Extech Instruments 407790 [ 1 / 3 OCT ] 01 / 07 / 27 14:26:39

100 W A L Hz 125 30 50 31.5 500 70 90 dB FAST 75.2 dB 8k 2k L W 80 60 40 User Manual Real Time Octave Band Analyzer Model 407790 Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 FAX 781.665.0780 - TestEquipmentDepot.com 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.TestEquipmentDepot.com Back to the Extech 407790 Product Page

Model 407790 Ver. 1.1 7/02L

2 whether written or oral, is expressed or implied. Introduction Congratulations on your purchase of the Extech 407790 Real Time Octave Band Analyzer. This auto ranging Type 2 integrating sound level meter features 1/1 and 1/3 Octave Band real-time noise analysis. The 407790 offers five measurement parameters: SPL (Sound Pressure Level), Leq (Equivalent Continuous Sound Pressure Level), LE (Sound Exposure Level), Lmax (Maximum
Model 407790 Ver. 1.1 7/02L 3 Display warning indicators: OVER indicator (Input signal exceeds the upper limit) UNDER indicator (Input signal under the lower limit) Memory capacity: See Table below (Manual/Auto storage blocks are separate) Storage type Mode Manual Auto Sound level meter 1024 data sets 10000 data sets 1/1-octave analysis 1024 data sets 6140 data sets 1/3-octave analysis 1024 data sets 3070 data sets AC output: 2 Vrms at FS (full scale); Ou
Model 407790 Ver. 1.1 7/02L 4 Meter Description Refer to the diagram below for the following descriptions. 1. Windscreen 2. Microphone 3. Display 4. Operation keys 5. Power switch 6. Calibration adjust knob 7. 6VDC adapter jack 8. DC analog output jack 9. AC analog output jack 10. LCD contrast adjust knob 11. LCD backlight on/off switch 12. Hand strap 13. RS-232 interface connector 14. Battery compartment 15. Tripod mounting screw FAST L 2k [ 1 / 3 OCT
Model 407790 Ver. 1.1 7/02L 5 Operation Buttons Refer to the diagram on the next page for the following button descriptions. MODE button Press to display the following: Sound pressure level (L), equivalent continuous sound pressure level (Leq), Sound exposure level (LE), maximum sound pressure level (Lmax), and minimum sound pressure level (Lmin). FREQ WGHT button Selects the frequency weighting. “A” weighting (A), “C” weighting (C), and flat fr
Model 407790 Ver. 1.1 7/02L 6 STORE button In manual dataloggin mode, press to store the measurement data. RECALL button In manual memory mode, press to recall the stored memory data. STORE + RECALL key (Erase memory data) Turn off the meter, press and hold the STORE and RECALL buttons and then turn on the meter. Release buttons when the LCD displays “All memory erased”.
Model 407790 Ver. 1.1 7/02L 7 Measurement Screens 1. Sound pressure level measurement screen [ SLM ] 01 / 07 / 27 14:26:39 A L 30 80 FAST dB 130 2. 1/1-octave analysis screen The staircase shape on the display represents the 1/1-octave analysis result. From left, the bars correspond to 31.5, 63, 125, 250, 500, 1k, 2k, 4k, 8kHz. To read the levels at these frequencies, use the �Cursor� key to move the cursor to the desired point. The f

[ 1 / 3 OCT ] 01 / 07/ 27 14:26:39

31.5 Hz 30 50 70 2k 500 125 8k L W A FAST 90 dB L W 75.2 dB 100 60 80 40 Calibration Use a standard acoustic calibrator such as the Extech 407744 or 407766. Acoustic calibration if often performed on a daily basis to compensate for any possible changes in microphone sensitivity. 1. Configure the following switch settings. Display: SLM (LA) Time weighting: FAST 2. Insert the microphone carefully into the insertion h
Model 407790 Ver. 1.1 7/02L 9 Setting the Calendar Clock Date and time information is stored with each record saved. Set the date and time as follows: 1. Press the 2nd button once and then press the MEAS TIME button to enter the date and time mode. The display screen in the second function location will flicker. 2. Press the �LEVEL� button to increment/decrement the values. 3. Use the �Cursor� button to scroll through year/month/day/hour/minute/