Extech Instruments RH300 Operation & User’s Manual
Text of Extech Instruments RH300 Operation & User’s Manual
Models RH300 and RH305 (kit)
Introduction
Congratulations on your purchase of the Extech RH300 Digital Psychrometer. This device
measures Humidity, Air Temperature, Dew Point temperature and Wet Bulb temperature.
Advanced features include External Temperature Measurement (with optional probe), Data
Hold, selectable Auto Power Off, and MAX/MIN/Average reading. Careful use of this meter
will provide years of reliable service
RH300 V1.4 11/04
2
Specifications
Function
Range and Resolution
Accuracy
Humidity
0.0 to 100.0% RH
±3% RH (10 to 90%) @ 23
oC
Temperature (internal)
-4.0 to 122.0
oF (-20 to 50
oC) ±1.8
oF (±1
oC)
Temperature (external)
-4.0 to 158.0
oF (-20 to 70
oC) ±1.8
oF (±1
oC)
Display
Triple LCD
Sensor Type
Humidity: Precision capacitance sensor
Temperature: Thermistor
Response Time
60 seconds typical
Dew Point
-90.4 to 122.0
oF (-68 to 50
oC) (calculated
RH300 V1.4 11/04 3
/WB
T2
T/DP
T1
1
2
3
4
5
6
7
8
9
10
11
Display Description
1.
Relative Humidity %
2.
T1
3.
T1-T2
4.
Dew Point
5.
Wet Bulb
6.
T2-Dew Point
7.
°C/°F
8.
T2
9.
Maximum
10. Minimum
11. Low Battery/Hold
Operation
Open the Protective Sensor Cap
Twist the protective sensor cap, located at the top of
the meter, counter-clockwise a quarter of a turn to
expose the sensor
Plug in the optional external temperature probe (if
needed).
Turn Pow
RH300 V1.4 11/04 4
RESET
MX/MN
RESET
MX/MN
RESET
MX/MN
RESET
MX/MN
/WB
T2
T/DP
T1
Wet Bulb and Dew Point measurements
1. To read the Wet Bulb measurement, press the button until the symbol Wb
appears in the lower right display. Note: T2 will not be displayed unless the external
probe is connected.
2. Wet Bulb temperature is now displayed.
3. To read the Dew Point measurement, press the button until the symbol DP
appears in the low
RH300 V1.4 11/04 5
RESET
MX/MN
RESET
MX/MN
CALIBRATION
The following verification and calibration procedures require the 33% and 75% RH reference
bottles which are supplied with the Model RH305 kit.
Accuracy Verification
Checking the 33% or 75% RH Calibration
1. Insert meter’s sensor into the 33% or 75% salt reference bottle
2. Check the reading after 10 minutes
3. Verify that the reading is within the accuracy specification
Relative Humidity Calib
RH300 V1.4 11/04 6
�
Support line (781) 890-7440
Technical support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User’s Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Maintenance
Cleaning and storage
1. The meter should be cleaned with a damp clot
Manual del usuario
Sicrómetro Digital
Modelo RH300 y RH305 (kit)
Introducción
Felicitaciones por su compra del Sicrómetro digital RH300 de Extech Este dispositivo mide
humedad, temperatura del aire, temperatura de punto de rocío y temperatura de
termómetro de bulbo húmedo . Las características avanzadas incluyen medición de
temperatura exterior (con sonda opcional), retención de datos, apagado automático
optativo y lecturas MÁX / MIN/Promedio. E
RH300 V1.4 11/04
2
Especificaciones
Función
Escala y Resolución
Precisión
Humedad
0.0 a 100.0% RH
±3% RH (10 a 90%) @ 23
oC
Temperatura (interior)
-20.0 a 50
oC (-4 a 122.0
oF) ±1
oC (±1.8
oF)
Temperatura (exterior)
-20 a 70
oC (-4.0 a 158.0
oF) ±1
oC (±1.8
oF)
Pantalla
LCD triple
Sensor tipo
Humedad: Sensor de precisión para capacitancia
Temperatura: Termistor
Punto de rocío
-68 a 50
oC (-90.4 a 122.0
oF)
Bulbo húmedo
-21.6 a 50
oC (-6.88 a 12