Keysight U1583B Operating Instructions
Text of Keysight U1583B Operating Instructions
Keysight U1583B Current Clamp
Operating Instructions
The U1583B current clamp is a dual range 40 A and 400 A clamp-on AC current clamp. This U1583B current
clamp is designed for Keysight handheld digital multimeter (DMM) and Keysight handheld oscilloscope. A
BNC-to-banana plug is required to connect the current clamp with the DMM. For the handheld oscilloscope,
use a BNC connector to connect the clamp.
Assistance
For technical assistance, contact your near
- Inspect the output cable without exposing the metal to ensure insulation.
- Do not operate the adapter around explosive gas, vapor, or dust.
- Do not exceed the rated voltage/current as marked on the adapter.
- Use with extreme caution when working around bare conductors or bus bars.
Accidental contact with the conductors could result in electric shock.
- Always keep your hand behind the finger guard of the clamp jaw.
- When servicing the adapter, use
• The current conductor must be centered within the jaw aperture and no influence from
adjacent currents.
• Connect to load impedance > 1 MW ||100 pF.
• Crest factor < 3.
Current Clamp with
BNC connector
BNC-to-banana plug
Handheld Digital
Multimeter (DMM)
Current Clamp with
BNC connector
Handheld
Oscilloscope
Figure 1: Connection Diagram
Alignment Marks
Put one conductor only within the jaws on the inside section of the
indicated marks as much as
AC 40 A range
1.
Lock the RANGE button of the U1583B current clamp to enable the 40 A mode.
2.
Connect the output BNC of the U1583B to a BNC-to-dual banana converter plug and proceed to
connect it to the output of the V (HI) and COM (LO) terminals of the multimeter.
3.
Set the multimeter to AC 500.0 mV or 1000.0 mV.
4.
Open the jaws of the current clamp and centrally place it around the 50 turns coil.
5.
Set the calibrator output to