TTI TSX Series Instruction Manual

Text of TTI TSX Series Instruction Manual

THURLBY THANDAR INSTRUMENTS

INSTRUCTION MANUAL TSX Series Programmable High Current DC Power Supplies

Table of Contents

Specification 3 Safety 5 EMC 6 Installation 7 Manual Operation 8 Maintenance and Repair 11 Remote Operation 12 ARC Interface 13 Remote Commands 22 Instructions en Francais Sécurité 34 Montage 35 Fonctionnement manuel 36 Entretien et Réparations 40 Fonctionnement à distance 41 Commandes à distance 52 Bedienungsanleitung auf Deutsch Sicherheit 64 Installation 65 Handbedienung 66 Wartung und Instandsetzung 70 Fernbetrieb 71 Fernbefehle

EC Declaration of Conformity

We Thurlby Thandar Instruments Ltd Glebe Road Huntingdon Cambridgeshire PE29 7DR England declare that the TSX3510P & TSX1820P Programmable Power Supplies meet the intent of the EMC Directive 2004/108/EC and the Low Voltage Directive 2006/95/EC. Compliance was demonstrated by conformance to the following specifications which have been listed in the Official Journal of the European Communities. EMC Emissions: a) EN61326-1 (2006)

Output Voltage Range:

0V to 35.3V (35V/10A); 0V to 18.15V (18V/20A). Output Current Range: 0.01A to 10.2A (35V/10A); 0.01A to 20.2A (18V/20A). Output Voltage Setting: Direct keyboard entry or quasi-analogue rotary control; setting resolution 10mV. Output Current Setting: Direct keyboard entry or quasi-analogue rotary control; setting resolution 10mA. Output Mode: Constant voltage or constant current with automatic cros

TTI TSX Series Note: All voltage and current levels set via the keyboard are displayed on a separate 0·3” 4-digit

display. This entry preview system ensures that the user can observe the value entered before it is effected thus avoiding possible error. The display is also used for setting additional functions. When the output switch is on and no other function is selected, the display shows output power in Watts. DIGITAL INTERFACES Operational Functions: Set voltage,

This power supply is a Safety Class I instrument according to IEC classification and has been

designed to meet the requirements of EN61010-1 (Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use). It is an Installation Category II instrument intended for operation from a normal single phase supply. This instrument has been tested in accordance with EN61010-1 and has been supplied in a safe condition. This instruct

This instrument has been designed to meet the requirements of the EMC Directive 2004/108/EC.

Compliance was demonstrated by meeting the test limits of the following standards: Emissions EN61326-1 (2006) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use. Test limits used were: a) Radiated: Class B b) Conducted: Class B c) Harmonics: EN61000-3-2 (2006) Class A; the instrument is Class A by product category.

Mains Operating Voltage

Check that the instrument operating voltage marked on the rear panel is suitable for the local supply. Should it be necessary to change the operating voltage, proceed as follows: 1. Ensure that the instrument is disconnected from the AC supply. 2. Remove the 6 screws holding the case upper and lift off the cover. 3. Change the appropriate zero-ohm links beside the transformer on the pcb: Link LK5 only for 230V operatio

The POWER switch is used to apply line voltage to the instrument. When switched to on (l) the

instrument will be powered and the start-up procedure will be executed; this will take approximately 5 second to complete. If all is well the settings from the last power-down will be installed and the instrument will be ready for use. The output terminals are located on the right of the front panel. The red terminal is positive and th

Once the required value is entered press the CONFIRM key and the set current will be updated

immediately. To exit without making any changes press the ESCAPE key. If a mistake is made during entry press the AMPS key and start again. The maximum and minimum values accepted will depend on the particular model, see the specification section for details. Set OVP The Over-Voltage Protection (OVP) value is not shown in the main display; however, it is active and

Current Meter Damping

The output current meter damping is toggled on and off by pressing the DAMPING key. When damping is on the DAMP LED will also be on. Output On/Off The output is alternatively turned on and off by pressing the OUTPUT key. The status is shown by the ON LED next to the key. Store Settings The instrument contains 25 stores each capable of holding the entire set-up. The stored data is non-volatile and is retained while power is off. To sav

The voltage drop in each output lead must not exceed 1V.

The shorting links should be re-made between the rear sense and output terminals when remote sensing is not being used. However, the sense connection is also made internally through a low value resistor and only a small error between the set and actual voltage will result if the links are left disconnected. Sense Miswiring Trip The output will be tripped off if the sense wires are wired to the wrong

The following sections detail the operation of the instrument via both GPIB and ARC. Where

operation is identical no distinction is made between the two. Where difference occur these are detailed in the appropriate sections or in some cases separate sections for GPIB and ARC. It is therefore only necessary to read the general sections and those sections specific to the interface of interest. Interface Selection All power supplies in the ra

TTI TSX Series ARC Interface Connections

The 9-way D-type serial interface connector is located on the instrument rear panel. The pin connections are as shown below: Pin Name Description 1 - No internal Connection 2 TXD Transmitted data from instrument 3 RXD Received data to instrument 4 - No internal connection 5 GND Signal ground 6 - No internal connection 7 RXD2 Secondary received data (see diagram) 8 TXD2 Secondary transmitted data (see diagram) 9 GND S

TTI TSX Series The ARC standard for the other interface parameters is as follows; in these power supplies (and

most other ARC instruments) they are fixed. Start bits 1 Data bits 8 Parity None Stop bits 1 ARC Character Set Because of the need for XON/XOFF handshake it is possible to send ASCII coded data only; binary blocks are not allowed. Bit 7 of ASCII codes is ignored, i.e. assumed to be low. No distinction is made between upper and lower case characters in command mn

Before a response can be read from an instrument it must be addressed to talk by sending the

Talk Address control code, 14H (TAD) followed by a single character which has the lower 5 bits corresponding to the unique address of the required instrument, as for the listen address control code above. Once addressed to talk the instrument will send the response message it has available, if any, and then exit the talk addressed state. Only one response message w

This instrument contains the following IEEE 488.1 subsets:

Source Handshake SH1 Acceptor Handshake AH1 Talker T6 Listener L4 Service Request SR1 Remote Local RL1 Parallel Poll PP1 Device Clear DC1 Device Trigger DT0* Controller C0 Electrical Interface E2 * Although no Device Trigger capability is included, the GET message will not cause a command error unless its position in the input stream dictates that it should; e.g. buried inside a

TTI TSX Series bit 7 =

X don't care bit 6 = 1 bit 5 = 1 Parallel poll enable bit 4 = 0 bit 3 = Sense - sense of the response bit; 0 = low, 1 = high bit 2 = ? bit 1 = ? bit position of the response bit 0 = ? Example. To return the RQS bit (bit 6 of the Status Byte Register) as a 1 when true and a 0 when false in bit position 1 in response to a parallel poll operation send the following commands ∗PRE 64, then PPC followed by 69H (PPE) The parallel poll resp

Standard Event Status and Standard Event Status Enable Registers

These two registers are implemented as required by the IEEE std. 488.2. Any bits set in the Standard Event Status Register which correspond to bits set in the Standard Event Status Enable Register will cause the ESB bit to be set in the Status Byte Register. The Standard Event Status Register is read and cleared by the ∗ESR? command. The Standard Event Status Enable register is set by the �

TTI Power Supply Owners Instructions: