EUTECH INSTRUMENTS AMMONIUM EPOXY General Instructions Manual
Text of the manual
EUTECH INSTRUMENTS PTE LTD.
AMMONIUM ION ELECTRODE
INSTRUCTION MANUAL
GENERAL INSTRUCTIONS
Eutech Instruments Ammonium Ion Electrode is used to measure ammonium ions in aqueous
solutions in a more simple, quick, accurate, and economical manner.
Required Equipment
1.
A pH/mV meter or an ion meter, either line operated
Remove any rubber cap(s) covering the electrode tip(s) and the rubber inserts covering the filling
holes of the reference electrode. Fill the combination electrode or the reference electrode with the
filling solution shipped with the electrode to a level just below the fill hol
indicated by a slope of 90-100%. See the TROUBLESHOOTING section if the slope is not within this
range.
MEASUREMENT
Measuring Hints
The sensing membrane is normally subject to water uptake and might appear milky. This has no
effect on performance. All samples and standards should be at the same temperature for precise
TABLE 1: Concentration Unit Conversion Factors
ppm NH4
+
ppm N moles/liter NH4
+
1.80
1.40
1x10-4
18.00
14.00
1x10-3
180.00
140.00
1x10-2
Measurement Procedure
Direct Measurement
A simple procedure for measuring a large number of samples. A single meter reading
A calibration curve is constructed on semi-logarithmic paper when using a
pH/mV meter in the direct measurement procedure. The measured electrode
potential in mV (linear axis) is plotted against the standard concentration (log
axis). In the linear region of the curve, only three standards are necessary to
determine a
Immerse the electrode tip(s) in the solution and wait for the reading to stabilize. Read the
concentration directly from the meter display.
12.
The electrodes should be re-calibrated every 1-2 hours. Simply repeat Steps 2-8 (2-9)
above.
Low Level Ammonium Determination (using a standard pH/mV meter)
This procedure
Measure out 100 ml of the sample into a 150 ml beaker, add 1 ml of low level ISA, and
place the beaker on the magnetic stirrer. Begin stirring. Lower the electrode tip(s) into the
solution.
10.
After the reading has stabilized, record the mV reading and determine the concentration
from the low level calibration cur
+Tris+1 is the cation of tris(hydroxymethyl)aminomethane
Interference
(ppm)
100 ppm NH4
+
10 ppm NH4
+
1 ppm NH4+
Cs+1
1.0X102
1.0X101
1.0
K4
+1
2.7X102
2.7X101
2.7
Tl+1
3.1X103
3.1X102
3.1X101
H+1
1.6 pH
2.6 pH
3.6 pH
Ag+1
2.7X105
2.7X104
2.7X103
+Tris+1
3.1X
Plotting the electrode mV potential against the ammonium concentration on semi-logarithmic paper
results in a straight line with a slope of about 56 mV per decade. Refer to Figure 1.
The time needed to reach 99% of the stable electrode potential reading, the electrode response time,
varies from one
Hydrogen ion interferes with measurements of low levels of ammonium ion although the electrode
can be used over a wide pH range. Table 3 should be used to determine the minimum pH at which
low level ammonium measurements can be made without more than a 10% error due to hydrogen
ion interference.
Electrode L