TTI PL154 Instruction Manual
Text of TTI PL154 Instruction Manual
TTI PL154 Table of Contents
Specification
2
Safety
4
EMC
6
Installation
7
Connections
8
Operation
9
Maintenance
12
Instructions en Francais
Sécurité
13
Installation
14
Fonctionnement
16
Entretien
19
Bedienungsanleitung auf Deutsch
Sicherheit
20
Installation
21
Betrieb
23
Wartung
27
1
Nominally 0-32V, (PL320/330); 0-15.5V (PL154).
Nominally 0-2·1A (PL320); 0-3·1A (PL330); 0-4A (PL154).
Output Voltage Setting:
By coarse and fine controls; resolution <5mV across the range.
Output Current Setting:
By single logarithmic control.
Output Mode:
Constant voltage or constant current modes with automatic cross-over.
Decimal points flash to indicate constant current mode.
Configuration Selection:
(QMD a
TTI PL154 < 20 µsec to within 50 mV of setting for 90% load change.
Temperature Coefficient:
Typically < 100 ppm/oC.
Metering (PL330QMT
only):
Meter Type:
3·75 digit (4095 count) with 12·5mm (0·5") LEDs. Reading rate 4Hz.
Meter Resolution:
Meter Accuracy:
10 mV and 10 mA
0·5% of reading + 1 digit.
GENERAL
AC Input Voltage:
Internally set for 110, 120, 220, 230 or 240VAC 50/60 Hz; operating range
±10% of setting. Installation Category I
TTI PL154 This instrument is Safety Class I according to IEC classification and has been designed to meet
the requirements of EN61010-1 (Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use). It is an Installation Category II instrument intended for operation
from a normal single phase supply.
This instrument has been tested in accordance with EN61010-1 and has been supplied in a safe
condition. This instruction manual cont
TTI PL154 EC Declaration of Conformity
We Thurlby Thandar Instruments Ltd
Glebe Road
Huntingdon
Cambridgeshire PE29 7DR
England
declare that the following power supplies:
PL154, PL320, PL320QMD, PL320QMT, PL330, PL330QMD, PL330QMT
meet the intent of the EMC Directive 2004/108/EC and the Low Voltage Directive 2006/95/EC.
Compliance was demonstrated by conformance to the following specifications which have been
listed in the Official Journal of the European Communitie
This instrument has been designed to meet the requirements of the EMC Directive 2004/108/EC.
Compliance was demonstrated by meeting the test limits of the following standards:
Emissions
EN61326-1 (2006) EMC product standard for Electrical Equipment for Measurement, Control and
Laboratory Use. Test limits used were:
a) Radiated: Class B
b) Conducted: Class B
c) Harmonics: EN61000-3-2 (2006) Class A; the instrument is Class A by product category.
Mains Operating Voltage
Check that the operating voltage of the instrument shown on the rear panel is suitable for the
local supply. Should it be necessary to change the operating voltage range proceed as follows:
1. Ensure that the instrument is disconnected from the AC supply.
2. Remove the screws holding the case upper and handle.
3. Lift off the case upper.
4. If the transformer primary taps are marked: A 0-110-120 B 0-110-1
Make sure that only fuses with the required rated current and of the specified type are used for
replacement. The use of makeshift fuses and the short-circuiting of fuse holders is prohibited.
Mains Lead
When a three core mains lead with bare ends is provided this should be connected as follows:
BROWN
-
MAINS LIVE
BLUE
-
MAINS NEUTRAL
GREEN/YELLOW
-
EARTH
Safety Earth Symbol
As the colours of the wires in the mains lead of this apparatus may not correspond
Setting Up the Output
With the power switch ON and the output switch OFF the output voltage and current limit may be
accurately set using the three output controls prior to connection to the load. The left hand meter
indicates voltage and the right hand meter indicates current.
With the output switch in the OFF (set) position, the current meter shows the value of the current
limit setting (indicated by all the decimal points coming o
WARNING! Such voltages are exceedingly hazardous and great care should be taken to shield
the output terminals for such use. On no account should the output terminals be touched when
the unit is switched on under such use. All connections to the terminals must be made with the
power switched off on all units.
It should be noted that the unit can only source current and cannot sink it, thus units cannot be
series connected in anti-phase.
The unit can be con
Additional Instructions for Quad Mode Triple Versions
The Quad-Mode Triple consists of a Quad-Mode Dual as described in the previous section, plus a
logic supply.
P310QMT Logic Supply
This has a fixed 5V 1·5A output which is isolated from the other supplies.
The logic supply is protected against short circuits. It is protected against external forward
voltages up to 16 Volts and reverse voltages by a diode: the reverse current must not exceed 3
Amps.
PL32
The unit has a very low output impedance, but this is inevitably increased by the resistance of the
connecting leads. At high currents, this can result in significant differences between the indicated
source voltage and the actual load voltage, (two 50mΩ connecting leads will drop 0·2V at 2 Amps,
for instance). This problem can be minimised by using short, thick, connecting leads, but where
necessary it can be completely overcome by using
Cet instrument est de Classe de sécurité 1 suivant la classification IEC et il a été construit pour
satisfaire aux impératifs EN61010-1 (Impératifs de sécurité pour le matériel électrique en vue de
mesure, commande et utilisation en laboratoire). Il s'agit d'un instrument d'installation Catégorie II
devant être exploité depuis une alimentation monophasée habituelle.
Cet instrument a été soumis à des essais conformément à EN610
Tension d'utilisation secteur
Vérifier que la tension opérationnelle de l'instrument indiquée sur le panneau arrière est
appropriée pour l'alimentation locale. Procéder de la manière ci-dessous pour changer la gamme
de tensions opérationnelles:
1. S'assurer que l'instrument est débranché de l'alimentation c.a.
2. Enlever les vis de retenue de la manette et de la partie supérieure du boîtier.
3. Retirer la partie supérieure d
Le fusible c.a. se trouve sur le panneau arrière; il faut noter que le PL320QMT et PL330QMT ont
un fusible c.a. supplémentaire pour leurs sections de sortie logiques. Le type de fusible correct
est 20 mm x 5 mm 250V, action différée HBC avec valeurs nominales suivantes:
Modèle
220/230/240V
110/115/120V
PL320 / PL154 (simple)
1·6A (T)
3·15A (T)
PL330 (simple)
2A (T)
4A (T)
PL320QMD / PL320QMT
3·15A (T)
6·3A (T)
PL330QMD
4A(T)
8A (T)
PL330QM
Réglage de la sortie
L'interrupteur d'alimentation allumé et le commutateur de sortie éteint, l'utilisation des trois
commandes de sortie permet de régler avec précision la limite de tension et de courant de sortie
avant d'effectuer une connexion à la charge. L'appareil de mesure gauche indique la tension et
l'appareil droit le courant. Le commutateur de sortie éteint (réglé), l'appareil de mesure courant
indique
Connexion en série ou en parallèle avec d'autres appareils
La sortie du bloc d'alimentation a une charge entièrement constante et elle peut être utilisée en
série avec d'autres blocs d'alimentation pour générer des hautes tensions c.a. pouvant atteindre
300V c.c. AVERTISSEMENT! Des tensions de ce genre sont extrêmement dangereuses et il
faut prendre soin de protéger les bornes de sortie pour une utilisation de ce genre. Il ne faut sous
aucun pr
Le commutateur de sortie de l'appareil asservi doit toujours être allumé. Le commutateur de
sortie de l'appareil principal assure alors la commutation des deux sorties. Il n'est pas possible
d'utiliser la télédétection en mode en série ou alignement et il faut donc laisser en place les
barres de court-circuitage.
Instructions supplémentaires pour les versions triples mode quadruple
La version triple mode quadruple est composée de la version double