OfiTE 151-00 Instruction Manual

Text of OfiTE 151-00 Instruction Manual

Garrett Gas Train
Part No. 151-00 Instruction Manual Updated 2/18/2020 Ver. 5 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A. Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com ©Copyright OFITE 2013
OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
1 Introduction.....................................................................................2 Specifications .................................................................................2 Components .........................................
OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
2 The Garrett Gas Train is the most widely used portable instrument for deter- mining the presence and quantity of soluble sulfides and carbonate concen- trations in drilling fluid. Soluble sulfides include H2S as well as sulfide (S-2) and bisulfide (
OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
3 Components #142-58 O-ring for HTHP Coupling #143-00 Regulator #143-02-10 CO2 Puncture Head Assembly #143-02-13 O-ring for Puncture Pin Holder Assembly, CO2 Cartridge #143-02-14 O-ring for Puncture Pin Holder Assembly #143-03 Barrel for CO2 Cartridge
OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
4 #151-00-SP Spare Parts Kit QTY #142-58 O-ring for HTHP Coupling, 4 #143-00-1 Diaphragm for Airco Regulator 1 #143-02-13 O-ring for Puncture Pin Holder Assembly, CO2 Car- tridge 4 #143-02-14 O-ring for Puncture Pin Holder Assembly 4 #143-05 EZ-Punctu
OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
5 Safety The Garrett Gas Train is designed to test drilling fluids for soluble sulfides, including hydrogen sulfide (H2S). Hydrogen sulfide is highly toxic and flam- mable. It smells like rotten eggs, but quickly deadens the sense of smell. You may no
OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
6 Setup 1. Be sure the gas train is clean and dry. Moisture in the train can cause the flowmeter to float erratically and may affect the accuracy of the Dräger tube reading. 2. Place the gas train on a level surface. Remove the top. 3. Add 20 mL of d
OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
7 7. Unscrew (counter-clockwise) the regulator T-screw completely. Remove the cartridge holder and place a new cartridge inside. Tighten the car- tridge holder until the cartridge inside is punctured. For sulfide analysis, use CO2 gas. For carbonate a

OfiTE Laboratory Equipment Owners Instructions: