SICK GM32 Operating Instructions Manual
Text of SICK GM32 Operating Instructions Manual
O PERATING INSTRU CTIO NS
Installation
Start-up
Maintenance
GM32
In-situ Gas Analyzer
Measuring Probe Version
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH
Document Information
Described Product
Product name:
GM32
Variants:
GM32 Probe (measuring probe)
Document Identification
Title:
Operating Instructions GM32
Part No.:
8012707
Version:
1.3
Release:
2009-03
Publisher
SICK MAIHAK GmbH
Nimburger Str. 11 · D-79276 Reute · Germany
Phone:
+49 7641 469-0
Fax:
+49 7641 469-11 49
E-Mail:
[email protected]
Guarantee Information
Specified product character
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH
3
Warning Symbols
Warning levels / Signal words
DANGER
Risk or hazardous situation which will result in severe personal
injury or death.
WARNING
Risk or hazardous situation which could result in severe personal
injury or death.
CAUTION
Hazard or unsafe practice which could result in personal injury or
property damage.
NOTICE
Hazard which could result in material damage.
Information Symbols
Hazard
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 1
Important Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.1
Main instructions for operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.2
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 5
4
Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.1
Necessary technical knowledge for start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.2
Required material (not included in the scope of delivery) . . . . . . . . . . . . . . . . . . . . . . .
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 8
Clearing Malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
8.1
General hazard caused by electrical voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8.2
Measured value blinks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 7
GM32
1
Important Information
Main safety information
Main instructions for operation
Intended use
Own responsibility
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH Important Information
1.1
Main instructions for operation
1.2
Intended use
1.2.1
Purpose of the device
The GM32 serves exclusively for emission and process monitoring of gases in industrial
plants.
GM32 measures continuously directly in the gas duct (in-situ).
1.3
Responsibility of user
Designated users
The GM32 may be operated by competent persons only who, based on their device-specific
train
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 9
1.4
Additional documentation/information
� Pay attention to the supplied documents.
Additional instructions
The following documents are applicable in addition to these Operating Instructions:
● Technical Information GM32 (option)
● Operating Instructions for purge air supply (for GMP probe)
● Technical Information “Modular I/O System” (option)
● Final inspect
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 11
GM32
2
Product Description
Product identification
Functional principle
Characteristics
Variants
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH Product Description
2.1
Product identification
2.2
Product description
The GM32 gas analyzer serves for continuous measurement of the gas concentrations in
industrial plants.
GM32 is an in-situ measuring system. Measurement is performed directly in the gas carry-
ing duct.
● Components: SO2, NO, NO2 and NH3 (device-specific) as well as the temperature and
pressure reference values.
● Measu
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 13
2.3
SOPAS ET (PC program)
SOPAS ET can be used to set the GM32 parameters and provides access to the GM32 log-
book.
SOPAS ET runs on an external PC connected via the Ethernet interface (→ p. 22, Figure 4) to
GM32.
2.4
Reference cycle
Correction of internal drifts in an adjustable interval (standard: 1 hour, setting: SOPAS ET),
via a command (with SOPAS ET) or via an exte
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH Product Description
Figure 1
Check point
● Output of measured values during the check cycle: Last valid measured value.
● Signal during the check cycle: Not_measuring. (Optionally digital output or OPC inter-
face).
● The determined zero and reference values can be output on analog outputs depending
on the parameter settings:
–
Directly after the check cycle.
–
On request (via a digi
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 15
2.6
Design of GM32
The GM32 Probe version comprises
● Sender/receiver unit (SR-unit)
The SR-unit contains optical and electronic subassemblies.
The concentration calculation of the sample gas according to the absorption spectro-
scope principle is performed in the SR-unit.
● Measuring probe with flange resp. purge air fixture (→ § 2.6.1)
● Purge air unit (for GMP m
Preparation on the Gas Duct Side
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH
17
GM32
3
Preparation on the Gas Duct Side
Setup
Installation
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH Preparation on the Gas Duct Side
3.1
Preparation of sampling point
� Determine the installation location.
Observe the ambient conditions of the GM32 (→ p. 67, §9.2).
� Observe the space requirements for the SR-unit (→ p. 72, §).
Provide for additional space for maintenance work (swiveling the housing door open,
pulling the measuring probe out).
� Determine the installation locatio
Preparation on the Gas Duct Side GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH
19
3.2
Overview of the installation steps (duct-side work)
Figure 3
Installation overview
3.2.1
Work steps (overview)
30
L
50
470
Ø 133
Weatherproof cover
“Flange with tube”: L = 240 mm (standard)
Attachment (e.g. 4 steel
pipes, 50 x 5 mm)
for purge air unit
SR-unit
Approx. 1.5 m
Min. 700 mm
Angle: Approx. 1°
Step
Procedure
Reference
1
Installing the “flan
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH Preparation on the Gas Duct Side
3.2.2
Installing the “flange with tube” on the gas duct
1
Make a cut-out in the gas duct for the “flange with tube”.
2
Insert the “flange with tube” so that the mark (TOP) ▴ points upwards vertically (irre-
spective of the gas duct angle) and attach the “flange with tube”.
–
The tube must project at least 30 mm into the gas duct.
–
Make su