The prediction may change
over time.
3
Dashed line
Predicted data.
4
Solid line
Measured data.
5
Control limit
colorations (only if
control limits are
enabled)
Prediction inclusive prediction uncertainty are:
•
within control limits: green
•
outside control limits: red
6
Save and Continue
button
See [Handling QuickPredict results } Page 66].
7
Elapsed
measurement time
8
Save and Stop
button
See [Handling QuickPredict results } Page 66].
9
Zoom-in (only if
control limits are
enabled)
Zooms in to a detailed view of the current prediction.
10
Current prediction
uncertainty
See below.
Messages on the prediction screen
During the measurement, the following messages can be displayed on the screen:
•
Prediction pending (1): This message is shown during the first few minutes of the measurement (initial-
ization time). Depending on the sample substance, the initialization time can be 2 or 4 minutes.
•
Prediction pending (x): If the initialization time is delayed (e.g. due to vibrations or differences in the
sample preparation or substance), this message is shown. QuickPredict is still trying to release a
prediction.
•
No prediction possible: QuickPredict cannot release a prediction. This can have several reasons, see
[Malfunctions shown on the display } Page 113]. If not aborted, the measurement will continue running
until the final result is determined.
Prediction uncertainty
Any prediction comes with a prediction uncertainty. The prediction uncertainty takes into account the repeata-
bility of the recorded reference curves and how much the currently measured curve is deviating from the
reference curves. This is why all predictions are stated i.e. as following: 5.00 ± 0.2 %MC. The final result is
therefore expected to be somewhere in the range of 4.80–5.20 %MC with a Gaussian probability distribution.
5.2
4.8
time
%MC
Display of control limits
If control limits are set for a method, the predicted results are colored according to their value. As with a normal
measurement, the display is colored according to the traffic light system, depending on whether a predicted
result is within or outside the control limits:
−
Red: Predicted result +/- prediction uncertainty surpass
the control limits:
−
Orange: Predicted result +/- prediction uncertainty
surpass the warning limits:
−
Green: Predicted result +/- prediction uncertainty are
within the control and warning limits:
No color: Control and warning limits are not enabled.
Note that the zoom function is also unavailable if no
limits are set:
See also
2 Malfunctions shown on the display } Page 113
7.2.5.2 Handling QuickPredict results
In routine operations, QuickPredict can be used in two ways once a prediction is available during a
measurement:
•
Save and Continue (typical for quality control): The user saves the current prediction and continues the
measurement. This option is most useful if you want to base a decision on a prediction but want to verify
the prediction at the end of the measurement. Both the acknowledged prediction, including the prediction
uncertainty, and the actual final value are documented.
Benefit: Fast decision making, verification of prediction, storage of actual end result.
•
Save and Stop (typical for in-process control): The user saves the current prediction and stops the
measurement. All decisions taken are based on the predicted value and the prediction uncertainty. Only
these values are documented.
Benefit: Short time to result, instrument is readily available for the next operator/measurement.
Depending on the specific use case of QuickPredict for a method, either none, just one or both of these two
options can be active. For relevant settings, see [QuickPredict } Page 44].
Activating Saving of prediction
§
QuickPredict is activated for the method.
§
The method settings are open.
1
Tap Result and value handling.
2
Tap QuickPredict.
ð The QuickPredict settings open.
3
Tap Measurement settings.
ð The Measurement settings open.
4
Check the box at Allow saving of prediction.
ð Allow saving of prediction is activated.
Activating Stopping of measurement
§
QuickPredict is activated for the method.
§
The method settings are open.
1
Tap Result and value handling.
2
Tap QuickPredict.
ð The QuickPredict settings open.
3
Tap Measurement settings
ð The Measurement settings open.
4
Check the box at Allow stopping of measurement.
ð Allow stopping of measurement is activated.
7.2.5.3 Working with uncertainty thresholds
Introduction
For both result handling options (Save and Stop and Save and Continue), a customizable threshold is
available. The threshold defines at what prediction uncertainty the corresponding option becomes available.
This means that it is possible to ensure that a measurement can, for example, only be stopped if the prediction
uncertainty reaches a certain value (e.g. is smaller than 0.03 %MC). Without the threshold, the activated
option is enabled as soon a prediction is available.
Thresholds should only be enabled after a reasonable amount of measurements with active results prediction
(> 20) have been carried out, after which mean values of prediction uncertainty for typical measurements are
at hand and reasonable values for thresholds can be determined. Usually and sensibly, the threshold to allow
Save and Continue would be put at a higher value than the one for Save and Stop.
Defining uncertainty thresholds
§
QuickPredict is activated for the method.
§
The result handling options (Save and Continue and / or Save and Stop) are set.
§
The Measurement settingsare opened.
1
• To set a threshold for the saving of a prediction, tap Threshold to allow saving.
• To set a threshold for the stopping of a prediction, tap Threshold to allow stopping.
2
A locked keyboard screen appears.
3
Tap the activation switch in the upper left corner.
ð The switch turns green and the keyboard is unlocked.
4
Enter the desired value for the threshold and confirm by tapping [OK].
ð The uncertainty threshold is now active. The threshold value is visible in the settings.
7.3 Results
7.3.1 Graphical evaluations of measurement results
Navigation: Home > Results
This function allows you to manage and evaluate measurement results.
Method list
Filter menu
Journal view
Chart view
Graphic view
Results of test measurements
1 Method list
−
Tap the method you want to evaluate.
ð The journal view appears.
2 Journal view
The journal view allows you to start different graphical evaluations from a measurement series. Note that results
that appear with an uncertainty (e.g. 9.19 +/- 9.63 %MC) are predicted results. You can perform the following
functions:
−
To start filter function, tap [
].
ð The filter menu appears.
−
To start chart view, tap [Chart].
ð The chart view of the series of measurement appears (4).
−
To start graphic view, tap a measurement result.
ð The graphic view appears (5).
3 Filter menu
The filter function allows you to evaluate the measurement series according to various criteria. You can select
the following criteria:
•
All (No filter)
•
Today's measurements /method tests
•
This week's measurements / method tests
•
Last 10 measurements / method tests*
•
Last 20 measurements / method tests
•
Advanced filters
* Factory setting
−
Tap [X] to close filter menu.
Advanced filters
The menu Advanced filters offers you a number of additional filter criteria.
The following settings can be defined:
Parameter
Description
Values
Time range
With the option Date range the results are being filtered by date.
Begin and end date must be defined.
With the option Last measurements / method tests the last
measurements and method tests that have been performed can
be shown. The number must be between 1 and 500.
No time range limit |
Date range | Last
measurements / method
tests
Measurement
types
With the option Measurements and method tests measurements
as well as method tests are being filtered.
With the option Only measurements only measurements are
being filtered.
With the option Only method tests only method tests are being
filtered.
Measurements and
method tests | Only
measurements | Only
method tests
4 Chart view
This view allows you to display the results of a series of measurements depending on the filter settings. If you
have defined the control limits in the method, they are diagramed in the results.
The
marked measuring point is displayed with date, time, and the measurement result. You can perform
following functions:
−
To jump forward to the next measurement result, tap [ -> ].
−
To jump backwards to the previous measurement result, tap [ <- ]. Alternatively, tap directly on the desired
measurement point.
−
To recall the full measurement curve of the corresponding measurement, tap [
].
Note
Only results with the same unit as the latest selected result can be displayed in the chart view.
With this function you can display a graphical view of the detailed results of a single measurement. If you have
defined the control limits in the method, passed, warned or failed status is shown, see [Control
limits } Page 43].
You can perform the following functions:
Printing result
−
To print the result, tap [
].
Exporting result
−
To export the result, tap [Export].
See Exporting Results.
Deleting result
−
To delete this measurement result, tap [Delete] (depending on user rights).
With the auto-delete functionality the oldest results will automatically be removed, see Result management.
More data
−
To display more measurement data and method parameters, tap [More data]
5b Displaying results of test measurements
For test measurements see [Test measurement } Page 51].
This view allows you to display the results of a test measurement. The screen shows the switch-off criterion as
well as the time and the result of the test measurement. Each criterion is marked as a measuring point and
displayed with its result.
7.3.2 Method versioning
Every method has a method versioning function that works in the background. Whenever the settings of a
method are changed, the versioning function automatically generates a new version of the method. This has
also an effect on the results: Every measurement overtakes the current version of the method.
Measurement and method test results can be filtered based on the method version as follows:
1
Open the menu Results.
2
Select a method.
ð A journal with the results of the performed measurements and method tests appears. If there are results
in several versions, the button Select versions... appears in the action bar.
3
Tap Select versions....
ð The dialogue Method versions appears.
4
Select the versions you want to filter.
5
Tap Save to confirm the selected versions.
ð The system returns to the result list. The results based on the selected versions appear on the list, the other
results are being grayed out.
Note
The results filtered by the method version can be exported or printed. If the results are being exported as PDF or
being printed, the results of the different versions are being listed based on their version.
7.3.3 Exporting results
Results can be exported to an external storage device or to an FTP server. The data can be exported as single
result exports or as multiple result exports.
Note
Results cannot be imported to another Moisture Analyzer.
7.3.3.1 Exporting results to an external storage device
1
Connect the storage device.
2
Select the section Results.
ð The method list appears.
3
Select a method.
ð A journal with the results of the performed measurements and method tests appears.
4
Select a single result and tap on [Export] to export this specific result.
- or -
Tap directly on [Export] to export all results.
ð The dialogue Export results appears.
5
Define the settings for the export.
6
Tap on [Export target] and select the option [Peripheral].
7
Tap on [Location].
8
Select the connected external storage device.
9
Select the location for the file and tap on [OK] to confirm.
10 Tap on [Export].
ð The chosen results are being exported to the USB storage device.
7.3.3.2 Exporting results via FTP server
1
Connect the device to the Ethernet.
2
Select the section Results.
ð The method list appears.
3
Select a method.
ð A journal with the results of the performed measurements and test measurements appears.
4
Select a single result and tap on [Export] to export this specific result.
- or -
Tap directly on [Export] to export all results.
ð The dialogue Export results appears.
5
Define the settings for the export.
6
Tap on [Export target] and select the option [Network (FTP)].
ð The dialogue Export results reappears with additional Ethernet options.
7
Define the Ethernet options.
8
Tap on [Export].
ð The chosen results are being exported to the FTP server.
7.3.3.3 Exporting results with EasyDirect Moisture
With EasyDirect Moisture, measurement and routine testing results are collected in a database. Results can be
viewed, exported and also routine tests can be scheduled. The PC software must be installed on an external PC
and the device must be connected to that PC with a USB cable or RS232 cable.
Before the function EasyDirect Moisture can be used, the following settings must be defined on the instrument:
Defining EasyDirect Moisture as peripheral device
Navigation: Settings > Instrument settings > Peripherals.
§
Peripherals is opened.
§
No printer is installed as a peripheral. Uninstall printer if necessary.
1
Tap [New...].
ð The list of possible peripherals appears.
Select [EasyDirect Moisture].
ð The parameters for EasyDirect Moisture appear.
3
Under Device Port, select the desired interface for the EasyDirect Moisture connection:
• For a RS232-interface, select COM1.
• For a USB-interface, select COM4.
4
Set the rest of the settings.
5
To save the settings, tap [Save].
ð The option EasyDirect Moisture has been defined as peripheral device on the instrument.
See also
2 Peripherals } Page 84
7.3.3.4 Single result export settings
The following parameters can be defined:
Parameter
Description
Values
File name
Defines the name of the new data file.
any
File type
Defines the type of the data file.
CSV* | PDF
Language
Defines the language for export files.
English | Deutsch |
Italiano | Français |
Español | Português |
Polski | Magyar |
Český | Pусский |
日本語 | 中文 |
한국어/조선말
Export target
Defines the destination of data transfer.
Peripheral |
Network (FTP)
Location
Finds the location on the storage device.
Browse and select
IP address
Defines the IP address of the receiving server.
any
Port
Defines the Port for communication with the server.
Note
Usually the default value is used.
21* | any
Password
Entry for the password needed for server login.
any
User name
Entry for the user name needed for server login.
any
* Factory setting
7.3.3.5 Multiple result export settings
The following parameters can be defined:
Parameter
Description
Values
File name prefix
Defines the name of the new data file.
any
File type
Defines the type of the data file.
CSV* | PDF
Language
Defines the language for export files.
English | Deutsch |
Italiano | Français |
Español | Português |
Polski | Magyar |
Český | Pусский |
日本語 | 中文 |
한국어/조선말
Export selection
Custom selection of multiple results for the export from selected
measurements. The selection can be made via the filter function.
Selection:
|
Defines the content of the multiple results export.
Summary export (one file, no intermediate values): Results as
summary report in one file.
Multiple export (one file per measurement): Results with inter-
mediate measurement values as separate files with the same
layout as for the single result export.
Summary export |
Multiple export
Export target
Defines the destination of data transfer.
Peripheral |
Network (FTP)
* Factory setting
7.4 Test/Adjust
Navigation: Home > Test/Adjust
Introduction
For optimum precision and repeatable results of the Moisture Analyzer, the function of the instrument should be
tested on a regular basis and, if necessary, adjusted. The frequency of testing and adjusting depends on the
use and location of the instrument.
Main menu
Submenu
Further information
Adjustments
Weight adjustment - external
see [Weight adjustment -
external } Page 77]
Temperature adjustment
see [Temperature
adjustment } Page 78]
Tests
Weight test - external
see [Weight test -
external } Page 73]
Temperature test
see [Temperature test } Page 74]
SmartCal x1 test
see Test Printout Examples
History
Weight test history
see [History } Page 80]
Weight adjustment history
Temperature adjustment history
Temperature test history
SmartCal x1 test history
7.4.1 Tests
Navigation: Home > Test/Adjust > Tests
Introduction
Routine tests are performed to ensure accurate results. Therefore the tests should be performed in fixed, regular
intervals depending on your Quality Management system and the results should be documented in a traceable
way.
Based on your process requirements, METTLER TOLEDO can help you to define the routine tests to be
performed. For further information, contact your local METTLER TOLEDO representative.
7.4.1.1 Weight test - external
Navigation: Home > Test/Adjust > Tests > Weight test - external
Target of the test
The external weight test determines if the integrated balance and the weight value display are working correctly
by placing a test weight. The test ensures correct and precise results.
Equipment
•
Test weights. To guarantee full traceability, METTLER TOLEDO recommends to use certified weights.
Test weight and test settings are defined.
1
Open sample chamber.
2
Tap the item Weight test - external.
ð The test screen appears.
3
Clear sample pan holder and tap Start test.
ð The required value flashes and prompts you to load
the weight.
4
Load the required test weight.
ð The balance indicates the test process, ––––––
appears.
5
When the display flashes 000.00 remove the test
weight.
ð The display shows the test results and indicates
whether the test has passed or failed according to
defined tolerances.
To print the results, tap [
].
Timeout
§
In case of a timeout of over 75 seconds, the error message appears Weight out of range. The test has
been aborted. The test was not executed.
1
Confirm with [OK].
2
Repeat the test.
See also
2 Accessories } Page 119
7.4.1.2 Temperature test
Navigation: Home > Test/Adjust > Tests > Temperature test
Target of the test
The temperature test checks if the temperature sensor is working correctly and if it delivers correct results. Using
the temperature kit, the difference of the measured temperature at the kit (target temperature 1) and the sensor
(target temperature 2) is used to calculate the precision of the sensor.
It is possible to start a test and subsequently transform it into an adjustment, if required.
Equipment
•
Temperature kit.
CAUTION
Burns due to hot surfaces!
Parts of the instrument can reach temperatures that cause injuries if touched.
1
Let the instrument cool down before setting up.
2
Hold the temperature kit by its handle only.
Installing the temperature kit
§
Instrument is cooled down to < 50 °C.
§
Instrument is switched on.
§
Sample chamber is open.
1
Remove sample pan handler (with pan) (1).
Remove sample pan holder (2).
3
Clean the contact areas (3) if needed.
4
Insert Temperature kit (4).
4
Testing procedure
§
Test settings are defined, see [Test settings } Page 97].
§
Temperature kit is installed.
1
Tap Temperature test.
ð Work screen Temperature test appears.
2
Close the sample chamber.
3
Tap [Start test].
ð The testing process starts: The instrument first heats up to target temperature 1 (lower) for 15 minutes,
then, for another 15 minutes, heats further up to target temperature 2 (higher). The screen shows the
target temperature values (Target), measured at the sensor, and the actual temperature (Actual),
measured at the temperature kit.
ð The instrument indicates whether the test has passed or failed according to defined tolerances.
4
If an adjustment is needed, decide if you want to perform it now, see Temperature adjustment.
5
• To return to the menu Tests, tap [Tests].
• To print the results, tap [
].
• To display or print the results at any time, tap [History].
• To return to the home screen, press [
].
6
Let the instrument cool down before removing the temperature kit.
See also
2 Accessories } Page 119
7.4.1.3 SmartCal test
Target of the test
A SmartCal test is a quick and integral test of the instrument functions. SmartCalTM is a granular test substance
whose moisture content value, when heated at a specific temperature for a specific time, is precisely defined.
By performing a measurement with SmartCalTM and comparing the results to the control limits, it can be tested if
both the integrated balance as well as the temperature sensor work well together and if the instrument functions
correctly.
However, SmartCal tests do not supersede the recommended periodic weight or temperature tests.
There are four SmartCal tests with four pre-defined temperatures and two additional tests where temperature and
control limits can be defined by the customer.
Equipment
•
SmartCalTM Moisture Analyzer Test Substance.
•
For best results and easy handling, use a RHT sensor.
If no RHT sensor is available, the room temperature and relative humidity must be entered manually.
Certified testo RHT sensors are supported.
Certified RHT sensors
The following RHT models from Testo are supported: testo 435-2, testo 435-4, testo 635-2, testo 735-2.
The testo RHT sensor has to be connected to the USB port on the terminal.
For more information about Testo, see
u http://www.testo.com
Rules for handling SmartCalTM
•
Store SmartCalTM at room temperature.
•
Remove SmartCalTM-stick from the blister pack right before use.
•
Do not use damaged or expired sticks (the expiry date is printed on the blister pack and on the stick (e.g.
Exp08.2021).
•
After a SmartCalTM test, the test substance can be disposed of as normal waste.
Additional information about SmartCalTM can be found under:
www.mt.com/smartcal
Testing procedure
The test is carried out in the same way as a measurement.
Note
Every SmartCal test has a maximum number of 30 test results. If there are more than 30 results for one test,
the oldest results will automatically be deleted.
§
The instrument has been connected to the power supply for at least 60 minutes in order to reach operating
conditions.
§
If available, a RHT sensor is connected.
§
The SmartCal package is acclimatized.
§
Test temperature and control limits are set.
§
The sample chamber is open.
1
Tap [SmartCal x1 test].
ð The method list is shown on display.
2
Select a method from the list.
ð The work screen appears.
3
If the test is performed without RHT sensor, tap [Enter room temperature and humidity], enter current room
temperature and humidity and confirm with [OK].
4
Enter SmartCalTM lot number if needed. Tap [OK]
Place the sample pan handler with the aluminum sample pan in the instrument.
6
Tap [->0/T<-] to tare the balance.
7
Remove a SmartCalTM stick from the blister pack, open it and distribute the entire contents evenly over the
sample pan. If necessary, carefully rotate and tilt the sample pan until it is fully and uniformly covered with
granulate.
8
Start the test immediately by tapping [Start drying].
ð The SmartCal test procedure starts.
ð When the test is finished, the instrument indicates whether the test has passed or failed. The display
shows the normalized result. If the SmartCal result has failed, see [Measures after failed
test } Page 77].
9
• To return to the menu Tests, tap [Tests].
• To print the results, tap [
].
• To display or print the results at any time, tap [History].
• To return to the home screen, press [
].
See also
2 Test settings } Page 97
2 Accessories } Page 119
7.4.1.4 Measures after failed test
In case of a failed test, take the following measures:
1
Repeat the test and ensure that all steps have been carried out correctly.
2
If the failure persists, perform an according adjustment.
• If a weight test has failed, perform a weight adjustment.
• If a temperature test has failed, perform a temperature adjustment.
• If a SmartCal test has failed, perform a weight adjustment first, then perform another SmartCal test and, if
the failure persists, perform a temperature adjustment.
3
If the failure persists, see Malfunctions when operating the instrument.
4
If the failure persists, contact your METTLER TOLEDO service representative.
7.4.2 Adjustments
Navigation: Home > Test/Adjust > Adjustments
Introduction
To guarantee correct results and high performance of the instrument, it is necessary to adjust its functions:
•
After setting up the balance for the first time.
•
After a change of location.
•
If a test indicates that an adjustment is due.
Test the instrument exclusively under operating conditions.
METTLER TOLEDO offers an adjustment service – please contact your METTLER TOLEDO representative.
7.4.2.1 Weight adjustment - external
Navigation: Home > Test/Adjust > Adjustments > Weight adjustment - external
Target of the adjustment
In an external weight adjustment, the integrated balance is adjusted by placing a predefined weight as a
reference for the instrument to re-calibrate itself.
Equipment
•
Test weights. To guarantee full traceability, METTLER TOLEDO recommends to use certified weights.
Test weight and Adjustment settings are defined.
1
Tap Weight adjustment - external.
ð The adjustment screen appears.
2
Clear sample pan holder and tap [Start adjustment].
ð The required value flashes and prompts you to load
the weight, e.g. "100.00 g".
3
Load the required test weight in the center of sample
pan holder.
ð The balance indicates the adjustment process,
"––––––" appears.
4
When the display flashes "000.00" remove the test
weight.
ð The balance is adjusted and ready to use for further
measurements. The display shows the adjustment
results.
Timeout
§
In case of a timeout of over 75 seconds, the error message appears Weight out of range. Adjustment
aborted. The test was not executed.
1
Confirm with [OK].
2
Repeat the adjustment.
See also
2 Accessories } Page 119
7.4.2.2 Temperature adjustment
Navigation: Home > Test/Adjust > Adjustments > Temperature adjustment
Target of the test
A temperature adjustment ensures that the temperature sensor in the sample chamber is correctly calibrated
and delivers the correct results. Using the temperature kit, the temperature value measured at the sensor is
recalculated and corrected.
Equipment
•
Temperature kit.
CAUTION
Burns due to hot surfaces!
Parts of the instrument can reach temperatures that cause injuries if touched.
1
Let the instrument cool down before setting up.
2
Hold the temperature kit by its handle only.
Installing the temperature kit
§
Instrument is cooled down to < 50 °C.
§
Instrument is switched on.
§
Sample chamber is open.
1
Remove sample pan handler (with pan) (1).
Remove sample pan holder (2).
3
Clean the contact areas (3) if needed.
4
Insert Temperature kit (4).
4
Adjustment procedure
§
Draft shield is installed.
§
Temperature kit is inserted.
§
Adjustment settings are defined, see Adjustment Settings.
1
Tap menu item Temperature adjustment.
ð Work screen Temperature adjustment appears.
2
Close sample chamber
3
Tap [Start adjustment].
ð The adjustment process starts: The instrument first heats up to target temperature 1 (lower) for 15
minutes, then, for another 15 minutes, heats further up to target temperature 2 (higher). The screen
shows the target temperature values (Target), measured at the sensor, and the actual temperature
(Actual), measured at the temperature kit.
ð The instrument adjusts the upper temperature and concludes the adjustment.
4
• To print the results, tap [
].
• To display or print out the results of previous adjustments, tap [History].
• To return to the home screen, tap [
].
5
Let the instrument cool down before removing the Temperature kit
Note
Your calibrated temperature kit can be re-calibrated. Please contact your local METTLER TOLEDO representative
for more information.
Navigation: Home > Test/Adjust > History
The instrument always records all test or adjustment operations that have been made and saves them in a
memory which is specially protected against power failure (limited to the last 50 records per history selection).
Note
Aborted adjustments or tests will not be saved.
When the History is selected, a list appears with the history selection as follows:
•
Weight test history
•
Weight adjustment history
•
Temperature test history
•
Temperature adjustment history
•
SmartCal x1 test history
The following information is displayed:
Weight test history
Weight test external
Calibration weight identification
Nominal weight
g
Actual weight
g
Difference
g
Tolerance
g
Temperature
°C
Weight adjustments history
Weight adjustment external
Calibration weight name
Calibration weight identification
Nominal weight
g
Temperature
°C
Temperature adjustment/test history
Temperature adjustment
Temperature test
Adjustment kit identification
Adjustment kit identification
Temperature 1
target - actual
Temperature 1
target - actual
Temperature 2
target - actual
Temperature 2
target - actual
Free temperature
target - actual
Tolerances for each temperature
°C
SmartCal x1 test history
Following parameters are available for each SmartCal method.
SmartCal test
User name
SmartCal identification
Start weight
g
Total test time
minutes
Dry weight
%MCN
Test result
passed / failed
Date and time
Leveling information
The selected individual results can be printed out by tapping [
].
7.5 Settings
Navigation: Home > Settings
In this section you will find information how to define the general settings of your instrument.
Main menu
Submenu
Further information
User preferences
Languages
see [Languages } Page 82]
Screen
see [Screen } Page 82]
Sound
see [Sound } Page 83]
Password
see [Password } Page 83]
Instrument settings
Regional Settings
see [Regional settings } Page 83]
Peripherals
see [Peripherals } Page 84]
Network interfaces
see [Network interfaces } Page 87]
Default settings
see [Default settings } Page 86]
Company information
see [Company infor-
mation } Page 86]
Touch screen adjustment
see [Touch screen
adjustment } Page 87]
Application settings
Identifications management
see [Identifications
management } Page 91]
Print and export management
see [Print and export
management } Page 92]
Method selection
see [Method selection } Page 95]
Quality management
Test / Adjustment settings
see [Test / Adjustment
settings } Page 96]
User management
Groups
see Groups
Users
see [Users } Page 100]
Account policies
see [Account policies } Page 101]
System and data management Result management
see Result management
Export / Import
see [Export / Import } Page 103]
Back up / Restore
see [Back up / Restore } Page 104]
Export event logs
see [Export event logs } Page 105]
Reset
see [Reset } Page 106]
Update
see [Update } Page 106]
Help and tutorials
Instrument tutorial
see [Instrument tutorial } Page 106]
7.5.1 User preferences
Navigation: Home > Settings > User preferences
This menu item defines the basic settings for each user. These settings are stored under the active user profile
and apply when working within this profile. When a user profile is called up (login user account), the relevant
settings are loaded automatically. See User Management.
The following parameters can be defined:
Menu item
Explanation
Further Information
Languages
Defines the preferred display and keyboard
language.
see [Languages } Page 82]
Screen
Defines the brightness and the color scheme of the
display
see [Screen } Page 82]
7.5.1.1 Languages
Navigation: Home > Settings > User preferences > Languages
Select your preferred language for operating the terminal.
Factory setting: Generally, the language of the destination country (if available) or English is set.
The following parameters can be defined:
Parameter
Description
Values
Display language Defines user-specific screen language.
English | Deutsch |
Italiano | Français |
Español | Português |
Polski | Magyar |
Český | Pусский |
日本語 | 中文 | 한국
어/조선말
Keyboard
language
Defines user-specific layout of the Keypad.
English | Deutsch |
Italiano | Français |
Español | Português |
Polski | Magyar |
Český
7.5.1.2 Screen
Navigation: Home > Settings > User preferences > Screen
You can define the following parameters:
Parameter
Description
Values
Brightness: x1 % Defines user-specific brightness of the display.
10…100 %
(80 %)*
Color scheme
Defines user-specific display color scheme.
Blue* | Green | Orange |
Grey
* Factory setting
Navigation: Home > Settings > User preferences > Sound
You can define the following parameters:
Parameter
Description
Values
Sound
Activate / deactivate of audible feedback and adjust the Volume:
x1 %.
Audible feedback:
•
when tapping a button.
•
upon measurement completed.
•
when an error message appears.
OFF* | ON |
0...100 %
(50 %)*
7.5.1.4 Password
Navigation: Home > Settings > User preferences > Password
In this menu item you can change the user password if it is activated by the administrator.
The following parameters can be defined:
Parameter
Description
Values
Password
Change an existing Password.
any
7.5.2 Instrument settings
Navigation: Home > Settings > Instrument settings
This menu item defines the basic settings for the instrument. The instrument settings apply to the entire system
and therefore to all user profiles and applications.
The following parameters can be defined:
Menu item
Explanation
Further Information
Regional
settings
Defines the settings of date & time, keypad and
country options.
see [Regional settings } Page 83]
Peripherals
Defines the configuration of printers and other
devices.
see [Peripherals } Page 84]
Network
interfaces
Defines network connections.
see [Network interfaces } Page 87]
Default settings
Defines the default settings of your instrument.
see [Default settings } Page 86]
Company infor-
mation
Defines company specific information.
see [Company infor-
mation } Page 86]
Touch screen
adjustment
Adjusts the touch screen.
see [Touch screen
adjustment } Page 87]
7.5.2.1 Regional settings
Navigation: Home > Settings > Instrument settings > Regional Settings
This menu item defines the settings of date & time.
The following parameters can be defined:
Parameter
Description
Values
Date
Sets the date of the clock of the instrument.
Day | Month | Year
Time
Sets the clock time of the instrument.
Hours | Minutes
Defines the date format.
D = Day
M = Month
Y = Year
D. MMM YYYY* |
MMM D YYYY |
DD.MM.YYYY |
MM/DD/YYYY |
YYYY/MM/DD |
YYYY-MM-DD
Time format
Defines the time format.
M = Minutes
24:MM* | 12:MM |
24.MM | 12.MM
* Factory setting
Note
Date and time of a connected printer will be ignored. Date and time of instrument is relevant.
7.5.2.2 Peripherals
Navigation: Home > Settings > Instrument settings > Peripherals
In this menu item you can configure one printer and one host connection.
7.5.2.2.1
Workstations (MT-SICS)
The MT-SICS host connection (remote client) is enabled by default and only usable with the METTLER
TOLEDOspecific protocol MT-SICS.
MT-SICS commands and functions can be transmitted via the RS232 interface (COM1 port), the USB device
interface (COM4 port) or via Network. The use of the USB device interface requires the installation of a USB
device driver on the PC, see [Installing USB device driver } Page 19].
Commands and further information please refer to the Reference Manual MT-SICS Interface Commands for
Halogen Moisture Analyzers downloadable from the internet.
u www.mt.com/hxhs
7.5.2.2.2
Connecting printers / EasyDirect Moisture
No printer is installed by default. To install a printer:
§
The desired printer is properly connected to the terminal.
Note
Do not attempt to connect a printer to the drying unit!
1
Tap [New...]
ð Peripheral type list appears.
2
Select device.
Supported devices from METTLER TOLEDO with preset parameters:
• RS-P2x Strip Printer, see RS232 parameters
• RS-P4x Strip Printer, see RS232 parameters
• RS-P5x Strip Printer, see RS232 parameters
• Local PCL printer, see USB parameters
• Network PCL printer, see network parameters
• EasyDirect Moisture, see RS232 parameters
3
Change the parameters if needed.
4
To store the settings, tap [Save].
5
To activate the interface, tap [ON].
ð Status changes from Disabled to Enabled.
Note
•
A printer or host connection can also be deleted.
•
For strip printer settings see [Recommended printer settings (Strip printer) } Page 94].
For defining the printouts see Print and Export Management.
•
Only one printer can be installed. To install another printer, you have to delete the old printer first.
RS232 parameters
The following parameters can be defined:
Parameter
Description
Values
Device Name
Defines a name for the device.
any
Device interface
Defines the device interface.
RS232* | USB |
Ethernet
Parameter
Description
Values
Device Port
Defines the device port.
COM1: RS232 / COM4: USB Device (Driver required)
Note
It is possible to use an USB-to-Serial adapter. If such an adapter
is connected, these devices are shown as e.g.SOU1, SOU2 (SOU
stands for serial over USB)
COM1* | COM 4
Baudrate
Defines the speed of the data transmission (data transfer rate /
Baud rate).
1200 | 2400 | 4800 |
9600* | 19200 |
38400 | 57600 |
115200
DataBits
Defines the number of data bits.
7 | 8*
Parity
Defines the variant of parity bit.
None* | Odd | Even
StopBits
Defines the stop bits of the transmitted data.
1 StopBit* |
2 StopBits
Flow Control
Defines the type of flow control (Handshake).
None | Xon/Xoff*
* Factory setting
USB devices
Note
A bar code reader can be connected to the USB port. With a connected bar code reader, bar codes can be
added in a keyboard dialog.
Note
•
Supported A4 Printer: OKI B411 USB
•
Supported printer command language: PCL 3 or higher
The following parameters can be defined:
Parameter
Description
Values
Device Name
Defines a name for the device.
any
Network parameters
The following parameters can be defined:
Parameter
Description
Values
Device Name
Defines a name for the device.
any
IP address
Defines the IP address of the printer.
Default address*: 192.168.000.004
Default address* | any
Port
Defines the printer's port.
Note
Usually the default value is used.
9100 * | any
Defines TCP port on device for MT-SICS communication.
1024...65635 (8001*)
* Factory setting
7.5.2.3 Default settings
Navigation: Home > Settings > Instrument settings > Default settings
7.5.2.3.1
Default languages
Navigation: Home > Settings > Instrument settings > Default settings > Default languages
The default language for operating the terminal can be changed. The default language has been defined with
the function Setup wizard when the device was switched on for the first time.
The following parameters can be defined:
Parameter
Description
Values
Default display
language
Defines default screen language.
English | Deutsch |
Italiano | Français |
Español | Português |
Polski | Magyar |
Český | Pусский |
日本語 | 中文 | 한국
어/조선말
Default keyboard
language
Defines default layout of the keyboard.
English | Deutsch |
Italiano | Français |
Español | Português |
Polski | Magyar |
Český
7.5.2.3.2
Default screen settings
Navigation: Home > Settings > Instrument settings > Default settings > Default screen settings
The following parameters can be defined:
Parameter
Description
Values
Brightness: x1 % Defines default brightness of the display.
10…100 %
(80 %)*
Color scheme
Defines default display color scheme.
Blue* | Green | Orange |
Grey
* Factory setting
7.5.2.4 Company information
Navigation: Home > Settings > Instrument settings > Company information
In this menu item you can define optional specific company informations for using on printouts and reports.
The following parameters can be defined:
Parameter
Description
Values
Company name
Entry of the company name.
any
Department
Entry of the company department.
any
Instrument ID
Entry of the company specific instrument identification
any
Instrument
location
Entry of the location of the instrument.
any
Instrument
responsible
Entry of the name of the responsible person for the instrument.
any
7.5.2.5 Touch screen adjustment
Navigation: Home > Settings > Instrument settings > Touch screen adjustment
If the alignment of the touch-sensitive areas of the screen correspond not exactly with the position of the buttons
e.g. after a software update was performed, then you can solve this error by executing this function.
NOTICE
Damage to the touch screen due to pointed or sharp objects
−
Operate the touch screen with your fingers.
§
Menu Instrument settings is opened.
1
Choose Touch screen adjustment and tap Start.
ð Adjustment screen with a target circle appears.
2
Tap the center of the target until it moves to the next position on the screen. Tap the center of the target as
accurately as possible and make sure not to touch any other locations on the screen and the fix key areas.
3
Repeat this procedure until a message appears (New
calibration settings have been measured).
4
Tap the screen to confirm or wait for 30 seconds for
canceling (timeout).
ð A message box appears.
5
To store the setting, tap OK.
7.5.2.6 Network interfaces
Navigation: Home > Settings > Instrument settings > Network interfaces
In this section you will find information how to define network connections and other terminal interfaces.
Menu item
Explanation
Further Information
Host name
Defines the host name of the instrument
see [Ethernet } Page 87]
Ethernet
Activates and sets Ethernet connection.
see [Ethernet } Page 87]
WLAN
Activates and sets wireless local area network
(WLAN) connection.
see [WLAN } Page 88]
7.5.2.6.1
Host name
Navigation: Home > Settings > Instrument settings > Network interfaces > Host name
The Host name is the name of your instrument and is a unique identifier on a network (Ethernet, WLAN). If
using multiple devices, you should ensure that the host name of each instrument is unique. If necessary,
contact the network administrator of your IT department or your IT support.
The following parameters can be defined:
Parameter
Description
Values
Host name
Defines the host name. The host name must be unique on a
network.
Default host name*: Model name and serial number of the
instrument.
any
* Factory setting
7.5.2.6.2
Ethernet
Navigation: Home > Settings > Instrument settings > Network interfaces > Ethernet
Ethernet interface for connecting the instrument to a TCP/IP network. The most simple network can be imple-
mented by connecting the instrument directly to the PC with a cross-over cable (RJ45 Socket). DHCP and fix IP
addresses are supported.
Note
•
We recommend to use the default setting DHCP (automatic network settings).
•
Ensure that the host name of the instrument is unique, see [Host name } Page 87].
•
For installing and configuring an Ethernet interface you need to have basic knowlege of TCP/IP-based
networks and of network technology in general. If necessary, contact the network administrator of your IT
department or your IT support.
The following parameters can be defined:
Parameter
Description
Values
Configuration
Defines whether the settings are made automatically (dynamic,
DHCP) or manually (static).
DHCP assignment |
Manual assignment
DHCP
assignment
Dynamic Host Configuration Protocol to assign all necessary
network parameters on this interface automatically (Dynamic
assignment).
none
Manual
assignment
Manually assign all the necessary network parameters on this
interface. (Static assignment)
IP address | Subnet
IP address
Defines the IP address of this instrument. Only settable when
DHCP is disabled.
Default address*: 192.168.002.100
The IP address unique numeric identifier for a device which
participates on a network.
If DHCP is enabled, the automatically assigned IP address is
displayed.
Default address* | any
Subnet
Defines the subnet mask within the ethernet interface is
addressable. Only settable when DHCP is disabled.
Default subnet mask*: 255.255.255.000
Default value* | any
Default Gateway
Defines the Default Gateway of this instrument. Only settable
when DHCP is disabled.
Default address*: 192.168.002.001
Default address* | any
* Factory setting
Note
The addresses are written as four dot-separated decimal integers in the range from 0 to 255. Leading zeros
need not be written.
7.5.2.6.3
WLAN
Navigation: Home > Settings > Instrument settings > Network interfaces > WLAN
Wireless LAN interface for connecting the instrument to a wireless TCP/IP network. DHCP and fix IP addresses
are supported.
Note
WLAN is not available in every country.
The following WLAN authentications are supported:
•
Open
•
WEP
•
WPA-PSK
•
WPA2-PSK (tkip only)
The following parameters can be defined:
Activates or deactivates WLAN radio transmitter.
OFF* | ON
* Factory setting
Note
After activating WLAN, it takes some time until WLAN is active.
WLAN network settings
Navigation: Home > Settings > Instrument settings > Network interfaces > WLAN > Settings > WLAN
configurations
•
We recommend to use the default setting DHCP (automatic network settings).
•
Ensure that the host name of the instrument is unique, see [Host name } Page 87].
•
For installing and configuring an WLAN interface you need to have basic knowledge of TCP/IP-based
networks and of network technology in general. If necessary, contact the network administrator of your IT
department or your IT support.
The following parameters can be defined:
Parameter
Description
Values
Configuration
Defines whether the settings are made automatically (dynamic,
DHCP) or manually (static).
DHCP assignment |
Manual assignment
DHCP
assignment
Dynamic Host Configuration Protocol to assign all necessary
network parameters on this interface automatically (Dynamic
assignment).
none
Manual
assignment
Manually assign all the necessary network parameters on this
interface. (Static assignment)
IP address | Subnet
IP address
Defines the IP address of this instrument. Only settable when
DHCP is disabled.
Default address*: 192.168.002.100
The IP address unique numeric identifier for a device which
participates on a network.
If DHCP is enabled, the automatically assigned IP address is
displayed.
Default address* | any
Subnet
Defines the subnet mask within the ethernet interface is
addressable. Only settable when DHCP is disabled.
Default subnet mask*: 255.255.255.000
Default value* | any
Default Gateway
Defines the Default Gateway of this instrument. Only settable
when DHCP is disabled.
Default address*: 192.168.002.001
Default address* | any
* Factory setting
Note
The addresses are written as four dot-separated decimal integers in the range from 0 to 255. Leading zeros
need not be written.
Networks
Receivable networks are displayed.
Entries for WLAN:
Parameter
Description
Values
WLAN name
(SSID)
Displays the name of the selected WLAN (SSID, service set
identifier)
not editable
Password
Entry for the required password.
any
Activate or deactivate automatic connecting to the preferred
network upon startup of the device.
Note
Only for one network possible.
(deactivated)* |
(activated)
Establish a connection:
1
Touch the desired network.
Note
If the device is connected to a network, the WiFi icon indicates the signal strength. The more bars displayed,
the stronger the connection.
2
Enter your password.
3
Enable automatic connection if needed.
4
Touch [connect] to establish the connection.
7.5.3 Applications settings
Navigation: Home > Settings > Application settings
The following parameters can be defined:
Menu item
Explanation
Further Information
Identifications
management
Defines identifications and their use in order to
comment measurements.
see [Identifications
management } Page 91].
Print and export
management
Defines the default settings for the printouts.
see [Applications
settings } Page 91].
Method selection Selects the method
see Method selection.
7.5.3.1 Identifications management
Navigation: Home > Settings > Application settings > Identifications management
Identifications (ID) contain descriptive text for individual measurements, which enables samples to be easily
allocated to specific tasks or customers. This feature allows you to define identifications in order to comment
measurements, such as company id, id badge, id sample. In the ID panel on the work screen, the values
(comments) can be entered or changed before measurement, see Work Screen. IDs have an impact on all
measurements, regardless of methods. The IDs are included on printouts or can be exported to a PC.
Main menu
Submenu
Further information
Identifications usage
Identifications input
see below
Clear input after measurement
Identifications (x1)
no submenu
The following parameters can be defined:
Parameter
Description
Values
Identifications
input
Defines the method for entering comments.
ON = activates ID panel on work screen.
Free input = the IDs appear after tapping on the ID panel (work
screen) for entries or changes of values (comments).
Guided input = each ID is displayed individually in order of
definition (ID1 to ID10) before measurement, to make entries or
changes of values (comments).
OFF* | ON |
Free input |
Guided input
Clear input after
measurement
Defines the ID whose value (comment) is deleted after
measurement.
OFF* | ON |
ID selection:
|
|
Identifications
(x1)
Defines general identifications in order to comment
measurements.
+ = New identifications (max. 10 IDs, 1…20 characters)
- = Remove identifications (selectable)
- | +* | any
* Factory setting
7.5.3.2 Print and export management
Navigation: Home > Settings > Application settings > Print and export management
The following parameters can be defined:
Menu item
Explanation
Further Information
Print
management
(Strip)
Defines the settings for printouts on strip printers.
see [Print
management } Page 92].
Print
management
(A4)
Defines the settings for ISO A4 printouts on PCL
printers.
see [Print
management } Page 92].
7.5.3.2.1
Print management
Print management (Strip)
Navigation: Home > Settings > Application settings > Print and export management > Print management
(Strip)
If a strip printer is connected and activated, results can be automatically printed out. The interim values are
printed out in the pre-selected intervals and the final results are logged at the completion of the measurement.
The printout contains information to the instrument, time, date as well as the method and results, depending of
the protocol. The level of detail depends on the selected type of printout and the activated protocol items
(settings or method), see Information on Printouts.
Note
•
The definitions in this menu item are the default settings for the printouts. These parameters are displayed at
every initialization of a Printout and can be modified again.
•
The setting for printout language in this menu item has no impact on test or adjustment printouts, see Print
and Output Settings.
•
Not all languages are compatible with some printers, and also depending on the settings of the printer, see
[Recommended printer settings (Strip printer) } Page 94].
•
For installing and defining a printer see [Peripherals } Page 84].
The following parameters can be defined:
Parameter
Description
Values
Printout type
Selects a detailed printout or in a short form.
Standard* | Short |
Custom 1 | Custom 2
Printout language Defines the language for general printouts.
Note
No impact on test or adjustment printouts.
English | Deutsch |
Italiano | Français |
Español | Português |
Polski | Magyar |
Český
Print results
automatically
after
measurement
Print results automatically after measurement if activated.
(deactivated)* |
(activated)
Include inter-
mediate results to
printout
Defines the time intervals at which intermediate results are trans-
ferred to the printer.
OFF* | 10 s | 30 s |
1 min | 2 min | 10 min
* Factory setting
With the options Custom 1 and Custom 2 the information that appear on the printout can be selected
individually. The information that appear on the printout can be defined in the sub-option Printout content.
The following information for Printout content can be enabled/disabled or defined:
Parameter
Description
Values
Head
METTLER TOLEDO | Halogen Moisture Analyzer | Blank line
S/N instrument
and software
Type | SNR (Drying unit) | SNR (Terminal) | SW(Drying Unit) | SW(Terminal) | Blank line
User name
User name | Blank line
Company infor-
mation
Company | Department | Instrument ID | Instrument location | Instrument responsible | Blank
line
Adjustment info
Weight Adjustment | Temperature Adjustment | Blank line
Method
parameters
Method Name | Drying Prog | Drying Temp | Switch Off | SOC delay | Display mode | Start
Weight | Start Weight Tol | Weigh-in aid | Control limits | Resolution | Start Mode | Free
factor | Preheating | Blank line
ID's
ID's | Blank line
Measurement
Preheating state skipped | Preheating Time | Inclinometer status | Start weight | Intermediate
results | Total time | Dry Weight | Moisture Content | Factor | Offset | Measured Result |
Measurement status | Note | Blank line
Signature 1
(deactivated)* |
(activated)
Signature 2
(deactivated)* |
(activated)
Note
The information Date / Time and the end result are always activated and appear at the end of the printout.
The settings for the option Intermediate results are still set but do only appear on short and standard printout.
Print management A4
Navigation: Home > Settings > Application settings > Print and export management > Print management
(A4)
Note
If a PCL printer is connected and activated, results can be manually or automatically printed out on ISO A4
paper size.
Note
•
The definitions in this menu item are the default settings for the printouts. These parameters are displayed at
every initialization of a printout and can be modified again.
•
The setting for printout language in this menu item has no impact on test or adjustment printouts, see Print
and Output Settings.
•
For installing and defining an A4 printer, see [Peripherals } Page 84].
The following parameters can be defined:
Printout language Defines the language for general printouts.
Note
No impact on test or adjustment printouts.
English | Deutsch |
Italiano | Français |
Español | Português |
Polski | Magyar |
Český | Pусский |
日本語 | 中文 |
한국어/조선말
Print results
automatically
after
measurement
Print results automatically after measurement if activated.
(deactivated)* |
(activated)
7.5.3.2.2
Recommended printer settings (Strip printer)
English, German, French, Spanish, Italian, Polish, Czech, Hungarian
Printer
Model
Char Set
Auto baudrate
Balance feature
RS-P2x
IBM/DOS
Off
Disable
RS-P4x
IBM/DOS 1)
—
—
Instrument / Printer
Model
Baudrate
Bit / Parity
Stop Bits
Handshake
RS-P2x
9600
8/NO
1
Xon/Xoff
RS-P4x
1200
8/NO
1
Xon/Xoff
Brazil Portuguese
Printer
Model
Char Set
Auto baudrate
Balance feature
RS-P2x
IBM/DOS
Off
Disable
RS-P4x
— 2)
—
—
Instrument / Printer
Model
Baudrate
Bit / Parity
Stop Bits
Handshake
RS-P2x
9600
8/NO
1
Xon/Xoff
RS-P4x
— 2)
— 2)
— 2)
— 2)
1) Printer settings not available.
2) Required font for this language not available.
7.5.3.2.3
Export management
Navigation: Home > Settings > Application settings > Print and export management > Export management
If a PC, WLAN or a storage device is connected and activated, final results can be automatically exported. The
export contains information to the instrument, time, date, method, result as well as the graphical curve.
The following parameters can be defined:
Parameter
Description
Values
File name prefix
Defines the name of the new data file.
any
File type
Defines the type of the data file.
CSV* | PDF
Printout language Defines the language for general printouts.
Note
No impact on test or adjustment printouts.
English | Deutsch |
Italiano | Français |
Español | Português |
Polski | Magyar |
Český | Pусский |
日本語 | 中文 |
한국어/조선말
Export
management
Export results automatically after measurement if activated.
Selection:
* |
Export target
Defines the destination of data transfer.
Peripheral |
Network (FTP)
* Factory setting
Export target: Peripheral
Parameter
Description
Values
Location
Finds the location on the storage device.
Browse and select
* Factory setting
Export target: Network (FTP)
For data transfer over a network (Ethernet and WLAN) the File Transfer Protocol FTP is used. The receiving
server is defined by entry of the appropriate IP address. The FTP server needs to be in the same subnet as the
device (according to Ethernet settings, see [Ethernet } Page 87]). If necessary, contact the network adminis-
trator of your IT department or your IT support.
Parameter
Description
Values
IP address
Defines the IP address of the receiving server.
any
Port
Defines the Port for communication with the server.
Note
Usually the default value is used.
21* | any
Password
Entry for the password needed for server login.
any
User name
Entry for the user name needed for server login.
any
* Factory setting
7.5.3.3 Method selection
Navigation: Home > Settings > Application settings > Method selection
In this menu, certain methods for different users can be defined. In the dialogue screen Select methods for
user a list of users appears. A user can be selected and certain methods can be activated/deactivated. With the
button [Copy from user] that can be found in the action bar, the method selection of one user can be copied
and used for another user.
NOTICE
Access to the section Method selection can be defined in the user access rights.
7.5.4 Quality management
Navigation: Home > Settings > Quality management
In this menu you can define all settings related to the adjustment and Test of the instrument. Periodic
adjustments or tests ensure accurate measurement results.
The following parameters can be defined:
Menu item
Explanation
Further Information
Test /
Adjustment
settings
Defines the equipment for adjusting and the
workflow for testing or adjusting.
see [Test / Adjustment
settings } Page 96]
Activates the QuickPredict function for a method and
defines the specifics for its use.
see [QuickPredict } Page 44]
7.5.4.1 Test / Adjustment settings
Navigation: Home > Settings > Quality management > Test / Adjustment settings
This section describes all the menu options available for defining the parameters relating to the adjustment and
testing of your instrument.
For performing test and adjustments see [Test/Adjust } Page 73].
Main menu
Submenu
Further information
Equipment
Test weight1
see [Equipment } Page 96]
Test weight2
Temperature kit
External humidity sensor
Test settings
Weight test - external
see [Test settings } Page 97]
Temperature test
SmartCal x1 test
Adjustment settings
Weight adjustment - external
see [Adjustment
settings } Page 98]
Temperature adjustment
Print and output settings
Printout language
see [Print and output
settings } Page 98]
Print results automatically after
test/adjustment
7.5.4.1.1
Equipment
Navigation: Home > Settings > Quality management > Test / Adjustment settings > Equipment
This menu item defines the equipment which is required for the following tasks:
•
Weight test
•
Weight adjustment
•
Temperature test
•
Temperature adjustment
•
SmartCal test (optional)
Test Weight 1 and Test Weight 2
The following parameters can be defined:
Parameter
Description
Values
Test weight name Defines a name that is in an easily recognizable form as an alter-
native to the ID. A name should be unique and unambiguous.
any
Nominal weight
Defines the nominal weight in grams. For adjustments, the weight
must be at least 30 g.
0.100…150.000 g
(100 g)*
Identification
Defines the identification (ID) of the weight. The ID of the weight
is provided on the Weight Certificate. The ID may contain your
company specific identification number. This ID is stored in the
adjustment or test history and shows up on the printouts.
any
* Factory setting
Temperature Kit
For performing temperature adjustment or temperature test, you need the optional temperature kit, see Acces-
sories and Spare Parts.
The following parameters can be defined:
Defines a name that is in an easily recognizable form as an alter-
native to the ID. A name should be unique and unambiguous.
any
Identification
Defines the identification (ID) of the Temperature kit. The ID is
provided on the Certificate. The ID may contain your company
specific identification number. This ID is stored in the adjustment
or test history and shows up on the printouts.
any
7.5.4.1.2
Test settings
Navigation: Home > Settings > Quality management > Test / Adjustment settings > Test settings
Weight test external
The following parameters can be defined:
Parameter
Description
Values
Test weight
Selects the predefined Test weight.
Test weight 1 / Test weight 2 = defined in the menu item
Equipment
Test weight 1 |
Test weight 2
Tolerance
Defines the measuring tolerance of the weight.
0.001…0.100 g
(0.001 g)*
* Factory setting
Temperature test
The following parameters can be defined:
Parameter
Description
Values
Temperature kit
Shows the name of the Temperature kit which was defined in the
menu item Equipment.
none
Temperature 1
Enables or disables the lower test temperature.
OFF | ON
(100 °C)
Tolerance
Defines the tolerance of the test temperature.
1…5 °C
(3 °C)*
Temperature 2
Enables or disables the upper test temperature.
OFF | ON
(160 °C)
Tolerance
Defines the tolerance of the test temperature.
1…5 °C
(3 °C)*
* Factory setting
SmartCal test
Note
There are 4 fixed SmartCal methods and 2 customizable SmartCal methods. All 6 methods are activated as
default values but the methods can individually be enabled or disabled.
The following parameters can be defined:
Parameter
Description
Values
Lot number input
Defines the type of input for the Lot number. The Lot number can
be found on the product label.
Free input: Input is not required
Guided input: Input is required.
ON* | OFF |
Free input |
Guided input *
Clear lot number
after
measurement
Enabled or disabled the function for clearing the Lot number after
test measurement.
(enabled) |
(disabled)*
Sets the control limits to the correct values for using of cSmartCal.
(enabled) |
(disabled)*
Temperature
Defines the test temperature.
Note
If temperature in customer SmartCal methods is set, the control
limits will be reset to default (off).
70 °C | 100 °C |
130 °C | 160 °C |
Custom temperature...
40 °C…230 °C
(105 °C*)
Control limits
Defines the control limits. The values of the control limits are in
dependence of the temperature setting. The recommended values
are automatically set. The values can also be changed.
The recommended values are (lower - upper limits):
70 °C: 3.2-4.4 %MCN (cSmartCal: 3.3 - 4.3 %MCN)
100 °C: 5.2-6.4 %MCN (cSmartCal: 5.3 - 6.3 %MCN)
130 °C: 7.4-8.8 %MCN (cSmartCal: 7.5 - 8.7 %MCN)
160 °C: 9.9-11.7 %MCN (cSmartCal: 10.0 - 11.6 %MCN)
Note
If temperature in customer SmartCal methods is set, the control
limits will be reset to default (off).
values see explanation |
OFF
* Factory setting
7.5.4.1.3
Adjustment settings
Navigation: Home > Settings > Quality management > Test / Adjustment settings > Adjustment settings
Weight adjustment external
The following parameters can be defined:
Parameter
Description
Values
Test weight
Selects the predefined Test weight.
Test weight 1 / Test weight 2 = defined in the menu item
Equipment
Test weight 1 |
Test weight 2
Temperature adjustment
The following parameters can be defined:
Parameter
Description
Values
Temperature kit
Shows the name of the Temperature kit which was defined in the
menu item Equipment.
none
7.5.4.1.4
Print and output settings
Navigation: Home > Settings > Quality management > Test / Adjustment settings > Print and output
settings
This menu item allows you to set specific printer functions for test or adjustment printouts.
•
The setting for printout language in this menu item has impact on test or adjustment printouts only. For
setting the general printout language, see [Print and export management } Page 92].
•
Not all languages are compatible with any printer, and also depending on the settings of the printer, see
[Recommended printer settings (Strip printer) } Page 94].
•
For installing and defining a printer see [Peripherals } Page 84].
The following parameters can be defined: