Extech Instruments VFM200 Operation & User’s Manual
| Brand: | Extech Instruments |
| Category: | Measuring Instruments |
| Model: | VFM200 (Operation & User’s Manual) |
| File Type: | |
| File Size: | — |
| Pages: | 5 |
Text of Extech Instruments VFM200 Operation & User’s Manual
Extech Instruments VFM200
User Manual
Formaldehyde (HCHO or CH20) and Total
Volatile Organic Compounds (TVOC) Tester
Model VFM200
Additional User Manual Translations available at www.extech.com
2 VFM200‐en‐GB_V1.2 12/16
1
6
5
7
2
3
4
8
Introduction
Thank you for selecting the Extech Instruments Model VFM200. The VFM200 measures HCHO
(formaldehyde) and TVOC (Total Volatile Organic Compounds).
Potential sources for TVOCs include paints, finishes, adhesives, cigarette smoke, pesticides,
personal care products, car exhaust, new
3 VFM200‐en‐GB_V1.2 12/16
mg/m
mg/m
3
3
1
2
3
4
5
6
7
LCD DISPLAY
1.
HCHO reading
2.
TVOC reading
3.
Selectable units of measure
4.
Alarm Beeper symbol
5.
Auto Power OFF reminder
6.
Rechargeable Battery status
7.
High and Low Alarm symbols
Operation
POWER THE METER
Press the
button to turn on the meter. The battery symbol informs the�
4 VFM200‐en‐GB_V1.2 12/16
CONTROL BUTTONS
1.
Short press the ▼SET button to silence the beeper (when it is sounding) or to
disable/enable the alarm beeper. This button is also used as a down arrow to change the
high and low Alarm threshold values in the Alarm Setup mode.
2.
Long press the ▼SET button to enter the Al
5 VFM200‐en‐GB_V1.2 12/16
Specifications
Sensor
Professional Fuel Cell technology
HCHO Range
0.00 to 5.00 mg/m³ (or PPM)
HCHO Resolution
0.01 mg/m3 (or PPM)
HCHO Basic Accuracy
±5% FS
TVOC Range
0.00 to 9.99 mg/m³ (or PPM)
TVOC Resolution
0.01 mg/m³ (or PPM)
TVOC Basic Accuracy
±5% FS
Response Time
≤ 2 seconds
Operating Temper