Honeywell BW Ultra Quick Reference Manual

Text of the manual

Honeywell BW ™ Ultra

Quick Reference Guide | Gas Detection Portable Five-gas Detector with Internal Pump ENGLISH

The Honeywell BW ™ Ultra gas detector warns of

hazardous gas at levels above user-defined alarm setpoints. The detector is a personal safety device. It is your responsibility to respond properly to the alarm. Monitored Gases The detector can monitor up to five gases at a time. Refer to the Honeywell BW ™ Ultra User Manual for the ful

Use the detector only as specified in this manual,

otherwise, the protection provided by the detector may be impaired. • Only the instrument capable of sounding the alarms and showing readings on a display should be used for immediate safety critical use. Wireless communication and infrastructure are only for informational monitoring. �

• Honeywell recommends that the sensors be bump tested before each day’s

use to confirm their ability to respond to gas. Manually verify that the audible, visual, and vibrator alarms are activated. Calibrate if the readings are not within the specified limits. • Portable safety gas detectors are life safety devices. The accuracy of ambient gas

• Warning: Substitution of components may impair Intrinsic Safety.

• To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing. • Protect the combustible sensor from exposure to lead compounds, silicones, and chlorinated hydrocarbons. Although certain organic vapors (such as leaded gasoline and halogenat

1. Visual alarm

indicator 2. Exhaust port 3. Screen 4. Button 5. Beeper Aperture 6. Pump inlet 7. Alligator clip 8. Pump assembly 9. Charging connector and IR Interface 5

Pump critical fail

Pairing failed Warning/failure/ error/low battery Stealth mode enabled Hole watch mode high Battery -three levels Hole watch mode low Low battery Hole watch -Oxygen bar IR LINK connection Correction factor Calibration passed Peak gas exposure Calibration failed Press

Screen Icons Target gas Inert mode Bump test cancelled Sensor failure Bump test passed STEL alarm Bump test failed TWA alarm Over limit alarm Firmware update in progress High alarm Sensor disabled Low alarm Heartbeat