Text of Extech Instruments RH350 Operation & User’s Manual
Extech Instruments RH350 RH350/355 V2.1 8/08
2
Introduction
Congratulations on your purchase of the Extech RH350 SuperHeat Psychrometer. This
device measures Differential Temperature (T1-T2, Air-T1, T1-DP), Humidity, Air
Temperature, Dew Point and Wet Bulb. Advanced features include built-in memory and
recall of up to 99 data points, Min/MAX/Average readings from the captured data, Data
Hold, selectable Auto Power Off, and an RS-232 interface for capturing readings to a PC
u
RH350/355 V2.1 8/08 3
Operation
Open the Protective Sensor Cap
Twist the protective sensor cap, located at the top of the
meter, counter-clockwise a quarter of a turn to expose the
sensors.
Turn Power ON or OFF
Press the [ON/OFF] button to turn power on or off. The
meter will perform a short self-test when turned on.
Taking measurements
1. Hold the probe in the area to be tested.
2. Allow adequate time for readings to stabilize.
3. The Relative Humid
RH350/355 V2.1 8/08 4
Backlight Display
Press the [M
in/AVRST] and [MEM] buttons together to turn on the backlight. It will remain on
for ten seconds.
Automatic Power OFF
With the meter off, press and hold down the [ON/OFFSET] + [RECHOLD] buttons for more than
1 second to enter auto power off settings. Press the [MEM] button to select non-sleep (n) or
2, 5, 10, 20, 40 and 60 minutes autosleep. Press [T2▲WB] to confirm the selection.
Manual Datalog
RH350/355 V2.1 8/08 5
CALIBRATION
The following verification and calibration procedures require the 33% and 75% RH reference
bottles which are supplied with the Model RH355 kit.
Accuracy Verification
Checking the 33% or 75% RH Calibration
1.
Insert meter’s sensor into the 33% or 75% salt reference bottle
2.
Check the reading after a minimum of 10 minutes
3.
Verify that the reading is within the accuracy specification
Relative Humidity Calibration (
RH350/355 V2.1 8/08 6
�
Support line (781) 890-7440
Technical support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User’s Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Specifications
Function
Range and Resolution
Accuracy
Humidity
0.0 to 100.0%