YSI Pro10 Operation & User’s Manual
File Specifications
| Brand: | YSI |
| Category: | Measuring Instruments |
| Model: | Pro10 (Operation & User’s Manual), prgvC6cHcjumY |
| File Type: | |
| File Size: | — |
| Pages: | 18 pages |
Download PDF
Download YSI Pro10 Manual (18 pages)Read Online:
Read the YSI Pro10 Operation & User’s Manual online using the built-in PDF viewer.
Key Pages of Document
Rev A, July 2014 For the latest version of this manual, visit ysi.com ©2014 YSI Incorporated. The YSI logo is a registered trademarks of YSI Incorporated. C O N T E N T S Warranty ................................................................................................i Introduction ...................................................... ...
Saving Data ...........................................................................15 Viewing and Erasing Saved Data ........................................15 Care, Maintenance and Storage ......................................................17 General Mainte ...
I N T R O D U C T I O N Thank you for purchasing the YSI Pro10, an instrument from the YSI Professional Series product family. The Pro10 measures temperature and either pH or ORP in water. The Pro10 features an impact resistant and waterproof (IP-67) case, a rugged MS-8 (military-spec) cable ...
battery symbols indicating polarities. The waterproof instrument case is sealed at the factory and is not to be opened, except by factory-authorized service technicians. Do not attempt to separate the two halves of the instrument case as this may damage the instrument, break the waterproof seal, and will ...
not water-proof. Do not submerge the cable without a sensor installed! Submerging the cable without a sensor installed may cause permanent damage to the cable that is not covered under warranty. Install a YSI ProSeries port plug into the port if not installing a pH or ORP sensor. Figure 3. T ...
will turn on the display backlight. The backlight will remain on until the key is pressed again or after two minutes of not pressing any key on the keypad. powerIng off To turn the instrument off, press and hold the power/backlight key for three seconds. navIgatIon The up and ...
aUdIo Audio can be enabled by highlighting Audio and pressing enter. When enabled, there will be an ‘X’ in the box next to Audio. When Audio is enabled, the Pro10 will beep twice to indicate stability when Auto Stable is enabled. The instrument will also beep when a key is pressed. When Audio is disable ...
the bottom of the display and press enter to close the Auto Stable submenu. To disable Auto Stable, set the % Change input to 0.0. ph BUffer set Highlight pH Buffer Set and press enter to open a submenu that allows you to select the Buffer Set used for auto buffer recognition during a pH ...
an older or dirty sensor, calibrate in buffer 7 or 6.86 first. For highest accuracy, use fresh, traceable pH buffers and ensure the sensor and calibration cup are clean. 1-POINT CALIBRATION 1. Place the sensor in pH buffer 7 or 6.86 and allow the temperature and pH readings to stabi ...