Teledyne 316RA Instruction Manual
Text of Teledyne 316RA Instruction Manual
TELEDYNE ANALYTICAL INSTRUMENTS
MODEL
316RA/RB/RAD/RBD
TRACE OXYGEN ANALYZER
INSTRUCTION MANUAL
TAI Sales Order Number:
Model Number:
Serial Number:
Micro-Fuel Cell Class:
Output Signal:
Ranges:
X1:
ppm Oxygen
X10:
ppm Oxygen
X100:
ppm Oxygen
X1000:
ppm Oxygen
Alarm Set Point #1:
Alarm Set Point #2:
P/N M34905.4
M38546.4
ECO# 95-348
06/06/95
TELEDYNE ANALYTICAL INSTRUMENTS Copyright © 1993 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted,
transcribed, stored in a retrieval system, or translated into any other language or com-
puter language in whole or in part, in any form or by any means, whether it be elec-
tronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written
consent of Teledyne Analytical Instrument
TELEDYNE ANALYTICAL INSTRUMENTS Table of Contents
Introduction ............................................................................... 1
Features .............................................................................. 2
Operational Theory ................................................................... 3
Micro-Fuel Cell Sensor........................................................ 3
Sampling System .......................................
TELEDYNE ANALYTICAL INSTRUMENTS 1
Trace Oxygen Analyzer
Model 316RA/RB/RAD/RBD
Introduction
The Teledyne Analytical Instruments (TAI) Series 316 Trace Oxygen
Analyzer uses a unique Micro-Fuel Cell to measure the concentration of
oxygen in a gas stream. The cell has an absolute zero and produces a linear
output from the low parts per million (ppm) level through 21% (210,000
ppm) oxygen. When coupled with a “state of the art” two stage amplifier
system t
TELEDYNE ANALYTICAL INSTRUMENTS 2
Model 316RA/RB/RAD/RBD Trace Oxygen Analyzer
Features
Maintenance-Free Sensor. Teledyne’s Micro-Fuel Cell oxygen
sensor is a sealed electrochemical transducer with no electrolyte to change
or electrodes to clean. When the cell reaches the end of its useful life (6
months minimum), simply remove and replace.
Temperature Compensation. The Series 316 includes a built-in
temperature compensation circuit for greater accuracy.
TELEDYNE ANALYTICAL INSTRUMENTS 3
Trace Oxygen Analyzer
Model 316RA/RB/RAD/RBD
Operational Theory
Micro-Fuel Cell Sensor
The Series 316 uses a Micro-Fuel Cell for the measurement of trace
oxygen. This Micro-Fuel Cell is an electrochemical transducer whose
features include:
•
Specificity for oxygen
•
Long interval between calibration
•
Long life
•
Low maintenance
•
Disposable configuration
The transducer functions as a fuel cell; in this instance,