Honeywell BW Ultra Quick Reference Manual
Text of the manual
Honeywell BW ™ Ultra
Quick
Reference
Guide
| Gas Detection
Portable Five-gas Detector
with Internal Pump
ENGLISH
The Honeywell BW ™ Ultra gas detector warns of
hazardous gas at levels above user-defined alarm
setpoints.
The detector is a personal safety device. It is your
responsibility to respond properly to the alarm.
Monitored Gases
The detector can monitor up to five gases at a time.
Refer to the Honeywell BW ™ Ultra User Manual for the ful
Use the detector only as specified in this manual,
otherwise, the protection provided by the detector may be
impaired.
• Only the instrument capable of sounding the alarms and showing readings
on a display should be used for immediate safety critical use. Wireless
communication and infrastructure are only for informational monitoring.
�
• Honeywell recommends that the sensors be bump tested before each day’s
use to confirm their ability to respond to gas. Manually verify that the audible,
visual, and vibrator alarms are activated. Calibrate if the readings are not
within the specified limits.
• Portable safety gas detectors are life safety devices. The accuracy of ambient
gas
• Warning: Substitution of components may impair Intrinsic Safety.
• To prevent ignition of flammable or combustible atmospheres, disconnect
power before servicing.
• Protect the combustible sensor from exposure to lead compounds, silicones,
and chlorinated hydrocarbons. Although certain organic vapors (such as
leaded gasoline and halogenat
1. Visual alarm
indicator
2. Exhaust port
3. Screen
4. Button
5. Beeper Aperture
6.
Pump inlet
7. Alligator clip
8. Pump assembly
9. Charging
connector and IR
Interface
5
Several icons are displayed in the detector’s screen as
conditions dictate.
BW ULTRA
Pump critical fail
Pairing failed
Warning/failure/
error/low battery
Stealth mode
enabled
Hole watch
mode high
Battery -three
levels
Hole watch
mode low
Low battery
Hole watch
-Oxygen bar
IR LINK
connection
Correction factor
Calibration
passed
Peak gas
exposure
Calibration
failed
Press