LeCroy AP034 Instruction Manual
Text of LeCroy AP034 Instruction Manual
LeCroy warrants this oscilloscope accessory for normal use and operation within
specifications for a period of three years from the date of shipment. Calibration after each 12
month interval is recommended to ensure performance to specification. Spare parts,
replacement parts and repairs are warranted for 90 days. The instrument’s firmware has
been thoroughly tested and thought to be functional, but is supplied without warranty of any
kind co
AP034-OM-E Rev D
ISSUED: January 2000
���
��������
Table of Contents
Overview
Description .................................................................................................................................. 1-1
Applications ................................................................................................................................. 1-1
Conventions used in this Manual ..................................
ISSUED: January 2000
AP034-OM-E Rev D
������������������
Incorrect Frequency Response........................................................................................5-2
DC Errors .........................................................................................................................5-3
Poor Common Mode Rejection..............................................................................
AP034-OM-E Rev D ISSUED: January 2000
�
��������
Specifications
Nominal Characteristics .............................................................................................................. 8-1
Warranted Electrical Characteristics ........................................................................................... 8-2
Typical Electrical Characteristics .......................................................................
ISSUED: January 2000 AP034-OM-E Rev D
������������������
BLANK PAGE
AP034-OM-E Rev D ISSUED: January 2000
���
��������
��������
DESCRIPTION
The AP034 is a wide-band, active differential probe. The probe
features low noise, low input capacitance, high common mode
rejection, and Field Effect Transistor (FET) buffered inputs in the
probe head. The user-selectable offset gives the probe the
flexibility to measure a large range of signal types. Plug-on
attenuators and AC coupling accessorie
ISSUED: January 2000 AP034-OM-E Rev D
������������������
CONVENTIONS USED IN THIS MANUAL
The following conventions may appear in this manual:
Note
A Note contains general information relating to
the use of the product.
Caution
A Caution contains information that should to be
followed to avoid possible damage to the
instrument or the device under test.
WARNING
A Warning alerts you to potential injury to
yourself
AP034-OM-E Rev D ISSUED: January 2000
���
������������������
������������������
OPERATOR SAFETY
The AP034 Active Differential Probe is intended to be used only
with instruments that are connected to earth ground through the
input BNC connector. When you are using the probe with an
ADPPS Power Supply Adapter, make sure the adapter is
connected to a BNC connector that is grounded b
ISSUED: January 2000 AP034-OM-E Rev D
������������������
BLANK PAGE
AP034-OM-E Rev D ISSUED: January 2000
���
���������
���������
Caution
The input circuits in the AP034 Active Differential
Probe incorporate components that protect the
probe from damage resulting from electrostatic
discharge (ESD). Keep in mind that this is an
active probe, and it should be handled carefully
to avoid damage. When using the AP034 Active
Differential Probe, you are advised to take
precautions against
ISSUED: January 2000 AP034-OM-E Rev D
������������������
connected to the test circuit. The ground connection is optional.
Positive voltages applied to the + input relative to the – input will
deflect the oscilloscope trace toward the top of the screen.
To maintain the high performance capability of the probe in
measurement applications, user care in connecting the probe to
the test circuit. Increasing the pa
AP034-OM-E Rev D ISSUED: January 2000
���
���������
characteristics of the input of the probe with
which
they
were
shipped.
The
input
characteristics of the external attenuator itself do
not exactly match those of the probe. Therefore,
using the ÷10 and ÷20 attenuators at the same
time is not recommended. The scale factor
encoding system will not operate correctly with
both attenuators installed simultaneously.
The input character
ISSUED: January 2000 AP034-OM-E Rev D
������������������
•
The larger (0.8 mm) mini clips can be used to
connect to through-hole leaded components.
•
The offset round pins can be used for hand-held
probing applications. Reposition the pins by rotating
them to obtain the required spacing.
PROBE INPUT LOADING
Attaching any probe to a test circuit will add some loading. In
most applications, the high impedance
AP034-OM-E Rev D ISSUED: January 2000
���
���������
In most cases it is not necessary to ground the probe to the
circuit under test. However, if the test circuit is isolated from earth
ground, it is usually necessary to connect the probe ground to a
point in the circuit. Grounding test circuits that are referenced to
earth ground may improve the fidelity of high-frequency
components in the waveforms. The potential for improvement
wi
ISSUED: January 2000 AP034-OM-E Rev D
������������������
SELECTING THE PROPER RANGE
The AP034 Active Differential Probe has a fixed gain of X1
(unity). Use of the ÷10 or ÷20 external attenuators reduces the
amplitude of the input signal when it is necessary to extend the
dynamic operating range of the probe. Attenuating the input
signal increases both the differential mode range and common
mode range of the p
AP034-OM-E Rev D ISSUED: January 2000
���
���������
When using a differential probe or amplifier, be careful to avoid
exceeding the common mode range. Because the common
mode signal is rejected by the differential probe, and is not
displayed, changes in the amplitude of the common mode
component are not apparent to the user. Exceeding the common
mode range may introduce distortion into the AP034 output.
Circuitry in the probe detect