The meter responds to program peaks� The PPM style of indicator was developed by

broadcasters to control the peak levels of audio signals and so prevent overmodulation of the transmitters� The PPM meter allows the user to verify in real time that an AFILS can provide sufficient field strength for good speech clarity� • Th

Instructions for use

Background noise measurement (Noise) NOTE: Do not use the monitor headphones on higher than volume 4 for this test, (i�e� on LO range setting) in order to maintain accuracy and stability in the instrument� 1� Switch off or disconnect from the mains power any audio induction loop equipment, (note that any other mains operated eq

6� Move the RANGE switch to HI position (7)�

7� Hold the meter in vertical position with the top at the listening height and measure the variation (field strength deviation) within the listening area� Confirm that the field strength does not deviate by more than ±3dB� 8� Measure the overspill distribution outside the loop area, if required�

Adjustment of field strength level (Field)

NOTE: Univox 1kHz_pulse�wav or itu�wav test signal should be used for adjustment of field strength level� A continuous sine wave is not recommended since the loop driver’s AGC might decrease the level� 1� Activate the audio file 1kHz_pulse�wav� 2� Adjust the input level according to the driver’