TTI PL154 Instruction Manual

Text of TTI PL154 Instruction Manual

TTI PL154 Table of Contents

Specification 2 Safety 4 EMC 6 Installation 7 Connections 8 Operation 9 Maintenance 12 Instructions en Francais Sécurité 13 Installation 14 Fonctionnement 16 Entretien 19 Bedienungsanleitung auf Deutsch Sicherheit 20 Installation 21 Betrieb 23 Wartung 27 1

Nominally 0-32V, (PL320/330); 0-15.5V (PL154).

Nominally 0-2·1A (PL320); 0-3·1A (PL330); 0-4A (PL154). Output Voltage Setting: By coarse and fine controls; resolution <5mV across the range. Output Current Setting: By single logarithmic control. Output Mode: Constant voltage or constant current modes with automatic cross-over. Decimal points flash to indicate constant current mode. Configuration Selection: (QMD a

TTI PL154 < 20 µsec to within 50 mV of setting for 90% load change.

Temperature Coefficient: Typically < 100 ppm/oC. Metering (PL330QMT only): Meter Type: 3·75 digit (4095 count) with 12·5mm (0·5") LEDs. Reading rate 4Hz. Meter Resolution: Meter Accuracy: 10 mV and 10 mA 0·5% of reading + 1 digit. GENERAL AC Input Voltage: Internally set for 110, 120, 220, 230 or 240VAC 50/60 Hz; operating range ±10% of setting. Installation Category I

TTI PL154 This instrument is Safety Class I according to IEC classification and has been designed to meet

the requirements of EN61010-1 (Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use). It is an Installation Category II instrument intended for operation from a normal single phase supply. This instrument has been tested in accordance with EN61010-1 and has been supplied in a safe condition. This instruction manual cont

TTI PL154 EC Declaration of Conformity

We Thurlby Thandar Instruments Ltd Glebe Road Huntingdon Cambridgeshire PE29 7DR England declare that the following power supplies: PL154, PL320, PL320QMD, PL320QMT, PL330, PL330QMD, PL330QMT meet the intent of the EMC Directive 2004/108/EC and the Low Voltage Directive 2006/95/EC. Compliance was demonstrated by conformance to the following specifications which have been listed in the Official Journal of the European Communitie

This instrument has been designed to meet the requirements of the EMC Directive 2004/108/EC.

Compliance was demonstrated by meeting the test limits of the following standards: Emissions EN61326-1 (2006) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use. Test limits used were: a) Radiated: Class B b) Conducted: Class B c) Harmonics: EN61000-3-2 (2006) Class A; the instrument is Class A by product category.

Mains Operating Voltage

Check that the operating voltage of the instrument shown on the rear panel is suitable for the local supply. Should it be necessary to change the operating voltage range proceed as follows: 1. Ensure that the instrument is disconnected from the AC supply. 2. Remove the screws holding the case upper and handle. 3. Lift off the case upper. 4. If the transformer primary taps are marked: A 0-110-120 B 0-110-1

Make sure that only fuses with the required rated current and of the specified type are used for

replacement. The use of makeshift fuses and the short-circuiting of fuse holders is prohibited. Mains Lead When a three core mains lead with bare ends is provided this should be connected as follows: BROWN - MAINS LIVE BLUE - MAINS NEUTRAL GREEN/YELLOW - EARTH Safety Earth Symbol As the colours of the wires in the mains lead of this apparatus may not correspond

Setting Up the Output

With the power switch ON and the output switch OFF the output voltage and current limit may be accurately set using the three output controls prior to connection to the load. The left hand meter indicates voltage and the right hand meter indicates current. With the output switch in the OFF (set) position, the current meter shows the value of the current limit setting (indicated by all the decimal points coming o

WARNING! Such voltages are exceedingly hazardous and great care should be taken to shield

the output terminals for such use. On no account should the output terminals be touched when the unit is switched on under such use. All connections to the terminals must be made with the power switched off on all units. It should be noted that the unit can only source current and cannot sink it, thus units cannot be series connected in anti-phase. The unit can be con

Additional Instructions for Quad Mode Triple Versions

The Quad-Mode Triple consists of a Quad-Mode Dual as described in the previous section, plus a logic supply. P310QMT Logic Supply This has a fixed 5V 1·5A output which is isolated from the other supplies. The logic supply is protected against short circuits. It is protected against external forward voltages up to 16 Volts and reverse voltages by a diode: the reverse current must not exceed 3 Amps. PL32

The unit has a very low output impedance, but this is inevitably increased by the resistance of the

connecting leads. At high currents, this can result in significant differences between the indicated source voltage and the actual load voltage, (two 50mΩ connecting leads will drop 0·2V at 2 Amps, for instance). This problem can be minimised by using short, thick, connecting leads, but where necessary it can be completely overcome by using

Cet instrument est de Classe de sécurité 1 suivant la classification IEC et il a été construit pour

satisfaire aux impératifs EN61010-1 (Impératifs de sécurité pour le matériel électrique en vue de mesure, commande et utilisation en laboratoire). Il s'agit d'un instrument d'installation Catégorie II devant être exploité depuis une alimentation monophasée habituelle. Cet instrument a été soumis à des essais conformément à EN610

Tension d'utilisation secteur

Vérifier que la tension opérationnelle de l'instrument indiquée sur le panneau arrière est appropriée pour l'alimentation locale. Procéder de la manière ci-dessous pour changer la gamme de tensions opérationnelles: 1. S'assurer que l'instrument est débranché de l'alimentation c.a. 2. Enlever les vis de retenue de la manette et de la partie supérieure du boîtier. 3. Retirer la partie supérieure d

Le fusible c.a. se trouve sur le panneau arrière; il faut noter que le PL320QMT et PL330QMT ont

un fusible c.a. supplémentaire pour leurs sections de sortie logiques. Le type de fusible correct est 20 mm x 5 mm 250V, action différée HBC avec valeurs nominales suivantes: Modèle 220/230/240V 110/115/120V PL320 / PL154 (simple) 1·6A (T) 3·15A (T) PL330 (simple) 2A (T) 4A (T) PL320QMD / PL320QMT 3·15A (T) 6·3A (T) PL330QMD 4A(T) 8A (T) PL330QM

Réglage de la sortie

L'interrupteur d'alimentation allumé et le commutateur de sortie éteint, l'utilisation des trois commandes de sortie permet de régler avec précision la limite de tension et de courant de sortie avant d'effectuer une connexion à la charge. L'appareil de mesure gauche indique la tension et l'appareil droit le courant. Le commutateur de sortie éteint (réglé), l'appareil de mesure courant indique

Connexion en série ou en parallèle avec d'autres appareils

La sortie du bloc d'alimentation a une charge entièrement constante et elle peut être utilisée en série avec d'autres blocs d'alimentation pour générer des hautes tensions c.a. pouvant atteindre 300V c.c. AVERTISSEMENT! Des tensions de ce genre sont extrêmement dangereuses et il faut prendre soin de protéger les bornes de sortie pour une utilisation de ce genre. Il ne faut sous aucun pr

Le commutateur de sortie de l'appareil asservi doit toujours être allumé. Le commutateur de

sortie de l'appareil principal assure alors la commutation des deux sorties. Il n'est pas possible d'utiliser la télédétection en mode en série ou alignement et il faut donc laisser en place les barres de court-circuitage. Instructions supplémentaires pour les versions triples mode quadruple La version triple mode quadruple est composée de la version double

TTI Power Supply Owners Instructions: