Low Temperature Incubator

Model 815 Precision 170 Marcel Drive Winchester, VA USA Phone: 540-869-8623 Toll Free: 800-621-8820 FAX: 540-869-8626 Manual P/N 36100114 (34001729) Rev D Dated 15 December, 2004 ...

THE MATERIAL IN THIS MANUAL IS FOR INFORMATION PURPOSES ONLY. THE

CONTENTS AND THE PRODUCT IT DESCRIBES ARE SUBJECT TO CHANGE WITHOUT NOTICE. PRECISION MAKES NO REPRESENTATIONS OR WARRANTIES WITH RESPECT TO THIS MANUAL. IN NO EVENT SHALL PRECISION BE LIABLE FOR ANY DAMAGES, DIRECT OR INCIDENTAL, ARISING OUT OF OR RELATED TO THE USE OF THIS MANUAL. ...

Update parts list for new kit #'s

C 9/01 Removed dual fans, updated parts list 18-20 D 12/04 Update for base model FFU2064DW1 18-20 ...

Introduction .......................................................................................................................... 1

Unpacking and Damage....................................................................................................... 1 General Information ................................................................... ...

Your satisfaction and safety are important to

PRECISION and a complete understanding of this unit is necessary to attain these objectives. As the user of this apparatus, you have the responsibility to understand the proper function and operational characteristics of your incubator. This instruction manual should be thoroughly read and all oper ...

The incubator is commonly referred to as the

standard Model 815. These instructions encompass the models listed below with their specific electrical characteristics. PERFORMANCE DATA Heat-Up Time: From -10°C to 50°C; 120 minutes (Compressor "ON"); Cool-Down Time: From 50° to -10°C; 100 minutes. NOTE THE ABOVE PERFORMANCE FIGURES A ...

EXPLANATION OF CONTROLS

-Front Panel- 1) HEATER ON LAMP - The "Heater On" lamp is illuminated when power is applied to the heater. 2) LED DISPLAY - Four digits are used to display the actual, set, calibrate, high limit and low limit values. 3) INDICATOR LAMPS - There are five (5) green indicator lamps that are used to inform the user which parameter the di ...

-Explanation of Auxiliary Panel-

On the left side of the incubator control panel there is an auxiliary panel which has the items shown in the figure below. They are described as follows: 1) FUSE - This fuse is in line with the main power cord that comes into the incubator. The rating of this fuse is printed above the fuse holder. The physical size of this ...

Located on the chart recorder printed circuit board

is an offset potentiometer R3. If for some reason the recorder output voltage does not match the temperature as shown in the chart, this potentiometer can be used to tune it in. When connecting a chart recorder to the recorder output connector, it is recommended to use a shielded cable with the shield gro ...

To assure proper ventilation, allow a minimum of

4" of clearance between the rear, top, and sides of the incubator and adjacent walls. If two or more incubators are positioned side by side, allow a minimum of 8" between cabinets. Adjust the front leveling feet of the incubator so that the front is higher than the rear. This will assis ...