O PERATING INSTRU CTIO NS

Installation Start-up Maintenance GM32 In-situ Gas Analyzer Measuring Probe Version

GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH

Document Information Described Product Product name: GM32 Variants: GM32 Probe (measuring probe) Document Identification Title: Operating Instructions GM32 Part No.: 8012707 Version: 1.3 Release: 2009-03 Publisher SICK MAIHAK GmbH Nimburger Str. 11 · D-79276 Reute · Germany Phone: +49 7641 469-0 Fax: +49 7641 469-11 49 E-Mail: [email protected] Guarantee Information Specified product character

GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH

3 Warning Symbols Warning levels / Signal words DANGER Risk or hazardous situation which will result in severe personal injury or death. WARNING Risk or hazardous situation which could result in severe personal injury or death. CAUTION Hazard or unsafe practice which could result in personal injury or property damage. NOTICE Hazard which could result in material damage. Information Symbols Hazard
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 1 Important Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 1.1 Main instructions for operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 1.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 5 4 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 4.1 Necessary technical knowledge for start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 4.2 Required material (not included in the scope of delivery) . . . . . . . . . . . . . . . . . . . . . . .
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 8 Clearing Malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 8.1 General hazard caused by electrical voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 8.2 Measured value blinks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 7 GM32 1 Important Information Main safety information Main instructions for operation Intended use Own responsibility
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH Important Information 1.1 Main instructions for operation 1.2 Intended use 1.2.1 Purpose of the device The GM32 serves exclusively for emission and process monitoring of gases in industrial plants. GM32 measures continuously directly in the gas duct (in-situ). 1.3 Responsibility of user Designated users The GM32 may be operated by competent persons only who, based on their device-specific train
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 9 1.4 Additional documentation/information � Pay attention to the supplied documents. Additional instructions The following documents are applicable in addition to these Operating Instructions: ● Technical Information GM32 (option) ● Operating Instructions for purge air supply (for GMP probe) ● Technical Information “Modular I/O System” (option) ● Final inspect
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 11 GM32 2 Product Description Product identification Functional principle Characteristics Variants
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH Product Description 2.1 Product identification 2.2 Product description The GM32 gas analyzer serves for continuous measurement of the gas concentrations in industrial plants. GM32 is an in-situ measuring system. Measurement is performed directly in the gas carry- ing duct. ● Components: SO2, NO, NO2 and NH3 (device-specific) as well as the temperature and pressure reference values. ● Measu
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 13 2.3 SOPAS ET (PC program) SOPAS ET can be used to set the GM32 parameters and provides access to the GM32 log- book. SOPAS ET runs on an external PC connected via the Ethernet interface (→ p. 22, Figure 4) to GM32. 2.4 Reference cycle Correction of internal drifts in an adjustable interval (standard: 1 hour, setting: SOPAS ET), via a command (with SOPAS ET) or via an exte
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH Product Description Figure 1 Check point ● Output of measured values during the check cycle: Last valid measured value. ● Signal during the check cycle: Not_measuring. (Optionally digital output or OPC inter- face). ● The determined zero and reference values can be output on analog outputs depending on the parameter settings: – Directly after the check cycle. – On request (via a digi
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 15 2.6 Design of GM32 The GM32 Probe version comprises ● Sender/receiver unit (SR-unit) The SR-unit contains optical and electronic subassemblies. The concentration calculation of the sample gas according to the absorption spectro- scope principle is performed in the SR-unit. ● Measuring probe with flange resp. purge air fixture (→ § 2.6.1) ● Purge air unit (for GMP m

Preparation on the Gas Duct Side

GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 17 GM32 3 Preparation on the Gas Duct Side Setup Installation
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH Preparation on the Gas Duct Side 3.1 Preparation of sampling point � Determine the installation location. Observe the ambient conditions of the GM32 (→ p. 67, §9.2). � Observe the space requirements for the SR-unit (→ p. 72, §). Provide for additional space for maintenance work (swiveling the housing door open, pulling the measuring probe out). � Determine the installation locatio
Preparation on the Gas Duct Side GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH 19 3.2 Overview of the installation steps (duct-side work) Figure 3 Installation overview 3.2.1 Work steps (overview) 30 L 50 470 Ø 133 Weatherproof cover “Flange with tube”: L = 240 mm (standard) Attachment (e.g. 4 steel pipes, 50 x 5 mm) for purge air unit SR-unit Approx. 1.5 m Min. 700 mm Angle: Approx. 1° Step Procedure Reference 1 Installing the “flan
GM32 Operating Instructions V1.3 8012707 © SICK MAIHAK GmbH Preparation on the Gas Duct Side 3.2.2 Installing the “flange with tube” on the gas duct 1 Make a cut-out in the gas duct for the “flange with tube”. 2 Insert the “flange with tube” so that the mark (TOP) ▴ points upwards vertically (irre- spective of the gas duct angle) and attach the “flange with tube”. – The tube must project at least 30 mm into the gas duct. – Make su