Teledyne MD300 Operating/service Instructions Manual
Text of Teledyne MD300 Operating/service Instructions Manual
Portable Oxygen Monitor
Teledyne Analytical Instruments
i
OPERATING & SERVICE INSTRUCTIONS FOR
AD300
PORTABLE OXYGEN ANALYZER
MD300
PORTABLE OXYGEN MONITOR
TYPE B EQUIPMENT:
Equipment providing a particular degree
of protection against electric shock,
particularly regarding—
• Allowable LEAKAGE CURRENT
• Reliability of the protective earth
connection (if present).
P/N M/AD/MD300
06-23-2010
Teledyne Analytical Instruments
ii
Copyright © 2010 Teledyne Analytical Instruments
All Rights Reserved
No part of this manual may be reproduced, transmitted, transcribed, stored in a
retrieval system, or translated into any other language or computer language in whole
or in part, in any form or by any means, whether it be electronic, mechanical, optical,
manual, or otherwise, without prior written consent of Teledyne Instruments,
Analytical I
Portable Oxygen Monitor Teledyne Analytical Instruments
iii
About This Manual
The AD300/MD300 Operator's Manual provides both introductory
and detailed information for configuring and operating these
instruments. The manual takes you from the time you unpack the
instrument until you complete the first gas analysis. The bulk of the
manual contains operating procedures and information. There are also
cautions, warnings, and guidelines to ensure that your m
Teledyne Analytical Instruments iv
Safety Messages
Your safety and the safety of others are very important. Please
carefully read the following safety messages.
Safety message are indented to alert the user of potential hazards.
Each safety message is associated with a safety alert symbol. These
symbols are found in the manual and on the instrument. The definition
of these symbols is described below:
CAUTION: Refer to the instructions for detai
Portable Oxygen Monitor Teledyne Analytical Instruments
v
Table of Contents
Safety Messages .......................................................................... iv
List of Figures .............................................................................. vii
List of Tables .............................................................................. viii
Introduction ...............................................................................
Teledyne Analytical Instruments vi
2.4 Do’s and Don’ts
27
Service Manual ............................................................................ 29
3.1 General Service Information
29
3.2 Overall Maintenance
29
3.3 Battery Maintenance
29
3.4 Sensor Maintenance
30
3.5 Gas Sampling
30
3.5.1 Discrepancy in Readings
30
3.6 Troubleshooting
31
3.7 Watchdog
34
3.8 Other Problems with the Instrument
36
3.9 Return Authorization for Service
36
Ap
Portable Oxygen Monitor Teledyne Analytical Instruments
vii
List of Figures
Figure 1-1: MD300 Front View ........................................................ 9
Figure 1-2: AD300 Front View ....................................................... 10
Figure 2-1: Installing the R17S Sensor .......................................... 14
Figure 2-2: Sensor Cable Connection to Monitor ......................... 14
Figure 2-3: Mounting the Sensor in the Tee Ada
Teledyne Analytical Instruments viii
List of Tables
Table 3-1 Troubleshooting ............................................................ 31
Table 3-2 Error Codes ................................................................... 35
Teledyne Analytical Instruments 9
Introduction
The Teledyne Analytical Instruments AD300 Oxygen Analyzer
here after referred to as AD300 and MD300 Oxygen Monitors with
alarms here after referred to as MD300 are portable instruments that
provide fast and accurate oxygen monitoring. In addition, the model
MD300 incorporates an audio/visual alarm capability. These instruments
are designed to monitor up to 100% oxygen concentration
Teledyne Analytical Instruments 10
Figure 1-2: AD300 Front View
The LCD display consists of up to 3.5 characters plus a decimal
point indicator capable of displaying up to 105%. (See Section 1.2
Options for alternate display configuration). An integral battery life
indicator is displayed continuously on the AD300 and ―on demand‖
with the MD300.
The instrument is powered by three AA alkaline batteries