EA80 V2.1H 11/06

3 Preparation for Use Power Supply The meter is powered by six AAA-size alkaline batteries or an AC adapter. Installing the Batteries Insert six AAA-size batteries as indicated by the diagram located on the inside of the battery compartment. When the battery voltage drop below the operating voltage, the “ + ” indicator will be displayed indicating that the batteries need to be changed. AC Adapter The AC adapter allows you to power t
EA80 V2.1H 11/06 4 Humidity Measurement 1. Press the “ ” button to power on the meter. 2. The display will show the humidity reading value (% RH) directly on the third display. 3. Hold the probe in the air in the area to be tested. Do NOT immerse in liquid. 4. Allow adequate time for readings to stabilize. 5. Read the measurements on the LCD. Temperature Measurement 1. Press the “ ” button to power on the meter. 2. Press the °C/°F button momen
EA80 V2.1H 11/06 5 MAX CO2 ppm d-h m:s 8 MIN CO2 ppm d-h m:s CO2 ppm d-h m:s CO2 Maximum and Minimum Recording Measurement 1. Press “MX/MN” button to enter the maximum / minimum recording mode, the “MAX” icon appears on the display. The maximum CO2 reading will be displayed and will be updated only when a higher reading occurs. 2. Press “TIME” button to enter time display mode. Displays 2 and 3 will indicate the time the maximum reading occur
EA80 V2.1H 11/06 6 SET ALM CO2 ppm ppm CO2 ppm CO2 CO2 Alarm Operation Setting the Alarm Limit Values 1. Press “ALARM” button to turn on the alarm function, the “ALM” icon, and current value are displayed. 2. Press “SET” button to enter High/Low limit value setting mode, the “SET” icon is displayed and hundred and thousand digits of the high limit value will flash. 3. Press “�” or “�” button to set desired value. 4. Press “�
EA80 V2.1H 11/06 7 Manual Dataogging Storing readings Press " MEM " button. LCD will display " M " and the memory address number. Total memory size is 99 readings. Viewing readings. Press " READ " button to enter READ mode. The LCD will display " R " and the memory address number. Press " �" or " �" button to scroll through stored readings Press " ↵ " enter button to exit this mode. Deleting stored data Press “ SET ” button three times, the L
EA80 V2.1H 11/06 8 Calibration CO2 Calibration 1. Press “ ” button to turn the meter on. 2. Place the probe in a known CO2 reference for 10 minutes. 3. Press the “SET” button 5 times until “C-01” is displayed in the second display. 4. Press “►” or “◄” button to select digits to adjust (flashing). 5. Press “▼” or “▲”button to adjust the display to the reference value. 6. Press “↵” to store the value and exit the c
EA80 V2.1H 11/06 9 Specifications Display : LCD triple display. Display Rate : One time per second. Low Battery Indication : The “ + ” is displayed when the battery voltage drops below the operating voltage. Power Supply : Six (6) AAA-size alkaline batteries. or Regulated AC adapter. Battery Life : Approx. 8 hours by using alkaline batteries (without backlight or Alarm function). If datalogging over a long period of time, use an AC adapter. Manual D
EA80 V2.1H 11/06 10 Maintenance Cleaning Periodically wipe the case with a dry cloth or a damp cloth with mild detergent. Do not use abrasives or solvents to clean this instrument. Battery Replacement When the “ ” symbol appears on the LCD, the six 1.5V ‘AAA’ must be replaced. 1. Turn the meter off. 2. Remove the meter’s battery cover 3. Replace the batteries observing polarity 4. Affix the battery cover and secure the rear screws. RS-