TTI TSX Series Instruction Manual
Text of TTI TSX Series Instruction Manual
THURLBY THANDAR INSTRUMENTS
INSTRUCTION MANUAL
TSX Series
Programmable High Current DC Power Supplies
Table of Contents
Specification
3
Safety
5
EMC
6
Installation
7
Manual Operation
8
Maintenance and Repair
11
Remote Operation
12
ARC Interface
13
Remote Commands
22
Instructions en Francais
Sécurité
34
Montage
35
Fonctionnement manuel
36
Entretien et Réparations
40
Fonctionnement à distance
41
Commandes à distance
52
Bedienungsanleitung auf Deutsch
Sicherheit
64
Installation
65
Handbedienung
66
Wartung und Instandsetzung
70
Fernbetrieb
71
Fernbefehle
EC Declaration of Conformity
We
Thurlby Thandar Instruments Ltd
Glebe Road
Huntingdon
Cambridgeshire PE29 7DR
England
declare that the
TSX3510P & TSX1820P Programmable Power Supplies
meet the intent of the EMC Directive 2004/108/EC and the Low Voltage Directive 2006/95/EC.
Compliance was demonstrated by conformance to the following specifications which have been
listed in the Official Journal of the European Communities.
EMC
Emissions:
a) EN61326-1 (2006)
Output Voltage Range:
0V to 35.3V (35V/10A); 0V to 18.15V (18V/20A).
Output Current Range:
0.01A to 10.2A (35V/10A); 0.01A to 20.2A (18V/20A).
Output Voltage Setting:
Direct keyboard entry or quasi-analogue rotary control; setting
resolution 10mV.
Output Current Setting:
Direct keyboard entry or quasi-analogue rotary control; setting
resolution 10mA.
Output Mode:
Constant voltage or constant current with automatic cros
TTI TSX Series Note: All voltage and current levels set via the keyboard are displayed on a separate 0·3” 4-digit
display. This entry preview system ensures that the user can observe the value entered before it
is effected thus avoiding possible error. The display is also used for setting additional functions.
When the output switch is on and no other function is selected, the display shows output power in
Watts.
DIGITAL INTERFACES
Operational Functions:
Set voltage,
This power supply is a Safety Class I instrument according to IEC classification and has been
designed to meet the requirements of EN61010-1 (Safety Requirements for Electrical Equipment
for Measurement, Control and Laboratory Use). It is an Installation Category II instrument
intended for operation from a normal single phase supply.
This instrument has been tested in accordance with EN61010-1 and has been supplied in a safe
condition. This instruct
This instrument has been designed to meet the requirements of the EMC Directive 2004/108/EC.
Compliance was demonstrated by meeting the test limits of the following standards:
Emissions
EN61326-1 (2006) EMC product standard for Electrical Equipment for Measurement, Control and
Laboratory Use. Test limits used were:
a)
Radiated: Class B
b)
Conducted: Class B
c)
Harmonics: EN61000-3-2 (2006) Class A; the instrument is Class A by product category.
Mains Operating Voltage
Check that the instrument operating voltage marked on the rear panel is suitable for the local
supply. Should it be necessary to change the operating voltage, proceed as follows:
1.
Ensure that the instrument is disconnected from the AC supply.
2.
Remove the 6 screws holding the case upper and lift off the cover.
3.
Change the appropriate zero-ohm links beside the transformer on the pcb:
Link LK5 only for 230V operatio
The POWER switch is used to apply line voltage to the instrument. When switched to on (l) the
instrument will be powered and the start-up procedure will be executed; this will take
approximately 5 second to complete. If all is well the settings from the last power-down will be
installed and the instrument will be ready for use.
The output terminals are located on the right of the front panel. The red terminal is positive and
th
Once the required value is entered press the CONFIRM key and the set current will be updated
immediately. To exit without making any changes press the ESCAPE key. If a mistake is made
during entry press the AMPS key and start again.
The maximum and minimum values accepted will depend on the particular model, see the
specification section for details.
Set OVP
The Over-Voltage Protection (OVP) value is not shown in the main display; however, it is active
and
Current Meter Damping
The output current meter damping is toggled on and off by pressing the DAMPING key. When
damping is on the DAMP LED will also be on.
Output On/Off
The output is alternatively turned on and off by pressing the OUTPUT key. The status is shown by
the ON LED next to the key.
Store Settings
The instrument contains 25 stores each capable of holding the entire set-up. The stored data is
non-volatile and is retained while power is off.
To sav
The voltage drop in each output lead must not exceed 1V.
The shorting links should be re-made between the rear sense and output terminals when remote
sensing is not being used. However, the sense connection is also made internally through a low
value resistor and only a small error between the set and actual voltage will result if the links are
left disconnected.
Sense Miswiring Trip
The output will be tripped off if the sense wires are wired to the wrong
The following sections detail the operation of the instrument via both GPIB and ARC. Where
operation is identical no distinction is made between the two. Where difference occur these are
detailed in the appropriate sections or in some cases separate sections for GPIB and ARC. It is
therefore only necessary to read the general sections and those sections specific to the interface
of interest.
Interface Selection
All power supplies in the ra
TTI TSX Series ARC Interface Connections
The 9-way D-type serial interface connector is located on the instrument rear panel. The pin
connections are as shown below:
Pin
Name
Description
1
-
No internal Connection
2
TXD
Transmitted data from instrument
3
RXD
Received data to instrument
4
-
No internal connection
5
GND
Signal ground
6
-
No internal connection
7
RXD2
Secondary received data (see diagram)
8
TXD2
Secondary transmitted data (see diagram)
9
GND
S
TTI TSX Series The ARC standard for the other interface parameters is as follows; in these power supplies (and
most other ARC instruments) they are fixed.
Start bits
1
Data bits
8
Parity
None
Stop bits
1
ARC Character Set
Because of the need for XON/XOFF handshake it is possible to send ASCII coded data only;
binary blocks are not allowed. Bit 7 of ASCII codes is ignored, i.e. assumed to be low. No
distinction is made between upper and lower case characters in command mn
Before a response can be read from an instrument it must be addressed to talk by sending the
Talk Address control code, 14H (TAD) followed by a single character which has the lower 5 bits
corresponding to the unique address of the required instrument, as for the listen address control
code above. Once addressed to talk the instrument will send the response message it has
available, if any, and then exit the talk addressed state. Only one response message w
This instrument contains the following IEEE 488.1 subsets:
Source Handshake
SH1
Acceptor Handshake
AH1
Talker
T6
Listener
L4
Service Request
SR1
Remote Local
RL1
Parallel Poll
PP1
Device Clear
DC1
Device Trigger
DT0*
Controller
C0
Electrical Interface
E2
*
Although no Device Trigger capability is included, the GET message will not cause a
command error unless its position in the input stream dictates that it should; e.g. buried
inside a
TTI TSX Series bit 7 =
X
don't care
bit 6 =
1
bit 5 =
1
Parallel poll enable
bit 4 =
0
bit 3 =
Sense - sense of the response bit; 0 = low, 1 = high
bit 2 =
?
bit 1 =
?
bit position of the response
bit 0 =
?
Example.
To return the RQS bit (bit 6 of the Status Byte Register) as a 1 when true and a 0 when
false in bit position 1 in response to a parallel poll operation send the following commands
∗PRE 64, then PPC followed by 69H (PPE)
The parallel poll resp
Standard Event Status and Standard Event Status Enable Registers
These two registers are implemented as required by the IEEE std. 488.2.
Any bits set in the Standard Event Status Register which correspond to bits set in the Standard
Event Status Enable Register will cause the ESB bit to be set in the Status Byte Register.
The Standard Event Status Register is read and cleared by the ∗ESR? command. The Standard
Event Status Enable register is set by the �