Uson Qualipak-770 Manual
Text of the manual
Portions of the software are copyright © 2012 Uson, L.P. and
copyright © 1991 Free Software Foundation, Inc.
59 Temple Place, Suite 330, Boston, MA 02111‐1307, USA
Product or trade names mentioned in this manual may be trademarks or registered trademarks of their
respective companies.
Uson reserves the right to
1
Table of Contents
Section 1: Safety ...................................................................................................................... 2
Safety Precautions (English) ............................................................................................................. 2
Safety Precautions (Español) .................
Section 1: Safety
Safety Precautions (English)
WARNING: Areas of the equipment and/or manual that display this
symbol, or variation of, are potentially hazardous. Consult the
appropriate section of the user manual in respect of the specific
hazard.
WARNINGS: The system operates on 24VDC from a power supply
(supplied by Uson or customer), which must be conn
Safety Precautions (Español)
WARNING/ADVERTENCIA
Las zonas del equipo que muestran este sRmbolo son potencialmente
peligrosas. Consulte la sección del manual del usuario
correspondiente a ese peligro en particular.
System funciona con 24VDC desde una fuente de alimentación
(suministrado por Usón o cliente), que debe estar conectado a una
fuente de alim
Section 2: Introduction
Overview
The Qualipak-770 is an easy-to-use leak testing fixture used with Uson’s Optima vT Leak
Tester to perform seal integrity and leak testing on packages/parts for the food, medical,
and pharmaceutical industries. See Figure 1.
The Qualipak-770 Packaging Test System instructions are outlined in two manuals:
• Qualipak-77
Air supply: 6 bar
Mains supply: 100v -240v
Weight: Chamber Packed 40 kg, unpacked 35 kg/Optima Packed 42 kg, unpacked 32 kg
Chamber, lid and internal components Anodised Aluminium.
Adjustment lock nut Black acetyl
Hinges and all Screws Stainless Steel
Latch and Striker plate Hardened Stainless Steel
Pressure plate sheet: Porvair
Sample Infil
Note: Conduit can be threaded through either side of the test
chamber.
Perform the following steps to get your system ready to use:
1. Unpack the Optima Package Test Fixture and the Optima vT Leak Tester.
2. Turn the QUALIPAK-770 upside-down and place on a sturdy workbench. Then
remove the fixture’s bracket by unscrewing its 2 screws. See
Figure 3:Package Tester Connections Completed
6. Turn the fixture upright, and then remove the two transit screws. See Figure 4.
Figure 4: Transit Screws
7. Connect power supply to the Optima vT’s 24vdc input. See Figure 5.
The systems power supply must be connected to an
earthed (grounded) power supply of 90 - 260V, 50 - 60Hz. Only
personnel trained to