RS PRO 1389 Instruction Manual

Text of RS PRO 1389 Instruction Manual

BRIX METER / EN

05/24/18 Version No. 00 CONTENTS Title Page 1. INFORMATION ....................................................................... 1 2. FEATURES ............................................................................. 2 3. SPECIFICATIONS .................................................................. 3 4. NAME OF PARTS AND POSITIONS .............................
BRIX METER / EN 02/24/21 Version No. 00 1 Internal 1. INFORMATION  The meter is a portable optical meter that uses the measurement of refractive index to determine the % Brix of sugar in aqueous solutions.  The measurement technique and temperature compensation use methodology recommended by ICUMSA (International Commission of Uniform Method of Sugar Analysis).  The meter is a waterproof brix meter with a rating of IP65 and has a memory of 99 data
BRIX METER / EN 2 02/24/21 Version No. 00 Internal 2. FEATURES  Wide measuring range (Brix 0 ~ 53%)  Refractive index measurement (RI)  Temperature measurement (°C/°F)  Ambient temperature compensation (ATC)  Brix or RI high/low limit value alarm function  Zero calibration with distilled or deionized water  Friendly operation and easy to use  Wide and dual display LCD and backlight function  Reliable water resistant (IP65) 
BRIX METER / EN 02/24/21 Version No. 00 3 Internal 3. SPECIFICATIONS  Sensor: Photodiode arrays  Measuring Function: Brix (%) Temperature (°C/°F) Refractive index (RI)  Brix Range: 0.0 ~ 53.0% (10 ~ 40°C/50 ~ 104°F Automatic temperature compensation)  RI Range: 1.3329 ~ 1.4265  Temperature Range: 4 ~ 60°C (39.2~140°F)  Accuracy (Water at 23°C /73.4°F): Brix: ±0.2% Temperature: ±1°C/°F RI: ±0.0003  Resolution: Brix: 0.1% Te
BRIX METER / EN 4 02/24/21 Version No. 00 Internal 4. NAME OF PARTS AND POSITIONS 4-1 Description of Parts & Control keys UNIT MEM SET ZERO READ FUNC TEST A.I.Alert 1. LCD Display: Dual display LCD is used for reporting the measurement Brix (%), Refractive index (RI), Temperature values and has several other marks to provide function guidance.
BRIX METER / EN 02/24/21 Version No. 00 5 Internal 2. ◄ MEM READ key: ◄ key: In the setting mode, press ◄ key to move flicking cursor to desired position. MEM key: After measurement, press “MEM” key to store the measured Brix (%), Refractive index (RI), Temperature (°C or °F) values and the memorized time (RTC) to memory. READ key: Press and hold down the “SET” key then press “READ” key to enter the memory data reading mode. Press “RE
BRIX METER / EN 6 02/24/21 Version No. 00 Internal ZERO key: Press and hold down the “SET” key then press “ZERO” key to perform the zero-calibration function. 6. SET key: key: In the setting mode, press key to confirm the setting item or stored the setting values. SET key: Press and hold down the “SET” key then press “READ” key to enter the memory data reading mode. Press and hold down the “SET” key then press “A.I.Alert” key to
BRIX METER / EN 02/24/21 Version No. 00 7 Internal 4-2 Description of Display : Measurement in progress indication. : Brix % and RI high/low limit value alarm function indication. RI: Refractive index reading indication. %: Brix % reading indication. CAL: Zero calibration indication. ATC: Automatic temperature compensation indication. : Manual memory address number indication. : Read mode indication. : Recall manual memory address number indication. ▲:
BRIX METER / EN 8 02/24/21 Version No. 00 Internal 5. OPERATING INSTRUCTIONS 5-1 Precautions  Handle the meter carefully. Do not drop or shake the meter.  The meter is used to measure sugar solutions. Do not expose the meter or prism to solvents that will damage it. This includes most organic solvents and extremely hot or cold solutions.  Never submerge the meter and do not spray water on the meter except the sample plate located over the prism. �
BRIX METER / EN 02/24/21 Version No. 00 9 Internal 4. Using plastic pipettes, drip a few drops of distilled or deionized water on the prism, make sure that water covers the prism surface completely. 5. Close the sample plate cover. 6. Press and hold down the “SET” key then press “ZERO” key to enter the zero-calibration mode, the “ CAL” symbol is displayed. When the “0.0 %” reading is displayed means the meter is calibrated. 7. Gently absor
BRIX METER / EN 10 02/24/21 Version No. 00 Internal 7. Press “MEM” key to store the measured values and the RTC to memory. Only one measurement can store at one time. 8. Remove sample from the sample plate by absorbing with a soft tissue. 9. Using plastic pipettes, rinse prism and sample plate with distilled or deionized water then wiper dry. The meter is ready for the next sample. 5-4 Changing Temperature Unit °C  °F 1. Press key to turn on the
BRIX METER / EN 02/24/21 Version No. 00 11 Internal 5. Press key to confirm the selection. 6. Press ◄ key to move the blinking cursor to the desired position then press ▲ or ▼ key to set the element to real-time. Repeat this step to all element. 7. Press key to store the setting and exit this mode. 5-6 Data Memory and Read Mode 5-6-1 To Memorize the Reading 1. After each measurement, Press “MEM” key to store the measured reading and the RTC to
BRIX METER / EN 12 02/24/21 Version No. 00 Internal 3. Press “FUNC” key to cycle display the Brix %, RI reading and the memorized RTC. 4. Press “READ” key again to exit this mode. 5-6-3 To Clear the Memorized Data 1. Press key to turn off the meter. 2. Press and hold down the “MEM” key then press key to turn on the meter, LCD will show “ ” symbol one time, it indicated that the memorized data is erased, the “ ” symbol is displayed. 5-
BRIX METER / EN 02/24/21 Version No. 00 13 Internal 5. Press key to confirm the selection and to enter the high limit value setting mode, the “ ▲” symbol and the previous high limit setting value are displayed. 6. Press ◄ key to move the blinking cursor to the desired position then press ▲ or ▼ key to set desired values, repeat this step to all digits. 7. Press key to store the setting value and to enter the low limit value setting mode, the
BRIX METER / EN 14 02/24/21 Version No. 00 Internal 6. EXPLANATION OF ERROR MESSAGES Error Code LCD Display Description Cover Sample plate cover not closed. No sample No sample exists on the prism surface. HHH Sample exceeds maximum measurement range LLL Sample is reading lower than the 0% standard used for meter zero calibration. Temperature reading are blinking Temperature measurement out of sampling range (<4.0°C)
BRIX METER / EN 02/24/21 Version No. 00 15 Internal Error Code LCD Display Description Temperature reading are blinking Temperature measurement out of sampling range (>60.0°C) ATC segment blinking Outside temperature compensation range (10°C to 40°C) CAL segment ON Wrong calibration used to zero the meter. Use deionized or distilled water. CAL segment ON Wrong calibration used to zero the meter. Use deionized or distilled water. CAL segment ON Temperatu