Extech Instruments 42515 Operation & User’s Manual
| Brand: | Extech Instruments |
| Category: | Thermometer |
| Model: | 42515 (Operation & User’s Manual) |
| File Type: | |
| File Size: | 0.18 Mb |
| Pages: | 9 |
Text of Extech Instruments 42515 Operation & User’s Manual
Wide Range InfraRed (IR) Thermometer
with Type K input and Laser Pointer
MODEL 42515
Shop for Extech products at:
1.877.766.5412
ShopExtech.com
42515-EU V4.1 2/09
2
Introduction
Congratulations on your purchase of the Model 42515 IR Thermometer. The Model 42515
IR thermometer measures and displays non-contact (infrared) temperature readings up to
1472
oF (800
oC). The built-in laser pointer increases target accuracy while the backlit LCD
and handy push-buttons combine for convenient, ergonomic operation. The Type K
thermocouple input provides contact measurements up to 1370
oC (2498
oF)*. The
42515-EU V4.1 2/09 3
Meter Description
1.
Laser pointer
2.
IR Sensor
3.
Measurement Trigger
4.
Battery and Switch compartment
5.
LCD Display
6.
Function Buttons
7.
Type-K thermocouple
8.
Temperature units (
oC or
oF) switch
9.
Test lock ON / OFF switch
10. Alarm ON / OFF switch
Note that switches are located in the battery compartment behind the battery.
1
2
3
4
8
Shop for Extech products at:
1.877.766.5412
ShopExtech.com
42515-EU V4.1 2/09 4
Operating Instructions
IR (non-contact) Measurements
1. Hold the meter by its handle and point it toward the surface to be measured.
2. Pull and hold the trigger to turn the meter on and begin testing. The temperature
reading, the flashing ‘SCAN’ icon, the emissivity, and the unit of measure will appear.
Note: Replace the 9V battery if the display does not switch on.
3. Release the Trigger and the reading will hold for approxim
42515-EU V4.1 2/09 5
The MODE button options
The MODE button is used to access the programming functions of the instrument. The
selected function is displayed on the bottom line of the LCD. Each parameter is listed
below with a detailed account of its use. Press the MODE button to step from one
parameter to the next.
EMS (Emissivity Value)
To change the emissivity value, use the UP and DOWN arrows (the range is 0.10 to
1.00). The current emissivity set
42515-EU V4.1 2/09 6
Over-range Indicator
If the temperature measurement exceeds the specified temperature range, the thermometer
will display dashes in place of a temperature reading.
High and Low Alarm Feature
The Model 42515 has an alarm feature whereas a High Alarm setting and a Low Alarm
setting can be programmed by the user. When either Alarm point is reached the meter will
alert the user via an audible beep and LCD display icon. Follow the steps
42515-EU V4.1 2/09 7
IR Measurement Notes
1. The object under test should be larger than the spot (target) size calculated by the field
of view diagram (printed on the side of the meter and in this guide).
2. Before measuring, be sure to clean surfaces that are covered with frost, oil, grime, etc.
3. If an object's surface is highly reflective, apply masking tape or flat black paint to the
surface before measuring. Allow time for the paint or tape to a
42515-EU V4.1 2/09 8
Emissivity and IR Measurement Theory
IR Thermometers measure the surface temperature of an object. The thermometer’s optics
sense emitted, reflected, and transmitted energy. The thermometer’s electronics translate
the information into a temperature reading which is then displayed on the LCD.
The amount of IR energy emitted by an object is proportional to an object's temperature
and its ability to emit energy. This ability is kn
42515-EU V4.1 2/09 9
Specifications
Infrared Thermometer (Non-contact) Specifications
Range / Resolution
-50.0 to 800.0
oC (-58.0 to 1472.0
oF)
Resolution: 0.1
o
Accuracy
-50 to -1°C (-58 to 30°F)
± 2.0%rdg or ±3°C/6°F
whichever is greater
-0.5 to 65°C (31°F to 150°F )
± 2.0%rdg or ±2.5°C/4.5°F
whichever is greater
66 to 315°C (151°F to 600°F)
± 2.0%rdg or ±2°C/4°F
whichever is greater
316 to 800°C (600F to 1472°F)
± (2.5%rdg +
Extech Instruments Thermometer Owners Instructions:
| Device Model | Type of Document | Pages | Size | Views | Downloads | Updated |
|---|---|---|---|---|---|---|
422324 |
Extech Instruments 422324 Manual (Instruction manual), @GGR26F |
21 | 569 | 120 | 28-04-2026 | |
IR400 |
IR400 (Thermometer ePDF Manual, #2ZPT21) |
6 | 0.22 Mb | 1363 | 219 | 03-10-2026 |
42580 |
42580 (Thermometer ePDF User Guide, #CEN5GQ) |
14 | 1294 | 259 | 25-08-2026 | |
4465CF |
4465CF (Thermometer ePDF User Guide, #2APLLF) |
8 | 0.23 Mb | 549 | 94 | 16-07-2026 |
42500 |
#N92P4R: 42500 Thermometer Operation & user’s manual |
4 | 0.15 Mb | 41 | 8 | 20-07-2026 |
AN200 |
AN200 (Thermometer ePDF User Guide, #1N5LO3) |
14 | 0.36 Mb | 689 | 152 | 15-04-2026 |