Univox FSM Basic Operation & User’s Manual
Text of Univox FSM Basic Operation & User’s Manual
Introduction ������������������������������������������������������������������������������������������� 2
Overview �����������������
The meter responds to program peaks� The PPM style of indicator was developed by
broadcasters to control the peak levels of audio signals and so prevent overmodulation
of the transmitters� The PPM meter allows the user to verify in real time that an AFILS
can provide sufficient field strength for good speech clarity�
•
Th
Instructions for use
Background noise measurement (Noise)
NOTE: Do not use the monitor headphones on higher than volume 4 for this test, (i�e� on LO
range setting) in order to maintain accuracy and stability in the instrument�
1� Switch off or disconnect from the mains power any audio induction loop equipment,
(note that any other mains operated eq
6� Move the RANGE switch to HI position (7)�
7� Hold the meter in vertical position with the top at the listening height and measure
the variation (field strength deviation) within the listening area� Confirm that the field
strength does not deviate by more than ±3dB�
8� Measure the overspill distribution outside the loop area, if required�
Adjustment of field strength level (Field)
NOTE: Univox 1kHz_pulse�wav or itu�wav test signal should be used for adjustment of field
strength level� A continuous sine wave is not recommended since the loop driver’s AGC
might decrease the level�
1� Activate the audio file 1kHz_pulse�wav�
2� Adjust the input level according to the driver’