Extech Instruments 407790 Operation & User’s Manual
| Brand: | Extech Instruments |
| Category: | Measuring Instruments |
| Model: | 407790 (Operation & User’s Manual) |
| File Type: | |
| File Size: | — |
| Pages: | 17 |
Text of Extech Instruments 407790 Operation & User’s Manual
Extech Instruments 407790 [ 1 / 3 OCT ] 01 / 07 / 27 14:26:39
100
W
A
L
Hz
125
30
50
31.5
500
70
90
dB
FAST
75.2 dB
8k
2k
L W
80
60
40
User Manual
Real Time Octave Band Analyzer
Model 407790
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Back to the Extech 407790 Product Page
Model 407790 Ver. 1.1 7/02L
2
whether written or oral, is expressed or implied.
Introduction
Congratulations on your purchase of the Extech 407790 Real Time Octave Band
Analyzer. This auto ranging Type 2 integrating sound level meter features 1/1 and 1/3
Octave Band real-time noise analysis. The 407790 offers five measurement parameters:
SPL (Sound Pressure Level), Leq (Equivalent Continuous Sound Pressure Level), LE
(Sound Exposure Level), Lmax (Maximum
Model 407790 Ver. 1.1 7/02L 3
Display warning indicators:
OVER indicator (Input signal exceeds the upper limit)
UNDER indicator (Input signal under the lower limit)
Memory capacity: See Table below (Manual/Auto storage blocks are separate)
Storage type
Mode
Manual
Auto
Sound level meter
1024 data sets
10000 data sets
1/1-octave analysis
1024 data sets
6140 data sets
1/3-octave analysis
1024 data sets
3070 data sets
AC output: 2 Vrms at FS (full scale); Ou
Model 407790 Ver. 1.1 7/02L 4
Meter Description
Refer to the diagram below for the following descriptions.
1.
Windscreen
2.
Microphone
3.
Display
4.
Operation keys
5.
Power switch
6.
Calibration adjust knob
7.
6VDC adapter jack
8.
DC analog output jack
9.
AC analog output jack
10. LCD contrast adjust knob
11. LCD backlight on/off switch
12. Hand strap
13. RS-232 interface connector
14. Battery compartment
15. Tripod mounting screw
FAST L
2k
[ 1 / 3 OCT
Model 407790 Ver. 1.1 7/02L 5
Operation Buttons
Refer to the diagram on the next page for the following button descriptions.
MODE button
Press to display the following: Sound pressure level (L), equivalent continuous sound
pressure level (Leq), Sound exposure level (LE), maximum sound pressure level
(Lmax), and minimum sound pressure level (Lmin).
FREQ WGHT button
Selects the frequency weighting. “A” weighting (A), “C” weighting (C), and flat
fr
Model 407790 Ver. 1.1 7/02L 6
STORE button
In manual dataloggin mode, press to store the measurement data.
RECALL button
In manual memory mode, press to recall the stored memory data.
STORE + RECALL key (Erase memory data)
Turn off the meter, press and hold the STORE and RECALL buttons and then turn
on the meter. Release buttons when the LCD displays “All memory erased”.
Model 407790 Ver. 1.1 7/02L 7
Measurement Screens
1. Sound pressure level measurement screen
[ SLM ] 01 / 07 / 27 14:26:39
A
L
30
80
FAST
dB
130
2. 1/1-octave analysis screen
The staircase shape on the display represents the 1/1-octave analysis result. From left,
the bars correspond to 31.5, 63, 125, 250, 500, 1k, 2k, 4k, 8kHz. To read the levels at
these frequencies, use the �Cursor� key to move the cursor to the desired point.
The f
[ 1 / 3 OCT ] 01 / 07/ 27 14:26:39
31.5
Hz
30
50
70
2k
500
125
8k
L
W
A
FAST
90
dB L
W
75.2 dB
100
60
80
40
Calibration
Use a standard acoustic calibrator such as the Extech 407744 or
407766. Acoustic calibration if often performed on a daily basis to
compensate for any possible changes in microphone sensitivity.
1. Configure the following switch settings.
Display: SLM (LA)
Time weighting: FAST
2. Insert the microphone carefully into the insertion h
Model 407790 Ver. 1.1 7/02L 9
Setting the Calendar Clock
Date and time information is stored with each record saved. Set the date and time as follows:
1. Press the 2nd button once and then press the MEAS TIME button to enter the date and
time mode. The display screen in the second function location will flicker.
2. Press the �LEVEL� button to increment/decrement the values.
3. Use the �Cursor� button to scroll through year/month/day/hour/minute/