RS PRO 1389 Instruction Manual
Text of RS PRO 1389 Instruction Manual
BRIX METER / EN
05/24/18 Version No. 00
CONTENTS
Title
Page
1. INFORMATION ....................................................................... 1
2. FEATURES ............................................................................. 2
3. SPECIFICATIONS .................................................................. 3
4. NAME OF PARTS AND POSITIONS .............................
BRIX METER / EN 02/24/21 Version No. 00
1
Internal
1. INFORMATION
The meter is a portable optical meter that uses the
measurement of refractive index to determine the % Brix of
sugar in aqueous solutions.
The measurement technique and temperature compensation
use methodology recommended by ICUMSA (International
Commission of Uniform Method of Sugar Analysis).
The meter is a waterproof brix meter with a rating of IP65 and
has a memory of 99 data
BRIX METER / EN 2
02/24/21 Version No. 00
Internal
2. FEATURES
Wide measuring range (Brix 0 ~ 53%)
Refractive index measurement (RI)
Temperature measurement (°C/°F)
Ambient temperature compensation (ATC)
Brix or RI high/low limit value alarm function
Zero calibration with distilled or deionized water
Friendly operation and easy to use
Wide and dual display LCD and backlight function
Reliable water resistant (IP65)
BRIX METER / EN 02/24/21 Version No. 00
3
Internal
3. SPECIFICATIONS
Sensor: Photodiode arrays
Measuring Function: Brix (%)
Temperature (°C/°F)
Refractive index (RI)
Brix Range: 0.0 ~ 53.0%
(10 ~ 40°C/50 ~ 104°F Automatic temperature
compensation)
RI Range: 1.3329 ~ 1.4265
Temperature Range: 4 ~ 60°C (39.2~140°F)
Accuracy (Water at 23°C /73.4°F): Brix: ±0.2%
Temperature: ±1°C/°F
RI: ±0.0003
Resolution: Brix: 0.1%
Te
BRIX METER / EN 4
02/24/21 Version No. 00
Internal
4. NAME OF PARTS AND POSITIONS
4-1 Description of Parts & Control keys
UNIT
MEM
SET
ZERO
READ
FUNC
TEST
A.I.Alert
1. LCD Display: Dual display LCD is used for reporting the
measurement Brix (%), Refractive index (RI),
Temperature values and has several other marks
to provide function guidance.
BRIX METER / EN 02/24/21 Version No. 00
5
Internal
2. ◄ MEM READ key:
◄ key: In the setting mode, press ◄ key to move flicking cursor
to desired position.
MEM key: After measurement, press “MEM” key to store the
measured Brix (%), Refractive index (RI),
Temperature (°C or °F) values and the memorized
time (RTC) to memory.
READ key: Press and hold down the “SET” key then press
“READ” key to enter the memory data reading
mode.
Press “RE
BRIX METER / EN 6
02/24/21 Version No. 00
Internal
ZERO key: Press and hold down the “SET” key then press
“ZERO” key to perform the zero-calibration function.
6.
SET key:
key: In the setting mode, press
key to confirm the setting
item or stored the setting values.
SET key: Press and hold down the “SET” key then press
“READ” key to enter the memory data reading mode.
Press and hold down the “SET” key then press
“A.I.Alert” key to
BRIX METER / EN 02/24/21 Version No. 00
7
Internal
4-2 Description of Display
: Measurement in progress indication.
: Brix % and RI high/low limit value alarm function indication.
RI: Refractive index reading indication.
%: Brix % reading indication.
CAL: Zero calibration indication.
ATC: Automatic temperature compensation indication.
: Manual memory address number indication.
: Read mode indication.
: Recall manual memory address number indication.
▲:
BRIX METER / EN 8
02/24/21 Version No. 00
Internal
5. OPERATING INSTRUCTIONS
5-1 Precautions
Handle the meter carefully. Do not drop or shake the meter.
The meter is used to measure sugar solutions. Do not
expose the meter or prism to solvents that will damage it.
This includes most organic solvents and extremely hot or
cold solutions.
Never submerge the meter and do not spray water on the
meter except the sample plate located over the prism.
�
BRIX METER / EN 02/24/21 Version No. 00
9
Internal
4. Using plastic pipettes, drip a few drops of distilled or
deionized water on the prism, make sure that water covers
the prism surface completely.
5. Close the sample plate cover.
6. Press and hold down the “SET” key then press “ZERO” key
to enter the zero-calibration mode, the “
CAL” symbol
is displayed. When the “0.0 %” reading is displayed means
the meter is calibrated.
7. Gently absor
BRIX METER / EN 10
02/24/21 Version No. 00
Internal
7. Press “MEM” key to store the measured values and the RTC
to memory. Only one measurement can store at one time.
8. Remove sample from the sample plate by absorbing with a
soft tissue.
9. Using plastic pipettes, rinse prism and sample plate with
distilled or deionized water then wiper dry.
The meter is ready for the next sample.
5-4 Changing Temperature Unit °C °F
1. Press
key to turn on the
BRIX METER / EN 02/24/21 Version No. 00
11
Internal
5. Press
key to confirm the selection.
6. Press ◄ key to move the blinking cursor to the desired position
then press ▲ or ▼ key to set the element to real-time. Repeat
this step to all element.
7. Press
key to store the setting and exit this mode.
5-6 Data Memory and Read Mode
5-6-1 To Memorize the Reading
1. After each measurement, Press “MEM” key to store the
measured reading and the RTC to
BRIX METER / EN 12
02/24/21 Version No. 00
Internal
3. Press “FUNC” key to cycle display the Brix %, RI reading and
the memorized RTC.
4. Press “READ” key again to exit this mode.
5-6-3 To Clear the Memorized Data
1. Press
key to turn off the meter.
2. Press and hold down the “MEM” key then press
key to
turn on the meter, LCD will show “
” symbol one
time, it indicated that the memorized data is erased, the “
” symbol is displayed.
5-
BRIX METER / EN 02/24/21 Version No. 00
13
Internal
5. Press
key to confirm the selection and to enter the high
limit value setting mode, the “
▲” symbol and the previous
high limit setting value are displayed.
6. Press ◄ key to move the blinking cursor to the desired
position then press ▲ or ▼ key to set desired values, repeat
this step to all digits.
7. Press
key to store the setting value and to enter the low
limit value setting mode, the
BRIX METER / EN 14
02/24/21 Version No. 00
Internal
6. EXPLANATION OF ERROR MESSAGES
Error Code
LCD Display
Description
Cover
Sample plate cover not closed.
No sample
No sample exists on the prism
surface.
HHH
Sample exceeds maximum
measurement range
LLL
Sample is reading lower than the
0% standard used for meter zero
calibration.
Temperature
reading are
blinking
Temperature measurement out of
sampling range (<4.0°C)
BRIX METER / EN 02/24/21 Version No. 00
15
Internal
Error Code
LCD Display
Description
Temperature
reading are
blinking
Temperature measurement out of
sampling range (>60.0°C)
ATC segment
blinking
Outside temperature
compensation range (10°C to
40°C)
CAL segment
ON
Wrong calibration used to zero the
meter. Use deionized or distilled
water.
CAL segment
ON
Wrong calibration used to zero the
meter. Use deionized or distilled
water.
CAL segment
ON
Temperatu