Models RH300 and RH305 (kit)

Introduction Congratulations on your purchase of the Extech RH300 Digital Psychrometer. This device measures Humidity, Air Temperature, Dew Point temperature and Wet Bulb temperature. Advanced features include External Temperature Measurement (with optional probe), Data Hold, selectable Auto Power Off, and MAX/MIN/Average reading. Careful use of this meter will provide years of reliable service

RH300 V1.4 11/04

2 Specifications Function Range and Resolution Accuracy Humidity 0.0 to 100.0% RH ±3% RH (10 to 90%) @ 23 oC Temperature (internal) -4.0 to 122.0 oF (-20 to 50 oC) ±1.8 oF (±1 oC) Temperature (external) -4.0 to 158.0 oF (-20 to 70 oC) ±1.8 oF (±1 oC) Display Triple LCD Sensor Type Humidity: Precision capacitance sensor Temperature: Thermistor Response Time 60 seconds typical Dew Point -90.4 to 122.0 oF (-68 to 50 oC) (calculated
RH300 V1.4 11/04 3 /WB T2 T/DP T1 1 2 3 4 5 6 7 8 9 10 11 Display Description 1. Relative Humidity % 2. T1 3. T1-T2 4. Dew Point 5. Wet Bulb 6. T2-Dew Point 7. °C/°F 8. T2 9. Maximum 10. Minimum 11. Low Battery/Hold Operation Open the Protective Sensor Cap Twist the protective sensor cap, located at the top of the meter, counter-clockwise a quarter of a turn to expose the sensor Plug in the optional external temperature probe (if needed). Turn Pow
RH300 V1.4 11/04 4 RESET MX/MN RESET MX/MN RESET MX/MN RESET MX/MN /WB T2 T/DP T1 Wet Bulb and Dew Point measurements 1. To read the Wet Bulb measurement, press the button until the symbol Wb appears in the lower right display. Note: T2 will not be displayed unless the external probe is connected. 2. Wet Bulb temperature is now displayed. 3. To read the Dew Point measurement, press the button until the symbol DP appears in the low
RH300 V1.4 11/04 5 RESET MX/MN RESET MX/MN CALIBRATION The following verification and calibration procedures require the 33% and 75% RH reference bottles which are supplied with the Model RH305 kit. Accuracy Verification Checking the 33% or 75% RH Calibration 1. Insert meter’s sensor into the 33% or 75% salt reference bottle 2. Check the reading after 10 minutes 3. Verify that the reading is within the accuracy specification Relative Humidity Calib
RH300 V1.4 11/04 6 � Support line (781) 890-7440 Technical support: Extension 200; E-mail: [email protected] Repair & Returns: Extension 210; E-mail: [email protected] Product specifications subject to change without notice For the latest version of this User’s Guide, Software updates, and other up-to-the-minute product information, visit our website: www.extech.com Maintenance Cleaning and storage 1. The meter should be cleaned with a damp clot

Manual del usuario

Sicrómetro Digital Modelo RH300 y RH305 (kit) Introducción Felicitaciones por su compra del Sicrómetro digital RH300 de Extech Este dispositivo mide humedad, temperatura del aire, temperatura de punto de rocío y temperatura de termómetro de bulbo húmedo . Las características avanzadas incluyen medición de temperatura exterior (con sonda opcional), retención de datos, apagado automático optativo y lecturas MÁX / MIN/Promedio. E

RH300 V1.4 11/04

2 Especificaciones Función Escala y Resolución Precisión Humedad 0.0 a 100.0% RH ±3% RH (10 a 90%) @ 23 oC Temperatura (interior) -20.0 a 50 oC (-4 a 122.0 oF) ±1 oC (±1.8 oF) Temperatura (exterior) -20 a 70 oC (-4.0 a 158.0 oF) ±1 oC (±1.8 oF) Pantalla LCD triple Sensor tipo Humedad: Sensor de precisión para capacitancia Temperatura: Termistor Punto de rocío -68 a 50 oC (-90.4 a 122.0 oF) Bulbo húmedo -21.6 a 50 oC (-6.88 a 12

Extech Instruments Measuring Instruments Owners Instructions: