Extech Instruments EX520 Operation & User’s Manual
Text of Extech Instruments EX520 Operation & User’s Manual
EX520-EU_ENG V2.3 1/08
2
Introduction
Congratulations on your purchase of the Extech EX520 True RMS Autoranging Multimeter.
This meter measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency
(electrical & electronic), Diode Test, and Continuity plus Thermocouple Temperature. It
features a rugged design for heavy duty use. Proper use and care of this meter will provide
many years of reliable service.
Safety
This symbol adjacent to anoth
EX520-EU_ENG V2.3 1/08 3
CAUTIONS
•
Improper use of this meter can cause damage, shock, injury or death. Read and
understand this user manual before operating the meter.
•
Always remove the test leads before replacing the battery or fuses.
•
Inspect the condition of the test leads and the meter itself for any damage before
operating the meter.
•
Use great care when making measurements if the voltages are greater than 25VAC
rms or 35VDC. These vo
EX520-EU_ENG V2.3 1/08 4
SAFETY INSTRUCTIONS
This meter has been designed for safe use, but must be operated with caution. The rules
listed below must be carefully followed for safe operation.
1.
NEVER apply voltage or current to the meter that exceeds the specified maximum:
Input Protection Limits
Function
Maximum Input
V DC or V AC
1000VDC/AC rms
mA AC/DC
800mA 1000V fast acting fuse
A AC/DC
10A 1000V fast acting fuse (20A for 30 seconds
max every 15
EX520-EU_ENG V2.3 1/08 5
Controls and Jacks
1. 6,000 count LCD
2. RANGE button
3. Hz and % button
4. Mode button
5. Function switch
6. mA, µA and 10A input jacks
7. COM input jack
8. Positive input jack
9. HOLD and Backlight button
10. RELATIVE button
11. MAX/MIN button
Note: Tilt stand and battery compartment are
on rear of unit.
Symbols and Annunciators
Auto power off
•)))
Continuity
Diode test
Battery status
n
nano (10
-9) (capacitance)
µ
micro
EX520-EU_ENG V2.3 1/08 6
Operating Instructions
WARNING: Risk of electrocution. High-voltage circuits, both AC and DC, are very
dangerous and should be measured with great care.
1. ALWAYS turn the function switch to the OFF position when the meter is not in use.
2. If “OL” appears in the display during a measurement, the value exceeds the range you
have selected. Change to a higher range.
DC VOLTAGE MEASUREMENTS
CAUTION: Do not measure DC voltages
EX520-EU_ENG V2.3 1/08 7
DC CURRENT MEASUREMENTS
CAUTION: Do not make 20A current measurements for longer than 30 seconds.
Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1. Insert black test lead banana plug into the negative COM jack.
2. For current measurements up to 6000µA DC, set the function
switch to the yellow µA position and insert the red test lead
banana plug into the µA/mA jack.
3. For current measurements up to
EX520-EU_ENG V2.3 1/08 8
RESISTANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and
discharge all capacitors before taking any resistance measurements. Remove the
batteries and unplug the line cords.
1. Set the function switch to the green Ω CAP position.
2. Insert the black test lead banana plug into the negative COM
jack. Insert the red test lead banana plug into the positive Ω jack.
3. Press the
EX520-EU_ENG V2.3 1/08 9
TEMPERATURE MEASUREMENTS
1. Set the function switch to the green Temp position.
2. Insert the Temperature Probe into the input jacks, making
sure to observe the correct polarity.
3. Press the MODE button to indicate ºF or ºC
4. Touch the Temperature Probe head to the part whose
temperature you wish to measure. Keep the probe touching
the part under test until the reading stabilizes (about 30
seconds).
5. Read the temperature i
EX520-EU_ENG V2.3 1/08 10
AUTORANGING/MANUAL RANGE SELECTION
When the meter is first turned on, it automatically goes into Autoranging. This automatically
selects the best range for the measurements being made and is generally the best mode
for most measurements. For measurement situations requiring that a range be manually
selected, perform the following:
1. Press the RANGE key. The “AUTO” display indicator will turn off.
2.Press the RANGE key to s
EX520-EU_ENG V2.3 1/08 11
Maintenance
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage
before removing the back cover or the battery or fuse covers.
WARNING: To avoid electric shock, do not operate your meter until the battery and fuse
covers are in place and fastened securely.
This MultiMeter is designed to provide years of dependable service, if the following care
instructions are performed:
1. KEEP THE METER DRY
EX520-EU_ENG V2.3 1/08 12
WARNING: To avoid electric shock, do not operate the meter until the battery cover is in
place and fastened securely.
NOTE: If your meter does not work properly, check the fuses and batteries to make sure that
they are still good and that they are properly inserted.
REPLACING THE FUSES
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage
before removing the fuse cover.
1. Disconnect the test le
EX520-EU_ENG V2.3 1/08 13
Specifications
Function
Range
Resolution
Accuracy
DC Voltage
600mV
0.1mV
±(0.09% reading + 2 digits)
6V
0.001V
60V
0.01V
600V
0.1V
1000V
1V
AC Voltage
50 to 60Hz
40Hz to 1KHz
6V
0.001V
±(1.0% reading + 3 dgts) ±(2.0% reading + 3 dgts)
60V
0.01V
600V
0.1V
1000V
1V
All AC voltage ranges are specified from 5% of range to 100% of range
DC Current
600μA
0.1μA
±(1.0% reading + 3 digits)
6000μA
1μA
60mA
0.01mA
600mA
0.1mA
6A
0
EX520-EU_ENG V2.3 1/08 14
Function
Range
Resolution
Accuracy
Resistance
600Ω
0.1Ω
±(0.3% reading + 4 digits)
6kΩ
0.001kΩ
60kΩ
0.01kΩ
600kΩ
0.1kΩ
6MΩ
0.001MΩ
40MΩ
0.01MΩ
±(0.3% reading + 20 digits)
Capacitance
60nF
0.01nF
±(3.5% reading + 4 digits)
600nF
0.1nF
6μF
0.001μF
60μF
0.01μF
600μF
0.1μF
1000μF
1μF
±(5% reading + 5 digits)
Frequency
(electronic)
9.999Hz
0.001Hz
±(0.1% reading + 1 digits)
99.99Hz
0.01Hz
999.9Hz
0.1Hz
9.999
EX520-EU_ENG V2.3 1/08 15
Enclosure
Double molded, waterproof (IP67)
Shock (Drop Test)
6.5 feet (2 meters)
Diode Test
Test current of 0.9mA maximum, open circuit voltage 2.8V DC
typical
Continuity Check
Audible signal will sound if the resistance is less than 100Ω
(approx.), test current <0.35mA
Temperature Sensor
Requires type K thermocouple
Input Impedance
>10MΩ VDC & >10MΩ VAC
AC Response
True rms
ACV Bandwidth
40Hz to 1kHz
Crest Factor
≤3 at